Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K993904

    Validate with FDA (Live)

    Date Cleared
    2000-02-01

    (76 days)

    Product Code
    Regulation Number
    876.5010
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The IntraStent™ DoubleStrut™ LD Biliary Stent is indicated as a palliative treatment for malignant neoplasms in the biliary tree.

    Device Description

    The IntraStent™ DoubleStrut™ LD is a balloon expandable stainless steel stent with an open cell design. The stent is provided unmounted, to be manually crimped onto an open lumen PTA balloon catheter for biliary stent expansion of choice by the physician. Upon balloon inflation the crimped stent expands to conform to the duct inner luminal surface and retains the expanded state upon balloon deflation.

    AI/ML Overview

    The provided text describes a 510(k) submission for the IntraStent™ DoubleStrut™ LD Biliary Stent, focusing on demonstrating substantial equivalence to predicate devices, rather than a study proving the device meets acceptance criteria from a clinical outcome perspective.

    Here's an analysis based on the information provided, specifically addressing the requested points:

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission does not define specific clinical acceptance criteria in terms of performance metrics (e.g., patency rates, complication rates). Instead, the "acceptance criteria" are implied as demonstrating substantial equivalence through bench tests to established predicate devices.

    Acceptance Criterion (Implied)Reported Device Performance
    Meet same performance characteristics as predicate IntraStent™ DoubleStrut™ (K991929)Bench tests were performed to verify this.
    Non-pyrogenicPyrogenicity testing performed on a lot-to-lot basis supports this claim.
    Identical materials to predicate devicesStated as identical to currently marketed IntraStent™ DoubleStrut™ and Cordis Corporation/Johnson & Johnson PALMAZ™ and PALMAZ-SCHATZ™ stents.
    Identical indication for use to predicate devicesStated as identical: "palliative treatment for malignant strictures of the biliary tree."
    Identical technological characteristics to predicate devicesStated as identical, providing a larger version of IntraStent™ DoubleStrut™ (9-12 mm expanded diameter), fabricated by laser-cutting 316L stainless steel hypotube.

    2. Sample Size Used for the Test Set and Data Provenance

    The submission describes bench tests and pyrogenicity testing, not clinical trials with patient data.

    • Test Set Sample Size: Not explicitly stated for bench tests (e.g., number of stents tested). For pyrogenicity, it's mentioned as "lot to lot basis," implying samples from manufacturing lots.
    • Data Provenance: Not applicable in the context of clinical data (no country of origin or retrospective/prospective clinical data mentioned). The data provenance is from internal bench testing conducted by the manufacturer.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. The ground truth for bench tests is the engineering specification and performance of the predicate device. There is no mention of experts establishing a "ground truth" for the test set in the sense of clinical diagnoses or interpretations.

    4. Adjudication Method for the Test Set

    Not applicable. This concept primarily applies to human reader studies or clinical trials where expert consensus is needed to resolve discrepancies. Bench tests rely on objective physical measurements and compliance with engineering specifications.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance?

    Not applicable. This is a medical device (stent), not an AI-powered diagnostic or decision-support system. Therefore, MRMC studies comparing human readers with and without AI assistance are not relevant to this submission.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable. This is a medical device (stent), not an algorithm or AI system.

    7. The Type of Ground Truth Used

    The "ground truth" for this 510(k) submission is the performance and characteristics of the legally marketed predicate devices (IntraStent™ DoubleStrut™ K991929 and Cordis Corporation PALMAZ™ and PALMAZ-SCHATZ™ stents). The testing (bench tests, pyrogenicity) aimed to demonstrate that the new device's performance aligns with these established predicates.

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI or machine learning device requiring a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As above, this is not an AI or machine learning device.

    Summary of the Study Proving Acceptance Criteria:

    The study proving the device meets the "acceptance criteria" (understood as demonstrating substantial equivalence to predicate devices) primarily involved:

    • Bench Tests: These were conducted to verify that the IntraStent™ DoubleStrut™ LD met the "same performance characteristics" as the predicate IntraStent™ DoubleStrut™ (K991929). The submission states "Performance testing (bench) further supports a substantial equivalence claim." No specific details about the types of bench tests (e.g., radial force, fatigue, crimp retention) or their results are provided in this summary.
    • Pyrogenicity Testing: This was performed on a "lot to lot basis" to support the claim that the device is non-pyrogenic.

    The conclusion is that "The collective evidence therefore provides assurance that the IntraStent™ meets the requirements that are considered acceptable for the intended use" by demonstrating its "identical" nature in materials, indication for use, and technological characteristics, and by showing similar performance in bench tests, to its predicate devices.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1