Search Filters

Search Results

Found 5 results

510(k) Data Aggregation

    K Number
    K192176

    Validate with FDA (Live)

    Device Name
    IntelePACS
    Date Cleared
    2020-04-02

    (234 days)

    Product Code
    Regulation Number
    892.2050
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    IntelePACS is a software application that receives digital images and data from various sources (such as, CT scanners, MR scanners, ultrasound systems, R/F units, computer and direct radiographic devices, secondary capture devices, scanners, imaging gateways, or other imaging sources). Images and data can be communicated, processed, manipulated, enhanced, stored, and displayed within the system and/or across computer networks at distributed locations. Postprocessing of the images can be performed using Multi Planar Reconstruction (MPR). Only preprocessed DICOM for presentation images can be interpreted for primary image diagnosis in mammographic images with lossy compression and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using an FDA cleared display that meets technical specifications reviewed and accepted by FDA.

    When used with a mobile device, IntelePACS is suitable for diagnostic image review only on tested devices as specified in Intelerad product documentation. IntelePACS is not intended for primary diagnostic image review on mobile devices. Mobile usage for Mammography is for reference and referral only.

    Device Description

    IntelePACS, (version 5), is based upon the predicate device IntelePACS, version 4.9.1 (K150707). Both the subject device and the predicate are from Intelerad Medical Systems Inc. IntelePACS is comprised of software modules that provide image capture, storage, distribution, enhancement, manipulation, and networking of medical images at distributed locations. In cases where DICOM images are not directly available to IntelePACS, the system can acquire medical images using a DICOM image gateway, which generates DICOM-type files. For example, film digitizers obtain images from original film and convert them to meet DICOM standards. Stored files are transmitted using a network and can be viewed or manipulated using InteleViewer viewing software.

    The main components are:

    • InteleViewer is an image viewing system, providing access to the tools required for reviewing images and searching to find studies and reports. InteleViewer features DICOM support, including DICOM Query/Retrieve, Store, and Print. The software module includes a thumbnail display, real-time cine playback, linked stacking, MPR, annotations and measurements, etc. The web-enabled InteleViewer software uses JPEG 2000 compression and an advanced streaming protocol to provide lossless or lossy image viewing and manipulation. The user can examine multiple studies simultaneously and define customized image layouts. With a mouse click, the User can stack, rotate, zoom, take measurements, and view reports.
    • InteleConnect Enhanced Viewer (used on Mobile Devices) is a web-based portal for medical images review that can be used on the desktop and on mobile devices without requiring software installation. This is a new module which is not in the Primary Predicate but is available in the Reference Predicate device.
    • InteleBrowser which is a web-based administrative interface for IntelePACS, and allows the facility System Administrator to customize and configure the facility radiological workflow and corporate branding. The System Administrator can administer hundreds or thousands of IntelePACS users by taking advantage of the user management and access granting tools.
    AI/ML Overview

    The provided text describes the 510(k) summary for IntelePACS, focusing on its substantial equivalence to predicate devices and performance data for mobile device use. However, it does not contain a table of acceptance criteria or a detailed study proving the device meets specific performance criteria beyond general statements of compliance with standards and limitations for mobile displays.

    Here's what can be extracted and what is missing based on your request:

    1. A table of acceptance criteria and the reported device performance:

    This information is not explicitly provided in the document in the format requested. The document mentions that "Validation testing indicated that as required by the risk analysis, designated individuals performed all verification and validation activities and that the results demonstrated that the predetermined acceptance criteria were met." However, the specific acceptance criteria themselves and the detailed performance metrics are not listed in a table.

    The closest to "reported device performance" is the section on mobile device testing, which presents observations and measurements regarding:

    • Spatial Resolution: "Devices tested provide high resolution displays with high pixels per inch counts, with pixel pitch well below 250 um. ... All devices tested, satisfy this criterion."
    • Pixel Defects: "Devices were visually inspected for any pixel defects and none were detected."
    • Artifacts: "At the end of the test, a visual observation was performed and confirmed that no such occurrence of either artifact, for all devices tested."
    • Temporal Response: "All devices tested present a reasonable response time in the 40-60% luminance transitions, namely around 30ms or better."
    • Luminance: "Maximum luminance is well above 350 cd/m² for all devices tested."
    • Conformance to DICOM GSDF: "All devices tested only conform to DICOM GSDF within 20% over subsets of the JND range. While some mobile devices performed better than others, any one of the mobile devices did not fully satisfy the criteria." (This is a non-conformance with a subsequent mitigation.)
    • Color Tracking: "All values for △D65(u',v') are within acceptable range for all devices tested."

    2. Sample size used for the test set and the data provenance:

    • Sample size: Four specific mobile devices were tested: iPad Pro 2017, iPhone 8, iPhone XS, and iPad Air (2019). It's unclear if "devices tested" refers to a single unit of each model or multiple units.
    • Data provenance: Not explicitly stated, but the testing seems to be internal ("A test lab evaluated...") to the manufacturer or a contracted lab. The document does not specify the country of origin for the data or whether it was retrospective or prospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Number of experts: Not mentioned.
    • Qualifications of experts: Not mentioned. The testing seems to be based on physical measurements and visual inspection against standards (AAPM TG18, AAPM report 270, AAPM report No.260), rather than expert consensus on diagnostic images.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Adjudication method: Not applicable/not mentioned. The testing described is physical display characteristic assessment, not diagnostic image review requiring adjudication among readers.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • An MRMC comparative effectiveness study was not conducted or described. This submission is for a PACS system, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • This is not applicable as IntelePACS is a PACS system for image management and viewing, not a diagnostic algorithm. The performance evaluation focuses on display quality.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The "ground truth" for the mobile device display testing appears to be objective physical measurements and visual assessments against established industry standards and guidelines (e.g., pixel pitch, luminance, temporal response, adherence to DICOM GSDF, pixel defects, artifacts). It is not based on clinical "ground truth" like pathology or expert consensus on a disease state, as the device itself is not making a diagnosis.

    8. The sample size for the training set:

    • Not applicable. As a PACS system, IntelePACS does not involve a "training set" in the context of machine learning. Its functionality is based on established software engineering principles and adherence to standards like DICOM.

    9. How the ground truth for the training set was established:

    • Not applicable for the same reason as above.

    Summary of what's provided for acceptance criteria and study:

    The document states that IntelePACS "passed all predetermined testing criteria" and "met all the requirements listed in the Standards." For mobile devices, it lists specific objective measurements and observations (e.g., luminance, resolution, pixel defects) and whether the devices met those criteria, with one notable exception (DICOM GSDF conformance) for which mitigations were implemented. The study described is a technical assessment of display characteristics for specific mobile devices against generally accepted industry standards for medical displays, rather than a clinical trial measuring diagnostic accuracy.

    Ask a Question

    Ask a specific question about this device

    K Number
    K150707

    Validate with FDA (Live)

    Device Name
    IntelePACS
    Date Cleared
    2015-05-27

    (69 days)

    Product Code
    Regulation Number
    892.2050
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    IntelePACS™ is a device that receives digital images and data from various sources (such as, CT scanners, ultrasound systems, R/F units, computer and direct radiographic devices, secondary capture devices, scanners, imaging gateways, or other imaging sources). Images and data can be communicated, processed, manipulated, stored, and displayed within the system and/or across computer networks at distributed locations. Post-processing of the images can be performed using Multi Planar Reconstruction (MPR).

    Mammographic images may only be interpreted using an FDA cleared monitor that meets technical specifications reviewed and cleared by FDA. Typical users of this system are trained professionals, physicians, nurses, and technicians.

    Contraindications: Lossy compressed images and digitized film screen images must not be reviewed for primary image interpretations.

    Device Description

    IntelePACS™ is comprised of software modules that provide image capture, storage, distribution, enhancement, manipulation, and networking of medical images at distributed locations.

    In cases where DICOM images are not directly available to IntelePACS™, the system can acquire medical images using a DICOM image gateway, which generates DICOM-type files. For example, film digitizers obtain images from original film and convert them to meet DICOM standards. Stored files are transmitted using a network and can be viewed or manipulated using InteleViewer viewing software.

    AI/ML Overview

    This document is a 510(k) premarket notification for the IntelePACS™ device. It asserts substantial equivalence to a predicate device (IntelePACS, Model Version 3.X.X, K083520). The primary purpose of this document is to demonstrate that the new device does not raise new questions of safety or efficacy compared to the predicate.

    Here's an analysis of the provided text in relation to acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document describes the device as a Picture Archiving and Communications System (PACS) and outlines its functionalities and technical characteristics in comparison to a predicate device. The acceptance criteria are broadly focused on demonstrating that the new IntelePACS™ performs equivalently to its predicate and that any differences do not impact safety or efficacy.

    Acceptance Criteria (Implied)Reported Device Performance (IntelePACS™ Subject Device)
    Intended Use Equivalence: Acquire, view, edit, and store radiographs and related patient images, with the same user types (radiologists).Same as predicate; no difference.
    Technical Characteristics Equivalence: Performance across various functions (network, monitor, user interaction, import/export, acquisition devices, imaging interfaces, image organization, search, storage, viewing, measurement, annotation, operations, security, DICOM 3.0 compatibility, generator control, RAW/post-image data processing, thumbnail viewing, image reset, panning, magnify glass, fit image).Same as predicate for all these functionalities; no difference.
    Operating System Compatibility: Workstation and Server OS updates (Windows 7 64-bit Professional, Red Hat Enterprise Linux 6 64-bit x86) are validated to ensure no impact on safety or efficacy.Confirmed: "The system has been tested and validated... and the results of testing verified that there is no impact on safety or efficacy and that no additional risks have been identified."
    Hardware Performance: Updated Server Memory (16GB vs 8GB), Server Database (Sybase and PostgreSQL), Workstation Processor (Intel 6-core Xeon Processor 3.2GHz vs Intel Core 2 Duo 2Ghz), and Workstation Memory (6GB vs 2GB) are validated to ensure no impact on safety or efficacy.Confirmed: "The system has been tested and validated... and the results of testing verified that there is no impact on safety or efficacy and that no additional risks have been identified."
    Software Feature Equivalence: Inclusion of "Blackford Registration" (K142337) does not affect image or intended use and has no impact on safety or efficacy.Confirmed: "The difference is a user preference and does not directly affect the image or intended use... no impact on safety or efficacy and that no additional risks have been identified."
    Overall Safety and Efficacy: The new device does not raise any new potential safety risks and is equivalent in performance to existing legally marketed devices.Confirmed: "Any difference between the two devices does not affect safety or efficacy." and "The modification to the subject device does not raise any new potential safety risks and is equivalent in performance to existing legally marketed devices."
    Manufacturing Standards Compliance: Manufactured in accordance with voluntary standards.Confirmed: "The subject device will be manufactured in accordance with the voluntary standards listed in the enclosed voluntary standard survey."
    Verification and Validation: All input, output, and actions performed by the software in each operational mode are assessed and tested.Confirmed: "The complete IntelePACS™ configuration has been assessed and tested at the factory and has passed all in-house testing criteria. The Verification Test Plan was designed to evaluate all input functions, output functions, and actions performed by the IntelePACS™ software in each operational mode and followed the process documented in the Validation Test Plan."

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not explicitly state the sample size used for the test set. It mentions "in-house testing criteria" and "all verification and validation activities" but without specific numbers of images, studies, or scenarios.

    The data provenance is implied to be retrospective as the testing is described as "in-house testing criteria" and "assessment at the factory." There is no mention of prospective data collection or specific countries of origin for the data, other than the company being based in Montreal, Quebec, Canada.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of those Experts:

    The document does not specify the number of experts or their qualifications used to establish ground truth for any test set. It clarifies that "A physician, providing ample opportunity for competent human interprets images and information being displayed and printed," implying that medical professionals are the end-users responsible for interpretation, but this doesn't detail their role in testing or establishing ground truth for the device's performance validation.

    4. Adjudication Method for the Test Set:

    The document does not specify any adjudication method (e.g., 2+1, 3+1, none) used for the test set. It mentions "designated individuals performed all verification and validation activities" but does not detail how discrepancies or consensus were managed.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    No, an MRMC comparative effectiveness study was not done. This 510(k) notification is for a Picture Archiving and Communications System (PACS), which is infrastructure software for managing and viewing medical images. It is not an AI algorithm designed to interpret images or assist human readers in diagnosis. Therefore, there is no discussion of human reader improvement with or without AI assistance, or any effect size.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    No, a standalone (algorithm only) performance study was not done in the context of an AI algorithm providing diagnostic interpretations. The device itself (IntelePACS™) is a standalone system in the sense that it performs its functions (image capture, storage, distribution, manipulation, viewing) without continuous human interaction for each specific task in its backend processing. However, it is not an AI algorithm whose performance is measured in terms of diagnostic accuracy without human intervention. The device's primary function is to enable human clinicians to view and interpret images.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.):

    The document does not specify the type of ground truth used for its validation. Given its nature as a PACS system, the "ground truth" for its testing would likely relate to the accurate storage, retrieval, display fidelity, manipulation functionality, and network performance of medical images, rather than diagnostic outcomes or pathology reports. The testing would ensure that the digital images are handled correctly and consistently to allow accurate human interpretation.

    8. The Sample Size for the Training Set:

    This document is for a PACS system, not an AI/ML algorithm that requires a training set. Therefore, the concept of a "training set" does not apply to the described device or its validation within this 510(k) summary.

    9. How the Ground Truth for the Training Set was Established:

    As there is no training set for this device, a method for establishing its ground truth is not applicable and therefore not described.

    Ask a Question

    Ask a specific question about this device

    K Number
    K083520

    Validate with FDA (Live)

    Date Cleared
    2009-06-23

    (209 days)

    Product Code
    Regulation Number
    892.2050
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K070080

    Validate with FDA (Live)

    Device Name
    INTELEPACS
    Date Cleared
    2007-01-24

    (15 days)

    Product Code
    Regulation Number
    892.2050
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    IntelePACS™ is a device that receives digital images and data from various sources (such as, CT scanners, MR scanners, ultrasound systems, R/F units, computer and direct radiographic devices, secondary capture devices, scanners, imaging gateways, or other imaging sources). lmages and data can be communicated, processed, manipulated, enhanced, stored, and displayed within the system and/or across computer networks at distributed locations. Postprocessing of the images can be performed using Multi Planar Reconstruction (MPP),

    Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mannographic images may only be interpreted using an FDA approved monitor that protateliar manimographio may only be medification specifications reviewed and accepted by FDA.

    Typical users of this system are trained professionals, physicians, nurses, and technicians.

    Device Description

    IntelePACS™ is comprised of software modules that provide image capture, storage, distribution, enhancement, manipulation, and networking of medical images at distributed locations. In cases where DICOM images are not directly available to IntelePACS™, the system can acquire medical imem mugos using a DICOM image gateway, which generates DICOM-type files. For example, and a Broom images from original film and convert them to meet DICOM standards and stored. Stored files are transmitted using a network and can be viewed or manipulated from imaging workstation.

    AI/ML Overview

    The provided text is a 510(k) summary for IntelePACS™. After a thorough review, it is clear that this document does not contain the acceptance criteria or a study proving that the device meets such criteria.

    The 510(k) summary focuses on establishing substantial equivalence to predicate devices, providing a device description, and outlining indications for use. It mentions that the system will be manufactured in accordance with voluntary standards, and a hazard analysis classified potential hazards as "minor," but these are not the detailed performance acceptance criteria and study results typically expected for a medical device.

    Therefore, I cannot provide the requested table or answer the specific questions about the study, sample sizes, ground truth, or expert involvement based on the provided text. The document is primarily a regulatory submission demonstrating equivalence, not a scientific study report.

    Ask a Question

    Ask a specific question about this device

    K Number
    K032533

    Validate with FDA (Live)

    Device Name
    INTELEPACS
    Date Cleared
    2003-10-16

    (62 days)

    Product Code
    Regulation Number
    892.2050
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    IntelePACS™ is a device that receives digital images and data from various sources (e.g. CT scanners, MR scanners, ultrasound systems, R/F units, computer & direct radiographic devices, secondary capture devices, scanners, imaging gateways or other imaging sources). Images and data can be communicated, processed, manipulated, enhanced, stored, and displayed within the system and or across computer networks at distributed locations.

    Device Description

    IntelePACS™ is comprised of software modules that provide image capture, storage, distribution, enhancement, manipulation, and networking of medical images at distributed locations. In cases where DICOM images are not directly available to IntelePACS™, the system can acquire medical images using a DICOM image gateway, which generates DICOM-type files. For example, film digitizers obtain images from original film and convert them to meet DICOM standards and stored. Stored files are transmitted using a network and can be viewed or manipulated from imaging workstation.

    AI/ML Overview

    This document is a 510(k) Summary of Safety and Effectiveness for the IntelePACS™ device, submitted by Intelerad Medical Systems Inc. It outlines the device details and establishes substantial equivalence to a predicate device. However, it does not include a study that proves the device meets specific acceptance criteria with performance metrics, sample sizes, or ground truth details.

    Therefore, I cannot provide the requested information. The document focuses on regulatory compliance and substantial equivalence, not on a detailed performance study with acceptance criteria.

    Information not available in the provided text:

    • Acceptance Criteria and Reported Device Performance: No specific performance metrics or acceptance criteria are reported. The document states that "adequate information and data" were provided to determine substantial equivalence but doesn't detail what that data contained regarding performance.
    • Sample Size for Test Set and Data Provenance: Not mentioned.
    • Number of Experts and Qualifications for Ground Truth: Not mentioned.
    • Adjudication Method: Not mentioned.
    • Multi-reader Multi-case (MRMC) Comparative Effectiveness Study: Not mentioned.
    • Standalone (algorithm only) Performance Study: Not mentioned.
    • Type of Ground Truth Used: Not mentioned.
    • Sample Size for Training Set: Not mentioned.
    • How Ground Truth for Training Set was Established: Not mentioned.

    Summary of what the document does include:

    • Device Name: IntelePACS™
    • Common Name: Picture Archiving Communications System
    • Classification: 21 CFR 892.2050, System, Image Processing (Class II)
    • Predicate Device: PACSPLUS (K023460) from MEDICAL STANDARD CO., LTD.
    • Indications for Use: Receives digital images and data from various sources; communicates, processes, manipulates, enhances, stores, and displays images and data within the system or across networks.
    • Technological Characteristics: Medical device image software for PACS, used with computer hardware, does not contact the patient, does not control life-sustaining devices, and images are interpreted by a physician with human intervention.
    • Conclusion: Device is substantially equivalent to the predicate. Compliance with voluntary standards, hazard analysis ("minor" concern level).
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1