Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K133933

    Validate with FDA (Live)

    Date Cleared
    2014-04-02

    (100 days)

    Product Code
    Regulation Number
    876.4300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Injection Gold ProbeTM Bipolar Electronemostasis Catheter is indicated for use in endoscopic injection therapy (to deliver pharmacological injection agents, such as vasoconstrictors) and endoscopic electrohemostasis (cauterization of tissue and coagulation of blood) of actual or potential bleeding sites in the gastrointestinal tract. These sites include the following:

    • Peptic Ulcers
    • Dieulafoy Lesions
    • Mallory-Weiss Tears
    • Bleeding Polyp Stalks
    • Arteriovenous Malformations (AVMs)
    • Angiomata

    The Injection Gold Probe Bipolar Electrohemostasis Catheter also has irrigation capability. Any other use is not recommended.

    Device Description

    The Injection Gold Probe Catheter is an injection therapy and bipolar electrohemostasis catheter with irrigation capabilities. The Injection Gold Probe Catheter is available in a working length of 210 cm with outer diameters of 7F (2.3 mm) and 10F (3.3 mm), compatible with a minimum of 2.8 mm and 3.7 mm scope channels, respectively. The catheter shaft provides firmness for persendicular and tangential use, while the Mediglide coating on the catheter promotes minimal friction during endoscope passage.

    I. Injection Therapy
    Each Injection Gold Probe Catheter contains a 25 gauge injection needle for injection therapy. The catheter handle contains the injection needle hub. Injection is performed by attaching a luerlock syringe to the "Injection" hub and advancing the needle.

    II. Bipolar Electrohemostasis
    The gold-spiral electrode on the probe tip provides bipolar electrohemostasis. The Injection Gold Probe Catheter connects directly to the Endostat™ family of Bipolar Electrosurgical Generators. Or utilize the Bipolar Cable Adapter supplied by Boston Scientific for other bipolar electrosurgical generators.

    III. Irrigation
    Irrigation with saline or sterile water is performed by connecting the irrigation hub to the irrigation port of the bipolar generator or to a 30 cm3 (30 cc) or 60 cm3 (60 cc) luer-lock syringe or mechanical irrigation system for generators without irrigation capability.

    AI/ML Overview

    This document describes a Special 510(k) Premarket Notification for the Injection Gold Probe™ Bipolar Electrohemostasis Catheter. The submission focuses on demonstrating substantial equivalence to a predicate device (K110887) by addressing material changes made to the proposed device.

    Here's an analysis of the acceptance criteria and supporting study information:

    Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Biocompatibility: Device materials are biocompatible."All biocompatibility tests conducted on the proposed Injection Gold Probe™ Bipolar Electrohemostasis Catheter passed. Therefore, the Injection Gold ProbeTM Bipolar Electrohemostasis Catheter is considered biocompatible."
    Bench Testing - Required specifications for the following tests:"All device bench tests results were acceptable. The data demonstrate that the Injection Gold Probe™ Bipolar Electrohemostasis Catheter meets design specifications and is suitable for its intended use."
    - Combination Therapy DeviceMet required specifications.
    - Device Irrigation Flow RateMet required specifications.
    - Injection Needle Flow RateMet required specifications.
    - Injection Needle Extension and RetractionMet required specifications.
    - Injection Needle Retraction LockMet required specifications.
    - Injection Hub to 22-gauge Hypotube Joint Tensile StrengthMet required specifications.
    - 25-gauge Injection Needle to the 22-Gauge Hypotube Joint Tensile StrengthMet required specifications.
    - Needle SpecificationMet required specifications.

    Study Information

    1. Sample size used for the test set and the data provenance:

      • The document does not specify sample sizes for the individual bench tests. It only states that tests were performed.
      • The data provenance is not explicitly mentioned (e.g., country of origin). However, as this is a 510(k) submission to the FDA (USA), the testing would typically be conducted under US or internationally recognized standards. The studies appear to be prospective in nature, designed to evaluate the modified device against predetermined specifications.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This document describes engineering bench testing and biocompatibility testing, not clinical studies involving expert assessment of images or patient outcomes. Therefore, no "experts" in the context of clinical ground truth establishment were utilized or mentioned. The "ground truth" for these tests would be the established engineering specifications and biocompatibility standards.
    3. Adjudication method for the test set:

      • Not applicable as this is not a study requiring adjudication of expert opinions (e.g., image interpretation). Bench tests and biocompatibility tests are evaluated against predefined, objective pass/fail criteria.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No, an MRMC comparative effectiveness study was not done. This is a submission for a medical device (a catheter) and not an AI/software device that assists human readers.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Not applicable. This is a physical medical device, not an algorithm.
    6. The type of ground truth used:

      • For biocompatibility: Adherence to established biocompatibility standards (e.g., ISO 10993 series).
      • For bench testing: Established engineering design specifications and performance requirements for the device's functions (e.g., flow rates, tensile strength, extension/retraction mechanisms).
    7. The sample size for the training set:

      • Not applicable. This is not a machine learning or AI-based device that requires a training set.
    8. How the ground truth for the training set was established:

      • Not applicable for the same reason as above.
    Ask a Question

    Ask a specific question about this device

    K Number
    K110887

    Validate with FDA (Live)

    Date Cleared
    2011-04-14

    (15 days)

    Product Code
    Regulation Number
    876.4300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Injection Gold Probe™ Bipolar Electrohemostasis Catheter is indicated for use in endoscopic injection therapy (to deliver pharmacological injection agents, such as vasoconstrictors) and endoscopic electrohemostasis (cauterization of tissue and coagulation of blood) of actual or potential bleeding sites in the gastrointestinal tract. These sites include the following: Peptic Ulcers, Dieulafoy Lesions, Mallory-Weiss Tears, Bleeding Polyp Stalks, Arteriovenous Malformations (AVMs), Angiomata. The Injection Gold Probe Bipolar Electrohemostasis Catheter also has irrigation capability. Any other use is not recommended.

    Device Description

    The Injection Gold Probe Catheter is an injection therapy and bipolar electrohemostasis catheter with irrigation capabilities. The Injection Gold Probe Catheter is available with two different style electrical connectors (plugs), Standard and Banana. The Injection Gold Probe Catheter is available in a working length of 210cm with outer diameter of 7Fr (2.3mm) and 10Fr (3.3mm) compatible with a minimum working channel of 2.8mm and 3.7mm.

    I. Injection Therapy
    Each Injection Gold Probe Catheter contains a 25gauge injection needle for injection therapy. The catheter handle contains the injection needle hub. Injection is performed by attaching a luerlock syringe to the "Injection" hub and advancing the needle.

    II. Bipolar Electrohemostasis
    The gold spiral electrode on the probe tip provides bipolar electrohemostasis. The Injection Gold Probe Catheter has a standard plug which adapts to Symmetry Endostat, Endostat II and Bicap bipolar electrosurgical generators or Banana plugs which are recommended for use with Valleylab bipolar electrosurgical generator containing female Banana plug receptacles. The Valleylab generators should be used with the appropriate Valleylab accessories, reference the operating manuals for specifications.

    III. Irrigation
    Irrigation with saline or sterile water is performed by connecting the irrigation hub to the irrigation port of the bipolar generator for the Standard plug or to a 30ml (30cc) or 60ml (60cc) luer-lock syringe for the Banana plug.

    AI/ML Overview

    The provided document is a 510(k) summary for the Boston Scientific Injection Gold Probe™ Bipolar Electrohemostasis Catheter. This type of submission is for medical devices that are substantially equivalent to a legally marketed predicate device, rather than a novel device requiring extensive clinical trials for effectiveness.

    Therefore, the document describes bench testing and biocompatibility testing to demonstrate that material and design changes to the proposed device do not negatively impact its safety and effectiveness compared to the predicate. It explicitly states it is not a study to prove device performance against clinical acceptance criteria often seen with AI/algorithm-based devices.

    Here's an analysis based on the provided text, addressing your questions where applicable, and noting when the information is not present as it pertains more to an AI/software device evaluation:


    Acceptance Criteria and Device Performance Study Analysis

    The document describes a 510(k) submission for a physical medical device (catheter) with material and design modifications. The "acceptance criteria" and "device performance" in this context refer to engineering specifications and safety requirements, not clinical diagnostic accuracy or effectiveness metrics typically associated with AI studies.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategoryDescription from DocumentReported Device Performance
    BiocompatibilityTo demonstrate that the material changes made to the proposed device are biocompatible. (Implicit criteria: meeting ISO standards for biocompatibility or similar industry standards for medical devices in contact with tissue)"All biocompatibility tests conducted on the proposed Injection Gold Probe™ Bipolar Electrohemostasis Catheter passed. Therefore, the Injection Gold Probe™ Bipolar Electrohemostasis Catheter is considered biocompatible."
    Bench Testing (Functional)To demonstrate that the modified device meets the required specifications of the completed tests. (Implicit criteria: performance metrics related to injection flow, electrohemostasis effectiveness, irrigation flow, mechanical integrity, dimensional tolerances, and electrical safety as per design specifications of the original device and relevant industry standards). Specifically mentioned changes: gold thick film conductor paste, coating, ceramic tip, radius tolerance, hypotube length."All device bench test results were acceptable. The data demonstrate that the Injection Gold Probe™ Bipolar Electrohemostasis Catheter meets design specifications and is suitable for its intended use."
    Substantial EquivalenceThe device must be substantially equivalent to a predicate device (K961349) in terms of intended use, fundamental design, and operating principle, with demonstrated safety and effectiveness of any modifications.Boston Scientific Corporation has demonstrated that the proposed Injection Gold Probe™ Bipolar Electrohemostasis Catheter is substantially equivalent to Boston Scientific Corporation's currently marketed predicate device.

    2. Sample Size Used for the Test Set and the Data Provenance

    • Test Set Sample Size: Not specified in the document. For bench and biocompatibility testing of a physical device, sample sizes are typically determined by statistical methods for engineering testing (e.g., AQL, confidence interval for reliability), but not explicitly detailed in a 510(k) summary in the same way clinical study sample sizes are.
    • Data Provenance: Not applicable in the context of clinical data provenance (e.g., country of origin, retrospective/prospective). The data comes from in vitro (bench) and in vivo (biocompatibility, likely animal or standardized material tests) testing.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • Not applicable as this is not an AI/algorithm device requiring expert-established ground truth for performance. The "ground truth" for the bench tests would be the established engineering specifications and performance benchmarks the device is designed to meet. Biocompatibility "ground truth" would be the standard material safety profiles and test protocols.

    4. Adjudication Method for the Test Set

    • Not applicable for the type of testing described (bench and biocompatibility tests). Adjudication methods like 2+1 or 3+1 are used for resolving disagreements in expert assessments for clinical ground truth, which is not relevant here.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No MRMC study was done. This type of study is specific to evaluating the impact of an AI system on human reader performance, which is not relevant for this physical medical device.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    • No standalone study (in the context of an algorithm's performance) was done. The "standalone" performance here would refer to the device's functional performance during bench testing, which was conducted to ensure it meets design specifications.

    7. Type of Ground Truth Used

    • For Biocompatibility: Standards-based toxicological and biological response profiles (e.g., ISO 10993 series).
    • For Bench Testing: Engineering design specifications, functional requirements, and established performance parameters of the predicate device and relevant industry standards for electrosurgical and injection catheters. This is effectively "engineering ground truth."

    8. Sample Size for the Training Set

    • Not applicable. This submission describes a physical medical device, not an AI/machine learning algorithm, so there is no "training set."

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable. As there is no training set for an algorithm, there's no ground truth to establish for it in this context.

    In summary: The provided document details a 510(k) submission for a modified physical medical device. The "study" described focuses on demonstrating biocompatibility and meeting design specifications through bench testing to establish substantial equivalence to a predicate device. It does not involve clinical studies or evaluations pertinent to AI performance, expert ground truth, or human-in-the-loop performance improvements typical of AI/CDSS submissions.

    Ask a Question

    Ask a specific question about this device

    K Number
    K961349

    Validate with FDA (Live)

    Date Cleared
    1996-05-15

    (37 days)

    Product Code
    Regulation Number
    876.4300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Modified Injection Gold Probe and Injection Gold Probe Direct are indicated for use in endoscopic injection therapy (to deliver pharmacological injection arents, such as vasoconstrictors) and endoscopic electrohemostasis (cauterization of tissue and coagulation of blood) of actual or potential bleeding sites in the gastrointestinal tract. These sites include the following: peptic ulcers; Mallory-Weiss tears; arteriovenous malformations (AVMs); Dieulafoy lesions; bleeding polyp stocks; and angiomata. The Modified Injection Gold Probe and Injection Gold Probe Direct also have irrigation capability.

    Device Description

    The Modified Injection Gold Probe is substantially equivalent to the currently-marketed Injection Gold Probe. Figure 9-1 compares the descriptive characteristics of these products.

    AI/ML Overview

    Table of Acceptance Criteria and Reported Device Performance:

    FeatureAcceptance Criteria (Modified Injection Gold Probe)Reported Device Performance (Injection Gold Probe - K942301)
    IndicationEndoscopic injection therapy and electrohemostasis of actual or potential bleeding sites in the gastrointestinal tract.Same
    Route of AdministrationEndoscopicSame
    Usable Length210 - 350 cmSame
    Shaft OD7 - 10 FSame
    Needle MaterialStainless SteelSame
    Needle Gauge25 GaugeSame
    Needle Extension Length4 - 6 mmSame

    Study to Prove Device Meets Acceptance Criteria:

    The provided document describes a 510(k) summary for the "Modified Injection Gold Probe." This submission claims substantial equivalence to a previously marketed device, the "Injection Gold Probe" (510(k) No. K942301). The study proving the device meets its acceptance criteria is not a clinical trial with human subjects, but rather a set of laboratory and biocompatibility tests designed to demonstrate that the new device is as safe and effective as the predicate device.

    1. Sample Size used for the test set and the data provenance:

      • Test Set Sample Size: The document does not specify a distinct "test set" in the context of clinical data for performance evaluation in the way one would for an AI/ML device. Instead, it refers to "laboratory testing regarding characteristics" performed on the Modified Injection Gold Probe. The number of units tested is not explicitly stated.
      • Data Provenance: The testing was conducted internally by Boston Scientific Corporation. The specific country of origin for the data is not mentioned beyond the company's address in Natick, MA, USA. This appears to be a retrospective analysis of laboratory test results.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This question is not applicable in the context of this 510(k) submission. The "ground truth" for demonstrating substantial equivalence is based on engineering specifications, material properties, and functional performance as verified through laboratory testing against established criteria for the predicate device, not through expert interpretation of clinical cases.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • This is not applicable. There was no clinical test set requiring expert adjudication.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No MRMC comparative effectiveness study was conducted or is relevant for this device. This device is a medical instrument (catheter), not an AI/ML diagnostic or assistive tool.
    5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

      • This is not applicable. The device is a physical medical instrument, not an algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The "ground truth" for this submission is established through:
        • Engineering specifications and design comparisons: The new device must meet the same design parameters (e.g., usable length, shaft OD, needle gauge) as the predicate device, as presented in Figure 9-1.
        • Functional performance testing: Laboratory tests were performed to verify safety and performance (e.g., electrohemostasis capabilities, injection therapy function).
        • Biocompatibility assessment: Performed on patient- and fluid-contact materials.
        • Sterilization and Pyrogenicity testing: To ensure the device is sterile and non-pyrogenic.
      • The acceptance criteria are essentially derived from the known safe and effective performance of the predicate device (Injection Gold Probe, K942301).
    7. The sample size for the training set:

      • This is not applicable. There is no "training set" in the context of this device's development or regulatory submission, as it is a medical instrument and not an AI/ML model.
    8. How the ground truth for the training set was established:

      • This is not applicable for the reasons stated above.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1