510(k) Data Aggregation

    K Number
    K200753

    Validate with FDA (Live)

    Date Cleared
    2020-06-09

    (78 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The IBS Implant System II is intended to replace missing teeth to restore chewing function. The IBS Implant System II can be placed in support of single or multiple-unit restorations including; cement retained, or overdenture restorations, and terminal or immediate abutment support for fixed bridgework. This system is for one or two stage surgical procedures. This system is intended for delayed loading.

    Device Description

    This submission is to add new magic Motion Abutment to the previously cleared device, IBS Implant System II (K162099).

    The Magic Motion Abutment system is composed of Magic Motion Abutment, Housing and O-ring. The Magic Motion Abutment is a device made of a material such as Ti-6AL-4V Eli (Conforming to ASTM Standard F-136). Magic Motion is an attachment-retained prosthetic product, which is used with patients with a fully edentulous maxilla. It is a one-piece structure. Various collar heights of the magic motions can be selected based on the gingival height. Magic motion does not have movement and it is a solid type.

    AI/ML Overview

    This FDA 510(k) K200753 letter is for a dental device, specifically an endosseous dental implant abutment called "IBS Implant System II" with an added component called "Magic Motion Abutment". As such, the approval is based on substantial equivalence to a predicate device, not on clinical performance criteria like those typically found in AI/ML device clearances. The document focuses on mechanical and material properties rather than AI model performance metrics.

    Therefore, many of the requested elements for an AI/ML device study are not applicable or cannot be extracted from this documentation.

    Here's an attempt to answer the questions based on the provided text, while highlighting what is not applicable:


    1. Table of Acceptance Criteria and Reported Device Performance

    Not directly applicable as this is a dental implant abutment, not an AI/ML device. The "acceptance criteria" here relate to meeting standards for mechanical properties, biocompatibility, and manufacturing, rather than statistical performance metrics like sensitivity/specificity for an AI model.

    The document discusses demonstration of substantial equivalence with the predicate device through design control activities and specific tests.

    Acceptance Criteria TypeReported Device PerformanceComments (Not AI/ML specific)
    Design/Dimensional ComparisonSubject device's diameter and hex are larger; neck and head diameter are larger and more round compared to the predicate. Magic Motion Housing height and diameter are larger.Stated that these differences "don't impact product's substantial equivalence" as they improve connection. This implies the design changes still fall within acceptable parameters for function and safety compared to the predicate.
    End User Steam Sterilization (AAMI TIR12:2010)Tests performed on predicate device and leveraged for subject device; "met the criteria of the standards".Confirms the device can be effectively sterilized by the end-user.
    Biocompatibility (ISO 10993-1, -10, -11)Tests performed on predicate device and leveraged for subject device; "met the criteria of the standards".Confirms the device material is safe for contact with biological tissues.
    Material Composition (Ti-6AL-4V Eli, ASTM F-136, Silicon rubber)Confirmed materials used.Demonstrates adherence to recognized material standards for medical implants.

    2. Sample Size Used for the Test Set and Data Provenance

    Not applicable for an AI/ML device. This is a mechanical device.

    • Test set sample size: Not specified in terms of clinical samples. Testing refers to mechanical and material property assessments on manufactured parts.
    • Data Provenance: Not mentioned in the context of clinical data. The tests performed are laboratory-based.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable as this is not an AI/ML device relying on human-established ground truth. Testing involves engineering standards and laboratory analyses.


    4. Adjudication Method for the Test Set

    Not applicable. This refers to methods for resolving discrepancies in expert labeling for AI/ML ground truth, which is not relevant here.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a MRMC study was not done. This type of study is specifically for evaluating the effectiveness of an AI system with human readers, which is not relevant for a dental implant abutment.


    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable. There is no "algorithm" or "standalone performance" for this mechanical dental device.


    7. The Type of Ground Truth Used

    Not applicable in the context of AI/ML. For this device, "ground truth" would correspond to established engineering standards, material specifications, and regulatory requirements that the device's characteristics (dimensions, material, sterility, biocompatibility) must meet.


    8. The Sample Size for the Training Set

    Not applicable for an AI/ML device. There is no "training set" for a mechanical dental implant.


    9. How the Ground Truth for the Training Set Was Established

    Not applicable for an AI/ML device. There is no "training set ground truth" for this mechanical device. The design, materials, and manufacturing processes adhere to established engineering principles and regulatory guidelines for dental implants.

    Ask a Question

    Ask a specific question about this device

    K Number
    K162099

    Validate with FDA (Live)

    Date Cleared
    2017-02-15

    (201 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The IBS Implant System II is intended to replace missing teeth to restore chewing function. The IBS Implant System II can be placed in support of single or multiple-unit restorations including; cement retained, or overdenture restorations, and terminal or immediate abutment support for fixed bridgework. This system is for one or two stage surgical procedures. This system is intended for delayed loading.

    Device Description

    An endosseous dental implant is a device made of a material such as Ti 6AL 4V Eli (Conforming to ASTM Standard F-136). The IBS Implant System II is consists of dental fixtures, screws and multiunit abutment cylinders.

    AI/ML Overview

    The provided document is a 510(k) summary for the IBS Implant System II, a dental implant device. It primarily focuses on demonstrating substantial equivalence to predicate devices based on design, materials, indications for use, and non-clinical testing. It does not describe an acceptance criteria study with the requested details (number of experts, sample size for test/training set, ground truth methods, MRMC study, or standalone performance).

    Instead, the document details a comparison of the new device to existing devices (predicates) regarding various technical specifications and non-clinical test results.

    Since the document does not contain the information for acceptance criteria and a study proving device meets it as requested, I cannot complete the table or answer the specific questions.

    No information for the requested table and study details is available in the provided text. The document focuses on demonstrating substantial equivalence through comparison with predicate devices and non-clinical tests (biocompatibility, endotoxin, sterilization validation, shelf-life validation, fatigue). These are not presented as a study with specific acceptance criteria that the device performed against in the way an AI/software performance study would be evaluated.

    Therefore, the requested table of acceptance criteria and reported device performance cannot be generated. Similarly, specific details about sample size (training/test), data provenance, expert involvement, adjudication, MRMC, or standalone performance are not provided as these types of studies are typically associated with performance claims for diagnostic or prognostic devices, not for substantial equivalence of an endosseous dental implant based on physical and material properties.