    Intended Use

    HydroPICC Single Lumen: Indicated for short-or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to the administration of fluids, medications, and nutrients; the sampling of blood; central venous pressure monitoring; and power injection of contrast media. Rated for maximum power injection flow rate of 5.0 ml/s

    HydroPICC Dual Lumen: Indicated for short- or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to the administration of fluids, medications, and nutrients; the sampling of blood; central venous pressure monitoring; and power injection of contrast media. Rated for a maximum power injection flow rate of 5.0 ml/s

    HydroMID Single Lumen: Indicated for short term access(< 30 days) to the peripheral venous access system for intravenous therapy, including but not limited to, the administration of fluids, medications, and the sampling of blood and blood products. Maximum Power Injection Flow Rate: 4Fr Single Lumen, 20cm: 6 ml/sec

    Device Description

    The HydroMID catheters are a family of midline catheters comprised of a radiopaque, hydrophilic catheter with a suture wing, Luer lock hub, and extension tubing made from materials commonly used in the manufacture of catheters. Catheters are provided packaged in kit configurations with the appropriate accessories for placement in the respective clinical environments. The maximum power injection flow rate for the lumen is indicated on the extension tube clamp. HydroMID has been shown to be effective in reducing thrombus accumulation and thrombotic occlusions. Reduction of thrombus accumulation and thrombotic occlusions were evaluated using an in vitro model. Pre-clinical in vitro evaluations do not necessarily predict clinical performance with respect to thrombus formation.

    The HydroPICC catheters are a family of peripherally inserted central catheter (PICC) comprised of a radiopaque hydrophilic catheter material with a suture wing, Luer lock hub, and extension tubing made from materials commonly used in the manufacture of catheters. Catheters are provided packaged in kit configurations with the appropriate accessories for placement in the respective clinical environments. The maximum power injection flow rate for the lumen is indicated on the extension tube clamp. HydroPICC has been shown to be effective in reducing thrombus accumulation and thrombotic occlusions. Reduction of thrombus accumulation and thrombotic occlusions were evaluated using an in vitro model. Pre-clinical in vitro evaluations do not necessarily predict clinical performance with respect to thrombus formation.

    The purpose of this 510(k) is to add an additional statement to the indications for use.

    AI/ML Overview

    The provided FDA 510(k) clearance letter (K243458) details the HydroPICC and HydroMID catheters. The focus of this submission is to add a statement to the indications for use regarding the anti-thrombogenic properties of the MIMIX® Technology incorporated into these catheters. The acceptance criteria and supporting studies are described below:

    Acceptance Criteria and Reported Device Performance

    The core acceptance criterion for this 510(k) submission is to demonstrate that the expanded Indications for Use statement, specifically concerning the anti-thrombogenic properties of the MIMIX® Technology, does not introduce new questions of safety or effectiveness and is supported by non-clinical data. The reported device performance is based on in vitro studies.

    Acceptance CriterionReported Device Performance (as per In vitro studies)
    Reduction of thrombus formation on device surfaces (external and internal fluid pathways)"both the external surfaces and internal fluid pathways of the catheter effectively reduce thrombus accumulation and thrombotic occlusions." This reduction is achieved due to the catheter's steric barrier.
    Device safety and efficacy with respect to predicate devices (for the anti-thrombogenic claims)All listed in vitro tests (Assessment of PICC Catheter Thrombosis in an in vitro Model, Thrombosis Accumulation Report, In Vitro Thrombosis Study with Saline Conditioning, In Vitro Assessment of PICC Thrombus, In Vitro Assessment of Catheter Thrombotic Occlusion, Blood Loop Analysis of HydroPICC Against Competitors, Exhaustive Recovery Assessment of Catheter Thrombosis in an In-Vitro Blood Flow Model) "passed".

    Note: The FDA letter explicitly states: "The clinical impact has not been evaluated in human clinical trials. This device is not intended for the treatment of existing vein thrombosis." This indicates that the acceptance criteria are based solely on in vitro performance for the anti-thrombogenic claims, not on in vivo clinical outcomes.

    Study Details for Acceptance Criteria

    The provided document does not fully delineate separate "test sets" for the in vitro studies in the way one might for an AI/algorithm-based device. Instead, the "studies" themselves are the performance evaluations for the stated acceptance criteria.

    1. Sample size used for the test set and the data provenance:

      • The document lists several in vitro tests.
      • Sample Size: The specific number of devices or experimental replicates used in each in vitro test is not explicitly stated in the provided text.
      • Data Provenance: All studies are indicated as "in vitro" (meaning conducted in a test tube, culture dish, or other controlled environment outside of a living organism). The country of origin is not specified but is presumably where Access Vascular Inc. conducts its research and development or contracts with testing facilities.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • For in vitro studies of this nature (evaluating properties like thrombus accumulation), the "ground truth" is typically established by the experimental setup, validated measurement techniques, and potentially statistical analysis.
      • No specific number of experts or their qualifications are mentioned as having established "ground truth" for these in vitro tests in the context of expert consensus, as might be the case for image-based diagnostic devices. The results are based on objective physical or biological measurements.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not applicable. Adjudication methods like "2+1" or "3+1" are typically used in clinical studies or studies involving subjective expert review (e.g., radiology image interpretation) to resolve discrepancies in independent assessments. As these are in vitro physical/biological tests, such adjudication methods are not relevant or mentioned.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not applicable. This device is a medical catheter, not an AI software/algorithm requiring human reader interpretation. No MRMC study was conducted or is relevant for this device.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Not applicable. This device is a physical medical device (catheter) with a material technology, not an AI algorithm. Therefore, "standalone" algorithm performance is not a relevant concept for this submission.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The ground truth for the in vitro studies described (e.g., "Assessment of PICC Catheter Thrombosis," "Thrombosis Accumulation Report") is based on objective measurements of physical and biological phenomena (e.g., quantification of thrombus accumulation, pressure required to remove occlusions) within controlled laboratory environments. This is a form of empirical scientific measurement, rather than expert consensus, pathology, or outcomes data.
    7. The sample size for the training set:

      • Not applicable. There is no mention of a "training set" because this is a physical medical device, not an AI/machine learning model that requires training data.
    8. How the ground truth for the training set was established:

      • Not applicable. As no training set is mentioned, the method for establishing ground truth for it is also not applicable.
    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Indicated for short-or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to the administration of fluids, medications, and nutrients; the sampling of blood; central venous pressure monitoring; and power injection of contrast media. Rated for maximum power injection flow rate of 5.0ml/s

    Device Description

    The HydroPICC 4F Single Lumen Catheter is comprised of a radiopaque hydrophilic catheter material with a suture wing, Luer lock hub, and extension tubing made from materials commonly used in the manufacture of catheters are provided packaged in kit configurations with the appropriate accessories for placement in the respective clinical environments. The maximum power injection flow rate for the lumen is indicated on the clamp. HydroPCC has been shown to be effective in reducing thrombus accumulation and thrombotic occlusions f thrombus accumulation and thrombotic occusions were evaluated using in vitro and in vivo models. Pre-clinical in vivo evaluations do not necessarily predict clinical performance with respect to thrombus formation.

    AI/ML Overview

    The provided text is a 510(k) summary for the HydroPICC 4Fr Single Lumen Marked catheter, where the purpose of the submission is to modify the power injection ratings in the indications for use. As such, it focuses on demonstrating substantial equivalence to a predicate device rather than providing details of a study against specific acceptance criteria for a new device's performance.

    Therefore, the document does not contain the information requested regarding acceptance criteria, a specific study proving device meets acceptance criteria, sample sizes, data provenance, expert qualifications, adjudication methods, MRMC studies, standalone performance, or ground truth details.

    The document discusses "comprehensive data demonstrating the safety and efficacy" and "rigorous testing protocols" including "in vitro" data to support the modified power injection rating, and states that "the form, fit, and function of these devices have not changed from their previous clearances." However, it does not provide specific details of these tests or their results in a measurable, quantifiable way that would allow for the completion of the requested table and study breakdown.

    Ask a Question

    Ask a specific question about this device

    K Number
    K213550

    Validate with FDA (Live)

    Date Cleared
    2022-05-04

    (177 days)

    Product Code
    Regulation Number
    880.5970
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Indicated for short-or long-term peripheral access to the central venous therapy, including but not limited to the administration of fluids, medications, and nutrients; the sampling of blood; central venous pressure monitoring; and power injection of contrast media.

    Rated for maximum power injection flow rate of 3.5ml/s

    Device Description

    The HydroPICC peripherally inserted central catheter (PICC) is a 5 French, dual lumen catheter comprised of a radiopaque hydrophilic catheter material with a suture wing, Luer lock hubs, and extension tubes made from materials commonly used in the manufacture of catheters are provided packaged in kit configurations with the appropriate accessories for placement in the appropriate clinical environments. The maximum power injection flow rate for each lumen is indicated on each extension tube clamp.

    HydroPICC has been shown to be effective in reducing thrombus accumulation. Reduction of thrombus accumulation was evaluated using in vivo models. Pre-clinical in vitro and in vivo evaluations do not necessarily predict clinical performance with respect to thrombus formation.

    AI/ML Overview

    This document is a 510(k) summary for the HydroPICC 5F Dual Lumen Catheter. It indicates that the device has undergone performance testing to demonstrate substantial equivalence to a predicate device. However, the document does not contain details about acceptance criteria, reported device performance, sample sizes for test or training sets, data provenance, number or qualifications of experts, adjudication methods, multi-reader multi-case studies, or standalone algorithm performance.

    The provided text focuses on the physical and technological characteristics of the medical device and its predicate, along with the regulatory and biocompatibility testing performed to establish substantial equivalence. It does not describe an AI/ML device or its performance study in the way the request specifies.

    Therefore, many of the requested fields cannot be filled from the provided text.

    Here's the breakdown of what can be extracted:

    1. A table of acceptance criteria and the reported device performance: Not available in the provided text. The document refers to conformance with recognized standards but does not list specific acceptance criteria or performance metrics for the device in a study context.

    2. Sample sized used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective): Not available in the provided text.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience): Not available in the provided text.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not available in the provided text.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not available in the provided text. This device is not an AI/ML product.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not available in the provided text. This device is not an AI/ML product.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc): Not available in the provided text. For the type of device (intravascular catheter), "ground truth" would typically refer to physical and mechanical properties, not diagnostic categories.

    8. The sample size for the training set: Not available in the provided text. This device is not an AI/ML product.

    9. How the ground truth for the training set was established: Not available in the provided text. This device is not an AI/ML product.

    What can be extracted from the document regarding "performance testing" related to the device:

    The document lists various performance tests conducted to demonstrate conformance with FDA-recognized standards for intravascular catheters. These tests are primarily focused on the physical, mechanical, and safety aspects of the catheter, rather than diagnostic or AI-driven performance.

    Acceptance Criteria (Implied by Standard Conformance)Reported Device Performance (Implied by "Conformance")
    Conformance to General Requirements for Intravascular Catheters (EN ISO 10555-1:2013)"tested and demonstrated to be in conformance" with this standard, indicating it meets general safety and performance requirements for intravascular catheters (e.g., freedom from defects, integrity of fluid pathways, connection security, etc.).
    Conformance for Central Venous Catheters (EN ISO 10555-3:2013)"tested and demonstrated to be in conformance" with this standard, indicating it meets specific requirements for central venous catheters.
    Kinking resistance for Single Lumen Catheters and Medical Tubing (EN ISO 13868:2014)"tested and demonstrated to be in conformance" with this standard, indicating it resists kinking to an acceptable degree.
    MRI Safety: Magnetically Induced Displacement Force (ASTM F2052:2015)"tested and demonstrated to be in conformance", confirming acceptable displacement force in an MRI environment. The device is stated to be "MRI Conditional."
    MRI Safety: Radio Frequency Induced Heating (ASTM F2182: 2011)"tested and demonstrated to be in conformance", confirming acceptable RF-induced heating in an MRI environment. The device is stated to be "MRI Conditional."
    MRI Safety: Magnetically Induced Torque (ASTM F2213: 2017)"tested and demonstrated to be in conformance", confirming acceptable magnetically induced torque in an MRI environment. The device is stated to be "MRI Conditional."
    MRI Safety: Evaluation of MR Image Artifacts (ASTM F2119: 2013)"tested and demonstrated to be in conformance", confirming acceptable levels of image artifacts from the device in an MRI environment. The device is stated to be "MRI Conditional."
    Biocompatibility (ISO 10993-1, FDA Guidance "Use of International Standard ISO 10993-1")"evaluation of the HydroPICC in accordance with ISO 10993-1 and FDA Guidance," covering direct blood path (catheter body) for >30 days and external contact (Luer lock hub, polyurethane extension tubing, suture wing) for >30 days, implying acceptable biocompatibility. The text also states, "HydroPICC has been shown to be effective in reducing thrombus accumulation. Reduction of thrombus accumulation was evaluated using in vivo models." (Pre-clinical, in vitro, and in vivo evaluations do not necessarily predict clinical performance with respect to thrombus formation - this is an important caveat).
    Sterility Assurance Level (ISO 11135-1, AAMI TIR28:2016)"Sterility Assurance Level (SAL) of 10-6 via a validated overkill Ethylene Oxide (EO) method." This meets requirements of ISO 11135-1 as determined through AAMI TIR28:2016.
    Package Integrity (ISTA 3A:2016, ASTM D4169)Package integrity testing, after environmental conditioning and simulated transportation, "complies to ISTA 3A:2016... and ASTM D4169," demonstrating protection of product and sterility maintenance.
    Labeling Conformance (ISO 15223-1:2016)"All labeling was evaluated according to ISO 15223-1:2016... and the subject device was determined to have the appropriate labeling."
    Sterile Barrier Packaging Testing (ASTM F88/F88-15, ASTM F20965)"complies to Seal strength ASTM F88/F88-15 and Detecting Gross Leaks ASTM F20965."
    Shelf Life Validation (ASTM F1980-16)"Shelf life 13 months is validated using the FDA recognized standard ASTM F1980-16 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices."

    Study Details (as much as can be inferred/is stated):

    • Sample Size, Provenance, Experts, Adjudication, MRMC, Standalone, Ground Truth (for diagnostic AI/ML): Not applicable or not provided. The study mentioned for "reducing thrombus accumulation" used "in vivo models," but specific details like sample size, provenance, control groups, or exact endpoints are not given.
    • Training Set Sample Size/Ground Truth: Not applicable for this type of medical device submission.
    Ask a Question

    Ask a specific question about this device

    K Number
    K193015

    Validate with FDA (Live)

    Date Cleared
    2020-02-20

    (114 days)

    Product Code
    Regulation Number
    880.5970
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    HydroPICC is indicated for short- or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to; the administration of fluids, medications, and nutrients; the sampling of blood; central venous pressure monitoring; and power injection of contrast media.

    HydroPICC is rated for a Maximum Power Injection Flow Rate of 3.5mL/sec.

    Device Description

    The HydroPICC is a peripherally inserted central catheter (PICC) intended for short or long-term use. The HydroPICC is a 4Fr, 55cm long catheter suitable for periodic access to the superior vena cava (SVC) while maintaining a reduced level of thrombus accumulation. HydroPICC is supplied in a kit with the necessary accessories to insert and maintain the catheter.

    AI/ML Overview

    This document is a 510(k) K193015 Premarket Notification for the HydroPICC (Model 142) intravascular catheter, asserting its substantial equivalence to a previously cleared predicate device, HydroPICC (Model 141, K172885).

    The submission focuses heavily on demonstrating equivalence and does NOT include a study involving AI assistance for human readers, nor a standalone AI algorithm performance study. Therefore, most of the requested information regarding AI study design, ground truth establishment for AI, expert involvement, and MRMC studies is not present in this document.

    The document describes performance testing for a medical device (intravascular catheter), not an AI algorithm. The acceptance criteria and performance data described relate to physical and mechanical properties of the catheter, as distinct from the performance metrics (e.g., sensitivity, specificity, AUC) typically associated with AI/ML medical devices.

    Here's the breakdown of the information that CAN be extracted, followed by a clear statement of what is NOT in the document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document states that "The subject devices met all predetermined acceptance criteria derived from the applicable guidance documents, standards and in-house protocols," but it does not provide a specific table with numerical acceptance criteria and reported performance values. Instead, it lists the types of tests performed and the standards followed.

    Test CategorySpecific Tests PerformedAcceptance Criteria (Not explicitly quantified in text, stated as "met all predetermined criteria")Reported Device Performance (Not explicitly quantified in text, stated as "met all predetermined criteria")
    Mechanical & Performance TestingPower Injection Flow RateDerived from standards/internal protocolsSuccessful/Compliant (implied)
    Static Burst StrengthDerived from standards/internal protocolsSuccessful/Compliant (implied)
    Multiple Power InjectionsDerived from standards/internal protocolsSuccessful/Compliant (implied)
    Catheter LengthDerived from standards/internal protocolsSuccessful/Compliant (implied)
    Dimensional Verification (including ID, OD, Length)Derived from standards/internal protocolsSuccessful/Compliant (implied)
    Catheter Kink/Flex ResistanceDerived from standards/internal protocolsSuccessful/Compliant (implied)
    Tensile Testing (of Catheter and Assembly)Derived from standards/internal protocolsSuccessful/Compliant (implied)
    Particulate testingDerived from standards/internal protocolsSuccessful/Compliant (implied)
    Shelf life testingDerived from standards/internal protocolsSuccessful/Compliant (implied)
    Sterility testingDerived from standards/internal protocolsSuccessful/Compliant (implied)
    Packaging distribution testingDerived from standards/internal protocolsSuccessful/Compliant (implied)
    MR Compatibility testingDerived from standards/internal protocolsSuccessful/Compliant (implies HydroPICC-142 is labeled with MRI Safety Information, unlike predicate)
    Evaluation of Humectant change on Thrombus accumulation compared to predicate (internal specifications)Derived from standards/internal protocolsSuccessful/Compliant (implied)
    Biocompatibility TestingRisk assessment evaluation that the proposed modifications did not impact biocompatibilityDerived from standards/internal protocolsSuccessful/Compliant (implied)

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the sample size used for the test set for each individual test. It merely states that "The subject devices met all predetermined acceptance criteria." The provenance of the data (e.g., country of origin, retrospective/prospective) is not mentioned for this engineering/performance testing.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    Not applicable. This document describes the testing of a medical device (catheter), not the performance of an AI algorithm based on expert-labeled data.

    4. Adjudication Method for the Test Set

    Not applicable. This is for medical device performance (physical, mechanical, safety), not for AI algorithm evaluation requiring expert consensus/adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study was done because this submission is not for an AI/ML medical device and does not involve human readers interpreting images with or without AI assistance.

    6. Standalone Algorithm Performance

    No standalone algorithm performance study was done relevant to AI/ML, as this is not an AI/ML device.

    7. Type of Ground Truth Used

    The "ground truth" for this device's performance testing is established by engineering standards and specifications (e.g., ISO, ASTM standards, FDA guidance documents, and internal protocols) for physical and mechanical properties, not expert consensus on medical images, pathology, or outcomes data in the context of an AI study.

    8. Sample Size for the Training Set

    Not applicable. This document describes testing for a manufactured medical device, not an AI/ML algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable for the same reason as point 8.


    Summary of what is NOT in the document regarding acceptance criteria and studies for an AI/ML device:

    The provided document is a 510(k) premarket notification for a physical medical device (intravascular catheter). It details the substantial equivalence to a predicate device and describes the various engineering and performance tests conducted to ensure the device meets safety and performance standards. It does not contain any information about an AI/ML device, its acceptance criteria, expert consensus, ground truth establishment, or any studies related to its performance (standalone or human-in-the-loop). Therefore, the specific requirements of the prompt regarding AI are not addressed by this document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K172885

    Validate with FDA (Live)

    Device Name
    HydroPICC
    Date Cleared
    2018-02-20

    (152 days)

    Product Code
    Regulation Number
    880.5970
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    HydroPICC is indicated for short- or long-term peripheral access to the central venous system for intrapy, including but not limited to; the administration of fluids, medications, and nutrients; the sampling of blood; central venous pressure monitoring; and power injection of contrast.

    HydroPICC is rated for a Maximum Power Injection Flow Rate of 3.5ml/sec.

    Device Description

    The HydroPICC is a peripherally inserted central catheter (PICC) intended for short or long-term use. The HydroPICC is a 4Fr, 55cm long catheter suitable for periodic access to the superior vena cava (SVC). HydroPICC has been shown to be effective in reducing thrombosis accumulation. Reduction of thrombosis accumulation was evaluated using in vitro and in vivo models. Preclinical in vitro and in vivo evaluations do not necessarily predict clinical performance with respect to thrombosis formation.. HydroPICC is supplied in a kit with the necessary accessories to insert and maintain the catheter.

    AI/ML Overview

    The provided text describes the 510(k) premarket notification for the HydroPICC device, focusing on its substantial equivalence to predicate devices rather than providing detailed acceptance criteria and study results in the format requested for an AI/algorithm-based device.

    The HydroPICC is a medical device (a peripherally inserted central catheter), and the performance data presented relates to its physical and biological properties. This document does not describe an AI or algorithm-driven device. Therefore, it does not contain the information required to answer the prompt regarding acceptance criteria and studies for an AI/algorithm.

    Specifically, the document focuses on:

    • Biocompatibility Testing: Cytotoxicity, intracutaneous reactivity, sensitization, intramuscular implant, pyrogenicity, chemical extractables, in vivo thrombogenicity.
    • Mechanical Testing: Power Injection Flow Rate, Static Burst Strength, Lifecycle Power Injections, Gravity Flow Rate, Catheter Length, Priming Volume, Dimensional Verification, Catheter Kink/Flex Resistance, Alcohol Compatibility, Catheter Marking & Identification/Radio Detectability Testing, Tensile Testing, Catheter Compatibility with Medications, Catheter Thrombus Accumulation Evaluation Using in Vitro Blood Loop, Catheter Collapse Resistance, Central Venous Pressure Monitoring, Pyrogens testing.
    • Sterilization: Validated overkill Ethylene Oxide (EO) method.

    These are standard tests for a physical medical device and do not involve AI/algorithm performance metrics such as sensitivity, specificity, AUC, or the involvement of expert readers for ground truth establishment.

    Therefore, I cannot fulfill the request as the provided text does not contain information about an AI/algorithm-driven device or its associated acceptance criteria and study proving its performance.