Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K200909

    Validate with FDA (Live)

    Date Cleared
    2020-06-12

    (67 days)

    Product Code
    Regulation Number
    864.5620
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Hemo Control is intended to be used for the quantitative determination of hemoglobin (Hb) concentrations in human blood.

    The Hemo Control Hemoglobin Microcuvettes are intended to be used with the Hemo Control photometer for the quantitative determination of hemoglobin (Hb) concentrations in human blood.

    For in-vitro diagnostic use only.

    Device Description

    Hemo Control consists of the Hemo Control photometer / analyzer and the Hemo Control Hemoglobin Microcuvettes, its accessories and consumables (i.e. Control Solution Hb-con).
    The Hemo Control photometer / analyzer is a semi-automated, spectrophotometric instrument, which provides instant quantitative total hemoglobin results.
    Using the reagent filled microcuvette a small amount of arterial, venous or capillary blood is taken up by capillary action. The filled microcuvette is inserted into the Hemo Control photometer. The color produced by chemical reaction in the microcuvette is measured and the Hb value is displayed.
    The measurement accuracy of the Hemo Control Hemoglobin Measurement System can be verified by use of Hb-con control solution, a quality control material with pre-determined hemoglobin concentration.
    As a second quality control measurement, the control cuvette as a physical standard is used for a comfortable and cheap check of the device.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided FDA 510(k) summary for the Hemo Control device:

    Device: Hemo Control (automated hemoglobin system)

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria/MetricReported Device Performance
    PrecisionWithin Run (CV) - Low (107 g/L)0.8%
    Within Run (CV) - Normal (129 g/L)0.6%
    Within Run (CV) - High (173 g/L)0.6%
    Total (CV) - Low (107 g/L)1.0%
    Total (CV) - Normal (129 g/L)1.0%
    Total (CV) - High (173 g/L)1.1%
    Single Observation 20 days (CV) - Low (107 g/L)0.9%
    Single Observation 20 days (CV) - Normal (129 g/L)0.8%
    Single Observation 20 days (CV) - High (173 g/L)1.0%
    Linearity/Assay RangeReportable range0 - 25.6 g/dL
    Method ComparisonComparison to NCCLS Reference Methody=1.0064x + 0.0234, r=0.0076, n=174
    Comparison to HemoCue hemoglobin measurement systemy=1.0005x - 0.2334, r=0.9962, n=286
    Comparison of Hemo Control Cuvettes in HemoCuey=0.9855x + 0.139, r=0.998, n=286
    Matrix ComparisonCapillary samples, 4 sitesy=0.96x + 0.3742, r=0.8256, n=275
    Arterial samples, 1 sitey=0.9868x - 0.0285, r=0.998, n=10

    The document does not explicitly state "acceptance criteria" but rather presents performance characteristics of the device. The reported performance is the demonstration that the device meets the implicit acceptance criteria for substantial equivalence to its predicate device. For example, the low %CV values for precision demonstrate acceptable reproducibility. The regression equations with r-values close to 1 and slopes close to 1 for method comparison studies demonstrate agreement with reference methods and predicate devices.

    2. Sample Sizes Used for the Test Set and the Data Provenance

    • Precision: The document provides Coefficient of Variation (CV) values for "Within Run," "Total," and "Single Observation 20 days" at low (107 g/L), normal (129 g/L), and high (173 g/L) hemoglobin concentrations. The specific number of samples or measurements for each of these precision studies is not explicitly stated in this summary, but typically, precision studies involve repeated measurements of control materials or patient samples.
    • Method Comparison (NCCLS Reference Method): n = 174 samples
    • Method Comparison (HemoCue hemoglobin measurement system): n = 286 samples
    • Method Comparison (Hemo Control Cuvettes in HemoCue): n = 286 samples
    • Matrix Comparison (Capillary samples): n = 275 samples
    • Matrix Comparison (Arterial samples): n = 10 samples

    Data Provenance: The document does not explicitly state the country of origin or whether the data was retrospective or prospective. Given that this is a 510(k) submission from a German company (EKF-diagnostic GmbH) and involves comparison to established reference methods and other commercially available devices, it is highly likely these were prospective studies conducted in a clinical or laboratory setting.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This device is an automated hemoglobin system that measures a quantitative value (hemoglobin concentration). It does not rely on expert interpretation for its output. Therefore, the concept of "experts establishing ground truth" in the way it applies to image interpretation or diagnostic classification (e.g., radiologists) is not relevant here.

    The "ground truth" for method comparison studies is established by:

    • NCCLS Reference Method: This refers to a standardized laboratory method (likely cyanmethemoglobin method, which is a gold standard for hemoglobin measurement). The reference method itself is the "ground truth."
    • Predicate Device (HemoCue hemoglobin measurement system): This is another commercially available, cleared device that serves as a comparison standard.

    4. Adjudication Method for the Test Set

    Not applicable. Since the device produces a quantitative numerical output for hemoglobin concentration, there is no need for an adjudication method among experts. The "ground truth" is determined by the reference method or comparison device.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic devices where human readers interpret output (e.g., medical images) to assess the impact of AI assistance on their performance. The Hemo Control is an automated quantitative device.

    6. If a Standalone (i.e., Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Yes, the analytical performance and method comparison studies described are inherently standalone performance studies. The Hemo Control device (photometer and microcuvettes) itself performs the measurement and provides the hemoglobin value. There is no "human-in-the-loop" interaction for interpreting the result, only for operating the device, taking the sample, and understanding the output.

    7. The Type of Ground Truth Used

    The ground truth used for performance evaluation included:

    • NCCLS Reference Method: A highly standardized, accepted laboratory method for hemoglobin measurement.
    • Predicate Device Performance: Comparison against the performance of another legally marketed device (HemoCue hemoglobin measurement system).

    8. The Sample Size for the Training Set

    The document does not explicitly mention a "training set" in the context of machine learning. This device operates based on spectrophotometric principles and chemical reactions, not on complex machine learning algorithms that require large training data sets in the typical sense. The fundamental physical and chemical principles are well-established.

    The studies described (precision, linearity, method comparison, matrix comparison) are typically considered verification and validation studies performed after the device design is largely finalized, not for "training" an algorithm.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as a "training set" in the machine learning sense is not mentioned or implied for this device's operation. The device's calibration and performance are established against physical standards, reference methods, and quality control materials, which form its inherent "ground truth." Specifically, the device is calibrated against the NCCLS reference method, and its accuracy can be verified using Hb-con control solutions (quality control material with pre-determined hemoglobin concentration) and a control cuvette as a physical standard.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1