510(k) Data Aggregation

    K Number
    K242651

    Validate with FDA (Live)

    Device Name
    GM85
    Date Cleared
    2024-10-01

    (27 days)

    Product Code
    Regulation Number
    892.1720
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The GM85 Digital Mobile X-ray Imaging System is intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications.

    Device Description

    The GM85 Digital Mobile X-ray Imaging System is used to capture images by transmitting X-ray to a patient's body. The X-ray passing through a patient's body is sent to the detector and then converted into electrical signals. These signals go through the process of amplification and digital data conversion in the signal process on the S-Station, which is the Operation Software (OS) of Samsung Digital Diagnostic X-ray System, and save in DICOM file, a standard for medical imaging. The captured images are tuned up by an Image Post-processing Engine (IPE) which is exclusively installed in S-Station, and send to the Picture Archiving & Communication System (PACS) sever for reading images.

    The GM85 Digital Mobile X-ray imaging System was previously cleared with K222353, and through this premarket notification, we would like to add more configurations in the previously cleared GM85 as a detector, accessories are newly added and software is updated for user convenience.

    The new detector added in the proposed device is designed to achieve a higher IP rating of Dust and Water and reduce weight while maintaining durability, functionality and operation like the detector of the predicate device. The new detector and predicate device's detector are both an x-ray conversion device using an amorphous silicon flat panel and absorb incident x-rays, converts it to a digital signal, and then transmits it to the Samsung Digital X-ray System like that of the predicate device.

    AI/ML Overview

    The provided text is a 510(k) Premarket Notification for the GM85 Digital Mobile X-ray Imaging System. It describes changes made to a previously cleared device (K222353) and argues for substantial equivalence.

    Based on the provided text, the device is the GM85 Digital Mobile X-ray Imaging System. It is an X-ray imaging system, and the study focuses on the performance of a new detector (F4343-AW) and other accessories and software updates compared to the predicate device, also named GM85.

    Here's an analysis of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document does not explicitly state "acceptance criteria" in a quantitative format for specific imaging metrics. Instead, it focuses on demonstrating that the proposed device, with its new detector, accessories, and software, is substantially equivalent to the predicate device (GM85, K222353). The performance is assessed by comparing technical specifications and qualitative evaluation by experts.

    The key performance characteristics and comparisons are as follows:

    AttributeAcceptance Criteria (Implied by Substantial Equivalence Goal)Reported Device Performance (Proposed Device)Comparison to Predicate (GM85, K222353)
    Detector CharacteristicsEquivalent or improved
    Detector TypeSame as predicate (CsI Indirect)CsI IndirectSame
    Detector AreaSame as predicate (17"X17")17"X17" (425mmX425mm)Same
    Number of pixelsSame as predicate (3036X3040)3036X3040Same
    Pixel Pitch (um)Same as predicate (140)140Same
    High Contrast Limiting Resolution (LP/mm)Same as predicate (3.57)3.57Same
    CommunicationSame as predicate (Wired / Wireless)Wired / WirelessSame
    Dust/Water-resistanceEquivalent or improved (IP54 for predicate)IP57Difference (Improved)
    Max. load capacitySame as predicate (400 kg/200 kg)400 kg/200 kgSame
    DQE (0lp/mm, Typical)Same as predicate (76%)76%Same
    MTF (0.5lp/mm, Typical)Same as predicate (86%)86%Same
    Weight (w/o Battery Set)Equivalent or improved (Approx. 3.4 kg for predicate)Approx. 2.5 kgDifference (Improved/Lighter)
    Image Quality (Phantom)Equivalent to predicate"Equivalent to the predicate devices""No significant difference in the average score of image quality evaluation"
    Safety and EffectivenessNo adverse impactVerified by standards and testing"does not contribute any adverse impact"

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: The document refers to "Anthropomorphic phantom images" but does not specify the number of phantom images used for evaluation. It also notes that clinical data was not required.
    • Data Provenance: The study is non-clinical. The "Anthropomorphic phantom images" would have been generated in a controlled testing environment, likely at the manufacturer's facility. The country of origin of the data is implicitly South Korea, where SAMSUNG ELECTRONICS Co., Ltd. is located. The study is a prospective evaluation of the new detector and modified system against the predicate.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • Number of Experts: The document states that phantom images "were evaluated by professional radiologists." It does not specify the number of radiologists who participated in this evaluation.
    • Qualifications of Experts: The experts are described as "professional radiologists." No further details on their experience level (e.g., years of experience, subspecialty) are provided.

    4. Adjudication Method for the Test Set

    The document states that phantom images "were evaluated by professional radiologists and found to be equivalent to the predicate devices" and that there was "no significant difference in the average score of image quality evaluation." This suggests a comparative scoring or assessment. However, the specific adjudication method (e.g., consensus, majority vote, independent reads with statistical comparison) is not described.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not explicitly done, nor was there a "human readers improve with AI vs without AI assistance" component. The study for substantial equivalence focused on comparing the image quality of the proposed device (with new detector) against the predicate device using phantom images, evaluated by radiologists. The purpose was to show equivalence of the device's output, not the improvement of human readers with AI assistance.

    6. Standalone Performance Study (Algorithm Only)

    The provided text describes the device as a "Digital Mobile X-ray Imaging System," which includes hardware (X-ray generator, detector) and software for image processing (IPE, S-Station). The evaluation primarily focuses on the entire system's ability to generate radiographic images with equivalent quality to the predicate, particularly with the new detector.

    While software features are mentioned (S-Share, S-Enhance, SimGrid, PEM, QAP, Bone Suppression, Remote View, Mirror View, RFID, Value-up Package), and software was updated for user convenience, the study does not report a standalone algorithm-only performance (without human-in-the-loop performance) in terms of diagnostic accuracy or reader improvement for specific diagnostic tasks. The "phantom image evaluation" evaluates the quality of the images produced by the overall system, not an AI algorithm's diagnostic output.

    7. Type of Ground Truth Used

    The ground truth for the phantom image evaluation was established based on expert consensus/evaluation of image quality metrics. The "professional radiologists" evaluated the anthropomorphic phantom images. This is not pathology, nor outcomes data.

    8. Sample Size for the Training Set

    The document does not provide information on the sample size for the training set. Training data would typically be applicable if this were an AI/CADe device with a specific machine learning model for diagnostic tasks, which is not the primary focus of this 510(k) for a basic X-ray imaging system with incremental changes. The "Image Post-processing Engine (IPE)" and features like "Bone Suppression" would have been developed using some form of training data previously, but details are absent here for this particular submission.

    9. How the Ground Truth for the Training Set Was Established

    As no training set sample size is provided, the method for establishing ground truth for a training set is not mentioned in this document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K222353

    Validate with FDA (Live)

    Device Name
    GM85
    Date Cleared
    2022-09-29

    (56 days)

    Product Code
    Regulation Number
    892.1720
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The GM85 Digital Mobile X-ray imaging System is intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications.

    Device Description

    The GM85 Digital Mobile X-ray Imaging System is used to capture images by transmitting X-ray to a patient's body. The X-ray passing through a patient's body is sent to the detector and then converted into electrical signals. These signals go through the process of amplification and digital data conversion in the signal process on the S-Station, which is the Operation Software (OS) of Samsung Digital Diagnostic X-ray System, and save in DICOM file, a standard for medical imaging. The captured images are tuned up by an Image Post-processing Engine (IPE) which is exclusively installed in S-Station, and send to the Picture Archiving & Communication System (PACS) sever for reading images.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the GM85 Digital Mobile X-ray Imaging System. The primary purpose of this notification is to add a new detector configuration to an already cleared device (K220175). As such, the study described focuses on demonstrating the substantial equivalence of the new detector and its integration, rather than establishing primary performance metrics from scratch for a completely new device.

    Here's a breakdown of the acceptance criteria and the study conducted based on the provided information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implicitly tied to demonstrating equivalence to the predicate device's existing performance, particularly for the new detector. The non-clinical and clinical (phantom image evaluation) data serve to show that the new detector maintains similar performance characteristics.

    Acceptance Criteria CategorySpecific Metric (Implicit/Explicit)Predicate Device PerformanceProposed Device (New Detector) PerformanceDiscussion/Acceptance
    DurabilityN/A (General statement)Assumed to be metRetains same durabilityMet
    FunctionalityN/A (General statement)Assumed to be metRetains same functionalityMet
    OperationN/A (General statement)Assumed to be metRetains same operationMet
    Detector TypeDetector TypeCsl, IndirectCsl, IndirectSame (Met)
    Detector AreaDetector Area14"X17" (345mmX425mm)14"X17" (345mmX425mm)Same (Met)
    Number of PixelsNumber of Pixels2466X30402466X3040Same (Met)
    Pixel PitchPixel Pitch140 um140 umSame (Met)
    High Contrast Limiting ResolutionHigh Contrast Limiting Resolution3.57 LP/mm3.57 LP/mmSame (Met)
    CommunicationCommunicationWired / WirelessWired / WirelessSame (Met)
    Dust/Water-resistanceDust/Water-resistance RatingIP54IP54Same (Met)
    Max. Load CapacityMax. Load Capacity400 kg/200 kg400 kg/200 kgSame (Met)
    DQEDQE (0lp/mm, Typical)76%76%Same (Met)
    MTFMTF (0.5lp/mm, Typical)86%86%Same (Met)
    WeightWeight (w/o Battery Set)Approx. 2.76 kgApprox. 2 kgDifference acknowledged as not impacting safety/effectiveness
    Image Quality EvaluationAverage Score of Image QualityBaseline (predicate)Equivalent to predicate devicesFound to be equivalent (Met)

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: The document mentions "Anthropomorphic phantom images were provided." It does not specify the exact number of phantom images or the number of distinct phantom cases used in the evaluation.
    • Data Provenance: The nature of phantom images means the data is synthetic/simulated clinical data (using phantoms to mimic human anatomy) rather than human patient data. The country of origin for the data is not explicitly stated, but it would have been generated as part of the manufacturer's internal testing as advised by FDA guidance. The study is prospective in the sense that the evaluations were specifically conducted for this submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • Number of Experts: The document states that phantom images "were evaluated by professional radiologists." It does not specify the exact number of radiologists involved.
    • Qualifications of Experts: They are described as "professional radiologists." Specific lengths of experience or subspecialties are not provided.

    4. Adjudication Method for the Test Set

    • The document implies a consensus or comparison approach: "They were evaluated by professional radiologists and found to be equivalent to the predicate devices. There is no significant difference in the average score of image quality evaluation between the proposed device and the predicate device." However, a specific adjudication method (e.g., 2+1, 3+1 decision rule) is not explicitly stated. It seems to be a comparative evaluation aiming for no significant performance degradation.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No MRMC comparative effectiveness study was done. This submission focuses on hardware modification (new detector) to an existing device, not an AI-assisted diagnostic tool. Therefore, the concept of improving human readers with AI assistance is not applicable here.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • This is not an AI/algorithm-only device. The GM85 is an X-ray imaging system. The "standalone" performance here refers to the device's ability to produce images with acceptable quality on its own. Non-clinical data (MTF, DQE measurements) and phantom images contribute to demonstrating this standalone image performance. The evaluation by radiologists of the phantom images is more about confirming the quality of the "output" of the device in a clinical context, rather than evaluating an algorithm's diagnostic capability.

    7. The Type of Ground Truth Used

    • For the phantom image evaluation, the "ground truth" is established by the known characteristics of the anthropomorphic phantoms. The purpose of the radiologists' evaluation is to determine if the images produced by the new detector (compared to the predicate's detector) adequately represent these known phantom structures and maintain diagnostic quality. In essence, the ground truth is the known physical properties of the phantom, and the evaluation confirms the device's ability to render these properties accurately in images.

    8. The Sample Size for the Training Set

    • This submission describes a modification to an existing medical device's hardware (a detector), not an AI or machine learning algorithm. Therefore, there is no "training set" in the context of AI model development that would typically be described here. The device itself is "trained" or designed through engineering and validated through performance testing.

    9. How the Ground Truth for the Training Set Was Established

    • As there is no AI training set as commonly understood in this context, the concept of establishing ground truth for a training set does not apply. The device's design and engineering would be based on established physics principles of X-ray imaging and medical imaging standards.
    Ask a Question

    Ask a specific question about this device

    K Number
    K220175

    Validate with FDA (Live)

    Device Name
    GM85
    Date Cleared
    2022-04-21

    (90 days)

    Product Code
    Regulation Number
    892.1720
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The GM85 Digital Mobile X-ray imaging System is intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications.

    Device Description

    The GM85 Digital Mobile X-ray Imaging System is used to capture images by transmitting X-ray to a patient's body. The X-ray passing through a patient's body is sent to the detector and then converted into electrical signals. These signals go through the process of amplification and digital data conversion in the signal process on the S-Station, which is the Operation Software (OS) of Samsung Digital Diagnostic X-ray System, and save in DICOM file, a standard for medical imaging. The captured images are tuned up by an Image Post-processing Engine (IPE) which is exclusively installed in S-Station, and send to the Picture Archiving & Communication System (PACS) sever for reading images.

    The GM85 Digital Mobile X-ray imaging System was previously cleared with K181626, and through this premarket notification, we would like to add more configurations in the previously cleared GM85 as an optional collapsible column type with a manual collimator, a tube, four detectors, and exposure switches (two wired and one wireless types) are optionally added, and software including the Image Post-processing Engine (IPE) is changed in order to support new hardware and apply new software features.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification for the Samsung GM85 Digital Mobile X-ray Imaging System. It describes the updated device and compares it to a legally marketed predicate device (K181626, also GM85). The document primarily focuses on demonstrating substantial equivalence rather than presenting detailed acceptance criteria and a comprehensive study for de novo device performance.

    However, based on the information provided, here's an attempt to extract and synthesize what is available regarding acceptance criteria and the supporting study, focusing on the changes made to the device:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of acceptance criteria with numerical targets and corresponding device performance for the overall system in a typical format found in clinical studies. Instead, it describes differences and then asserts that these differences do not "contribute any adverse impact to the device's safety and effectiveness" or "do not impact safety and/or effectiveness."

    The closest to acceptance criteria are the statements made about the new detectors and software features, aiming for equivalent or improved performance compared to the predicate device.

    Aspect of Performance/CharacteristicAcceptance Criteria (Implicit)Reported Device Performance (Implicit)
    Detectors: Image Characteristics (MTF, DQE)Equivalent to existing (predicate) detectors, without adverse impact on safety and diagnostic effectiveness.New detectors have "equivalent image characteristics as the existing ones." MTF is "slightly higher." DQE is "similar" or "a little lower" but "do not contribute any adverse impact to the device's safety and diagnostic effectiveness."
    Detectors: Spatial Resolution (Pixel Pitch, High Contrast Limiting Resolution, Number of pixels)No adverse impact on safety and/or effectiveness.New detectors have "same or lower pixel pitch," "same or higher pixel number," and "same or higher resolution" compared to predicate. These "do not impact safety and/or effectiveness."
    Detectors: Mechanical/Environmental (Dust/Water-resistance, Max. load capacity)No adverse impact on safety and effectiveness.New detectors have "same or better dust/water-resistance" and "same or higher max load capacity" than predicate. These changes "do not contribute any adverse impact to the device's safety and effectiveness."
    SimGrid (Image Processing Software)No adverse impact on the device's safety and effectiveness. Improved functionality.Updated SimGrid provides a parameter for controlling strength, which "does not contribute any adverse impact to the device's safety and effectiveness."
    IPE (Image Post-processing Engine)No impact to the device's safety and effectiveness.Upgraded IPE (Clinical Parameter Control) allows simultaneous comparison of editing image with current image, and this "does not contribute any impact to the device's safety and effectiveness."
    Overall System (Clinical Equivalence)Equivalent to the predicate device.Phantom image evaluations were found to be "equivalent to the predicate devices." "There is no significant difference in the average score of image quality evaluation between the proposed device and the predicate device."

    2. Sample Size for Test Set and Data Provenance

    • Sample Size for Test Set: Not explicitly stated as a numerical sample size for "cases" in the traditional sense of a clinical study. The document mentions "Anthropomorphic phantom images were provided." The number of phantom images or specific phantom types isn't detailed. For software elements, "Software System Test Case for verification and validation" was performed, but no numerical count is given.
    • Data Provenance: The study used "Anthropomorphic phantom images" and non-clinical data (MTF, DQE measurements, Software System Test Cases). This indicates testing in a controlled environment (laboratory/phantom studies). The country of origin for the data is not specified, but the manufacturer is based in the Republic of South Korea. The data is non-clinical/phantom-based, so it's not strictly "retrospective or prospective" in the human clinical trial sense.

    3. Number of Experts and Qualifications for Ground Truth

    • Number of Experts: "A professional radiologist" (singular)
    • Qualifications of Experts: "professional radiologist" (no specific years of experience or subspecialty mentioned beyond being a radiologist).

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not applicable or not specified. A single "professional radiologist" evaluated the phantom images, implying no consensus or multi-reader adjudication process as there was only one reviewer mentioned.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • MRMC Study: No, a multi-reader multi-case (MRMC) comparative effectiveness study was not explicitly stated or described. The evaluation involved a single radiologist reviewing phantom images.
    • Effect Size: Not applicable, as no MRMC study was conducted.

    6. Standalone (Algorithm only without human-in-the-loop performance) Study

    • Standalone Study: Yes, performance data such as MTF and DQE measurements (IEC 62220-1) are inherent standalone technical performance metrics of the detectors. Software System Test Cases for verification and validation would also be considered standalone testing. The "anthropomorphic phantom images" evaluation by a single radiologist could be considered a form of standalone performance evaluation for the system's image quality output, as it assesses the device's generated images directly.

    7. Type of Ground Truth Used

    • Type of Ground Truth: The ground truth for the phantom image evaluation was implicitly the expected characteristics and quality of images of anthropomorphic phantoms, as assessed by a professional radiologist for equivalence to predicate devices. For technical metrics like MTF and DQE, the ground truth is against the measurement standards (IEC 62220-1). There was no pathology or outcomes data.

    8. Sample Size for the Training Set

    • Sample Size for Training Set: The document does not provide information regarding the sample size of a training set for any machine learning components. While the device includes an "Image Post-processing Engine (IPE)" and features like "SimGrid," if these involve learned algorithms, their training data is not discussed. This submission is for an X-ray system, and the focus is on hardware and general image processing functionality rather than an AI/CADe device where training data would typically be detailed.

    9. How the Ground Truth for the Training Set Was Established

    • How Ground Truth for Training Set was Established: Not applicable or not provided, as training set details are absent.
    Ask a Question

    Ask a specific question about this device

    K Number
    K171119

    Validate with FDA (Live)

    Device Name
    GM85
    Date Cleared
    2017-05-12

    (28 days)

    Product Code
    Regulation Number
    892.1720
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The GM85 Digital X-ray imaging System is intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications.

    Device Description

    The GM85 Digital Mobile X-ray imaging System is the equipment that captures images by transmitting X-ray to a patient's body. The X-ray passing through a patient's body is sent to the detector and then converted into electrical signals. These signals go through the process of amplification and digital data conversion in the signal process device before being sent to the S-Station (Operation Software) and saved in DICOM file, a standard for medical imaging. The captured images are sent to the Picture Archiving & Communication System (PACS) server, and can be used for reading images.

    The Bone Suppression Image (BSI) is a post-image processing software option which provides companion images to assist diagnosis in addition to the images obtained from normal diagnosis protocol. The BSI software option suppresses bone anatomy to enhance visualization of chest pathology in a companion image that is delivered in addition to the original diagnostic image.

    The BSI option is designed to be exclusively installed in S-station, SAMSUNG digital Xray operation software, while it uses the radiological image as an input and do not depend on how the image is acquired or which radiology device is used.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the GM85 Digital X-ray imaging System, specifically focusing on the integration of the Bone Suppression Image (BSI) feature. However, the document explicitly states that the application of BSI to the proposed device does not require clinical data and therefore, clinical studies to establish acceptance criteria or device performance are not detailed within this document.

    The document primarily focuses on demonstrating substantial equivalence to predicate devices based on non-clinical data and technical characteristics, rather than presenting a performance study with acceptance criteria.

    Therefore, many of the requested details about acceptance criteria and the study that proves the device meets them cannot be extracted from this particular submission.

    Here's what can be inferred and what is explicitly stated as not applicable or not provided:


    1. A table of acceptance criteria and the reported device performance

    • Acceptance Criteria: Not explicitly stated in terms of clinical performance metrics (e.g., sensitivity, specificity, AUC). The acceptance criteria for this submission appear to be focused on substantial equivalence to predicate devices through technical comparisons and non-clinical testing.
    • Reported Device Performance: No clinical performance metrics are reported because clinical data was deemed not required and thus not provided. The performance is assessed based on the device having the same detectors and therefore showing "no difference in non-clinical testing data such as MTF and DQE measurements from the predicate device #1 and #2."

    2. Sample size used for the test set and the data provenance

    • Sample Size: Not applicable/not provided. No clinical test set.
    • Data Provenance: Not applicable/not provided. No clinical test data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Number of Experts: Not applicable/not provided. No clinical test set.
    • Qualifications of Experts: Not applicable/not provided.

    4. Adjudication method for the test set

    • Adjudication method: Not applicable/not provided. No clinical test set.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • MRMC Study: No. The document explicitly states "The application of BSI... to the proposed device GM85 does not require clinical data." Therefore, an MRMC study was not performed for this submission.
    • Effect Size: Not applicable/not provided.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Standalone Performance: Not explicitly detailed as a performance study for the BSI feature itself. The BSI software option "provides companion images to assist diagnosis," implying it's an aid, not a standalone diagnostic tool without a human. However, its verification and validation were done through a "Software System Test Case" (non-clinical).

    7. The type of ground truth used

    • Type of Ground Truth: Not applicable for a clinical performance study. For the BSI software's non-clinical verification and validation, the ground truth would be against expected software output based on its design specifications, rather than clinical outcomes or expert consensus on patient data.

    8. The sample size for the training set

    • Sample Size for Training Set: Not applicable/not provided. The document focuses on performance relative to predicate devices and non-clinical validation. Details regarding the training of the BSI algorithm itself are not included in this submission.

    9. How the ground truth for the training set was established

    • Ground Truth for Training Set: Not applicable/not provided as training set details are not included.

    Summary of what is provided regarding acceptance and study:

    The acceptance for this 510(k) submission is based on demonstrating substantial equivalence to predicate devices. This is achieved through:

    • Technical Characteristics Comparison: Showing that the proposed device (GM85 with BSI) is identical or equivalent to Predicate Device #1 (GM85 without BSI, K162278) in its core components (HVG, X-ray Tube Assembly, Detectors) and shares the BSI feature with Predicate Device #2 (GC70 with BSI, K163115).
    • Safety, EMC, and Performance Data (Non-clinical):
      • Electrical, mechanical, environmental safety, and performance testing according to standards (ES 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-2-28, IEC 60601-2-54, ISO14971, 21CFR1020.30, 21CFR1020.31).
      • EMC testing according to IEC 60601-1-2.
      • Wireless function testing according to guidance on "Radio frequency Wireless Technology in Medical Devices."
      • All these tests "were satisfying the standards."
    • Non-clinical Data for Imaging Performance:
      • Conformance to FDA "Guidance for the Submission of 510(k)'s for Solid-State X-ray Imaging Devices," including MTF and DQE measurements as tested by IEC 62220-1.
      • The proposed device uses the same detectors as the predicate devices, "so it shows no difference in non-clinical testing data such as MTF and DQE measurements from the predicate device #1 and #2."
    • Software Verification and Validation: "BSI installed in S-Station of GM85 has been verified and validated through GM85 Software System Test Case." (This is a non-clinical, internal software validation).

    Conclusion from the document: "The non-clinical demonstrates that the proposed device is as safe, as effective, and performs as well as the legally marketed predicate devices."

    The critical point for this document is that the FDA determined, and the submission reflects, that for the integration of the already cleared BSI software into the GM85 mobile X-ray system, no new clinical data was required to demonstrate substantial equivalence.