Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K213611

    Validate with FDA (Live)

    Device Name
    Fenom Pro
    Date Cleared
    2023-06-07

    (569 days)

    Product Code
    Regulation Number
    862.3080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Fenom Pro is a portable, non-invasive device to measure fractional exhaled nitric oxide (FENO) in human breath. FENO is increased in some airway inflammatory processes, such as asthma, and decreases in response to anti-inflammatory treatment. Fenom Pro measures fractional exhaled nitric oxide according to guidelines established by the American Thoracic Society.

    Measurement of FENO by Fenom Pro is a non-invasive quantitative method to measure the decrease in FENO concentration in asthma patients that often occurs after treatment with anti-inflammatory pharmacological therapy as an indication of therapeutic effect in patients with elevated FENO measurements are to be used as an adjunct to established clinical assessments. Fenom Pro is suitable for adults and children 6 years and older.

    Fenom Pro should be used in a point-of-care healthcare setting under professional supervision. Fenom Pro should not be used in critical care, emergency care or in anesthesiology.

    Device Description

    Fenom Pro is a non-invasive point-of-care breath analyzer for the quantitative measurement of fractional exhaled nitric oxide (FENO) in expired human breath.

    Fenom Pro is comprised of four major components. The main unit contains a touch screen interface for the user and houses the nitric oxide sensor and pneumatics needed to sample the patient's breath. The patient exhales into the Fenom Pro though a single use disposable mouthpiece which is attached to the handpiece. The handpiece is connected to the main unit via a breath tube.

    For testing, the patient inhales to total lung capacity and then slowly exhales into the mouthpiece for 6 or 10 seconds, depending on the mode of operation. Approximately 28 seconds after a successful breath maneuver, the FENO concentration is displayed in parts per billion (ppb) on the device touch screen. The device has a daily quality control procedure to ensure consistent device performance.

    AI/ML Overview

    The Fenom Pro device, a breath nitric oxide test system, was evaluated for performance against specified acceptance criteria.

    1. Table of Acceptance Criteria and Reported Device Performance:
    CriteriaAcceptance CriteriaReported Device Performance
    Precision<5 ppb NO for concentrations ≤50 ppb. <10% for NO concentrations >50 ppb.<5 ppb NO for concentrations ≤50 ppb. <10% for NO concentrations >50 ppb.
    Accuracy± 5 ppb NO for concentrations ≤50 ppb. ± 10% for NO concentrations >50 ppb.± 5 ppb NO for concentrations ≤50 ppb. ± 10% for NO concentrations >50 ppb.
    LinearitySlope: 1.0 ± 0.05, r2: ≥ 0.998Slope: 1.0 ± 0.05, r2: ≥ 0.998
    Detection Level10 ppb10 ppb
    Measurement Range10-200 ppb NO10-200 ppb NO
    Analysis TimeApproximately 30 secondsApproximately 30 seconds
    Breath Sample ConditioningNot explicitly stated as acceptance criteria, but a difference in technology was noted compared to the predicateNot explicitly stated as accepted, but the device's technological difference is "None" compared to predicate's "Reduced humidity"
    Test PrincipleNot explicitly stated as acceptance criteria, but a difference in technology was noted compared to the predicateNot explicitly stated as accepted, but the device uses "Amperometric Sensor Technology" compared to predicate's "Potentiometric Sensor Technology"
    Measurement ModeNot explicitly stated as acceptance criteria, but a difference in technology was noted compared to the predicateNot explicitly stated as accepted, but the device has "6 second and 10 second breath maneuver" compared to predicate's "10 second breath maneuver"
    1. Sample size used for the test set and data provenance:

      • The document states that "Clinical studies were performed to evaluate the clinical precision and clinical accuracy of the candidate device." However, specific sample sizes for these clinical studies are not provided in the summary.
      • The provenance of the data (e.g., country of origin, retrospective or prospective) is not explicitly mentioned.
    2. Number of experts used to establish the ground truth for the test set and their qualifications: Not provided in the summary.

    3. Adjudication method for the test set: Not provided in the summary.

    4. Multi-Reader Multi-Case (MRMC) comparative effectiveness study: No MRMC study was mentioned. The device measures fractional exhaled nitric oxide (FENO) which is a quantitative measurement, not typically subject to subjective interpretation by multiple human readers in the same way as imaging studies. The indications for use state that FENO measurements are to be used as an adjunct to established clinical assessments, implying a standalone measurement rather than an AI-assisted human reading.

    5. Standalone (algorithm only without human-in-the-loop performance) study: Yes, the device is described as measuring FENO quantitatively. The non-clinical and clinical studies performed (precision, accuracy, linearity, etc.) relate to the device's standalone performance in measuring FENO.

    6. Type of ground truth used: For quantitative measurements like FENO, the ground truth would typically be established by highly accurate reference methods or calibration gases with known nitric oxide concentrations for non-clinical studies. For clinical studies, comparison to a gold standard FENO measurement device or clinical outcomes (e.g., response to anti-inflammatory treatment) would be used to establish clinical accuracy and precision. The summary implies that "studies were performed to evaluate performance with regards to the specificity (interference), precision, accuracy, linearity, effect of temperature and humidity, detection limit and stability" for non-clinical data, and "clinical precision and clinical accuracy" for clinical data. The exact nature of the "ground truth" (e.g., comparison to a specific reference method) for each of these studies is not detailed.

    7. Sample size for the training set: Not applicable based on the provided information. This device is a measurement system and the information provided does not indicate the use of machine learning models requiring a specific "training set" in the common sense for AI/ML devices.

    8. How the ground truth for the training set was established: Not applicable, as no training set for an AI/ML model is indicated.

    Ask a Question

    Ask a specific question about this device

    K Number
    K182874

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-02-13

    (124 days)

    Product Code
    Regulation Number
    862.3080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Fenom Pro™ Nitric Oxide Test is a portable, non-invasive device to measure fractional exhaled nitric oxide (FeNO) in human breath. FeNO is increased in some airway inflammatory processes, such as asthma, and often decreases in response to anti-inflammatory treatment of FeNO by Fenom Pro™ is a method to measure the decrease in FeNO concentration in asthma patients that often occurs after treatment with anti-inflammatory pharmacological therapy as an indication of therapeutic effect in patients with elevated FeNO levels. FeNO measurements are to be used as an adjunct to established clinical assessments. Fenom Pro™ is suitable for children, approximately 7-17 years, and adults 18 years and older.

    Testing using the Fenom Pro™ should only be done in a point-of-care healthcare setting under professional supervision. Fenom Pro™ should not be used in critical care, emergency care or in anesthesiology.

    Device Description

    Fenom Pro™ is a point-of-care breath analyzer that uses solid-state electrochemical technology to measure the fraction of exhaled nitric oxide (FeNO), a marker for airway inflammation, in human exhaled breath. Measurement of FeNO by Fenom Pro is a quantitative and non-invasive method to indicate therapeutic effects of anti-inflammatory pharmacological therapy in patients with elevated FeNO levels. Fenom Pro™ is suitable for children, approximately 7-17 years, and adults 18 years and older.

    Fenom Pro uses solid state, potentiometric, sensor technology sensitive to nitric oxides (NO) compounds. The solid state sensor is fluidly preceded by a reactive filter material that renders (oxidizes) potentially confounding species such as carbon monoxide (CO), ammonia (NH4), and methanol (CH4O) inactive, or inert, to the NO sensor. Fenom Pro provides visual and audible feedback during its use. The visual and audible feedback is especially important during the FeNO measurement such that the user can modulate their breath speed within the flow parameters required by the American Thoracic Society (ATS) and the European Respiratory Society (ERS) standards.

    Fenom Pro is comprised of four major components. The main unit contains a touch screen interface for the use as well as houses the nitric oxide sensor and pneumatics needed to sample the patient's breath. The patient interfaces with Fenom though the mouthpiece which is attached to the handpiece. The handpiece is connected to the main unit via a breath tube. The handpiece contains a breath conditioning cartridge which prepares the breath sample from the patient for proper analysis in the main unit. Both the mouthpiece and the breath conditioning cartridge are consumables.

    AI/ML Overview

    Here's an analysis of the provided text, extracting the acceptance criteria and the study details for the Fenom Pro™ Nitric Oxide Test.

    Device: Fenom Pro™ Nitric Oxide Test (K182874)
    Intended Use: Portable, non-invasive device to measure fractional exhaled nitric oxide (FeNO) in human breath. FeNO measurements are used as an adjunct to established clinical assessments to indicate therapeutic effect in patients with elevated FeNO levels after anti-inflammatory pharmacological therapy. Suitable for children (approx. 7-17 years) and adults (18+ years).


    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document details non-clinical (analytical) and clinical studies. We will parse the acceptance criteria and performance from these sections.

    Non-Clinical (Analytical) Performance:

    Study CategoryAcceptance CriteriaReported Device PerformancePass/Fail (based on stated performance)
    AccuracyFor 15 ppb: +/- 5 ppbAcross 5 environmental conditions for 2 devices and 5 replicates each, all results for 15 ppb target concentration were within +/- 5 ppb. (e.g., Ambient T/RH: GP18 had upper 95% error limit of 0.17ppb and lower of -1.97ppb; GP35 had upper 0.44ppb and lower -3.84ppb).Pass
    For 75 ppb & 200 ppb: +/- 10%Across 5 environmental conditions for 2 devices and 5 replicates each, all results for 75 ppb and 200 ppb target concentrations were within +/- 10%. (e.g., Ambient T/RH: GP18 had upper 1.83% and lower -2.12% for 75ppb; High T/Low RH: GP35 had upper -1.63% and lower -3.38% for 200ppb).Pass
    Precision (Repeatability)Not explicitly stated as a pass/fail criterion in the table, but reported as SD (ppb) and %CV.For 10 ppb: SDs ranged from 1.1 to 1.9 ppb. For 25 ppb: SDs ranged from 1.2 to 2.3 ppb. For 75 ppb: %CVs ranged from 3.2% to 7.1%. For 200 ppb: %CVs ranged from 3.0% to 7.3%.N/A (Data Reported)
    Precision (Within-Device)Not explicitly stated as a pass/fail criterion in the table, but reported as SD (ppb) and %CV.For 10 ppb: SDs ranged from 1.2 to 2.3 ppb. For 25 ppb: SDs ranged from 1.2 to 3.3 ppb. For 75 ppb: %CVs ranged from 4.3% to 7.5%. For 200 ppb: %CVs ranged from 3.4% to 8.5%.N/A (Data Reported)
    LinearitySlope between 0.95 and 1.05 and R (correlation coefficient) presumably close to 1.0.Device GammaPrime42: Slope 1.03, Intercept 2.32, R 0.999. Device GammaPrime49: Slope 1.02, Intercept 0.231, R 0.999. Combined: Slope 1.02, Intercept 1.27, R 0.998.Pass
    Limit of Detection (LoD)< 10 ppbDevice #1: LoD = 1.8 ppb. Device #2: LoD = 4.6 ppb.Pass
    Interference (Other Gases)Change in response <= 4 ppb NO equivalentFor acetaldehyde, acetone, ammonia, carbon dioxide, carbon monoxide, ethanol, hydrogen, hydrogen sulfide, isoprene, oxygen: all interferences were <= 4 ppb NO equivalent. Acetonitrile showed 120.8 ppb interference, but is only present in exhaled breath when recently smoked, and the device is labeled for use not after smoking.Pass (with labeling mitigation for acetonitrile)
    Interference (Exogenous Substances)Mean difference between 60 minutes and baseline values <= 5 ppb. Lower and upper 95% confidence intervals around the means must include zero.All 7 tested substances (Alcohol Free Mouthwash, Caffeinated Soda, Caffeine Free Soda, Menthol Lozenge, Mouthwash with Alcohol, Non-Menthol Lozenge, Toothpaste) showed mean differences <= 5 ppb and 95% CIs that included zero at 60 minutes.Pass

    Clinical Performance:

    Study CategoryAcceptance CriteriaReported Device PerformancePass/Fail (based on stated performance)
    Clinical Precision (User Bias)No explicit numerical acceptance criteria given, but the objective was to confirm no user bias. The text implies a "very high quantitative agreement."Maximum mean bias observed was only 1.2 ppb between pairs of HCPs. Deming regression, correlation (Pearson R > 0.9873 for all HCP pairs), and all bias analyses demonstrated very high quantitative agreement.Pass (Implied)
    Clinical Precision (Within-subject Variability)Less than 5 ppb by mean standard deviation for FeNO values below 50 ppb. %CVs for FeNO values greater than 50 ppb were maintained at less than 10%, "unless sample size was small."Across both clinical precision studies and all age groups: - For FeNO < 50ppb (excluding 0-<10ppb where N=0): Mean SDs were generally less than 5 ppb. Exceptions occurred in small N subgroups (e.g., All ages Visit 1 20-<30ppb where N=5, Mean SD 4.97). - For FeNO >= 50ppb: Mean CVs were generally less than 10%. Exceptions also occurred with small N (e.g., 5-17 Visit 1 75-<100ppb where N=3, Mean CV 18.73%). The document mitigates these by stating "unless sample size was small".Pass (with noted small N exceptions)
    Clinical Efficacy (Concordance with established measures)Implicitly, significant concordance between Fenom Pro and other established asthma-related outcome measures (FEV1, ACQ/pACQ).Significant differences between the two visits were achieved for all three modalities (FeNO, spirometry, asthma questionnaires). Kendall's Tau p-values were 0.0370 for FeNO vs. FEV1 and 0.0125 for FeNO vs. ACQ/pACQ, indicating statistically significant correlations (concordance) given typical alpha levels of 0.05.Pass

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Non-Clinical (Analytical):

      • Accuracy and Environmental Testing: 2 devices, 3 concentrations (15, 75, 200 ppb), 5 replicates per concentration, 5 environmental conditions. Total of 150 tests.
      • Precision: 5 devices, 5 operating days, 2 sessions per day, 4 runs per session, 2 replicates per concentration (10, 25, 75, 200 ppb). This calculates to 80 replicates per sample per device according to CLSI EP05-A3.
      • Linearity: 2 devices, 8 NO concentration levels (5, 10, 15, 30, 50, 100, 150, 200 ppb), 5 replicates per concentration. Total 80 tests.
      • Limit of Detection: 2 devices, 0 ppb (60 replicates), 5 ppb (30 replicates), 10 ppb (30 replicates) over three days. Total 120 replicates.
      • Interference (Other Gases): Tested in a laboratory setting. No specific number of tests/replicates provided, but states "The applicable concentration of each substance was generated... and the sensor signal was measured."
      • Interference (Exogenous Substances): Minimum of 10 volunteers per substance (7 substances). This means at least 70 volunteers.
      • Data Provenance: The document does not explicitly state the country of origin for the non-clinical data. It is implied to be laboratory-based and likely retrospective as it's presented as completed tests.
    • Clinical:

      • Clinical Precision (User Bias Study): 127 subjects (44 pediatric, 83 adults). Each subject yielded 6 evaluations (2 Fenom Pro measurements with assistance of 3 HCPs). Data provenance not explicitly stated (e.g., country), but implied prospective data collection for this study.
      • Clinical Precision (Within-subject Variability/Longitudinal Study): 82 subjects (37 pediatric, 45 adults). Each subject provided replicate (n=2) FeNO measurements at Visit 1 (baseline) and Visit 2 (after approx. two weeks). Data provenance not explicitly stated, but implied prospective data collection.
      • Clinical Efficacy (Concordance Study): 82 subjects in longitudinal study (matches the precision study's subject count) - 37 pediatric, 45 adults. 80 subjects for regression analysis (due to missing asthma symptom scores for 2 subjects). Data provenance not explicitly stated, but implied prospective data collection.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • Non-Clinical Studies:
      • Accuracy, Precision, Linearity, LoD: Ground truth was established using calibrated reference instruments/methods. Specifically, a chemiluminescence device calibrated against a NIST (National Institute of Standards and Technology) traceable NO tank was used to measure the actual concentration of nitric oxide in simulated breath mixtures. This does not involve human experts in the same way as, for example, image interpretation.
      • Interference (Other Gases): Ground truth was based on the controlled generation of specific gas concentrations and measurement by the device.
      • Interference (Exogenous Substances): The ground truth for this was the subjects' baseline Fenom Pro levels before exposure to the substances, acting as their own control for comparison.
    • Clinical Studies:
      • Clinical Precision (User Bias Study) and Clinical Precision (Within-subject Variability/Longitudinal Study): The ground truth was effectively the FeNO measurements themselves, as performed by the device. The study evaluated agreement between HCPs and within subjects. The study used three healthcare professionals (HCPs) to assist with measurements. Their specific qualifications (e.g., type of healthcare professional, years of experience) are not provided in the summary.
      • Clinical Efficacy (Concordance Study): The "ground truth" for showing clinical efficacy was the change in FeNO, which was correlated with changes in established asthma-related outcome measures (spirometry - FEV1, and asthma questionnaires - ACQ/pACQ). These established measures themselves serve as the comparative ground truth. The expertise in interpreting FEV1 and ACQ/pACQ would be inherent in the clinicians applying these standard assessments but is not explicitly detailed as part of "ground truth establishment" for the device's performance.

    4. Adjudication Method for the Test Set

    • Non-Clinical Studies: No adjudication method is applicable as these are instrumental measurements against calibrated standards.
    • Clinical Studies: No formal adjudication method involving multiple experts resolving discrepancies is described for the clinical studies. The "user bias" study quantifies bias between HCPs rather than adjudicating a "true" FeNO value. The other clinical studies compare the device's output to standard clinical measures rather than relying on a panel for ground truth adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, What was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    This device is a breath test system, not an imaging AI system that assists human readers (like radiologists). Therefore, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study focusing on human reader improvement with AI assistance is not applicable to this device.


    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    The Fenom Pro™ is a point-of-care device that directly measures FeNO. Its non-clinical (analytical) performance tests (Accuracy, Precision, Linearity, LoD, Interference) are essentially standalone performance of the device's measurement algorithm/sensor. While human operators perform the physical test, the output (FeNO value) is generated by the device's internal algorithms from the sensor data, without human interpretation or intervention in the measurement result itself.

    The "user bias" clinical study specifically explores human-in-the-loop variance (i.e., how different operators' assistance impacts measurements), which is a crucial aspect for a point-of-care device. However, the core measurement itself, once the breath sample is collected, is standalone.


    7. The Type of Ground Truth Used

    • Non-Clinical (Analytical) Studies:
      • Calibrated Reference Standard: For Accuracy, Precision, and Linearity, the ground truth was established using a chemiluminescence device calibrated against a NIST traceable NO tank measuring specific concentrations of nitric oxide mixed in simulated breath.
      • Controlled Input: For Limit of Detection and Interference (other gases), known, controlled concentrations of gases were presented to the device.
      • Self-reference (Baseline): For Interference (Exogenous Substances), each subject's own baseline Fenom Pro measurement before exposure served as the ground truth for evaluating the impact of the substances.
    • Clinical Studies:
      • Comparative Assessment: For the Clinical Efficacy study, the "ground truth" for demonstrating the device's utility was established by comparison with established clinical assessments: FEV1 (spirometry) and ACQ/pACQ (asthma questionnaires).
      • Internal Consistency: For the Clinical Precision studies, the "ground truth" was essentially the device's own measurement output, and the studies focused on the consistency and variability of these measurements across different operators and occasions.

    8. The Sample Size for the Training Set

    The document does not explicitly describe a "training set" in the context of an machine learning model that would require separate training data. The device appears to be a physical sensor-based measurement system rather than a machine learning algorithm that learns from a dataset.

    The listed studies refer to:

    • Non-clinical testing: Performed on a limited number of devices (e.g., 2 devices for accuracy, 5 for precision) against controlled gas mixtures. This is instrument verification/validation, not AI model training.
    • Clinical studies: These are validation studies on human subjects to demonstrate performance characteristics like precision and concordance with established clinical measures, after the device's design is largely finalized. They do not constitute a "training set" for an AI algorithm.

    Therefore, the concept of a "training set sample size" as typically understood for AI/ML devices is not applicable here.


    9. How the Ground Truth for the Training Set Was Established

    As explained in point 8, a "training set" for an AI/ML model is not applicable to this device description. The device's fundamental measurement principle relies on solid-state electrochemical sensor technology, calibrated using reference standards in a laboratory setting, rather than being "trained" on a large dataset of patient results.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1