510(k) Data Aggregation

    K Number
    K140372

    Validate with FDA (Live)

    Date Cleared
    2014-04-07

    (52 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Skeletal Dynamics Protean Fragment Plate System is intended for stabilization and fixation of small bone fragments in fresh fractures, revision procedures, joint fusion and reconstructions of small bones of the hand, foot, wrist, ankle, humerus, scapula, finger, toe, and pelvis, particularly in osteopenic bone.

    Device Description

    The Skeletal Dynamics Protean Fragment Plate System is a set of titanium bone plates designed for stabilization and repair of small bone fragments. Included in the set are titanium bone screws, fixation pegs, cobalt chrome cannulated polyaxial screws, and specialized instrumentation. The Fragment Plates are available in 3 configurations and are made of medical grade titanium alloy. The system is provided non-sterilized in the user facility. The Skeletal Dynamics Protean Fragment Plate System is comprised of: Titanium alloy plates and screws, Cobalt chrome polyaxial screws, Stainless steel K-wires (for provisional fixation not for implantation), System specific instrumentation.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device called the "Skeletal Dynamics Protean Fragment Plate System." This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving novel effectiveness through clinical studies with detailed acceptance criteria as one might find for new drug applications or high-risk medical devices requiring PMA.

    Therefore, many of the requested elements for describing "acceptance criteria and the study that proves the device meets the acceptance criteria" are not applicable in this context. The primary "study" is a preclinical mechanical testing study to show equivalence, not a clinical trial evaluating human performance or AI capabilities.

    Here’s a breakdown based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Criterion TypeAcceptance Criterion (Not explicitly stated as numerical, but implied as "substantially equivalent")Reported Device Performance
    Mechanical PerformanceSubstantially equivalent to the predicate device (DePuy Small Bone Locking Plate System (K081546)) in static and dynamic testing."Preclinical analysis and testing demonstrated that the Skeletal Dynamics Protean Fragment Plate System is substantially equivalent to the predicate device currently marketed. Mechanical testing which established equivalency included static and dynamic testing." The exact numerical results are not provided.
    Intended UseIdentical or highly similar to the predicate device."The Skeletal Dynamics Protean Fragment Plate System is intended for stabilization and fixation of small bone fragments in fresh fractures, revision procedures, joint fusion and reconstructions of small bones of the hand, foot, wrist, ankle, humerus, scapula, finger, toe, and pelvis, particularly in osteopenic bone." This is also considered substantially equivalent to the predicate.
    MaterialsSimilar to predicate device; medical grade.Titanium alloy plates and screws, Cobalt chrome polyaxial screws, Stainless steel K-wires. This is stated to be substantially equivalent to the predicate.
    DesignSimilar fundamental scientific technology to predicate device.No specific design details against a criterion are given, but the overall design is stated to be substantially equivalent.
    SterilityNot explicitly detailed but implied to be manageable post-manufacturing by the user."The system is provided non-sterilized in the user facility." Stated as substantially equivalent to the predicate.
    PackagingSubstantially equivalent to the predicate device.Not detailed, but stated as substantially equivalent to the predicate.

    2. Sample size used for the test set and the data provenance

    • Sample Size: Not specified. For mechanical testing of medical devices, "samples" typically refer to the number of devices or components subjected to a particular test. The document only states "Preclinical analysis and testing."
    • Data Provenance: Not applicable in the context of clinical data. This refers to in-vitro mechanical testing conducted by the manufacturer.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not Applicable. This is a mechanical device evaluation, not a clinical study involving expert interpretation of data or images. "Ground truth" would relate to precisely measured mechanical properties or failure points, not expert consensus on diagnoses.

    4. Adjudication method for the test set

    • Not Applicable. There is no adjudication method described because there's no clinical data or expert interpretation involved in determining a "ground truth" through consensus.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. This device is a bone plate system, not an AI or imaging diagnostic tool. No MRMC study was conducted.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable. This is a hardware medical device, not an algorithm or AI.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • For mechanical testing, the "ground truth" would be established by the physical testing parameters and measurements (e.g., force applied, displacement, cycles to failure, yield strength) as per recognized engineering standards (e.g., ASTM, ISO). The document does not specify which standards were followed, but these would define how performance is "measured."

    8. The sample size for the training set

    • Not Applicable. This is a mechanical device, not an AI system that requires a "training set" of data.

    9. How the ground truth for the training set was established

    • Not Applicable. No training set exists for this type of device submission.
    Ask a Question

    Ask a specific question about this device

    K Number
    K061748

    Validate with FDA (Live)

    Date Cleared
    2006-07-17

    (26 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The proposed Mini Fragment Plates are intended for essentially non load bearing stabilization and fixation of small bone fragments in fresh fractures, revision procedures, joint fusion, and reconstructions of small bones of the hand, foot, wrist, ankle, humerus, scapula, finger, toe, pelvis, and craniomax, illofacial skeleton.

    Device Description

    The proposed Mini Fragment Plates are a set of orthopedic plates and fasteners supplied in a sterilization tray together with several reusable and disposable tools. The set of orthopedic plates and fasteners provided in the sterilization tray consists of the following implantable devices: Fragment Plates that are provided in the following design configurations: Straight, Y, Right Hockey Stick, and Left Hockey Stick. Pegs; Screws; and K-Wires. Included in the sterilization tray are the following re-useable instruments: Peg drivers, Other standard surgical tools. In addition, the following are non-reusable instruments included in the sterilization tray: F.A.S.T. Guide Technology™ Drill Guide, Drill bits.

    AI/ML Overview

    The provided text is for a 510(k) premarket notification for "Mini Fragment Plates", a type of orthopedic implant. It focuses on demonstrating substantial equivalence to a predicate device rather than presenting a performance study with acceptance criteria in the context of device performance metrics like sensitivity, specificity, or reader studies.

    Therefore, the requested information, such as acceptance criteria, reported device performance, sample sizes for test and training sets, expert qualifications, adjudication methods, MRMC studies, standalone performance, and ground truth types and establishment, are not present in the provided document.

    Instead, the document details:

    • Device Name: Mini Fragment Plates
    • Manufacturer: Hand Innovations, LLC
    • Predicate Device: Fragment Plate System (K041081)
    • Classification: Class II
    • Indications for Use: Stabilization and fixation of small bone fragments in fresh fractures, revision procedures, joint fusion, and reconstructions of small bones of the hand, foot, wrist, ankle, humerus, scapula, finger, toe, pelvis, and craniomaxillofacial skeleton.
    • Device Description: Orthopedic plates and fasteners (Straight, Y, Right Hockey Stick, Left Hockey Stick), pegs, screws, K-wires, reusable instruments (peg drivers, standard surgical tools), and non-reusable instruments (F.A.S.T. Guide Technology™ Drill Guide, drill bits).
    • Biocompatibility: Not required as the material (Ti-6Al-4V ELI per ASTM F 136-00) is the same as the previously approved predicate device and has a long history of safe use.
    • Summary of Substantial Equivalence: The proposed Mini Fragment Plates have the same indications for use, intended use, basic shape/design, operating principle, and materials as the predicate device. The changes do not alter the fundamental scientific technology.

    This document is a regulatory submission focused on confirming that the new device is "substantially equivalent" to an already approved device, meaning it doesn't raise new questions of safety and effectiveness. It does not involve performance testing against specific clinical or system-level acceptance criteria in the way a novel diagnostic or AI-driven device would.

    Ask a Question

    Ask a specific question about this device

    K Number
    K041081

    Validate with FDA (Live)

    Date Cleared
    2004-07-01

    (66 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Fragment Plate System is intended for essentially non load-bearing stabilization and fixation of small bone fragments in fresh fractures, revision procedures, joint fusion, and reconstruction of small bones of the hand, foot, wrist, ankle, humerus, scapula, finger, toe, pelvis, and craniomaxillofacial skeleton.

    Device Description

    The Fragment Plate System is comprised of a variety of titanium plates with shapes and sizes designed for internal fixation of small bone fragments. The set also includes screws, pins, and k-wires. The screws and pins are have a center drive head and are 2.5 mm in diameter. The plates will include "Y," straight, right, and left configurations. Manual surgical instruments are supplied with the system to facilitate implantation.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the Fragment Plate System:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria / StandardReported Device Performance
    Material/Mechanical Standards:
    ASTM F366-82 (Reapproved 2000), Standard Specification for Fixation Pins and WiresThe Hand Innovations, Inc. Fragment Plate Systems comply with this standard.
    ASTM F136-96, Standard Specification for Wrought Titanium-6 Aluminum-4 Vanadium ELI (Extra Low Interstitial) Alloy (R56401) for Surgical Implant ApplicationsThe Hand Innovations, Inc. Fragment Plate Systems comply with this standard.
    Substantial Equivalence:
    Substantial equivalence to a legally marketed predicate device (Distal Volar Fracture Repair System)Demonstrated through a 13-point technological comparison of features and bench testing. The devices were found to be substantially equivalent in terms of technological characteristics. The Fragment Plate System is deemed to be as safe and effective as, or better than, the predicate device.
    Manufacturing Practices:
    Good Manufacturing Practices (GMP)The Fragment Plate System will be manufactured per specifications using good manufacturing practices to ensure safety and effectiveness for its intended use.
    Biocompatibility:
    Implantable and tissue contact materials have a long history of safe usage in medical devices.Materials used in the Fragment Plate System have a long history of safe usage in medical devices.

    2. Sample Size Used for the Test Set and Data Provenance

    This document describes a premarket notification (510(k)) where substantial equivalence is claimed. It does not present clinical study data with a "test set" in the traditional sense for evaluating device performance in patients.

    Instead, the "test set" equivalent primarily involves:

    • Bench Testing: Used for the 13-point technological comparison to demonstrate substantial equivalence to the predicate device. The specific sample sizes for these bench tests are not provided in the document.
    • Compliance with Standards: The document states compliance with ASTM material and fixation standards. These standards themselves define sample sizes and testing methodologies, but the report doesn't detail the specific number of samples tested by Hand Innovations to meet these standards.

    Data Provenance: Not applicable in the context of a clinical test set. The data presented is from internal bench testing and compliance assessments, not patient data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. This device's clearance is based on substantial equivalence to a predicate device and compliance with material standards, not on a clinical ground truth established by medical experts for a test set of patient data.

    4. Adjudication Method for the Test Set

    Not applicable. No clinical ground truth requiring adjudication by experts was established for a test set.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    Not applicable. This is a medical device (fragment plate system) for surgical fixation, not an AI-powered diagnostic or assistive tool for human readers (e.g., radiologists). Therefore, no MRMC study or AI-related comparative effectiveness was conducted or reported.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable. This is a physical medical device, not an algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" in this context is established by:

    • Predicate Device: The performance and safety profile of the "Distal Volar Fracture Repair System" by Hand Innovations, Inc. is the primary ground truth for comparison.
    • Industry Standards: The specifications and performance requirements outlined in ASTM F366 and ASTM F136 serve as a ground truth for material properties and fixation characteristics.
    • Historical Biocompatibility: The "long history of safe usage" of the implantable materials establishes a form of historical ground truth for biocompatibility.

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI/machine learning device, so there is no "training set."

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. No training set for an algorithm was used.