Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K173181

    Validate with FDA (Live)

    Device Name
    ExSpiron 1Xi
    Date Cleared
    2018-03-16

    (168 days)

    Product Code
    Regulation Number
    868.1850
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ExSpiron 1Xi is indicated for use by healthcare professionals in healthcare facilities, such as post-operative care and critical care units, to monitor breathing in patients at least one year of age.

    ExSpiron 1Xi is a non-invasive monitor that graphically displays lung volume against time and reports an approximate value of:

    • Minute Ventilation (MV)
    • Tidal volume (TV)
    • Respiratory rate (RR)
      ExSpiron 1Xi measurements are used as an adjunct to other clinical information.
    Device Description

    The ExSpiron 1Xi is a noninvasive respiratory monitoring system that graphically displays lung volume against time and reports Minute Ventilation, Tidal Volume and Respiratory Rate.

    The ExSpiron 1Xi system consists of:
    • Monitor:
    The Monitor contains a bioimpedance measurement system and a tablet PC housed within a single enclosure.
    o Bioimpedance measurement system: The Monitor incorporates a stabilized high frequency current generator and an adaptive circuit that conditions the resulting voltage signal and converts it to digital form. Firmware within the Monitor performs signal acquisition and relays data to the tablet PC.

    o Computer: A tablet PC performs signal processing and calibration, and runs the graphical user interface (GUI). The PC takes user input from a touch screen through a virtual keyboard and mouse. The GUI is used for recording patient data and displaying the respiratory trace as well as scalar values and trends for minute ventilation, tidal volume, and respiratory rate.

    There are no hardware changes included in this submission.

    Single-Patient Use ExSpiron Electrode PadSet:
    • A Single-Patient Use Electrode PadSet is placed on the skin of the patient's torso. It delivers current to, and records impedance measurements from, the skin. The Electrode PadSet for the ExSpiron 1Xi is identical to that cleared for the Predicate device.

    AI/ML Overview

    The acceptance criteria and study proving the device meets them are summarized below:

    1. Table of Acceptance Criteria and Reported Device Performance:

    ParameterAcceptance Criteria (Equivalence to NM3 within)Reported Device Performance (Mean Measurement Accuracy Error)
    Minute Ventilation (MV)+/- 10%11.9%
    Tidal Volume (TV)+/- 10%12.0%
    Respiratory Rate (RR)+/- 10%4.2% (0.6 breaths/min)

    Note: The document states that the equivalence tests rejected the null hypothesis that the ExSpiron 1Xi and NM3 have different mean values and concluded with 90% power that the measurements of MV, TV, and RR from the ExSpiron 1Xi and NM3 are equivalent within +/-10%. This indicates that the device met the acceptance criteria despite the accuracy error values being slightly outside the 10% for MV and TV in the table. The "accuracy error" values seem to represent the deviation, while the "equivalence tests" are the primary mechanism for meeting the criteria.

    2. Sample Size and Data Provenance for the Test Set:

    • Sample Size: 72 pediatric patients.
    • Data Provenance: Prospective observational study conducted at Children's Hospital of Boston, USA.

    3. Number of Experts and Qualifications for Ground Truth:

    • The study utilized continuous respiratory data collected simultaneously from the ExSpiron 1Xi and a monitoring spirometer (Philips NM3, referred to as the reference device) placed in the respiratory circuit. The NM3 itself essentially served as the "expert" ground truth (or established reference) in this comparative study.
    • No human experts were explicitly mentioned as establishing a separate ground truth for the test set; the NM3 measurements were the reference.
    • The qualifications of the individuals operating the devices or interpreting the NM3 data are not specified, beyond general "healthcare professionals."

    4. Adjudication Method:

    • No explicit adjudication method (like 2+1, 3+1) was mentioned. The study directly compared measurements from the ExSpiron 1Xi to the reference device (NM3).

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • A MRMC comparative effectiveness study was not conducted in the context of human readers improving with or without AI assistance. This study was focused on the standalone performance of the ExSpiron 1Xi against a reference device, not on assisting human readers.

    6. Standalone (Algorithm Only) Performance:

    • Yes, a standalone study was performed. The clinical performance testing described is a standalone evaluation of the ExSpiron 1Xi's ability to measure MV, TV, and RR compared to a recognized reference device (Philips NM3) without a human-in-the-loop component.

    7. Type of Ground Truth Used:

    • The ground truth used was a reference device comparison. Specifically, the measurements from the Philips NM3 Respiratory Profile Monitor, a monitoring spirometer placed in the respiratory circuit, served as the reference standard for MV, TV, and RR.

    8. Sample Size for the Training Set:

    • The device's algorithm uses a "proprietary, non-linear, regressive model trained with a dataset containing respiratory data from a large historical cohort."
    • The exact sample size for the training set is not provided in the document, but it is described as a "large historical cohort."

    9. How the Ground Truth for the Training Set Was Established:

    • The document states that the software algorithm calculates respiratory parameters using a model "trained with a dataset containing respiratory data from a large historical cohort."
    • It does not explicitly detail how the ground truth for this historical training dataset was established. However, given the context of a medical device measuring physiological parameters, it's highly probable that this training data would have been collected using established, accurate measurement methods, likely similar to (or including) reference spirometry or other validated respiratory monitoring techniques.
    Ask a Question

    Ask a specific question about this device

    K Number
    K162131

    Validate with FDA (Live)

    Device Name
    ExSpiron 1Xi
    Date Cleared
    2017-05-09

    (281 days)

    Product Code
    Regulation Number
    868.1850
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ExSpiron 1Xi is indicated for use by healthcare professionals in healthcare facilities, such as post-operative care and critical care units, to monitor breathing in adult (at least 21 years old) patients.

    ExSpiron 1Xi is a non-invasive monitor that graphically displays lung volume against time and reports an approximate value of:

    • Minute Ventilation (MV)

    • Tidal volume (TV)

    • Respiratory rate (RR)

    ExSpiron 1Xi measurements are used as an adjunct to other clinical information.

    Device Description

    The ExSpiron 1Xi is a noninvasive respiratory monitoring system that graphically displays lung volume against time and reports Minute Ventilation, Tidal Volume and Respiratory Rate.

    The ExSpiron 1Xi system consists of:
    Monitor:
    The Monitor contains a bioimpedance measurement system and a tablet PC housed within a single enclosure.
    Bioimpedance measurement system: The Monitor incorporates a stabilized high frequency current generator and an adaptive circuit that conditions the resulting voltage signal and converts it to digital form. Firmware within the Monitor performs signal acquisition and relays data to the tablet PC. Computer: A tablet PC performs signal processing and calibration, and runs the graphical user interface (GUI). The PC takes user input from a touch screen through a virtual keyboard and mouse. The GUI is used for recording patient data and displaying the respiratory trace as well as scalar values and trends for minute ventilation, tidal volume, and respiratory rate.
    A Single-Patient Use Electrode PadSet is placed on the skin of the patient's torso. It delivers current to, and records impedance measurements from, the skin. The Electrode PadSet for the ExSpiron 1Xi is identical to that cleared for the Predicate device.

    AI/ML Overview

    Acceptance Criteria and Device Performance for ExSpiron™ 1Xi

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document compares the proposed device (ExSpiron 1Xi) to a predicate device (ExSpiron™ 1Xi, K130170). The comparison focuses on various characteristics, including accuracy. While explicit "acceptance criteria" for accuracy are not stated as numerical thresholds, the document presents the accuracy results of the proposed device. The implied acceptance is that the proposed device performs at least as well as, or better than, the predicate device.

    CharacteristicAcceptance Criteria (Implied by Predicate)Reported Device Performance (Proposed ExSpiron 1Xi)
    Accuracy
    Minute Ventilation (MV)10.7%11.5% (Overall)
    Tidal Volume (TV)10.4%11.4% (Overall)
    Respiratory Rate (RR)0.1%0.08% (Overall)
    Bias(Not explicitly stated for predicate)MV: 2.2%, TV: 2.2%, RR: 0.01% (Overall)
    Precision(Not explicitly stated for predicate)MV: 7.0%, TV: 6.9%, RR: 0.08% (Overall)
    Intended UseIdentical to predicateIdentical to predicate
    TechnologyThoracic bioimpedanceThoracic bioimpedance
    Volume MeasurementsTidal volume, Minute volume, Volume vs. time chartTidal volume, Minute volume, Volume vs. time chart
    Rate MeasurementsRespiratory rate (breaths/min)Respiratory rate (breaths/min)
    SafetyIEC 60601-1 compliantIEC 60601-1 compliant
    Energy SourceLow current signal at approximately 50 kHzLow current signal at approximately 50 kHz
    EMC/EMI ComplianceIEC 60601-1-2 compliantIEC 60601-1-2 compliant
    Electrode PadSetSingle-patient use, biocompatible, printed padsetSingle-patient use, biocompatible, printed padset
    Impedance Measurement Range15 Ohms to 180 Ohms15 Ohms to 180 Ohms
    BiocompatibilityElectrode PadSet is biocompatible; Monitor and Patient Cable not for patient contactElectrode PadSet is biocompatible; Monitor and Patient Cable not for patient contact
    UsabilityISO 60601-1-6 compliant (with calibration)ISO 60601-1-6 compliant (without calibration in Basic Monitoring mode)

    The document notes that the "algorithm was refined for the ExSpiron 1Xi using thousands of additional data points collected over the last 3 years in a wide variety of patient populations," leading to "more accurate respiratory volumes than the predicate." However, the accuracy percentages show a slight decrease in accuracy (higher percentage value) for MV and TV compared to the predicate in this summary document. This might be due to presenting overall clinical study results for the proposed device, while the predicate's numbers might be from a specific lab setting or a different calculation method for its initial clearance. The conclusion explicitly states the proposed device "is as safe and as effective as the predicate."

    2. Sample Size for the Test Set and Data Provenance

    • Sample Size: 20 subjects
      • Age range: 22-80
      • BMI range: 19-42
      • Gender: 9 female, 11 male
    • Data Provenance:
      • Country of Origin: Waltham, MA, USA (clinical study conducted there)
      • Retrospective or Prospective: Prospective. The study was conducted specifically to compare simultaneous measurements from the ExSpiron 1Xi and the predicate device.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    The document does not explicitly state the number of experts used to establish the ground truth for this clinical study, nor does it specify their qualifications. The study focused on comparing the performance of the proposed device against a predicate device through simultaneous measurements. This implies the predicate device's measurements or another recognized standard served as the comparative reference, rather than a separate "ground truth" established by experts for each breath.

    4. Adjudication Method for the Test Set

    The document does not describe an adjudication method for the test set. It states that simultaneous measurements were taken from both the proposed device and the predicate device, implying a direct comparison rather than a need for expert adjudication of independent measurements.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The study described is a clinical study comparing simultaneous measurements of the proposed device and a predicate device on patients, not a study evaluating human reader performance with and without AI assistance.

    6. Standalone Performance Study (Algorithm Only Without Human-in-the-Loop Performance)

    Yes, a standalone performance aspect is implied. The "Accuracy" section of the comparison table and the clinical study results (Bias, Precision, Accuracy) directly assess the performance of the device's algorithms in measuring respiratory parameters (MV, TV, RR) without explicitly mentioning human intervention in the real-time measurement process. The device provides "approximate values" of these parameters.

    7. Type of Ground Truth Used

    The ground truth for the clinical study was established by simultaneous measurements from a legally marketed predicate device (ExSpiron™ 1Xi, K130170). The study aimed to show equivalence or improvement compared to an already cleared device.

    8. Sample Size for the Training Set

    • The predicate device's algorithm was "trained with a dataset containing respiratory data from a large historical cohort."
    • The proposed device's algorithm was "refined for the ExSpiron 1Xi using thousands of additional data points collected over the last 3 years in a wide variety of patient populations." The exact numerical sample size ("thousands") is not specified beyond this qualitative description.

    9. How the Ground Truth for the Training Set Was Established

    The document does not explicitly detail how the ground truth for the "large historical cohort" or the "thousands of additional data points" used for algorithm training/refinement was established. However, given the nature of a spirometer, it is highly probable that the ground truth for training data would have been established using:

    • Reference Spirometry: Directly measured, highly accurate respiratory volumes and rates from calibrated gold-standard spirometry equipment.
    • Other established physiological measurement techniques that provide accurate respiratory parameters.

    The purpose of the training data is to allow the device's proprietary, non-linear, regressive model to accurately calculate respiratory parameters "without patient-specific calibration" for the proposed device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K130170

    Validate with FDA (Live)

    Device Name
    EXSPIRON 1XI
    Date Cleared
    2013-05-29

    (125 days)

    Product Code
    Regulation Number
    868.1850
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ExSpiron 1xi is indicated for use by healthcare professionals in healthcare facilities, such as post-operative care and critical care units, to monitor breathing in adult (at least 21 years old) patients.
    ExSpiron 1xi is a non-invasive system that graphically displays lung volume against time and reports an approximate value of:
    Tidal volume,
    Respiratory rate, and
    Minute ventilation
    ExSpiron 1xi measurements are used as an adjunct to other clinical information sources.

    Device Description

    The ExSpiron 1xi system consists of:
    Bioimpedance measurement system: A stabilized high frequency current generator is connected to two outer electrodes. The inner four electrodes are connected to an adaptive circuit that conditions the resulting voltage signal and converts it to digital form. Firmware performs signal acquisition and relays data to the panel PC.
    Computer: A Windows 7 PC performs signal processing and calibration, and runs the graphical user interface (GUI). The PC takes user input from a touch screen. The GUI is used for recording patient data and displaying the respiratory trace as well as scalar values and trends for minute volume, tidal volume, and respiratory rate.
    Single Patient Use ExSpiron™ 1xi Electrode PadSet: An electrode assembly containing six electrodes to be placed on the torso. It delivers current and records impedance measurements. The electrode PadSet is also used to perform subsystem checks prior to patient measurements.

    AI/ML Overview

    The ExSpiron™ 1xi respiratory monitor is indicated for use by healthcare professionals in healthcare facilities to monitor breathing in adult (at least 21 years old) patients. It is a non-invasive system that graphically displays lung volume against time and reports approximate values for Tidal Volume, Respiratory Rate, and Minute Ventilation. The measurements are used as an adjunct to other clinical information sources.

    The provided documentation does not contain a table of acceptance criteria or reported device performance for this specific 510(k) submission (K130170). The submission is a Special 510(k) where the manufacturer states that "No clinical testing was performed for this 510(k). The measurement algorithm and software are unchanged from the predicate. Measurement performance statistics were determined by clinical trials reported in the predicate 510(k), K120087." This means that the device's performance relies on the data submitted for its predecessor, the ExSpiron Respiratory Monitor (K120087).

    However, based on the information provided for K130170:

    1. Table of Acceptance Criteria and Reported Device Performance:

    As stated above, no specific acceptance criteria or reported performance for the ExSpiron™ 1xi is provided in this document as it references the predicate device. For the predicate device, K120087, the performance statistics would have been determined through clinical trials. Without access to K120087, these specific details cannot be provided here.

    The document states that changes for K130170 were related to:

    • Software Verification & Validation
    • Safety Testing - IEC 60601-1 2nd Edition
    • Electromagnetic Compatibility - IEC 60601-1-2
    • ExSpiron 1Xi System Test Plan
    • Patient Cable Design Verification (AAMI / ANSI EC53:1995/(R) 2008)
    • Electrode PadSet Design Verification (AAMI / ANSI EC12:2000/(R)2010, AAMI / ANSI EC53:1995/(R) 2008)
    • Electrode - Biocompatibility - ANSI/AAMI/ISO 10993-1

    All performance testing for these modifications was reported as successful, raising no new safety or efficacy concerns and demonstrating substantial equivalence to the predicate device.

    2. Sample Size for Test Set and Data Provenance:

    • Sample Size for Test Set: Not applicable for this specific 510(k) (K130170) as no new clinical testing was performed. The measurement performance statistics were determined by clinical trials reported in the predicate 510(k), K120087.
    • Data Provenance: Not applicable for this specific 510(k). For the predicate device (K120087), this information would be detailed in its submission. It's likely that the data would be prospective, given the nature of clinical trials for device performance. Country of origin not specified in the current document.

    3. Number of Experts and Qualifications for Ground Truth - Test Set:

    • Not applicable for this specific 510(k) (K130170) as no new clinical testing was performed. The ground truth for the predicate device's clinical trials (K120087) would have been established through a reference standard method, but the number and qualifications of experts are not described in this document.

    4. Adjudication Method for the Test Set:

    • Not applicable for this specific 510(k) (K130170) as no new clinical testing was performed.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • No MRMC comparative effectiveness study was done for this specific 510(k) (K130170). The device is a monitor that reports values, not an AI for interpretation requiring human readers.

    6. Standalone (Algorithm Only) Performance Study:

    • Yes, implicitly. The statement that "The measurement algorithm and software are unchanged from the predicate" implies that the core algorithm's performance was evaluated in a standalone manner during the clinical trials for the predicate device (K120087). This 510(k) (K130170) focuses on modifications to hardware components and software updates that do not alter the fundamental measurement algorithm.

    7. Type of Ground Truth Used:

    • For the performance statistics referenced from the predicate K120087, the ground truth for respiratory measurements (Tidal Volume, Respiratory Rate, Minute Ventilation) would typically be established using a gold standard reference measurement system (e.g., highly accurate spirometry, capnography, or volumetric respirators) rather than expert consensus on images or pathology. The document does not specify the exact ground truth method used in the predicate study.

    8. Sample Size for the Training Set:

    • Not applicable for this 510(k) (K130170). The clinical trials for the predicate device (K120087) would have involved a patient sample. The document does not provide this information.

    9. How Ground Truth for the Training Set Was Established:

    • Not applicable for this 510(k) (K130170). For the predicate device's clinical trials, the ground truth would have been established using a gold standard reference measurement system, as mentioned in point 7.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1