    Intended Use

    Endoform™ Reconstructive Template – PLGA is intended for use as a surgical mesh to reinforce and/or repair soft tissue where weakness exists. Indications for use include the repair of hernias and/or abdominal wall defects that require the use of reinforcing material to obtain the desired surgical outcome.

    Device Description

    Endoform™ Reconstructive Template – PLGA is a surgical mesh comprised of multiple layers of ovine-derived extracellular matrix reinforced with poly(lactic-co-Glycolic) acid (PLGA). The device design includes a range of shapes, sizes, and thicknesses in surface areas up to 400cm², to give a range of strengths as required for a particular implant procedure. Devices are terminally sterilized by ethylene oxide (EO) sterilization.

    AI/ML Overview

    This FDA 510(k) clearance letter and supporting summary for the Endoform™ Reconstructive Template – PLGA device does not involve artificial intelligence (AI) or machine learning (ML). Therefore, many of the requested elements for AI/ML device validation are not applicable.

    The device is a traditional medical device (surgical mesh) and its substantial equivalence is demonstrated through non-clinical performance data and comparison to a predicate device. There is no mention of an algorithm, AI assistance, or human readers.

    However, I can extract the information relevant to its acceptance criteria and the study that proves it meets those criteria, based on the provided document.


    Acceptance Criteria and Study Proof for Endoform™ Reconstructive Template – PLGA

    Since this is a traditional medical device (surgical mesh) and not an AI/ML device, the concept of "acceptance criteria" is related to meeting performance specifications and demonstrating biocompatibility and safety comparable to a predicate device, rather than diagnostic accuracy metrics. The "study" refers to non-clinical testing.

    Here's a breakdown based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document describes the types of tests conducted and the conclusion that the device meets specifications, rather than providing specific numerical acceptance criteria and results in a table format common for AI/ML.

    For this device, the "acceptance criteria" are implied by the performance characteristics demonstrated to be "substantially equivalent" to the predicate device and meeting recognized standards.

    Acceptance Criteria CategoryDescription (Implied/Direct)Reported Device Performance (Summary)
    Material CompositionConsistency with ovine-derived extracellular matrix with PLGA reinforcement; comparable to predicate's ovine-derived ECM with PGA.Comprised of multiple layers of ovine-derived extracellular matrix reinforced with poly(lactic-co-Glycolic) acid (PLGA). "Maintains the same fundamental technological characteristics as the predicate device with respect to material types."
    Physical DimensionsRange of shapes, sizes, and thicknesses for surface areas up to 400cm².Designs include a range of shapes, sizes, and thicknesses in surface areas up to 400cm². "The subject device is offered in the same range of shapes, sizes, and thicknesses as the predicate device."
    Sterilization Method & SALEthylene Oxide sterilization with a Sterility Assurance Level (SAL) of 10⁻⁶."Devices are terminally sterilized by ethylene oxide (EO) sterilization." SAL of 10⁻⁶ achieved.
    Endotoxin ContentEndotoxin content less than 20 EU/device.Endotoxin Content: <20 EU/device.
    PackagingDouble Tyvek 2FS and PET/PE pouch inside a cardboard zip-style envelope."Double Tyvek 2FS and PET/PE pouch inside a cardboard zip-style envelope" (same as predicate).
    UsageSingle Use."Single Use" (same as predicate).
    Use SettingFor use in a surgical setting by or under the supervision of a medical professional."For use in a surgical setting by or under the supervision of a medical professional" (same as predicate).
    BiocompatibilityCompliance with ISO 10993-1, assuring no unacceptable risk to patients/users."Biocompatibility assessment and testing was performed in accordance with the FDA Guidance Use of International Standard ISO 10993-1... The results of all testing and the accompanying biocompatibility assessments have demonstrated that Endoform™ Reconstructive Template - PLGA devices are safe for use and will not cause any unacceptable risk to patients or users when used as intended."
    Mechanical StrengthAdequate mechanical properties for intended use (reinforce/repair soft tissue).Non-clinical testing included "mechanical strength," "suture retention," and "elastic modulus." "Results of testing confirm that the proposed device meets all product specifications and supports substantial equivalence to the predicate device."
    Other PerformanceCompliance testing, collagen differential scanning calorimetry, bioburden.Non-clinical testing included "collagen differential scanning calorimetry," "endotoxin," "bioburden," and "compliance testing." "Results of testing confirm that the proposed device meets all product specifications."

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify a "test set" in the context of an AI/ML algorithm evaluation with patient data. Instead, it refers to non-clinical performance data.

    • Sample Size: Not explicitly stated for each non-clinical test (e.g., number of devices tested for mechanical strength, biocompatibility, etc.). These are typically described in a detailed test report.
    • Data Provenance: Not applicable in the context of patient data. The "data" comes from in vitro and in vivo (animal) lab testing according to recognized standards (e.g., ISO 10993-1) for medical device performance and safety.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    • Not applicable. This is not an AI/ML diagnostic device requiring human expert ground truth for imaging or clinical assessments. The ground truth for device performance is based on established engineering and materials science principles, and compliance with recognized standards.

    4. Adjudication Method for the Test Set

    • Not applicable. This is not an AI/ML diagnostic device where human expert disagreement needs adjudication for a test set.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • Not applicable. This is a physical medical device, not an AI/ML diagnostic tool that assists human readers. No MRMC study was done.
    • The document explicitly states: "Substantial equivalence was not based on an assessment of clinical performance data."

    6. Standalone (Algorithm Only) Performance Study

    • Not applicable. This is not an AI/ML algorithm.

    7. Type of Ground Truth Used

    • Physical and Chemical Properties: The "ground truth" for this device's performance relies on established scientific and engineering principles for material properties (e.g., tensile strength, elasticity), biocompatibility (e.g., cellular response, toxicity), and sterility. This is typically established through:
      • Validated Test Methods: Adherence to international standards (e.g., ISO 10993 for biocompatibility).
      • Comparative Analysis: Demonstrating that the subject device's properties are equivalent or superior to a legally marketed predicate device with a known safety and effectiveness profile.

    8. Sample Size for the Training Set

    • Not applicable. This is not an AI/ML device that requires a training set of data.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable. As there's no training set for an AI/ML model, there's no ground truth established in that context.
    Ask a Question

    Ask a specific question about this device

    K Number
    K181935

    Validate with FDA (Live)

    Date Cleared
    2018-12-04

    (138 days)

    Product Code
    Regulation Number
    878.3300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Endoform Reconstructive Template – Non-Absorbable is intended for use as a surgical mesh to reinforce and/or repair soft tissue where weakness exists.

    Endoform Reconstructive Template – Non-Absorbable is indicated for use in the repair of hernias and/or abdominal wall defects that require the use of reinforcing or bridging material to obtain the desired surgical outcome.

    Device Description

    Endoform Reconstructive Template - Non-Absorbable is a surgical mesh manufactured by layering sheets of ovine forestomach matrix to create multi-layer configurations of devices laminated with non-resorbable polypropylene (PP) suture material in a variety of sizes for use in soft tissue reconstruction. Devices up to 400 cm² are available in thicknesses from 1- through 10- ply, and larger devices up to 1000cm2 are available in 4-, 6-, and 8-ply presentations.

    AI/ML Overview

    This document describes the Endoform Reconstructive Template - Non-Absorbable, a surgical mesh. The submission (K181935) is a 510(k) premarket notification for a line extension to increase the maximum device size.

    Here's the information about the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (Performance Specifications)Reported Device Performance
    Material compositionRemains the same as predicate
    BiocompatibilityRemains the same as predicate
    SterilizationRemains the same as predicate
    Packaging materials and processesRemains the same as predicate
    Mechanical StrengthMeets all product specifications for the intended use
    Endotoxin levelsMeets all product specifications for the intended use
    Dimensional verificationMeets all product specifications for the intended use
    Performance (general)Performance specifications have not changed and are met

    The document states that the performance specifications for the device have not changed and that the device meets these specifications. The specific quantitative values of these specifications are not provided in this summary.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    • Sample size for test set: Not explicitly stated. The document refers to "Bench testing and validation" and "results of verification and validation testing" without providing specific sample numbers for each test.
    • Data provenance: Not explicitly stated, but the company is Aroa Biosurgery Ltd. located in New Zealand. These would be non-clinical, laboratory-based tests.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    Not applicable. This device is a surgical mesh, and the testing described is non-clinical (bench testing) for physical properties, not a clinical study involving radiologists or medical image interpretation.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Not applicable. Adjudication methods are typically for clinical studies involving multiple reviewers/experts, which is not the case here.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI/software device, and no MRMC study was conducted.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a physical medical device (surgical mesh), not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the non-clinical performance data, the "ground truth" would be established by:

    • Engineering specifications and standards for mechanical strength and dimensional properties.
    • Validated laboratory assays for endotoxin levels and biocompatibility (as carried over from the predicate device).

    8. The sample size for the training set

    Not applicable. There is no concept of a "training set" for this type of physical device testing.

    9. How the ground truth for the training set was established

    Not applicable. There is no training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K153632

    Validate with FDA (Live)

    Date Cleared
    2016-06-16

    (181 days)

    Product Code
    Regulation Number
    878.3300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Endoform® Reconstructive Template is intended for use as a surgical mesh to reinforce and/ or repair soft tissue where weakness exists. Indications for use include the repair of hernias and/or abdominal wall defects that require the use of reinforcing or bridging material to obtain the desired surgical outcome.

    Device Description

    Endoform Reconstructive Template - Non-Absorbable is a surgical mesh manufactured by layering sheets of ovine forestomach matrix (OFM) to create 1- through 10- ply embroidered devices for soft tissue reconstruction. The device design includes thicknesses from 1- through 10- ply to give a range of strengths as required for a particular implant procedure. The construction of Endoform™ Reconstructive Template - Non-Absorbable devices includes lyophilization of single sheets of OFM followed by embroidery with polypropylene. Devices are terminally sterilized by ethylene oxide (EO) sterilization.

    AI/ML Overview

    This document is an FDA 510(k) premarket notification decision letter for a surgical mesh device, Endoform® Reconstructive Template - Non-Absorbable. It does not describe an AI medical device or a study involving AI. Therefore, it is impossible to answer the questions about acceptance criteria and study proving an AI device meets those criteria based on the provided text.

    The document discusses the substantial equivalence of the surgical mesh to a predicate device based on:

    • Biocompatibility Data: Additional cytotoxicity testing, toxicological assessment, and viral inactivation assessment were performed.
    • Performance Data: Biomechanical testing (uniaxial tensile strength, ball burst strength, suture retention strength) and an in-vivo study in a soft tissue reinforcement model.
    • Clinical Data: No clinical data was submitted.

    The provided text pertains to a physical medical device (surgical mesh) and its substantial equivalence determination, not an AI/ML medical device. Therefore, the questions about acceptance criteria for an AI device, sample sizes for AI test/training sets, experts for ground truth, adjudication methods, MRMC studies, standalone performance, and ground truth establishment for AI models are not applicable to the content of this document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K153633

    Validate with FDA (Live)

    Date Cleared
    2016-03-15

    (88 days)

    Product Code
    Regulation Number
    878.3300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Endoform Reconstructive Template is intended for usc as a surgical mesh to reinforce and/or repair soft tissue where weakness exists. Indications for use include the repair of hernias and/or abdominal wall defects that require the use of reinforcing or bridging material to obtain the desired surgical outcome.

    Device Description

    The Aroa Endoform Reconstructive Template ("ERT") is a surgical mesh manufactured by layering sheets of ovine forestomach matrix (OFM) to create 1- through 10- ply embroidered devices for implant applications as part of soft tissue reconstruction. The device design includes a range of thicknesses from 1- through 10- ply to give a range of strengths as required for a particular implant procedure. The construction of ERT devices includes lyophilization of single sheets of OFM followed by embroidery with polyglycolic acid (PGA). Devices are terminally sterilized by ethylene oxide (EO) sterilization. The device has received FDA clearance (K130547) for sizes up to 200cm2.

    AI/ML Overview

    This document is a 510(k) premarket notification for a medical device called the "Endoform® Reconstructive Template" (ERT), which is a surgical mesh. The purpose of this notification is to declare that a modified version of the ERT is substantially equivalent to a previously cleared predicate device. The only change proposed is an increase in the maximum device size.

    Here's an analysis of the acceptance criteria and study information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Measured Parameter)Reported Device Performance (Goal)Study Finding
    Bench Testing"meets all product specifications"
    Mechanical StrengthAdequate for intended useEnsured through bench testing
    EndotoxinWithin acceptable limitsEnsured through bench testing
    Dimensional VerificationAccurate dimensionsEnsured through bench testing
    Validation Testing"meets all product specifications"
    Manufacturing Process ValidationRobust and consistentAchieved
    Labeling ValidationAccurate and compliantAchieved
    Packaging Transportation ValidationMaintains integrity during transitAchieved
    Sterilization (SAL)10⁻⁶Achieved

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state specific numerical sample sizes for each test set (e.g., how many devices were tested for mechanical strength, endotoxin, etc.). It only broadly refers to "bench testing and validation testing."

    • Test Set Sample Size: Not explicitly stated (e.g., number of units tested per bench test).
    • Data Provenance: The studies were conducted by Aroa Biosurgery, Limited, which is based in New Zealand. The studies appear to be prospective for the modified device based on the phrase "Performance and validation testing executed based on risk analysis of the proposed change supports the substantial equivalence..."

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This information is not applicable in this context. The "acceptance criteria" here refer to engineering and manufacturing specifications (e.g., mechanical strength, sterility, dimensions) for a physical medical device. Ground truth, in the context of expert consensus, is typically relevant for diagnostic devices or those requiring human interpretation of results. For this surgical mesh, the "ground truth" is defined by the established engineering standards and specifications.

    4. Adjudication Method for the Test Set

    This information is not applicable. Adjudication methods (like 2+1, 3+1) are used to resolve disagreements among multiple human readers or experts when establishing a ground truth, often for image interpretation or clinical diagnosis. Since the "acceptance criteria" for this device are based on objective physical and chemical measurements (bench testing, sterility), there is no human interpretation or adjudication involved in meeting these criteria.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The document explicitly states: "No clinical data was submitted to support the safety and effectiveness of the modified Endoform Reconstructive Template." MRMC studies inherently involve human readers and clinical outcomes.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This question is not applicable. A "standalone" performance typically refers to the performance of an algorithm or AI system without human intervention. The Endoform Reconstructive Template is a physical surgical mesh, not an algorithm or AI-powered device.

    7. The Type of Ground Truth Used

    The "ground truth" (or basis for acceptance) for this device's performance is based on established product specifications, engineering standards, and regulatory requirements. This includes:

    • Pre-defined mechanical strength targets.
    • Specified endotoxin limits.
    • Dimensional tolerances.
    • Sterility Assurance Level (SAL) of 10⁻⁶.
    • Biocompatibility as demonstrated by the predicate device.

    8. The Sample Size for the Training Set

    This information is not applicable. The concept of a "training set" is relevant for machine learning or AI models. The Endoform Reconstructive Template is a physical medical device. Its development and testing involve manufacturing processes, material science, and engineering validation, not an AI training paradigm.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable for the same reasons as point 8. There is no "training set" in the context of this device's development.

    Ask a Question

    Ask a specific question about this device

    K Number
    K130547

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2013-05-31

    (88 days)

    Product Code
    Regulation Number
    878.3300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Endoform™ Reconstructive Template is intended for use as a surgical mesh to reinforce and/or repair soft tissue where weakness exists. Indications for use include, but are not limited to the following procedures; hernioplasty and repair of body wall defects. The device allows reinforcement or bridging of a deficit to obtain the desired surgical outcome. Endoform™ Reconstructive Template is intended for single use only.

    Device Description

    The Mesynthes Endoform™ Reconstructive Template (ERT) is a surgical mesh comprising either a single layer of decellularized extracellular matrix (ECM) or alternatively, multiple layers (2- to 10- layers) of ECM bonded together by dehydrothermal lamination and polyglycolic acid (PGA) suture material.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the Mesynthes Endoform™ Reconstructive Template surgical mesh. This type of submission relies on demonstrating substantial equivalence to a legally marketed predicate device, rather than explicit acceptance criteria and performance against those criteria in the way a novel device might.

    Therefore, the information you're requesting regarding "acceptance criteria" and "study that proves the device meets the acceptance criteria" in terms of specific performance metrics, sample sizes for test sets, expert ground truth, adjudication methods, MRMC studies, or standalone algorithm performance does not apply directly to this 510(k) summary.

    Instead, the submission focuses on demonstrating that the new device has "technological characteristics and substantial equivalence" to existing predicate devices.

    Here's an breakdown based on the information provided, keeping in mind the nature of a 510(k) submission:

    1. Table of Acceptance Criteria and Reported Device Performance

    As this is a 510(k) submission based on substantial equivalence, there are no explicit "acceptance criteria" in the format of defined metrics and targets for clinical performance that the device directly "meets." Instead, the acceptance is based on demonstrating similarity to predicate devices through a series of non-clinical performance tests and animal testing.

    The "performance" described is about demonstrating similarity or equivalence to predicates, rather than achieving specific quantitative clinical thresholds.

    Test TypeComparison/Performance Description
    Non-Clinical Biophysical TestsThe Endoform™ Reconstructive Template performed similarly to its predicate devices in:
    - Ball burst strength
    - Uniaxial strength
    - Suture retention strength
    Biocompatibility TestsThe product meets biocompatibility requirements of the ISO standard (ISO10993-1) for:
    - Cytotoxicity
    - Sensitization
    - Irritation
    - Systemic Toxicity
    - Subacute Toxicity
    - Genotoxicity (Ames, chromosomal aberration and mouse lymphoma)
    - Implantation
    Animal Testing (In vivo studies)The performance was substantially equivalent to the predicate device based on:
    - Incorporation into new tissue
    - Inflammatory response
    - Strength of the graft

    2. Sample size used for the test set and the data provenance

    The document does not specify the exact sample sizes for the individual biophysical, biocompatibility, or animal tests. It only states that these tests were "performed on finished, terminally sterilized product" and in "a model of soft tissue reinforcement" for animal studies. The data provenance is implied to be from internal lab testing by Mesynthes Ltd.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable/not provided in the context of this 510(k) submission. "Ground truth" established by experts is typically relevant for studies validating diagnostic algorithms or clinical evaluations, which were not conducted here as per Section 9.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable/not provided. Adjudication methods are typically relevant for clinical studies involving human interpretation or subjective assessments, which were not part of this 510(k) submission.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    A multi-reader multi-case (MRMC) comparative effectiveness study was not done. This type of study focuses on the impact of a device on human performance, often in diagnostic imaging, which is not relevant for a surgical mesh device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This concept is not applicable to a surgical mesh device. Standalone performance refers to the accuracy of an automated system without human intervention, typically in AI/diagnostic applications.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the non-clinical and animal studies, "ground truth" implicitly refers to the established scientific and engineering standards for measuring device properties (e.g., strength, biocompatibility), and histopathological or physiological assessment in animal models for tissue integration and inflammatory response. These are not typically established via expert consensus in the same way as clinical diagnostic accuracy.

    8. The sample size for the training set

    This information is not applicable/not provided. "Training set" is a concept related to machine learning and AI development, which is outside the scope of this medical device submission.

    9. How the ground truth for the training set was established

    This information is not applicable/not provided for the same reasons as point 8.