Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K983940

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1999-07-15

    (252 days)

    Product Code
    Regulation Number
    870.1340
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Adam Spence Percutaneous Catheter Introducer is intended to provide a means for percutaneous vascular access while minimizing the back flow of blood during the introduction of cardiovascular devices such as catheters and guide wires.

    Device Description

    The Adam Spence Corp. Percutaneous Catheter Introducer (PCI) is a disposable device intended for use in diagnostic angiographic procedures.

    AI/ML Overview

    Here's an analysis of the provided text regarding the Adam Spence Percutaneous Catheter Introducer, focusing on acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Feature/Test)Reported Device Performance (Adam Spence PCI)Predicate Device Performance (Daig Fast-Cath™)Acceptance Threshold/Standard (Implied/Explicit)
    Tensile PropertiesExceeds RequirementExceeds RequirementManufacturer's internal requirement (not specified quantitatively)
    Valve/Sheath LeakNo Leakage4 Fr. leaked after dilationNo leakage
    Insertion/WithdrawalSubstantially Equivalent to PredicateAcceptablePerformance comparable to predicate device
    Air AspirationGreatly Exceeded Predicate DeviceAcceptableGreater or equivalent to predicate device
    Tip PeelbackNo Damage or Peelback; Lower average insertion force required to puncture skin than predicate device.No Damage or Peelback; Higher average insertion force required to puncture skin than ASC PCI.No damage/peelback; Insertion force comparable to or better than predicate.
    Sterile PackagingSterile and Bulk PackagedSterile PackSterile
    Sterilization MethodEthylene Oxide GasEthylene Oxide GasEthylene Oxide Gas
    Shelf Life3 Years3 Years3 Years
    Available Sizes4 though 9 French size4 through 9 French size4 though 9 French size
    BiocompatibilityAll appropriate biocompatibility tests have been performed per ISO 10993(Not directly stated for predicate)ISO 10993 compliance

    Notes on Acceptance Criteria:

    • Many acceptance criteria are comparative against a predicate device rather than absolute quantitative standards.
    • The document explicitly states that "Performance standards have not been established under Section 514 of the Food, Drug and Cosmetic Act."
    • Referenced standards like "ANSI MD70.1-1983" and "ISO/DIS 11070.2" likely define specific physical and performance requirements that the tests above would then demonstrate compliance with, but the specific numerical thresholds from these standards are not detailed in the provided text.

    2. Sample Sizes and Data Provenance

    The provided document describes a summary of technological characteristics and performance testing for the Adam Spence Percutaneous Catheter Introducer compared to a predicate device. However, it does not provide specific details on the sample sizes used for the test set or the data provenance (e.g., country of origin, retrospective/prospective nature). The testing appears to be primarily bench testing (laboratory-based) comparing physical properties and performance aspects rather than clinical trials with patient data.

    3. Number of Experts and Qualifications for Ground Truth

    The document does not mention the use of experts to establish ground truth for a test set. The evaluation is based on engineering and performance testing.

    4. Adjudication Method

    No adjudication method is mentioned, as the evaluation is based on objective performance testing rather than expert-based assessments requiring adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study was done. The device is a physical medical device (catheter introducer), not an imaging or diagnostic AI tool that would typically involve human readers.

    6. Standalone Performance Study

    Yes, a standalone performance evaluation was done for the Adam Spence PCI. The "Summary of Technological Characteristics" table presents the performance of the Adam Spence PCI (algorithm/device only) in various tests (e.g., tensile properties, valve/sheath leak, air aspiration, tip peelback) independent of human-in-the-loop performance, though often in direct comparison to the predicate device.

    7. Type of Ground Truth Used

    The "ground truth" for the performance claims appears to be established through engineering measurements, physical testing, and laboratory experiments (e.g., measuring tensile strength, observing leakage, quantifying insertion force). For biocompatibility, it's compliance with established ISO standards.

    8. Sample Size for the Training Set

    The concept of a "training set" is not applicable here as the device is a physical medical device, not an AI/machine learning algorithm.

    9. How Ground Truth for the Training Set Was Established

    Not applicable for a physical medical device as there is no "training set" in the machine learning sense. The device's design and manufacturing are based on established engineering principles and materials science, with performance validated through direct testing against predetermined specifications or predicate device performance.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1