Search Filters

Search Results

Found 5 results

510(k) Data Aggregation

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Emprint™ ablation system is intended for use in percutaneous, and intraoperative coagulation (ablation) of soft tissue, including partial or complete ablation of non-resectable liver tumors.

    The Emprint™ ablation system is not intended for use in cardiac procedures.

    The Emprint™ ablation system overlapping technique is only intended for use in the liver.

    Device Description

    The Emprint™ Ablation System with Thermosphere™ Technology, Covidien's microwave ablation system, was released for commercial distribution in the United States in April 2014.

    The Emprint™ Ablation System is a microwave-based ablation system intended to deliver energy through an antenna inserted into soft tissue for the purpose of coagulating (ablating) a defined volume of that tissue. The Emprint™ Ablation System utilizes a 2450 MHz 100W, or 150W generator to deliver power to a single microwave ablation antenna. The Emprint™ Ablation 100W Generator is composed of analog and digital circuits with no software or firmware. The Emprint™ Ablation 150W Generator utilizes software and firmware, however both generators' settings are controlled in the same fashion. Emprint™ Ablation Generators provide for user setting of ablation time (0-10 minutes) and ablation power (5 to 150W). With an optional temperature probe, the ablation generators can be set to monitor the temperature of a desired target and to automatically shut the generator off when the target reaches a pre-set temperature.

    The Emprint™ Ablation System uses circulating room temperature normal saline to cool the non-radiating portion of the antenna shaft and to provide a more consistent ablation zone. The normal saline is pumped from an IV bag through the antenna shaft and back to the IV bag in a closed system. No saline is in contact with the patient.

    The 510(k) Cleared Emprint™ Ablation System consists of the following components:

    • Emprint™ Ablation Generator (2450 MHz) ●
    • Emprint™ HP Ablation Generator (2450 MHz) .
    • EmprintTM Percutaneous Antenna (sterile, single use) ●
    • Emprint™ Ablation Reusable Cable ●
    • Emprint™ Ablation Pump ●

    The system also includes the following optional equipment/accessories:

    • . Emprint™ Ablation Cart (with Isolation Transformer)
    • Emprint™ HP Ablation Cart (with Isolation Transformer)
    • Ablation Footswitch ●
    • Remote Temperature Probe (sterile, single use) ●

    The system must be used with a standard IV bag of sterile normal saline (not provided with the system).

    AI/ML Overview

    The provided text does not contain information about the acceptance criteria or a study that proves a device (specifically an AI/machine learning device) meets acceptance criteria, as the Emprint Ablation System is a microwave ablation system, not an AI/machine learning device. The 510(k) submission for the Emprint Ablation System describes updates to the instructions for use (IFU) to include "overlapping" and "ramp-up" ablation techniques, which were already observed in published literature and clinician practice.

    Therefore, I cannot fulfill the request to describe the acceptance criteria and study proving the device meets them for an AI/machine learning device based on the given input, as the device in question is a traditional medical device and the submission focuses on updating its IFU, not on AI/ML performance.

    The "Performance Data" section (9.9) explicitly states:

    • "Clinical Studies in human subjects were not required to demonstrate the performance and safety of the Emprint Ablation System with Thermosphere Technology." (9.9.2)
    • "Studies or testing in animal subjects were not required to demonstrate the performance and safety of the Emprint Ablation System with Thermosphere Technology." (9.9.3)
    • Bench testing was conducted to generate zone charts for updated techniques (9.9.1). This is a standard ex-vivo engineering characterization, not a clinical study involving AI/ML methods or human readers.

    None of the points outlined in the prompt (sample size for test set, data provenance, number of experts for ground truth, adjudication method, MRMC study, standalone performance, type of ground truth, training set sample size, ground truth for training set) are relevant or discussed in the provided document, as it pertains to a physical medical device and not an AI/ML system.

    Ask a Question

    Ask a specific question about this device

    K Number
    K200796

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-08-26

    (153 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Emprint™ Ablation System is intended for use in percutaneous, laparoscopic, and intraoperative coagulation (ablation) of soft tissue, including partial or complete ablation on nonresectable liver tumors.

    The Emprint™ Ablation System is not intended for use in cardiac procedures

    Device Description

    The Emprint™ Ablation System with Thermosphere™ Technology, Covidien's microwave ablation system, was released for commercial distribution in the United States in April 2014.

    The 510(k) Cleared Emprint™ Ablation System consists of the following components:

      1. Emprint™ Ablation Generator (2450 MHz)
      1. Emprint™ Percutaneous Antenna (sterile, single use)
      1. Emprint™ Ablation Reusable Cable
      1. Emprint™ Ablation Pump

    The system also includes the following optional equipment/accessories:

      1. Emprint™ Ablation Cart (with Isolation Transformer)
      1. Ablation Footswitch
      1. Remote Temperature Probe (sterile, single use)

    The current Emprint™ Ablation System utilizes a 2450 MHz 100W generator (CAGEN1) to deliver power to a dedicated single microwave ablation antenna. The optional cart, Emprint™ Ablation Cart is intended to be used with the Emprint™ Ablation Generator.

    The new system component, Emprint™ HP Ablation Generator (CAGENHP) [subject device] is similar to the device, Emprint™ Ablation Generator (CAGEN1) [predicate device] as described in K193232 with two exceptions. The new system component, Emprint™ HP Ablation Generator (CAGENHP), will extend the power setting range from 5 to 100W (Emprint™ Ablation Generator - CAGEN1) to 5 to 150W. The increased power will allow clinicians to create larger ablation zones when ablating soft tissue lesions, compared to use of the system with the existing 100-watt microwave generator (CAGEN1). Further the Emprint™ HP Ablation Generator (CAGENHP) is composed of analog and digital circuits with software whereas the predicate Emprint™ Ablation Generator (CAGEN1), is composed of analog and digital circuits with no software or firmware. The Emprint™ HP Generator will be compatible with all current Emprint™ ablation system components and accessories except the cart. The new optional Emprint™ HP Cart (CARTHP) has been added to the system and is compatible with the Emprint™ HP Ablation Generator (CAGENHP).

    The optional cart holds the Emprint™ HP Generator (CAGENHP) and the Emprint™ Ablation Pump (CAPUMP1) securely in one location. The optional footswitch can be placed on the cart's lower shelf. There are hooks on both sides of the cart which are meant to hold either the normal saline bag during a procedure, or to loop (and store) the reusable cable over while not in use.

    Emprint™ HP Ablation Cart (CARTHP) is similar to the previously cleared (K193232) Emprint™ Ablation Cart (CART1), which holds the Emprint™ Generator (CAGEN1) and the Emprint™ Ablation Pump (CAPUMP1) securely in one location.

    AI/ML Overview

    The provided document is a 510(k) summary for the Covidien Emprint™ Ablation System with Thermosphere™ Technology. It describes a submission for new components, specifically the Emprint™ HP Ablation Generator (CAGENHP) and its optional accessory, the Emprint™ HP Ablation Cart (CARTHP).

    This document does not describe an AI medical device or a study proving its performance against acceptance criteria for an AI or imaging-based diagnostic device. Instead, it focuses on demonstrating substantial equivalence of a microwave ablation system (a physical medical device used for soft tissue coagulation) to a predicate device.

    Therefore, many of the questions asked in the prompt, such as those related to AI-specific performance criteria (e.g., sample size for training/test sets, expert consensus, MRMC studies, effect size of AI assistance), are not applicable to the content of this document.

    However, I can extract information relevant to the device's acceptance criteria and the studies performed to prove its substantial equivalence as described in the document.


    Analysis of Device Acceptance Criteria and Performance (Based on the provided non-AI device document):

    Since the document describes a microwave ablation system and not an AI/imaging diagnostic device, the acceptance criteria and supporting studies are focused on the safety and effectiveness of the physical device and its new components, primarily demonstrating substantial equivalence to a predicate device.

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't present a formal "acceptance criteria" table in the way one might for an AI device's performance metrics (e.g., sensitivity, specificity thresholds). Instead, the "acceptance criteria" are implied by the various tests and verification activities conducted to demonstrate the new components are safe and effective and do not alter the system's fundamental performance compared to the predicate.

    Implied Acceptance Criterion / Test ObjectiveReported Device Performance / Outcome (Conclusion)
    Bench Verification Testing:
    1. System meets all design requirements.Demonstrated.
    2. Ablation performance design requirements met; equivalent ablation zone dimensions at identical power settings (HP Generator vs. Predicate Generator).Demonstrated that the ablations zone dimensions created with the Emprint™ HP Ablation Generator (subject) and Emprint™ Ablation Generator (predicate) are equivalent for identical power settings.
    3. Compatibility with previously approved Emprint™ Ablation System accessories (except cart).Verified as compatible. (Note: New HP Cart is compatible with HP Generator).
    4. Meets IEC 60601-1-2:2014 EMC regulatory requirements.Met.
    5. Meets IEC 60601-1:2005 and IEC 60601-2-6:2016 safety requirements.Met.
    6. Performs as specified within storage and use conditions.Verified.
    7. Usability of Emprint™ HP Ablation Generator.Verified.
    8. Packaging protects generator and cart during simulated transportation.Verified.
    Animal Testing (In Vivo Porcine Model):
    Equivalent performance in terms of ablation zone creation relative to predicate control device.Demonstrated equivalent performance and procedural safety of the Emprint™ Ablation System with CAGENHP generator in an in vivo porcine model undergoing soft tissue ablation procedures.
    Overall Substantial Equivalence:
    Differences (extended power range, software control) do not impact safety and effectiveness.Performance data demonstrate that these differences do not impact the safety and effectiveness of the device. (Implies all tests support this).
    Addition of optional accessory (HP Ablation Cart) has no impact on safety and effectiveness.Verification and validation activities conducted have shown that the addition of the optional accessory, Emprint™ HP Ablation Cart, has no impact on the safety and effectiveness of the Emprint™ Ablation System. (Implies all tests support this).

    The following points are not applicable as the document describes a physical medical device, not an AI/imaging diagnostic device. I'll explicitly state "Not Applicable" or "Not Provided/Relevant" for these.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size for Test Set: Not explicitly specified for discrete tests. For the animal study, it mentions "an in vivo porcine model" but doesn't give a specific number of animals. For bench testing, typical engineering verification involves numerous tests without a "sample size" in the clinical trial sense.
    • Data Provenance (Country of origin, retrospective/prospective): Not specified. The animal study is stated as "GLP-compliant pivotal study," suggesting a prospective, controlled experimental design, but location is not mentioned.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    • Not Applicable. This is not an AI/diagnostic device where human expert "ground truth" for image interpretation would be established. The ground truth for this device is based on physical measurements (e.g., ablation zone dimensions), engineering specifications, and safety standards.

    4. Adjudication Method for the Test Set:

    • Not Applicable. No human interpretation for "ground truth" to be adjudicated.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

    • No. This is not a diagnostic device involving human readers.
    • Effect size of human readers improve with AI vs. without AI assistance: Not Applicable.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Not Applicable. This is a hardware device, not an algorithm.

    7. The Type of Ground Truth Used:

    • For the ablation performance, the "ground truth" appears to be physical measurements of ablation zone dimensions (compared directly between the subject and predicate devices under identical power settings) and compliance with engineering design requirements, international safety and EMC standards (IEC 60601-1, IEC 60601-1-2, IEC 60601-2-6), and preclinical (animal) study results demonstrating equivalent therapeutic effect and safety.

    8. The Sample Size for the Training Set:

    • Not Applicable. This is not an AI device trained on a dataset.

    9. How the Ground Truth for the Training Set Was Established:

    • Not Applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K193232

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-02-13

    (80 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Emprint™ Ablation System is intended for use in percutaneous, laparoscopic, and intraoperative coagulation (ablation) of soft tissue, including partial or complete ablation of nonresectable liver tumors.

    The Emprint™ Ablation System is not intended for use in cardiac procedures.

    The Emprint™ SX Ablation Platform with Thermosphere™ Technology is intended for use in percutaneous, laparoscopic, and intraoperative coagulation (ablation) of soft tissue, including partial or complete ablation of nonresectable liver tumors.

    The Emprint™ SX Ablation Platform is not intended for use in cardiac procedures.

    The system's optional, 3-D navigation feature assists in the placement of the Emprint™ SX Navigation Antenna with Thermosphere™ Technology using real-time image guidance during intraoperative and laparoscopic ablation procedures. The navigation feature enhances the output of a compatible medical ultrasound imaging system and displays an image of the antenna and its predicted traiectory on a computer monitor. The size and shape of the predicted ablation zone relative to the position of the antenna is displayed on the enhanced ultrasound image.

    Device Description

    The Emprint™ Ablation System with Thermosphere™ Technology, Covidien's microwave ablation platform, was released for commercial distribution in the United States in April 2014. The 510(k) Cleared Emprint™ Ablation System consists of the following components: 1. Emprint™ Ablation Generator (2450 MHz), 2. Emprint™ Percutaneous Antenna (sterile, single use), 3. Emprint™ Ablation Reusable Cable, 4. Emprint™ Ablation Pump. The system also includes the following optional equipment/accessories: 5. Emprint™ Ablation Cart, 6. Ablation Footswitch, 7. Isolation Transformer, 8. Remote Temperature Probe (sterile, single use). The ablation platform must be used with a standard IV bag of sterile, normal saline (not provided with the device).

    The Emprint™ SX Ablation Platform with Thermosphere™ Technology is a microwave ablation system with an optional navigation feature. It relies on the same principles of operation, fundamental technology, and performance characteristics as its predicate, the Emprint™ Ablation System with Thermosphere™ Technology (K133821, K163105). The main differences between the Emprint™ SX Ablation Platform with Thermosphere™ Technology and the predicate device include software (vs. manual) control of the generator and pump, as well as the addition of an optional navigation feature. Both systems include a 2.45 GHz microwave generator, a cooling pump, and accessories. A bag/bottle of sterile fluid that connects to the antenna's tubing and to the peristaltic pump is required for proper system operation. The 510(k) Cleared Emprint™ Percutaneous Antenna (K163105) can be used with either system to create predictable, spherical ablation zones in soft tissue in percutaneous, laparoscopic, and intraoperative procedures.

    The Emprint™ antenna, the applied part of the Emprint™ ablation system with Thermosphere™ Technology and Emprint™ SX ablation platform with Thermosphere™ Technology that delivers microwave energy to the patient, has a sharp, engineered ceramic tip (called "trocar") at the distal end of the device that allows a user to puncture tissue. The shaft of the disposable antenna is inserted directly into the tissue intended to be ablated. The radiating tip of the antenna shaft delivers microwave energy to the target tissue.

    Modifications have been made to the Emprint™ Ablation System and Emprint™ SX Ablation Platform's IFU's and Antenna's. Updates were made to the IFU to reduce the possible trocar detachment post ablation resulting from atypical use. Changes made to the bond curing schedule is to enhance the bond strength between the trocar and the fiberglass shaft. Additionally, the clear non-stick heat shrink was extended further on the antennae to reduce tissue adherence post ablation.

    AI/ML Overview

    The information provided does not describe a study involving a device that uses AI or reports its performance using metrics such as sensitivity, specificity, or AUC. Instead, it describes a medical device (an ablation system) and modifications made to its antenna and instructions for use. The acceptance criteria and performance data discussed are related to the physical and functional characteristics of the ablation antenna, not AI performance. Therefore, I cannot fulfill the request as it pertains to AI-based device performance and study design.

    However, I can extract the acceptance criteria and performance data related to the physical modifications of the Emprint Ablation System with Thermosphere Technology and Emprint SX Ablation Platform with Thermosphere Technology antennae based on the provided text, and present it in a modified format relevant to the device's physical changes.

    Accepted Criteria and Device Performance (for physical modifications)

    Acceptance CriteriaReported Device Performance
    Trocar bond strength greater than the predicate.Verification testing demonstrated the modified antenna's trocar bond strength was greater than the predicate.
    No impact on the reliability of the device (thermal ablations in soft tissue during extended use test).Verification testing confirmed that the change did not impact the reliability of the device by performing thermal ablations in soft tissue during the extended use test.
    Quantify and compare frictional force between modified antenna and adjustable depth guide.Verification testing quantified and compared the frictional force between the modified antenna and the adjustable depth guide. (Specific quantitative results not provided in this summary, but the test was conducted.)
    Quantify and compare insertion force (ex-vivo liver tissue) of the modified antenna relative to cleared devices.Verification testing quantified and compared the insertion force (ex-vivo liver tissue) of the modified antenna relative to the 510(k) cleared devices. (Specific quantitative results not provided in this summary, but the test was conducted.)
    Quantify and compare extraction force (ex-vivo liver tissue) of the modified antenna relative to cleared devices.Verification testing quantified and compared the extraction force (ex-vivo liver tissue) of the modified antenna relative to the 510(k) cleared devices. (Specific quantitative results not provided in this summary, but the test was conducted.)
    The modifications do not affect the system's essential function, performance, or underlying principles of operation.The modification to the antennae trocar bonding and clear non-stick heat shrink extension have improved parameters such as tensile bond strength, antenna insertion force, and antenna extraction force compared to the current device (predicate) without affecting the design requirements and the essential performance of the antenna and the system as a whole. The antenna changes do not affect the Emprint™ Ablation System with Thermosphere™ Technology and Emprint™ SX Ablation Platform with Thermosphere™ Technology's essential function, performance, or underlying principles of operation. Modifications made to the Antenna's do not alter the Indications for use for the Emprint™ Ablation system and the Emprint™ SX ablation platform.

    Study Details (for physical modifications)

    1. Sample size used for the test set and the data provenance: Not explicitly stated for specific tests (e.g., how many antennas were tested for bond strength or insertion force). The verification testing included "thermal ablations in soft tissue during the extended use test" and "ex-vivo liver tissue." The data provenance is implied to be from bench testing conducted by Covidien LLC/Medtronic Inc. (the manufacturer). The study is retrospective in the sense that it evaluates modifications to an existing, cleared device.

    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience): Not applicable. The ground truth here is related to physical measurements (bond strength, forces, reliability during ablation) demonstrated through bench and ex-vivo testing, not expert consensus on medical images or diagnoses.

    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable. This was bench and ex-vivo testing, not a clinical study requiring adjudication of expert opinions.

    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This document does not describe an AI medical device or a MRMC study.

    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This document does not describe an AI medical device.

    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc): The "ground truth" for the performance claims in this submission is derived from bench verification testing (measuring physical parameters like bond strength, frictional forces, insertion/extraction forces) and ex-vivo testing (thermal ablations in soft tissue, ex-vivo liver tissue).

    7. The sample size for the training set: Not applicable. This is not an AI/machine learning device.

    8. How the ground truth for the training set was established: Not applicable. This is not an AI/machine learning device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K163105

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-12-06

    (29 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Emprint™ Ablation System is intended for use in percutaneous, laparoscopic, and intraoperative coagulation (ablation) of soft tissue, including partial or complete ablation of non-resectable liver tumors.

    The Emprint Ablation System is not intended for use in cardiac procedures.

    Device Description

    Covidien has modified the Emprint™ Ablation System's Percutaneous Antenna (Catalog Numbers: CA15L1, CA20L1 and CA30L1). The modified ablation antenna is identical to the device described in K133821, with two exceptions. The antenna's tissue contacting cooling jacket is manufactured from a new, biocompatible material that significantly increases the antenna's stiffness. Additionally, an adhesive has been added at the antenna shaft's proximal joint to strengthen the seal. By making the antenna stiffer, the cooling jacket material change is expected to enhance the antenna's usability; it does not affect the Emprint Ablation System's essential function, performance or underlying principles of operation.

    Like its predicate, the modified antenna (Catalog Numbers CA15L2, CA20L2 CA30L2) is provided sterile (EO) and is intended for use with the Emprint Ablation System in health care facilities.

    AI/ML Overview

    The provided text describes a 510(k) submission for a modified medical device, the Emprint™ Ablation System's Percutaneous Antenna. The modification involves a change in the cooling jacket material to increase stiffness and an improved adhesive seal. This submission focuses on demonstrating that the modified device is substantially equivalent to its predicate.

    However, the document does not contain the kind of detailed information typically found in a study proving a device meets specific acceptance criteria in terms of diagnostic performance or clinical outcomes, as would be relevant for an AI/ML-driven diagnostic device. Instead, it focuses on verifying that the modification does not negatively impact the device's original performance and safety.

    Given the context of the document (a 510(k) summary for a physical medical device modification), I will adapt the request to what information is available about the performance criteria and the study conducted to demonstrate equivalence for this specific type of device.

    Here's the breakdown based on the provided text:

    Acceptance Criteria and Device Performance (Adapted to a Physical Medical Device Modification)

    The "acceptance criteria" here are implicitly related to maintaining the performance characteristics and safety of the predicate device despite the modification. The "reported device performance" refers to the verification that these characteristics are maintained or improved without introducing new risks.

    Acceptance Criteria (Implied for Device Modification)Reported Device Performance
    Mechanical Robustness: The modified antenna must maintain or improve its resistance to mechanical stresses.Verification testing demonstrated the modified antenna's continued robustness to mechanical stresses.
    Printed Depth Markings Integrity: The new cooling jacket material should not affect the clarity or accuracy of the antenna's printed depth markings.Verification testing confirmed that the antenna's printed depth markings were not affected by the underlying cooling jacket material change.
    Stiffness: The modified antenna's shaft must demonstrate increased stiffness as intended by the material change.Verification testing quantified the stiffness of the antenna's shaft relative to the 510(k) cleared device, indicating an increase.
    Essential Function (Thermal Ablations): The modification must not impact the system's ability to perform thermal ablations in soft tissue correctly and safely.Verification testing confirmed that the change did not impact the system's essential function – the performance of thermal ablations in soft tissue.
    Biocompatibility: The new materials must be biocompatible and not introduce new biological risks.Biocompatibility testing of the new fiberglass material and the modified cooling jacket subassembly was conducted in accordance with FDA Blue Book Memorandum #G95-1 and ISO 10993-1. The modified cooling jacket subassembly also underwent a full chemical characterization. (Presumably, these tests showed acceptable biocompatibility).
    Stability/Performance Over Time: The modification should not negatively affect the antenna's performance or integrity over its shelf life or intended use time.Product stability testing evaluated the effect of the cooling-jacket material change on the antenna's performance over time. The testing included mechanical stress testing, cytotoxicity testing, and performance testing. (Presumably, these tests showed acceptable stability and performance retention).
    Usability & Safety (User Interaction): The altered stiffness should not introduce new use-related risks or negatively impact usability, and should meet usability requirements.Representative end users participated in a simulated-use validation study. Objective and subjective test results confirmed that (1) essential and safety-critical tasks were performed with the stiffer antenna in a safe and effective manner, (2) the modified device met usability requirements and, (3) there were no new, use-related risks, associated with the change.

    Here's the additional information requested, based on the provided document:

    1. Sample size used for the test set and the data provenance:

      • Test Set Sample Size: The document does not specify a numerical sample size for "bench verification testing," "product stability testing," "biocompatibility testing," or "validation testing." For the validation study, it mentions "Representative end users participated," but not the number of users.
      • Data Provenance: The studies were conducted by Covidien LLC as part of their 510(k) submission to the FDA. The location of the testing is not explicitly stated, but it would presumably be internal to Covidien or conducted by contract research organizations on their behalf. This is a prospective set of tests and analyses designed specifically for this regulatory submission.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This question is not applicable in the context of this device and its modification. The "ground truth" here isn't a diagnostic label established by medical experts, but rather the objective physical and functional properties of the device and its safety characteristics. The "experts" involved would be engineers and scientists performing the verification and validation tests, and possibly human factors specialists for the usability study, but their number and specific qualifications beyond being "representative end users" are not detailed.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • This concept is not applicable here. Adjudication methods like "2+1" or "3+1" are used to resolve disagreements among multiple human readers on diagnostic labels. The studies described are physical and functional performance tests for a medical device modification, not diagnostic reads.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No, an MRMC comparative effectiveness study was not done. This device is a physical ablation system, not an AI-driven diagnostic tool. Therefore, the concept of human readers improving with or without AI assistance is not relevant.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • No, a standalone (algorithm-only) performance study was not done. This is a physical medical device, not an algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The "ground truth" for this device's performance is based on established engineering principles, material science standards, biocompatibility regulations (ISO 10993 series), and functional performance specifications of the predicate device. For example:
        • Mechanical Integrity: Ground truth is defined by engineering specifications (e.g., force required to cause deformation, tear strength).
        • Biocompatibility: Ground truth is defined by regulatory standards (e.g., ISO 10993) and the absence of adverse biological reactions.
        • Functional Performance: Ground truth is the ability to generate the same type, size, and shape of thermal ablations as the predicate device, as measured in test media.
        • Usability: Ground truth is the successful completion of tasks by users and the absence of new use-related hazards.
    7. The sample size for the training set:

      • There is no "training set" in the context of this device. Training sets are typically used for machine learning algorithms.
    8. How the ground truth for the training set was established:

      • Not applicable, as there is no training set for this type of device submission.
    Ask a Question

    Ask a specific question about this device

    K Number
    K133821

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-04-28

    (133 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Covidien Emprint™ Ablation System is intended for use in percutaneous, laparoscopic, and intraoperative coagulation (ablation) of soft tissue, including partial or complete ablation of non-resectable liver tumors.

    The Covidien Emprint™ Ablation System is not intended for use in cardiac procedures.

    Device Description

    The Emprint™ Ablation System is a microwave-based system intended to deliver energy through an antenna inserted into soft tissue for the purpose of coagulating (ablating) a defined volume of that tissue. The Emprint™ Ablation System utilizes a 2450 MHz 100W generator to deliver power to a single microwave ablation antenna.

    The Emprint™ Ablation Generator is composed of analog and digital circuits with no software or firmware. The Emprint™ Ablation Generator provides for user setting of ablation time and ablation power (5 to 100W). With an optional temperature probe, the Emprint™ Ablation Generator can be set to monitor the temperature of a desired target and to automatically shut the generator off when the target reaches a pre-set temperature.

    The Emprint™ Ablation System uses circulating room temperature normal saline to cool the non-radiating portion of the antenna shaft and to provide a more consistent ablation zone. The normal saline is pumped from an IV bag through the antenna shaft and back to the IV bag in a closed system. No saline is in contact with the patient.

    The Emprint™ Ablation System consists of the following components: Emprint™ Ablation Generator (2450 MHz), Emprint™ Ablation Antenna (sterile, single use), Emprint™ Ablation Reusable Cable, and Emprint™ Ablation Pump. The system also includes the following optional components: Emprint™ Ablation Cart (with Isolation Transformer), Remote Temperature Probe (sterile, single-use), and Footswitch.

    The system must be used with a standard IV bag of sterile normal saline (not provided with the system).

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the Emprint™ Ablation System, a microwave-based system for soft tissue ablation. However, the document does not contain the detailed information necessary to fully answer all aspects of your request regarding acceptance criteria and a specific study proving the device meets those criteria.

    Here's an analysis based on the available information:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document broadly states: "The performance of the proposed device was characterized using the same methods used for one of the predicate devices. Coagulations were conducted on several ex vivo and in vivo tissue types at various power and duration settings. The shapes of the coagulation zones created were analyzed and compared to that of the predicate. Bench testing also included verification of dimensional characteristics, surface temperatures, and skin penetration."

    However, specific numerical acceptance criteria (e.g., minimum ablation zone size, maximum temperature deviation, etc.) and the precise reported performance values against these criteria are not provided in this summary. The comparison to the predicate device is mentioned as the primary method for demonstrating substantial equivalence, but the detailed results of that comparison are not given.

    Therefore, a table cannot be fully constructed.

    2. Sample Size Used for the Test Set and Data Provenance:

    The document mentions "several ex vivo and in vivo tissue types" but does not specify the sample sizes for these tests.

    The data provenance is not explicitly stated in terms of country of origin or whether it was retrospective or prospective. It implies an internal study conducted by the manufacturer (Covidien llc).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

    The document does not mention the use of experts to establish ground truth. The performance data focuses on physical characteristics of the ablation zone.

    4. Adjudication Method for the Test Set:

    Not applicable, as no expert review or human interpretation of cases is described.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

    No, an MRMC study was not done. The document describes performance testing of the device itself (ablation characteristics) and not a study involving human readers or the AI's impact on human performance.

    6. If a Standalone (algorithm only without human-in-the-loop performance) was done:

    This device is a physical medical instrument (microwave ablation system), not an AI algorithm. Therefore, the concept of "standalone (algorithm only)" does not apply in this context. The performance data presented relates to the device's physical capabilities and effects on tissue.

    7. The Type of Ground Truth Used:

    The "ground truth" for the performance evaluation appears to be the physical measurements and characteristics of the ablation zones created by the device in ex vivo and in vivo tissue, and a comparison of these to a predicate device. This is based on direct observation and measurement of the device's physical output.

    8. The Sample Size for the Training Set:

    The concept of a "training set" is relevant for machine learning algorithms. This device is a hardware system, not an AI algorithm, so there is no training set in the machine learning sense. The device's "training" or development would involve engineering design, testing, and optimization based on physical principles, not data input for an algorithm.

    9. How the Ground Truth for the Training Set was Established:

    Not applicable, as there is no training set for an AI algorithm. The device's design and optimization would be based on engineering specifications and empirical testing.


    In summary, the provided 510(k) summary focuses on demonstrating substantial equivalence to predicate devices through performance testing related to the physical characteristics of the ablation system. It does not provide the level of detail regarding acceptance criteria, sample sizes, expert involvement, or specific ground truth methodologies that would typically be found in an AI/software device submission.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1