Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K191667

    Validate with FDA (Live)

    Date Cleared
    2019-12-18

    (180 days)

    Product Code
    Regulation Number
    870.1875
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Electronic Stethoscope is intended for the detection and amplification of sounds from the heart, lungs and other internal organs with the use of a selective frequency. It can be used on any person undergoing a physical assessment for medical diagnostic purposes only.

    Device Description

    The Mintti Electronic Stethoscope, model smartho-D2 is a healthcare device that picks up sounds of the heart, lungs and other internal organs with the use of selective frequency ranges from 10-2000Hz. After amplification and filtering, the sounds are transferred to the user's ears via an wired-connected headset.

    The user interface includes a 5-button keypad and an 1.3' OLED display with a blue back-light. Sound processing is carried out with the aid of a digital signal processor.

    The smartho-D2 does not incorporate any off-the-shelf (OTS) software, and it incorporates embedded software which controls all the features of the product, including sound capture, digital signal processing, volume control, OLED display and Bluetooth function.

    The smartho-D2 operates on a certificated rechargeable 3.7V/2000mAh lithium ion battery.

    AI/ML Overview

    The provided document is a 510(k) summary for the "Electronic Stethoscope - smartho-D2". It outlines the device's characteristics and its comparison to a predicate device, 3M™ Littmann® Electronic Stethoscope Model 3200 (K083903), to demonstrate substantial equivalence for regulatory clearance.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" and "reported device performance" in a separate table format. However, it provides a "Specific Comparison to Predicate" table which implicitly serves this purpose by comparing the proposed device's specifications to those of the predicate. The "Comment" column in this table indicates whether the characteristic is "Same" or "Different." For differences, the justification is that they "do not affect the basic design principle, usage, effectiveness and safety of the subject device." The performance testing section further supports these claims.

    Here’s a table summarizing the implicit acceptance criteria based on the predicate comparison and the performance demonstrated by the proposed device:

    Feature/SpecificationAcceptance Criteria (from Predicate)Reported Device Performance (smartho-D2)Status (Comment in Document)
    Regulatory
    ClassificationStethoscope, Electronic (Class II)Stethoscope, Electronic (Class II)Same
    Regulation21 CFR 870.187521 CFR 870.1875Same
    Product CodeDQDDQDSame
    Indications for UseDetection and amplification of sounds from heart, lungs, arteries, veins, and other internal organs for medical diagnostic purposes.Detection and amplification of sounds from heart, lungs, and other internal organs for medical diagnostic purposes.Different (justified as not raising different questions of safety/effectiveness)
    ContraindicationsNoneNoneSame
    Power Supply TypeBatteryBatterySame
    Battery TypeAA alkaline, lithium, or NiMHRechargeable 3.7V/2000mAh lithium ionDifferent (justified)
    Battery Operation Time50-60 Hours48 HoursDifferent (justified)
    Binaural HeadsetYESYESSame
    Chest-pieceYESYESSame
    Sound ProcessingDigital signal processorDigital signal processorSame
    Display ScreenLCD1.3' Color OLEDDifferent (justified)
    Low Battery IndicatorYESYESSame
    Automatic Power OffYESNoDifferent (justified)
    Volume Control1-9 level1-4 levelDifferent (justified)
    Sound AmplifierUp to 24XUp to 100XDifferent (justified)
    Signal Sampling Rate4 kHz8 kHzDifferent (justified)
    Frequency ResponseBell (20-1000 Hz), Diaphragm (20-2000 Hz), "Extended Range" (50-500 Hz)20-2000Hz (Heart Sound Mode: 20-500Hz, Lung Sound Mode: 200-2000Hz)Different (justified)
    Intuitive KeypadNoNOSame
    A/P Chest KeyNoNoSame
    Direct ListeningYes (through attached binaurals)Only via headsetDifferent (justified)
    Recording and PlaybackYes (stores 12 30-sec tracks on device)Not on the device itselfDifferent (justified)
    Wireless TechnologyYes (Bluetooth 2.4 GHz)Yes (Bluetooth 2.4 GHz)Same
    Ambient & Frictional Noise Reduction TechnologyYESYESSame

    The study that "proves" the device meets these criteria is the "Performance Testing For Substantial Equivalence Evaluation" section, which includes:

    • Non-Clinical Data: Biocompatibility testing, Reprocessing (Cleaning and Disinfection) validation, Electrical safety and electromagnetic compatibility (EMC) testing, Bench Testing, Software Verification and Validation Testing, Risk Analysis, and Wireless Function Testing.
    • Clinical Data: None submitted.

    The overall conclusion is that "The Mintti smartho-D2 Electronic Stethoscope is Substantially Equivalent (SE) to the 3M™ Littmann® Electronic Stethoscope Model 3200 cleared by K083903." This implies that the differences observed do not raise new questions of safety or effectiveness and are addressed by the performance testing.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: The document does not specify a distinct "test set" sample size in terms of patient data. The non-clinical testing refers to testing on the device itself and its components. For example:
      • Biocompatibility tests (Cytotoxicity, Skin Sensitization, Skin Irritation) were conducted on the "smartho-D2 Electronic Stethoscope." The sample size refers to the number of devices or components tested, not human subjects.
      • Electrical safety and EMC testing were conducted on "the smartho-D2 Electronic Stethoscope device, consisting of all the modules and accessories."
      • Bench testing was conducted on "the smartho-D2 Electronic Stethoscope device, consisting of all the accessories."
      • The document explicitly states: "No animal or clinical testing was submitted in this 510(k)." This means there was no human or animal test set for performance on biological subjects.
    • Data Provenance: Not explicitly stated for specific test data, but the submitter is Hefei Mintti Medical Technology Co., Ltd. China. Given the nature of the non-clinical testing (lab-based), the "country of origin" would likely be the location where these tests were performed, presumably China. The data would be prospective with respect to the tests conducted on the device, as they were performed to support this submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • As "No animal or clinical testing was submitted in this 510(k)," there was no human or animal test set requiring experts to establish ground truth in a medical diagnostic context. The ground truth for electrical safety, biocompatibility, and functional performance (e.g., frequency response, amplification) is established by adherence to recognized international standards (e.g., ISO, IEC) and internal company specifications, with the assessment of experts in those respective engineering and scientific fields (e.g., electrical engineers, biocompatibility specialists) who perform the tests and interpret the results against the standard's criteria. Their qualifications are not detailed in this summary.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Since no human or animal clinical test set with medical diagnostic outcomes was conducted, there was no adjudication method involving multiple human readers/experts comparing device output to clinical ground truth. The "adjudication" for non-clinical tests would involve comparison of test results against predefined pass/fail criteria from international standards or design specifications.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No MRMC study was done. The device (Electronic Stethoscope - smartho-D2) is a simple electronic stethoscope for sound detection and amplification. It does not appear to incorporate AI, nor does the submission mention any AI features or claims of improving human reader performance. The device's function is to assist physicians by amplifying and filtering sounds, not to provide diagnostic interpretations or AI-assisted readings.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • The device is a hardware product (an electronic stethoscope) with embedded software for control and signal processing, not a standalone algorithm providing diagnostic output. Therefore, no standalone algorithm performance study was conducted in the sense of an AI algorithm making a diagnosis without human input. The "standalone" performance testing refers to the device's functional performance (e.g., amplification, frequency response) as a physical instrument.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • For the non-clinical performance testing:
      • Biocompatibility: Ground truth is established by the pass/fail criteria specified in ISO 10993-1.
      • Reprocessing: Ground truth is established by validation results demonstrating effective cleaning and disinfection, presumably against microbial reduction criteria.
      • Electrical safety and EMC: Ground truth is established by compliance with IEC 60601 series standards' requirements.
      • Bench Testing (functionality): Ground truth is established by design specifications (e.g., 10-2000Hz operating range, 100X amplification) that the device must meet, measured against calibrated test equipment.
      • Software V&V: Ground truth is established by functional requirements and specifications that the software must fulfill, validated through testing.
      • Risk Analysis: Ground truth is adherence to ISO 14971 principles and documented mitigation of identified risks.
      • Wireless Function Testing: Ground truth is demonstrated by compliance with FDA Guidance on Wireless Technology, ensuring no adverse effect on safe and effective use.
    • No clinical ground truth (expert consensus, pathology, or outcomes data from human/animal subjects) was used, as no clinical studies were submitted.

    8. The sample size for the training set

    • The document does not mention any "training set" as no machine learning or AI models are described for this device. The device is an electronic stethoscope that amplifies and filters sounds, not an algorithm that learns from data to make predictions or classifications.

    9. How the ground truth for the training set was established

    • Since there is no mention of a training set or machine learning, this question is not applicable to the information provided in the document.
    Ask a Question

    Ask a specific question about this device

    K Number
    K173663

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-08-17

    (261 days)

    Product Code
    Regulation Number
    870.1875
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The IMEDIPLUS Electronic Stethoscope DS3011A is intended for the detection, amplification and recording of sounds from the heart, lungs, anterior and posterior chest, abdomen, neck, limbs, arteries, veins and other internal organs with selective frequency ranges. And the stethoscope chest-piece is designed for use with child, adolescent and adult patients. It is used for any subject undergoing a physical examination and intended diagnostic purposes in clinic or hospital.

    Device Description

    The IMEDIPLUS Electronic Stethoscope DS3011A picks up the sounds from the heart, lung, anterior/posterior chest, abdomen, neck, limbs, arteries, veins and other internal organs from a patient's body. When picking up the sounds, the phonogram of sounds could simultaneously display after pushing the button 'OK' or 'REC' for recording. When you auscultate with DS3011A, the sounds are conducted simultaneously to the user's ears bilaterally by active speaker embedded at the bottom of the IMEDIPLUS Electronic Stethoscope DS3011A. At the meantime, sound processing is operated with the aid of a digital signal processor. The IMEDIPLUS Electronic Stethoscope DS3011A could identify the recording number by 1-D barcode reader, indicate the sound location by intuitive keypad, and record the sounds from different sites. The one-hand user interface includes a full-color OLED display, an intuitive keypad at the anterior part, a barcode reader at the posterior part, a chest-piece at the superior part, a tube connector for output of sounds at the inferior part, and a recording button at the left part. After opening the Electronic Stethoscope DS3011A with your setting password and connection with a wireless device by Bluetooth, the IMEDIPLUS Electronic Stethoscope DS3011A could transmit digital data of recorded sounds under the operation of IMEDIPLUS DS3011A_DM Software (DS3011A Data Management) installed in the connected device. The IMEDIPLUS DS3011A_DM Software (DS3011A Data Management) will only operate under the user's authorization with the seething password. The IMEDIPLUS DS3011A-DM Software (IMEDIPLUS DS3011A-Datamanagement Software) installed in mobile device provides the users to connect with the IMEDIPLUS Electronic Stethoscope DS3011A, receive the recorded data of sounds, storage the data, display the phonograms, and replay the recorded audio data with the speaker of effective frequency range from 20 to 1000Hz. The IMEDIPLUS Electronic Stethoscope DS3011A could also exchange audio data with an external personal computing device by micro SD card. Every single audio file stored in the micro SD card was labeled with the user's ID, recording number and indicated position. The IMEDIPLUS Electronic Stethoscope DS3011A does not incorporate any other off-the-shelf (OTS) software. The recorded audio data only can be replayed by the IMEDIPLUS Electronic Stethoscope DS3011A, the IMEDIPLUS Electronic Stethoscope DS301, and the IMEDIPLUS DS3011A_DM Software (DS3011A_Data Management) installed in the mobile device with the speaker of effective range from 20 to 1000Hz. The IMEDIPLUS Electronic Stethoscope DS3011A operates on one (1) NP-120 lithium battery with an included power management system to prolong the battery life.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the studies that prove the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Performance AspectAcceptance CriteriaReported Device Performance
    BiocompatibilityAssessed low risk due to short contact duration and use of common medical device materials (based on FDA #G95-1 and ISO10993-1). No laboratory tests needed.Compliant: Risk assessed as low for biological hazards due to short patient skin contact (< 5 minutes), short doctor skin contact (< 3 hours), and materials commonly used in medical devices. No biocompatibility laboratory tests were deemed necessary.
    Electrical Safety & EMCCompliance with IEC 60601-1 and EN 60601-1-2, CISPR 11, IEC 60601-1-2, IEC 61000-3-2, IEC 61000-3-3, IEC 61000-4-2, IEC 61000-4-3, IEC 61000-4-4, IEC 61000-4-5, IEC 61000-4-6, IEC 61000-4-8, IEC 61000-4-11 and FCC Part 15.Compliant: Test reports issued by SGS Taiwan Limited confirm compliance with IEC 60601-1 (safety) and a comprehensive list of EMC standards (EN 60601-1-2, CISPR 11, etc., including FCC Part 15).
    Software V&VCompliance with "guidance for the content of premarket submissions for medical devices software, May 11, 2005" and IEC/TR80002-1:2009. Software level of concern classified as "moderate."Compliant: Software verification and validation comply with the specified FDA guidance and IEC/TR80002-1:2009. The software's level of concern is "moderate" (failure or flaw could indirectly cause minor injury).
    Reliability & MTBFDemonstrates conditions for operation, storage, and transportation, and shelf life.Compliant: Reports from DEKRA Integrated Service Technology, Inc. show the device's reliability under various conditions and its shelf life.
    Battery SafetyCompliance with IEC 62133:2012 and UN38.3.Compliant: Battery test reports from SGS Taiwan Limited confirm compliance with IEC 62133:2012 (Second Edition) and UN38.3.
    Photobiological SafetyCompliance with IEC 62471:2006 and EN 62471:2008.Compliant: Photobiological test report from SGS Taiwan Limited confirms compliance with IEC 62471:2006 (First Edition) and EN 62471:2008.
    Usability EvaluationCompliance with Clause 5.3 and 5.7.3 of IEC 62366:2015 and IEC 60601-1-6.Compliant: Usability evaluation complies with the specified clauses of IEC 62366:2015 and IEC 60601-1-6.
    Clinical EvaluationMedical equipment (DS3011A) can meet the basic principles of safety and effectiveness when used in Child, Adolescent, Adult and other ethnic groups for heart, chest, intestines and neck auscultation.Compliant: A statistical analysis of a questionnaire from a clinical evaluation demonstrates the device's safety and effectiveness for its intended use across target patient populations (child, adolescent, adult) and body regions (heart, chest, intestines, neck).
    Wireless CoexistenceCompliance with ANSI C63.27-2017.Compliant: The wireless coexistence report shows compliance with ANSI C63.27-2017.
    Heart Rate DetectionSubstantially equivalent to predicate device 3M 3200 and DS301 for heart rates between 30-180 bpm.Compliant: Bench testing shows the heart rate detection of DS3011A is substantially equivalent to the predicate devices (3M 3200 and DS301) for heart rates between 30-180 bpm.
    Audio PerformanceSubstantially equivalent to reference device 3M 3200 for audio frequency below 1 kHz. Substantially equivalent to predicate device IMEDIPLUS Electronic Stethoscope DS301 for bell, diaphragm, and wide frequency band mode.Compliant: Bench testing demonstrates that the audio performance of DS3011A and DS3011A_DM software are substantially equivalent to the reference device 3M 3200 for audio frequencies below 1 kHz, and substantially equivalent to the predicate device DS301 for bell, diaphragm, and wide frequency band modes.

    2. Sample Sizes used for the Test Set and Data Provenance

    • Clinical Evaluation:
      • Sample Size: 15 subjects (4 children, 4 adolescents, and 7 adults).
      • Data Provenance: Not explicitly stated, but given the submission is from Taiwan, it is likely the data originated from Taiwan. The study is described as a "clinical evaluation," implying it was prospectively conducted.
    • Bench Testing (Heart Rate Detection & Audio Performance):
      • Sample Size: Not specified for the number of measurements, but the comparison involves three devices: IMEDIPLUS Electronic Stethoscope DS3011A, 3M 3200, and IMEDIPLUS Electronic Stethoscope DS301.
      • Data Provenance: Retrospective comparison against established predicate and reference devices.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    • Clinical Evaluation:
      • Number of Experts: 4 clinical doctors.
      • Qualifications: Described as "clinically certified and qualified professionals." No further details on their specific specialties or years of experience are provided in this document.

    4. Adjudication Method for the Test Set

    • Clinical Evaluation: The text mentions "statistical analysis of the questionnaire" and that the medical equipment can meet the "basic principles of safety and effectiveness." It implies that the experts assessed the device's performance based on their clinical judgment of safety and effectiveness, likely through the questionnaire. It does not explicitly state a formal adjudication method like "2+1" or "3+1" for discrepancies.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    • No, a formal Multi-Reader Multi-Case (MRMC) comparative effectiveness study as typically understood for diagnostic accuracy (e.g., comparing human readers with and without AI assistance on a case set) was not explicitly described.
    • The clinical evaluation involved 4 clinicians using the device on 15 subjects, but the objective was to confirm safety and effectiveness, not to measure an improvement in human reader performance with AI assistance.
    • The comparison studies were primarily bench tests against predicate/reference devices for specific technical functionalities (heart rate detection, audio performance), not human reader performance.

    6. If a Standalone (algorithm only without human-in-the-loop performance) was done

    • Yes, for certain aspects. The "Bench Testing" for Heart Rate Detection and Audio Performance appears to be standalone algorithm performance, as it compares the DS3011A's objective measures against those of predicate and reference devices, without involving human interpretation or decision-making beyond operating the device for measurement.

    7. The Type of Ground Truth Used

    • Clinical Evaluation: The ground truth was established through the "clinical evaluation" conducted by "4 clinical doctors." This suggests a form of expert consensus/clinical judgment on the device's safety and effectiveness during actual use or simulated use. It's not explicitly tied to pathology or outcomes data from the subjects in this summary.
    • Bench Testing (Heart Rate & Audio): The ground truth for these tests would be the established performance characteristics or specifications of the predicate and reference devices (3M 3200 and IMEDIPLUS DS301), and objective measurement standards for heart rate and audio frequency.

    8. The Sample Size for the Training Set

    • The document describes the device itself and its performance, but does not mention any AI or machine learning components that would require a "training set." The device is an electronic stethoscope that amplifies, records, and processes sounds, not an AI diagnostic tool that learns from data. Therefore, the concept of a training set is not applicable here.

    9. How the Ground Truth for the Training Set was Established

    • As there's no mention of AI/ML components requiring a training set, this question is not applicable to the provided information.
    Ask a Question

    Ask a specific question about this device

    K Number
    K160023

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-09-30

    (268 days)

    Product Code
    Regulation Number
    870.1875
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The IMEDIPLUS Electronic Stethoscope DS301 is intended for the detection, amplification and recording of sounds from the heart, lung, anterior and posterior chest, abdomen, neck, limbs, arteries, veins and other internal organs with selective frequency ranges. And the stethoscope chest-piece is designed for use with child, adolescent and adult patients. It is used for any subject undergoing a physical examination and intended only for medical diagnostic purposes in clinic or hospital.

    Device Description

    The IMEDIPLUS Electronic Stethoscope DS301 picks up the sounds from the heart, lung, anterior/posterior chest, abdomen, neck, limbs, arteries, veins and other internal organs from a patient's body. After detection and amplification, the sounds are transferred to the user's ears via an active speaker and passive sound tubes. It could also identify the recording number by 1-D barcode reader, indicate the sound location by intuitive keypad, and simultaneously record the sounds from different sites.

    The one-hand user interface includes a full-color OLED display and an intuitive keypad at the anterior part, a barcode reader at the posterior part, a chest-piece at the superior part, a tube connector for output of sounds at the inferior part, and a recording button at the left part. Sound processing is operated with the aid of a digital signal processor.

    The IMEDIPLUS Electronic Stethoscope DS301 could also exchange audio data with an external personal computing device using micro SD card. Every single audio file stored in micro SD card was labeled with the user's ID, recording number and indicated position. And the recorded audio data only can be replayed by IMEDIPLUS Electronic Stethoscope DS301, but cannot be replayed by personal computing device.

    The IMEDIPLUS Electronic Stethoscope DS301 does not incorporate any off-theshelf (OTS) software.

    The IMEDIPLUS Electronic Stethoscope DS301 operates on one NP-120 lithium battery with an included power management system to prolong the battery life.

    The associated accessories include:

    • Rechargeable Lithium-ion Battery.
    • Battery charger which include adaptor and charger. ●
    • Micro SD card for exchange audio data.
    AI/ML Overview

    Here's an analysis of the acceptance criteria and study information for the IMEDIPLUS Electronic Stethoscope DS301, based on the provided text:

    Important Note: The provided document is a 510(k) summary for a medical device, which focuses on demonstrating substantial equivalence to a predicate device. It typically does not contain the detailed, granular level of information often found in a full clinical study report, especially regarding specific acceptance criteria and detailed statistical outcomes for each performance metric. The information below is extracted and inferred from the available text.


    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't explicitly state quantitative acceptance criteria for each functional aspect. Instead, it relies on demonstrating substantial equivalence to the predicate device (3M™ LITTMANN® ELECTRONIC STETHOSCOPE MODEL 3200). The "performance" is generally reported as achieving substantial equivalence or compliance with relevant standards.

    Feature / Performance MetricAcceptance Criteria (Implicitly Substantial Equivalence to Predicate)Reported Device Performance (IMEDIPLUS DS301)
    BiocompatibilityLow risk based on FDA #G95-1 and ISO10993-1.Assessed as low risk; no laboratory tests performed.
    Electrical SafetyCompliance with IEC 60601-1.Complies with IEC 60601-1.
    EMCCompliance with EN 60601-1-2, CISPR 11, etc.Complies with relevant EMC standards.
    Software V&VCompliance with IEC 62304, AAMI TIR45, IEC/TR80002-1.Complies with all specified software standards.
    Software Level of Concern"Moderate" for minor injury.Considered "moderate" level of concern.
    Reliability & MTBFDemonstrated shelf life and operating/storage conditions.Reports demonstrate shelf life and operational conditions.
    Battery SafetyCompliance with IEC 62133, UN38.3.Complies with IEC 62133, UN38.3.
    Photobiological SafetyCompliance with IEC 62471.Complies with IEC 62471.
    UsabilityCompliance with IEC 62366, IEC 60601-1-6.Complies with specified usability standards.
    Heart Rate CalculationSubstantially equivalent to predicate in 30-180 bpm range.Substantially equivalent within 30-180 bpm range.
    Audio PerformanceSubstantially equivalent to predicate for audio frequency below 1 kHz.Substantially equivalent for audio frequency below 1 kHz.
    Clinical Evaluation (Safety & Effectiveness)Meets basic principles of safety and effectiveness for stated indications.Questionnaire results showed the device meets basic principles of safety and effectiveness for child, adolescent, and adult patients for heart, chest, bowel, and neck auscultation.

    2. Sample Size Used for the Test Set and Data Provenance

    • Heart Rate & Audio Performance Bench Testing: The document doesn't specify a "test set" sample size for these, as these were bench tests comparing the device's output directly against the predicate or known signal inputs. It implies these were direct comparative measurements.
    • Clinical Evaluation:
      • Sample Size: 3 children, 3 adolescents, and 3 adults. This is a very small sample size.
      • Data Provenance: Not explicitly stated (e.g., country of origin, retrospective/prospective). Given "4 testers" are involved, it's likely a prospective observational study in a clinical setting.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    • Clinical Evaluation:
      • Number of Experts: "4 testers" were used.
      • Qualifications: Described as "clinically certified and qualified professionals." No further detail (e.g., specific specialty like cardiologist, years of experience) is provided.
    • Bench Testing (Heart Rate & Audio): Ground truth for these technical tests would typically be established by calibrated instruments or the predicate device itself, not human experts in the same way clinical ground truth is established.

    4. Adjudication Method for the Test Set

    • Clinical Evaluation: Not explicitly stated. Given the small sample size and "questionnaire" mention, it's unlikely a formal adjudication process (like 2+1 reading) was used for ground truth. The "4 testers" might have individually evaluated the device for each subject, and then their feedback collected via questionnaire. It's unclear if there was any consensus building among these testers.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

    • No, a multi-reader multi-case (MRMC) comparative effectiveness study of human readers with vs. without AI assistance was not conducted or reported. This device is an electronic stethoscope that amplifies and records sounds, not an AI-assisted diagnostic tool in the typical sense for MRMC studies.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • This question is less directly applicable to an electronic stethoscope. The device itself is the "algorithm" in a sense, as it processes sound.
    • The bench tests for heart rate calculation and audio performance can be considered "standalone" performance evaluations as they test the device's technical specifications independently. The results for these were:
      • Heart Rate Detection: Substantially equivalent to the predicate (3M™ LITTMANN® MODEL 3200) within the 30-180 bpm range.
      • Audio Performance: Substantially equivalent to the predicate for audio frequencies below 1 kHz.

    7. The Type of Ground Truth Used

    • Bench Testing (Heart Rate & Audio): The ground truth was comparative. The performance of the IMEDIPLUS DS301 was compared to the predicate device (3M™ LITTMANN® MODEL 3200) under specified conditions, implying the predicate served as the reference standard.
    • Clinical Evaluation: The ground truth was based on the subjective evaluation and professional assessment of "4 clinically certified and qualified professionals" using a questionnaire regarding the "safety and effectiveness" during auscultation. This is closer to an expert consensus/opinion rather than objective pathology or outcomes data.

    8. The Sample Size for the Training Set

    • The document does not mention a "training set" in the context of machine learning or AI. This device appears to be a traditional electronic stethoscope with digital signal processing, not a device that uses AI models that require specific training data. The "performance data" describes technical and clinical validations, not AI training.

    9. How the Ground Truth for the Training Set Was Established

    • As no training set is mentioned (see point 8), this question is not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K121144

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-09-26

    (163 days)

    Product Code
    Regulation Number
    870.1875
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BM-520 Electronic Stethoscope is intended for medical diagnostic purposes only. It may be used for the detection and amplification of sounds from heart, lungs, blood vessels, and other internal organs. It can be used on any person undergoing a physical assessment.

    Device Description

    The Sharp Electronic Stethoscope Model BM-520 is an electronic stethoscope to auscultate sounds from heart, lung, blood vessels, and other internal organs. It can be used on adults undergoing a physical assessment. The sounds can also be transmitted to an appropriately configured receiving device via Bluetooth® wireless communication.

    AI/ML Overview

    The provided text describes the Sharp Electronic Stethoscope Model BM-520. However, it does not contain details about specific acceptance criteria or an analytical study that proves the device meets such criteria in terms of diagnostic performance (e.g., sensitivity, specificity, accuracy for detecting specific conditions).

    Instead, the submission document focuses on establishing substantial equivalence to predicate devices based on:

    • Intended Use: "medical diagnostic purposes only... detection and amplification of sounds from heart, lungs, blood vessels, and other internal organs... on any person undergoing a physical assessment."
    • Principles of Operation: "operates continuously to provide sounds from heart, lungs, and blood vessels. The sounds detected by the chest piece are output through the ear tips after being amplified and digitally filtered. At the same time, the sounds can be transmitted to an appropriately configured receiving device via Bluetooth® wireless communication. The BM-520 offers two selectable modes, Bell and Diaphragm. Volume is adjustable between 0 and 48 dB."
    • Functional Design: Not explicitly detailed but implied through the description of the device's operation.
    • Technological Characteristics: Similar to principles of operation, emphasizing the features like digital filtering, amplification, adjustable volume, and wireless communication.

    The "Summary of Test Data" section states: "The Sharp BM-520 was developed and is produced under consideration of all applicable technical standards and internal specifications. The performance of the BM-520 has been verified in the course of bench testing and software validation testing."

    This indicates that internal testing was conducted to ensure the device met its design specifications and general safety/performance standards, rather than a clinical study evaluating its diagnostic accuracy against a specific ground truth or clinical outcome.

    Therefore, many of the requested details related to clinical performance studies cannot be extracted from this document.

    Here's a breakdown of what can be extracted and what cannot:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Implied from Submission)Reported Device Performance
    Compliance with "applicable technical standards and internal specifications"Verified by "bench testing and software validation testing"
    Substantial equivalence to predicate devices (K050159, K041934, K083903 3M Littmann Stethoscope models)."BM-520 to be substantially equivalent to the predicate devices listed above. This conclusion is based on the similarities in primary intended use, principles of operation, functional design, and established medical use."
    Functionality as an electronic stethoscope (amplification, digital filtering, selectable modes, adjustable volume)Device description outlines these functionalities.
    Bluetooth® wireless communication capabilityDevice description confirms this capability.

    Note: The document does not provide specific numerical acceptance criteria (e.g., minimum amplification, frequency response ranges, signal-to-noise ratio) or corresponding numerical performance results from the bench testing.

    2. Sample size used for the test set and the data provenance:

    • Not Applicable. The document describes bench testing and software validation, not a clinical test set with patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not Applicable. No clinical test set or ground truth establishment by experts is described.

    4. Adjudication method for the test set:

    • Not Applicable. No clinical test set or adjudication process is described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This is not an AI/CAD device. It's an electronic stethoscope. No MRMC study is mentioned.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • Not Applicable. This is a medical device for sound amplification, not an algorithm. Bench testing for device functionality was performed, which is a form of standalone performance verification, but not in the context of diagnostic accuracy.

    7. The type of ground truth used:

    • Not Applicable. The submission focuses on substantial equivalence based on technical specifications and functionality, not clinical ground truth for diagnostic performance validation.

    8. The sample size for the training set:

    • Not Applicable. This device does not use a "training set" in the context of machine learning.

    9. How the ground truth for the training set was established:

    • Not Applicable.

    In summary: The 510(k) submission for the Sharp Electronic Stethoscope Model BM-520 relies on demonstrating substantial equivalence to previously cleared predicate devices through similarities in intended use, principles of operation, and functional design. It references internal bench testing and software validation to confirm the device meets "applicable technical standards and internal specifications," but it does not detail a clinical study with specific acceptance criteria related to diagnostic performance or efficacy for a particular condition.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1