    Intended Use

    The Amplify Surgical DualXSLIM® T/PLIF is indicated for intervertebral body fusion of the lumbar spine, from L2 to S1, in skeletally mature patients. The device is intended for use at either one or two contiguous levels for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis or retrolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device is intended for use with supplemental fixation and autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft. The devices are to be used in patients who have had at least six months of non-operative treatment. The Amplify Surgical DualXSLIM® T/PLIF is indicated for unilateral or bilateral implantation. The DualXSLIM® T/PLIF implants are intended for single use.

    Device Description

    The purpose of this 510(k) is to offer the DualXSLIM® T/PLIF as additional sizes and lordosis options to the previously cleared DualX (K181397). The DualX implants are a family of expandable interbody fusion devices that expand sequentially in lateral and then vertical directions. The implants are designed to be used in minimally invasive spine surgery. The DualXSLIM® T/PLIF implants are offered in a variety of sizes to accommodate patient anatomical needs. All implant components are manufactured from Ti-6A1-4V ELI per ASTM F136.

    AI/ML Overview

    This document is a 510(k) summary for the Amplify Surgical DualXSLIM® T/PLIF intervertebral body fusion device. It focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than providing a detailed study of the device's diagnostic performance against specific acceptance criteria relevant to an AI/ML powered medical device.

    Therefore, the information requested about acceptance criteria, study details, sample sizes, ground truth establishment, expert qualifications, and MRMC studies for an AI/ML device is not applicable or available in this document.

    This document describes a traditional medical device (implant) and its mechanical performance testing, not a diagnostic AI/ML device.

    Here's a breakdown of why each requested point is not applicable:

    1. A table of acceptance criteria and the reported device performance: Not applicable. This document is for a physical implant, not an AI/ML diagnosis. It focuses on mechanical testing relative to standards (ASTM F2077, F2267) rather than diagnostic performance metrics (e.g., sensitivity, specificity).

    2. Sample size used for the test set and the data provenance: Not applicable. This refers to physical "tests" (e.g., static/dynamic compression) on the device, not a test set for an algorithm. The "test sets" would be the manufactured devices themselves.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. There is no concept of "ground truth" established by experts in the context of mechanical testing for an implantable medical device.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable. This relates to expert consensus for diagnostic ground truth, which is not relevant here.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is for AI-assisted diagnostic devices.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This device does not have a standalone algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable. For mechanical tests, the "truth" is determined by the physical properties measured against engineering standards.

    8. The sample size for the training set: Not applicable. This device is not an AI/ML model that requires a training set.

    9. How the ground truth for the training set was established: Not applicable, for the same reason as above.

    In summary, this 510(k) document is for an intervertebral body fusion device and details mechanical performance testing, not the diagnostic performance of an AI/ML algorithm. Therefore, the specific questions related to AI/ML device performance and validation cannot be answered from this provided text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K211740

    Validate with FDA (Live)

    Device Name
    DualX
    Date Cleared
    2021-09-29

    (114 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ALIF System is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2 to -S1.

    DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies. These patients should be skeletally mature and six months of non-operative treatment prior to treatment with the devices. These DDD patients may also have up to Grade I spondylolisthesis at the involved level(s).

    The ALIF System 10° - 20° lordotic cages may be used as a standalone system. The ALIF System 25° - 40° lordotic cages must be used with supplemental internal spinal fixation systems pedicle screw and rod system) that are cleared by the FDA for use in the lumbar spine.

    The ALIF System implants can also be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis; however, when used in these patients at multiple levels and for patients with degenerative spondylolisthesis, the ALIF System must be used with a supplemental internal spinal fixation system (e.g., pedicle screw system) cleared by the FDA for use in the lumbar spine in addition to the integrated screws.

    For use with autogenous and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion.

    Device Description

    The Axis Spine Technologies ALIF is an interfixated interbody system consisting of a modular interbody spacer manufactured from Ti6Al4V. The system is designed to be assembled in vivo. The spacer comprises of two endplates and a central core. Each component is available in a variety of shapes and sizes to allow the assembled device to suit the individual pathology and anatomical conditions of the patient.

    The Axis Spine Technologies ALIF System intervertebral fusion device is designed to address lumbar pathologies utilizing placement through an anterior approach. The device's hollow core or graft aperture allows for packing of autogenous and/or allogeneic bone graft to help promote a solid fusion. A rough surface on the device endplates serves to grip the adjacent vertebrae to resist migration and expulsion of the device, alternately models with protrusions on the device endplates grip the adjacent vertebrae to resist migration and aid in expulsion resistance. The ALIF System 10° - 20° lordotic cages may be used as a standalone system. The ALIF System 25° - 40° lordotic cages must be used with supplemental internal spinal fixation systems (i.e., posterior pedicle screw and rod system) that are cleared by the FDA for use in the lumbar spine.

    The ALIF System implants can also be used as an adjunct to fusion in patients diagnosed with multilevel deqenerative scoliosis; however, when used in these patients at multiple levels and for patients with degenerative spondylolisthesis, the ALIF System must be used with a supplemental internal spinal fixation system (e.g., pedicle screw system) cleared by the FDA for use in the lumbar spine in addition to the integrated screws.

    The Axis Spine Technologies ALIF implant is composed of:

    • one (1) inferior endplate manufactured from titanium alloy (Ti-6Al-4V ELI) conforming to ASTM F136 and ISO 5832-3.
    • one (1) inferior endplate manufactured from titanium alloy (Ti-6Al-4V ELI) conforming to ASTM F136 and ISO 5832-3.
    • one (1) core manufactured from titanium alloy (Ti-6Al-4V ELI) conforming to ASTM F136 and ISO 5832-3.
    • three (3) bone screws manufactured from titanium alloy (Ti-6Al-4V ELI) conforming to ASTM F136 and ISO 5832-3.
    • one (1) cover plate manufactured from PEEK-Optima HA Enhanced.
    AI/ML Overview

    This FDA 510(k) summary provides information for a medical device called the "Axis Spine Technologies ALIF," an intervertebral body fusion device.

    Here's an analysis of the provided text in relation to your request:

    1. A table of acceptance criteria and the reported device performance:

    The document doesn't explicitly state "acceptance criteria" in a numerical or pass/fail threshold format that would typically be found in a performance study for, for example, diagnostic accuracy. Instead, it states that non-clinical testing was performed to demonstrate substantial equivalence to predicate devices. The reported device performance is that it "meets or exceeds the performance of the predicate device and does not introduce any new risks."

    Therefore, the "acceptance criteria" can be inferred as meeting the established standards for intervertebral body fusion devices, and the reported performance is that the device successfully met these standards.

    Acceptance Criteria (Inferred from testing standards)Reported Device Performance
    Demonstrate performance per ASTM F2077 for static and dynamic compression.Met or exceeded the performance of the predicate device.
    Demonstrate performance per ASTM F2077 for static and dynamic compression shear.Met or exceeded the performance of the predicate device.
    Demonstrate performance per ASTM F2267 for subsidence.Met or exceeded the performance of the predicate device.
    Demonstrate performance for expulsion and axial pushout of retention device.Met or exceeded the performance of the predicate device.
    Not introduce any new risks compared to predicate devices.Does not introduce any new risks.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    The document does not provide specific sample sizes for the non-clinical tests. It refers to "nonclinical testing" without detailing the number of specimens tested for each type of evaluation (e.g., number of interbody spacers for compression testing).

    The studies performed are non-clinical (mechanical/physical tests), not involving human or animal subjects, so terms like "country of origin of the data," "retrospective," or "prospective" are not applicable.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    This information is not applicable. The studies conducted are non-clinical mechanical tests, not clinical studies requiring expert interpretation or ground truth establishment in a medical context.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    This information is not applicable, as the tests are non-clinical mechanical validations.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    No MRMC comparative effectiveness study was done. The submission explicitly states: "No clinical studies were conducted as part of submission to prove substantial equivalence." This device is a physical implant, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    This information is not applicable. This device is a physical implant, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    This information is not applicable. For non-clinical mechanical tests, "ground truth" would refer to the physical properties and performance metrics established by the ASTM standards, not clinical or pathological diagnoses. The "ground truth" for these tests are the specified parameters and methodologies defined by the ASTM standards (F2077, F2267).

    8. The sample size for the training set:

    This information is not applicable. There is no "training set" for physical device testing as there would be for an AI algorithm.

    9. How the ground truth for the training set was established:

    This information is not applicable, as there is no training set for this type of submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K192434

    Validate with FDA (Live)

    Date Cleared
    2019-10-15

    (40 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DualX™ PLIF is indicated for intervertebral body fusion of the lumbar spine, from L2 to S1, in skeletally mature patients. The device is intended for use at either one or two contiguous levels for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis or retrolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device is intended for use with supplemental fixation and autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft. The DualX™ PLIF is indicated for unilateral implantation.

    The DualX™ LLIF is indicated for intervertebral body fusion of the lumbar spine, from L2 to L5, in skeletally mature patients. The device is intended for use at either one or two contiguous levels for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis or retrolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device is intended for use with supplemental fixation and autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft.

    The DualX™ TLIF is indicated for intervertebral body fusion of the lumbar spine, from L2 to S1, in skeletally mature patients. The device is intended for use at either one or two contiguous levels for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis or retrolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device is intended for use with supplemental fixation and autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft.

    Device Description

    The DualX™ Lumbar Intervertebral Body Fusion Device implants are an expanding family of interbody fusion devices that expand sequentially in lateral and then vertical directions. The implants are designed to be used in minimally invasive spine surgery and, in the collapsed configuration, are between 12.3 mm wide (TLIF and PLIF) and 13.3mm wide (LLIF).

    After entering the joint space, the implants expand laterally to maximize the contact area around the cortical rim of the vertebral bodies. The implants then expand vertically to maintain distraction of the vertebral bodies and maximize clearance for exiting nerve roots. The implants are binary, and each implant can only be locked out in the fully expanded state. Specific sizes are included in the system with various expanded heights to optimize the fit for the patient.

    The DualX™ Lumbar Intervertebral Body Fusion Device is made from titanium that conforms to ASTM F136.

    AI/ML Overview

    This looks like a 510(k) premarket notification for a medical device called the DualX™ Lumbar Intervertebral Body Fusion Device. The provided document does not contain information typically found in a study proving a device meets acceptance criteria, especially not for an AI/algorithm-based device.

    The document describes a modification to the sterilization trays of the device, allowing for non-sterile implants to be sterilized by the end user. It explicitly states: "mechanical testing was not required to demonstrate substantial equivalent mechanical strength since there are no modifications to the implants and instruments. The substantial equivalence of mechanical strength for the DualX™ Lumbar Intervertebral Body Fusion Device was established in the clearance of K181397."

    This means the submission is not about proving the initial performance of the device itself (the implant), but rather about a change in how it is processed (sterilization trays), and relies on the previously established substantial equivalence of the device's mechanical strength under K181397.

    Therefore, I cannot provide the requested information about acceptance criteria and study details for an AI-based device, as this is a physical medical implant called an "intervertebral body fusion device" and the submission is for a minor modification to its packaging/sterilization.

    If you have a document related to K181397 or a different type of medical device (like an AI/algorithm-based one) that includes performance study data, I would be happy to analyze that.

    Ask a Question

    Ask a specific question about this device

    K Number
    K181397

    Validate with FDA (Live)

    Device Name
    DualX
    Manufacturer
    Date Cleared
    2018-09-27

    (121 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DualX™ PLIF is indicated for intervertebral body fusion of the lumbar spine, from L2 to S1, in skeletally mature patients. The device is intended for use at either one or two contiguous levels for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis or retrolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device is intended for use with supplemental fixation and autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft. The DualXTM PLIF is indicated for unilateral or Bilateral implantation.

    The DualX™ LLIF is indicated for intervertebral body fusion of the lumbar spine, from L2 to L5, in skeletally mature patients. The device is intended for use at either one or two contiguous levels for the treatment of degenerative disc disease DDD) with up to Grade I spondylolisthesis or retrolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device is intended for use with supplemental fixation and autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft.

    The DualX™ TLIF is indicated for intervertebral body fusion of the lumbar spine, from L2 to S1, in skeletally mature patients. The device is intended for use at either one or two contiguous levels for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis or retrolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device is intended for use with supplemental fixation and autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft.

    Device Description

    The DualX Implants are an expanding family of Interbody Fusion Devices that expand sequentially in lateral and then vertical directions. The implants are designed to be used in minimally invasive spine surgery and, in the collapsed configuration, are between 12.3 mm wide (TLIF and PLIF) and 13.3mm wide (LLIF). After entering the joint space, the implants expand laterally to maximize the contact area around the cortical rim of the vertebral bodies. The implants then expand vertically to maintain destraction of the vertebral bodies and maximize clearance for exiting nerve roots. The implants are binary, and each implant can only be locked out in the fully expanded state. Specific sizes are included in the system with various expanded heights to optimize the fit for the patient. DualX Lumbar Intervertebral Body Fusion Device is made from titanium that conforms to ASTM F136.

    AI/ML Overview

    The provided text is a 510(k) premarket notification for the DualX™ intervertebral body fusion device. This document focuses on demonstrating substantial equivalence to predicate devices based on indications for use, technological characteristics, and performance data, primarily mechanical testing. It does not describe a study involving a medical AI device for diagnosis or prognosis, nor does it present clinical study data with acceptance criteria for such a device.

    Therefore, I cannot extract the information required to answer your prompt, as the prompt's questions pertain to:

    1. Acceptance criteria and device performance for an AI/diagnostic device. The document discusses mechanical test results for an implantable device, not accuracy metrics for an AI.
    2. Sample size and data provenance for a test set. The data discussed is from physical, mechanical testing of the implant, not patient data for AI model evaluation.
    3. Number of experts and their qualifications for ground truth. Not applicable to mechanical testing.
    4. Adjudication method for the test set. Not applicable to mechanical testing.
    5. Multi-reader multi-case (MRMC) comparative effectiveness study. Not applicable.
    6. Standalone (algorithm only) performance. Not applicable.
    7. Type of ground truth used. "Ground truth" in this context refers to the defined test conditions for mechanical properties, not expert consensus or pathology on patient data.
    8. Sample size for the training set. Not applicable, as this is not an AI device.
    9. How ground truth for the training set was established. Not applicable.

    The "Performance Data" section explicitly states that "The DualX Lumbar Intervertebral Body Fusion Device has been tested in the following test modes: Static axial compression per ASTM F2077, Static compressive shear per ASTM F2077, Dynamic axial compression per ASTM F2077, Dynamic compressive shear per ASTM F2077, Subsidence per ASTM F2267, Expulsion. The results of this non-clinical testing show that the strength of the DualX Lumbar Intervertebral Body Fusion Device is sufficient for its intended use and is substantially equivalent to legally marketed predicate devices." This is all mechanical testing, not a clinical or AI performance study.

    In summary, the provided document is not relevant to the type of device or study described in your prompt.