Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K221231

    Validate with FDA (Live)

    Device Name
    Drug Relief v1
    Manufacturer
    Date Cleared
    2022-06-06

    (38 days)

    Product Code
    Regulation Number
    882.5896
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Drug Relief v1 is a percutaneous nerve field stimulator (PNFS) system, that can be used as an aid to reduce the symptoms of opioid withdrawal, through application to branches of cranial nerves V, VII, IX and X, and the occipital nerves identified by transillumination.

    Device Description

    The Drug Relief v1 is a wearable, battery-operated device that is designed to administer periodical low level electrical pulses to the ear over five days / 120 hours from the time of activation of the device. The electrical pulse from the device will be delivered to the branches of cranial nerves on the ear through a set of wire assembly and stimulation needles. Three zinc air batteries with 1.4 V each provide the required stimulation energy for 120 hours. There are three stimulation electrode and one ground electrode which constitute of a titanium needle and lead/ wire with the snap-fit ring. The stimulation needles are inserted at three specific points, which have the ability to stimulate the cranial nerves. The ground electrode is inserted at one specific point (constant in all treatments) which forms the functional earthing to the device. The device is rectangular in shape and designed to fit comfortably on the neck. The biocompatible adhesive ensures that the device maintains contact with the skin during normal use. The adhesive fasteners ensure that the electrode needles and the entire device stay in place in a secure manner. This constant current source guarantees equivalent stimulation energy regardless of the individual impedance of the skin. The stimulation pattern consists of rectangular pulses with differing interpulse intervals. A 3-pin connector is provided, which is used to check the output voltage of the device once it is activated and before applying to the patient with any one of the voltage measuring devices available in the market with the appropriate regulatory compliance.

    AI/ML Overview

    The provided text describes the regulatory clearance of the "Drug Relief v1" device (K221231) by DyAnsys, Inc. It explicitly states that the device is substantially equivalent to a previously cleared predicate device, also named "Drug Relief v1" (K211971). The document primarily focuses on demonstrating this substantial equivalence rather than presenting an independent clinical study to prove the device meets specific performance criteria against a predefined ground truth in a clinical setting.

    Therefore, many of the requested items related to clinical study design, expert involvement, and statistical measures of performance (like effect size with AI assistance) are not directly applicable or available in this regulatory submission. The "acceptance criteria" discussed are largely in the context of demonstrating equivalence to the predicate and safety/performance through non-clinical testing.

    Here's an attempt to answer the questions based solely on the provided text, noting where information is not available:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present a table of quantitative acceptance criteria and corresponding reported device performance against a clinical ground truth for the subject device (K221231). Instead, it focuses on demonstrating that the subject device has "equivalent performance specifications" and "identical fundamental technological and operational characteristics" to the predicate device (K211971). The "Performance Data" section discusses non-clinical testing.

    Acceptance Criteria (Implicit - related to predicate equivalence and non-clinical performance):

    Acceptance Criteria (Implicit)Reported Device Performance (K221231)
    Intended Use / Indications for Use: Identical to predicateIdentical to predicate: "aid to reduce the symptoms of opioid withdrawal, through application to branches of cranial nerves V, VII, IX and X, and the occipital nerves identified by transillumination."
    Technological Characteristics: Equivalent to predicate"identical intended use, indications for use, population, application of the device, environment of use" to predicate
    Product Dimensions (mm x mm x mm): 38 * 21 * 1038 * 21 * 10
    Mass (g): 8g (including battery)8g (including battery)
    Frequency (Hz): 1 - 101 - 10
    Waveform: Biphasic with Rectangular PulseBiphasic with Rectangular Pulse
    Battery Type: P13 Zinc Air batteries, non-rechargeableP13 Zinc Air batteries, non-rechargeable
    Battery Capacity: 310 mAh310 mAh
    No. x Voltage (V): 3 x 1.4 V3 x 1.4 V
    Pulse Width (ms): 0.9800.980
    Duty Cycle: 2 hours ON / 1 min OFF2 hours ON / 1 min OFF
    Typical Battery Operating Time (hours): 120120
    Operating Temperature: 5 °C to 45 °C5 °C to 45 °C
    Operating Humidity: 40% to 80%40% to 80%
    Environmental Use: Clinics, Hospital, HomeClinics, Hospital, Home
    Sterilization of Electrodes: EtO SterilizationEtO Sterilization
    Re-use: Single use DeviceSingle use Device
    Max Charge Density (µC / cm²) per needle: 67.16 @ 1 kΩ, 8.01 @ 10 kΩ67.16 @ 1 kΩ, 8.01 @ 10 kΩ
    Max Average Power Density (W / cm²): 0.363 @ 1 kΩ, 0.0516 @ 10 kΩ0.363 @ 1 kΩ, 0.0516 @ 10 kΩ
    Net Charge (µC per pulse): 0 (biphasic)0 (Due to biphasic nature of the waveform)
    Software Controlled: YesYes
    Shelf Life: Increased from 6 months (predicate) to 12 months12 months (demonstrated through sterilization and non-clinical performance testing)
    Biocompatibility: PassSubjected to biocompatibility testing (implied pass for clearance)
    Electrical Safety (EMC and safety): PassSubjected to electrical safety (electromagnetic compatibility and safety) testing (implied pass for clearance)
    Performance Bench Testing (120 hours continuous performance): Pass for pulse width, pulse duration, amplitude, current valuesFunctional test performed for 120 hours to monitor continuous performance. Captured pulse width, pulse duration, amplitude, and current values. Performance equivalent to predicate.
    Software Validation: PassSubjected to software validation (implied pass for clearance)

    2. Sample size used for the test set and the data provenance
    The document does not describe a clinical test set with human subjects or patient data. The "test set" discussed refers to non-clinical bench testing.

    • Sample size: Not specified for individual bench tests, but implies a sufficient number of devices were tested for validation.
    • Data provenance: Non-clinical bench testing data, likely conducted by DyAnsys, Inc.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
    Not applicable. There was no clinical test set requiring expert ground truth establishment for diagnostic performance.

    4. Adjudication method for the test set
    Not applicable. No clinical test set and no expert adjudication were described.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
    Not applicable. This device is a nerve stimulator, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
    Not applicable. This device is a nerve stimulator, not an algorithm. Its function is direct stimulation.

    7. The type of ground truth used
    For the non-clinical performance data, the "ground truth" was likely defined by engineering specifications and direct measurements against those specifications (e.g., measuring actual pulse width against the specified 0.980 ms). For establishing substantial equivalence, the "ground truth" was the characteristics, intended use, and indications for use of the predicate device (K211971).

    8. The sample size for the training set
    Not applicable. There is no mention of a training set for an algorithm in this submission.

    9. How the ground truth for the training set was established
    Not applicable. No training set was described.

    Ask a Question

    Ask a specific question about this device

    K Number
    K211971

    Validate with FDA (Live)

    Device Name
    Drug Relief v1
    Manufacturer
    Date Cleared
    2021-11-05

    (134 days)

    Product Code
    Regulation Number
    882.5896
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Drug Relief v1 is a percutaneous nerve field stimulator (PNFS) system, that can be used as an aid to reduce the symptoms of opioid withdrawal, through application to branches of cranial nerves V, VII, IX and X, and the occipital nerves identified by transillumination.

    Device Description

    The Drug Relief v1 is designed to be used as an aid to reduce opioid withdrawal symptoms by the method of cranial electrical stimulation at the auricular stimulation points. The Drug Relief v1 is a wearable, battery-operated device that is designed to administer periodical low level electrical pulses to the ear over five days / 120 hours from the time of activation of the device. The electrical pulse from the device will be delivered to the branches of cranial Nerves on the ear through a set of wire assembly and stimulation needles. Three zinc air batteries with 1.4 V each provide the required stimulation energy for 120 hours. There atimulation electrode and one ground electrode which constitute of a needle and lead/ wire with the snap-fit ring. The stimulation needles are inserted at three specific points, which have the ability to stimulate the cranial nerves. The ground electrode is inserted at one specific point (constant in all treatments) which forms the functional earthing to the device. This constant current source guarantees equivalent stimulation energy regardless of the individual impedance of the skin. The stimulation pattern consists of rectangular pulses with differing inter-pulse intervals and a duty cycle of 2 hours ON/1 minute OFF. A 3-pin connector is provided, which is used to check the output voltage of the device once it is activated and before applying to the patient with any one of the voltage measuring devices available in the market with the appropriate regulatory compliance.

    AI/ML Overview

    The provided text describes the Drug Relief v1 device and its substantial equivalence to a predicate device, Drug Relief (K173861). The document primarily focuses on demonstrating that the new device has similar performance specifications and safety characteristics to the previously cleared predicate.

    Here's an analysis of the provided information concerning acceptance criteria and studies:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of predetermined acceptance criteria with corresponding performance results. Instead, it relies on demonstrating that the "Drug Relief v1 has equivalent performance specifications when compared to the predicate device." This is a common approach for 510(k) submissions, where the new device's performance is compared to a legally marketed predicate rather than against absolute, pre-defined acceptance metrics for efficacy in a clinical setting.

    However, the core of the performance evaluation for this type of device, as indicated, revolves around:

    • Electrical Safety Parameters: Max Charge Density (µC/cm²) and Max Average Power Density (W/cm²).
    • Functional Performance: Pulse width, duty cycle, amplitude, and current values over the intended operating time (120 hours).
    • Biocompatibility: Ensuring materials are safe for patient contact.
    • Software Verification and Validation: Confirming the software functions as intended.
    • Sterilization: Achieving a defined sterility assurance level (SAL).

    The document states that the reported device performance for the Drug Relief v1 shows these characteristics are "identical or near identical" to the predicate device. Specifically, the electrical output parameters are presented in "Table 13.3.2 Comparison of Technical System Characteristics" (pages 6-7).

    System Characteristic / ParameterPredicate Device (Drug Relief K173861) PerformanceSubject Device (Drug Relief v1) PerformanceJustification / Outcome
    Max Charge Density (µC / cm²) per needle65.67 @ 1 kΩ; 7.96 @ 10 kΩ67.16 @ 1 kΩ; 8.01 @ 10 kΩNear identical; minor technological differences do not impact safety/effectiveness.
    Max Average Power Density (W / cm²)0.346 @ 1 kΩ; 0.0509 @ 10 kΩ0.363 @ 1 kΩ; 0.0516 @ 10 kΩNear identical; minor technological differences do not impact safety/effectiveness.
    Pulse ShapeRectangleRectangleIdentical
    Frequency (HZ)1 - 10 (Pulse with modulating frequency)1 - 10 (Pulse with modulating frequency)Identical
    WaveformRectangular PulseRectangle pulseIdentical
    Pulse Width (ms)0.9800.980Identical
    Duty Cycle2 hours ON / 1 min OFF2 hours ON / 1 min OFFIdentical
    Battery Operating Time (hours)120120Identical
    BiocompatibilityPassed (Implied, as K211971 passed)PassedDemonstrated via nonclinical testing.
    Electrical Safety (EMC and Safety)Passed (Implied, as K211971 passed)PassedDemonstrated via nonclinical testing, shown as "near identical" to predicate.
    Software Verification and ValidationPerformedPerformedDemonstrated via nonclinical testing.
    Sterility Assurance Level (SAL)Achieved (Implied, as K211971 passed)AchievedEvidenced by bio-burden, EtO sterilization, and sterility tests per ISO standards.

    2. Sample Size Used for the Test Set and Data Provenance

    The document describes nonclinical testing, including "performance bench testing" and "electrical safety (electromagnetic compatibility and safety)" for the Drug Relief v1 device (page 7). It also mentions "software verification and validation" and "biocompatibility testing."

    • Sample Size: The specific sample sizes for these tests (e.g., how many devices were tested for electrical safety or how many batches for sterilization) are not explicitly stated in the provided text.
    • Data Provenance: The studies are described as nonclinical (bench testing) and likely conducted by the manufacturer (DyAnsys, Inc.). There is no mention of clinical data, human subjects, country of origin related to human data, or retrospective/prospective study design, as this submission is based on demonstrating substantial equivalence through technical and performance comparisons to a predicate device.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable and not provided because the evaluation presented is based on nonclinical testing (bench tests, electrical safety, etc.) and direct comparison to a predicate device's technical specifications. There is no "ground truth" derived from expert interpretation of clinical data in this document.

    4. Adjudication Method for the Test Set

    This is not applicable and not provided as the submission does not involve clinical studies with human readers or diagnostic interpretations requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable and not provided. The Drug Relief v1 is a percutaneous nerve stimulator, not an AI-based diagnostic imaging or analysis device. There is no mention of human readers, AI assistance, or MRMC studies.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This is not applicable and not provided. The device is a nerve stimulator, not an algorithm, and the submission does not refer to AI or algorithmic performance evaluation.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    As this is a nonclinical performance and electrical safety submission, the "ground truth" refers to established engineering standards, electrical measurement protocols, and material science specifications. For example:

    • Electrical Safety: Standards like those for current leakage, output parameters, and EMC.
    • Biocompatibility: ISO standards for biological evaluation of medical devices.
    • Sterilization: ISO standards for sterilization processes and sterility assurance.
    • Functional Performance: The established specifications of the predicate device (Drug Relief K173861) serve as the benchmark for "equivalent performance."

    8. The Sample Size for the Training Set

    This is not applicable and not provided, as there is no mention of an algorithm or AI model requiring a training set in this submission.

    9. How the Ground Truth for the Training Set was Established

    This is not applicable and not provided, for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K173861

    Validate with FDA (Live)

    Device Name
    Drug Relief
    Manufacturer
    Date Cleared
    2018-05-02

    (133 days)

    Product Code
    Regulation Number
    882.5896
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Drug Relief is a percutaneous nerve field stimulatory (PNFS) system, that can be used as an aid to reduce the symptoms of opioid withdrawal, through application to branches of Cranial Nerves V, VII, IX, and X, and the occipital nerves identified by transillumination.

    Device Description

    The Drug Relief™ is designed to aid in the treatment of opiate withdrawal symptoms by the method of electrical stimulation at the auricular stimulation points. The Drug Relief is a wearable, battery-operated device that is designed to administer periodical low-level electrical pulses to the ear over five days / 120 hours (2 hours ON/1 minute OFF) from the time of activation of the device. The electrical pulse from the device will be delivered to the branches of Cranial Nerves on the ear through a set of wire assembly and Stimulation needles. Three zinc air batteries with 1.4 V each provides the required stimulation energy for a maximum of 120 hours. There are three Stimulation electrodes and one ground electrode - which constitute of a needle and lead/ wire with the snap-fit ring. The stimulation needles are inserted at three specific points, which have the ability to stimulate the cranial nerves. The ground electrode is inserted at one specific point (constant in all treatments) which forms the functional earthing to the device. This constant current source guarantees equivalent stimulation energy regardless of the individual impedance of the skin. The stimulation pattern consists of rectangular pulses with differing inter-pulse intervals. A 3-pin connector is provided, which is used to check the output voltage of the device once it is activated and before applying to the patient with any one of the voltage measuring devices available in the market with the appropriate regulatory compliance.

    AI/ML Overview

    The provided text is a 510(k) summary for the medical device "Drug Relief," a percutaneous nerve stimulator for opioid withdrawal. This document focuses on demonstrating substantial equivalence to a predicate device (NSS-2 BRIDGE) rather than presenting a clinical study with detailed acceptance criteria and performance metrics for the device itself.

    Therefore, the information required to answer your request fully, particularly regarding specific acceptance criteria for "device performance" in a clinical setting, sample sizes for test sets, expert ground truth establishment, MRMC studies, and effect sizes, is not present in the provided document. The document primarily describes non-clinical performance and summarizes testing done to show equivalence to a predicate device.

    However, I can extract and present the available information to address as many of your points as possible:


    Summary of Acceptance Criteria and Device Performance (Based on Available Data)

    Based on the provided 510(k) summary, the "acceptance criteria" appear to be primarily focused on demonstrating substantial equivalence to the predicate device (NSS-2 BRIDGE) through performance testing of the device's electrical characteristics and compliance with recognized standards, rather than clinical efficacy metrics.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategoryReported Device Performance / Comparison to Predicate
    Electrical Performance
    Pulse Width0.980mSec (Equivalent to Predicate)
    Operating Time120 hrs (5 days), 2 Hours ON (Periodically) (Equivalent to Predicate)
    FrequencyPulses with modulating frequency (1 Hz - 10 Hz) (Equivalent to Predicate)
    WaveformRectangular pulse (Equivalent to Predicate)
    Max Charge Density (per needle)65.67 @ 1K ohm; 7.96 @ 10K ohm (Compared to Predicate: 48.91 @ 1K ohm; 5.79 @ 10K ohm)
    Max Average Power Density (per needle)0.346 @ 1K ohm; 0.0509 @ 10K ohm (Compared to Predicate: 0.2645 @ 1K ohm; 0.0371 @ 10K ohm)
    Output Voltage Measurement Feasibility3-pin connector provided to measure output voltage. (Feasible on subject device, not on predicate)
    Mechanical/Physical Equivalence
    Device ShapeRectangle (Equivalent to Predicate)
    Device AdhesiveFoam pad connected to adhesive – gel pad (Different from Tegaderm on predicate)
    Needle Dimensions0.42 mm (widthlength) (Slightly different from predicate: 0.5mm width x 2mm length)
    Wire Assembly4 units of wire with snap-fit ring (Different configuration than predicate)
    Sterilization MethodEtO Sterilization (Different from Irradiation (Gamma) on predicate)
    Shelf Life6 Months (Different from 12 months on predicate)
    Compliance with Standards
    Electrical SafetyIEC 60601-1 (Compliance stated)
    EMCIEC 60601-1-2 (Compliance stated)
    BiocompatibilityISO 10993 (Compliance stated for several parts)
    SterilizationISO 11135 and other related ISO standards (Compliance stated)
    Functional Performance
    Continuous PerformanceMonitored for 120 hours (Assumed to meet internal specifications for equivalence)
    Effectiveness"Validate the effectiveness of each unit" (Implied through functional test and comparison to predicate's "effectual output")

    Note: The document explicitly states that "These differences do not affect the safety and effectiveness of the device." This implies that the observed differences were deemed acceptable for substantial equivalence.

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Test Set Sample Size: The document does not specify a sample size for a "test set" in the context of clinical performance or a human study. The "performance testing" described appears to be bench testing on the device units themselves.
    • Data Provenance: Not applicable as no clinical study data for device performance or efficacy is presented. The comparison is against the specifications of a predicate device.

    3. Number of Experts and Qualifications for Ground Truth

    • Not applicable. The ground truth for this submission is based on the technical specifications and performance of a legally marketed predicate device, and compliance with recognized standards. There is no mention of expert-established ground truth for clinical performance.

    4. Adjudication Method for the Test Set

    • Not applicable. No clinical test set requiring adjudication is described.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No evidence of an MRMC study. The document focuses on device characteristics and substantial equivalence, not on human reader performance with or without AI assistance. The device itself is a stimulate, not an AI diagnostic tool.

    6. Standalone (Algorithm Only) Performance

    • Not applicable for an AI algorithm. This device is a percutaneous nerve stimulator. Its "performance" is based on its electrical output and how it delivers stimulation. The summary states that "The Drug Relief device and its components are subjected to performance testing to validate the effectiveness of each unit." This describes a standalone device performance test, but not an algorithm's performance.

    7. Type of Ground Truth Used

    • The "ground truth" for this 510(k) submission is primarily:
      • Predicate Device Specifications: The technical and performance characteristics of the legally marketed predicate device (NSS-2 BRIDGE).
      • Industry Standards: Compliance with recognized international standards (IEC, ISO) for medical devices, electrical safety, EMC, biocompatibility, and sterilization.
      • Bench Testing Results: Measured electrical outputs and functional performance over time.

    8. Sample Size for the Training Set

    • Not applicable. This document describes a physical medical device, not a machine learning or AI model that requires a "training set."

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable. As above, there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1