Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K240961

    Validate with FDA (Live)

    Date Cleared
    2024-08-15

    (129 days)

    Product Code
    Regulation Number
    880.5860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Disposable Pen Injector Assembly is a disposable pen injector designed for single patient use by diabetics for the subcutaneous self-injection of a desired dose insulin. The disposable pen injector assembly uses 3 mL cartridge of HUMALOGTM (U-100, insulin lispro for injection), and a single use detachable and disposable insulin pen needle (supplied separately). The pen injector allows the user to dial the desired dose up to 80 units in 1 unit increments. It is intended for general population.

    Device Description

    The Disposable Pen Injector Assembly is a disposable single-patient use pen injector capable of injecting a dose of up to 80 units of insulin in 1 unit increments. The pen injector is intended for use with 3mL insulin cartridges and single-use, disposable insulin pen needles (supplied separately). After the intended insulin cartridge is used up, the Disposable Pen Injector Assembly should be disposed. The Disposable Pen Injector Assembly is intended for single user only and provided non-sterile.

    AI/ML Overview

    This document describes the FDA's decision regarding the 510(k) premarket notification for the "Disposable Pen Injector Assembly" from Wuxi NEST Biotechnology Co., Ltd. It outlines the device's characteristics, comparison to a predicate device, and performance data used to establish substantial equivalence.

    Here's an analysis of the acceptance criteria and study information provided:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the Disposable Pen Injector Assembly are primarily based on meeting the requirements of ISO 11608-1:2022 Needle-based injection systems for medical use - Requirement and test methods - Part 1: Needle-based injection systems.

    Acceptance Criteria / Performance AspectStandard / RequirementReported Device Performance
    Dose AccuracyISO 11608-1:2022 requirementsMeets ISO 11608-1:2022 requirements. Performance test with FDA-cleared insulin cartridges (3mL) demonstrated compliance.
    BiocompatibilityISO 10993-1: Evaluation and testing within a risk management process ISO 10993-5:2009: Test for in vitro cytotoxicity ISO 10993-10:2021: Tests for skin sensitization ISO 10993-23:2021: Tests for irritationBiocompatibility tests were conducted and verified that the proposed devices are safe for contacting human skin. The device is categorized for skin contact, duration A-limited (<24hrs). Specific results for each test are not detailed but the conclusion is positive.
    Shelf LifeASTM F1980-16: Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical DevicesA five-year shelf life is claimed. Aging studies conducted in accordance with ASTM F1980-16 verified this shelf life and maintained functionality.
    Simulated TransportationASTM D4169:2022A test was conducted according to ASTM D4169:2022 and demonstrated that the device could function as intended after transportation.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify sample sizes for any of the performance tests (Dose Accuracy, Biocompatibility, Shelf Life, Simulated Transportation).

    The data provenance is from non-clinical performance tests conducted by the manufacturer, Wuxi NEST Biotechnology Co., Ltd., in China. The data is retrospective, as it was submitted as part of the 510(k) premarket notification.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This information is not provided in the document. The tests performed are primarily engineering and technical performance tests based on established ISO and ASTM standards, rather than clinical studies requiring expert ground truth for interpretation of medical images or patient outcomes.

    4. Adjudication Method for the Test Set

    This information is not applicable/not provided. The studies are non-clinical performance tests against defined international standards, not studies involving human perception or interpretation where adjudication would be necessary.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The submission explicitly states: "No clinical study is included in this submission."

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study was Done

    This question is not applicable as the device is a physical medical device (pen injector) and not an AI/algorithm-based software device. The performance tests are for the physical device's mechanics and material properties.

    7. The Type of Ground Truth Used

    The ground truth used for these studies is based on:

    • Compliance with international standards: ISO 11608-1:2022 for dose accuracy, ISO 10993 series for biocompatibility, and ASTM standards for shelf life and simulated transportation. These standards define the acceptable performance parameters and test methods.
    • Material properties and physical functionality: The device's ability to precisely deliver insulin, its biological safety with human contact, and its durability under various conditions.

    8. The Sample Size for the Training Set

    This information is not applicable. The device is a physical product, not an AI/machine learning model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable as there is no training set for a physical medical device like this.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1