510(k) Data Aggregation

    K Number
    K230542

    Validate with FDA (Live)

    Date Cleared
    2023-05-31

    (92 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Disposable Nitrile Glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

    Device Description

    The Disposable Nitrile Gloves are non-sterile, single use only, disposable examination gloves intended for medical purposes to be worn by examiners to prevent contamination between the patient and the examiner. The gloves are blue or black color, powder free, nitrile ambidextrous gloves. The gloves are offered in large (L) size, packed in a paper box. The gloves are designed and manufactured in accordance with the ASTM D6319-19 standard.

    AI/ML Overview

    The provided text describes the 510(k) premarket notification for Disposable Nitrile Gloves (ABC-DG01). It outlines the device's characteristics, its intended use, and a comparison to a predicate device (K171422). The submission focuses on demonstrating substantial equivalence through non-clinical testing.

    Here's an analysis of the provided information to address your request:

    Based on the provided document, the "device" in question is Disposable Nitrile Gloves (ABC-DG01). This is a Class I medical device, meaning it is considered low risk and subject to general controls. The acceptance criteria and "study" described are for physical and chemical performance, not for a diagnostic algorithm or AI.

    Therefore, many of the questions regarding AI-specific studies (e.g., sample size for test set, data provenance, expert consensus, MRMC studies, standalone algorithm performance, training set details) are not applicable to this type of medical device submission.

    The "study" refers to non-clinical performance testing conducted to demonstrate that the Disposable Nitrile Gloves meet established standards for patient examination gloves.


    1. Table of Acceptance Criteria and Reported Device Performance

    The relevant information is found in "Table 4. Performance Comparison" and "Summary of Non-Clinical Testing" (specifically the table under section G).

    Test Method (Acceptance Criteria)PurposeAcceptance Criteria (K230542)Reported Device Performance (K230542) (Blue & Black)
    Dimensions (ASTM D6319-19)Evaluate the physical dimension of the gloveWidth 110 ±10mm Length 230mm min Palm - 0.05mm min. Finger - 0.05mm min.Pass 106mm min width 250mm min length Palm – 0.06mm min. Finger – 0.12mm min
    Physical properties (ASTM D6319-19)Evaluate the tensile strength and ultimate elongation before and after agingTensile Strength: Before Aging ≥ 14 MPa, min. After Aging ≥ 14 MPa, min. Elongation: Before Aging 500%, min. After Aging 400%, min.Pass Tensile Strength: Before Aging 19.07MPa, min. After Aging 18.11 MPa, min. Elongation: Before Aging 586.47%, min. After Aging 469.20%, min.
    Freedom from holes (ASTM D5151-19)Detect holes in the glovesNo leakage at sampling level of G-1, AQL 2.5Pass No leakage, 125 of 125 passed of each color
    Residual Powder (ASTM D6124-06)Detect the powder residue in the glove<2mg per glovePass average 0.07 mg per glove

    2. Sample size used for the test set and the data provenance

    The document provides some direct sample sizes for specific tests:

    • Freedom from holes: "125 of 125 passed of each color" (meaning 125 blue gloves and 125 black gloves were tested).
    • For other tests like Dimensions, Physical properties, and Residual Powder, specific sample sizes are not explicitly stated in the provided text, but the results indicate that a sample was tested according to the ASTM standards.
    • Data provenance: Not explicitly stated, but given the company location (Huizhou, China), the testing was likely conducted in China, often by accredited labs specializing in material testing. The data is retrospective in the sense that it was collected as part of product verification before submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not Applicable. This is a physical product (gloves), not a diagnostic algorithm where expert consensus establishes "ground truth." The "ground truth" for these tests are the objective measurements against the specified ASTM standards.

    4. Adjudication method for the test set

    Not Applicable. There is no "adjudication" in the sense of reconciling expert opinions for a diagnostic outcome. The tests are objective measurements (e.g., measuring thickness, tensile strength, liquid leakage, residual powder mass). The results pass or fail based on the numerical comparison to the acceptance criteria.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not Applicable. This is for a physical medical product (gloves), not an AI/Software as a Medical Device (SaMD). No "human readers" or "AI assistance" are involved in the performance testing of gloves.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not Applicable. This is for a physical medical product (gloves), not an AI/SaMD. There is no algorithm.

    7. The type of ground truth used

    The "ground truth" for these physical and chemical tests is defined by the objective measurement against established engineering standards (ASTM D6319-19, ASTM D5151-19, ASTM D6124-06). For example, for "freedom from holes," the ground truth is simply whether liquid leaks or not according to the specified test method. Similarly, for "tensile strength," the ground truth is the measured force required to break the glove.

    8. The sample size for the training set

    Not Applicable. This is not an AI/machine learning device. No "training set" is used.

    9. How the ground truth for the training set was established

    Not Applicable. As above, no training set or AI model is involved.

    Ask a Question

    Ask a specific question about this device

    K Number
    K211781

    Validate with FDA (Live)

    Date Cleared
    2021-09-29

    (112 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Disposable nitrile gloves are disposable devices intended for medical purposes that are worn on the examiner's hands to prevent contamination between patient and examiner.

    Device Description

    The proposed device is Powder Free Disposable nitrile gloves. The proposed device is blue. The design of proposed device is addressing the standards as ASTM D6124, ASTM D5151, and ASTM D6319. The proposed device is non-sterile.

    AI/ML Overview

    The provided text is a 510(k) Premarket Notification from the FDA for "Disposable Nitrile Gloves." This document primarily focuses on demonstrating that the new device is "substantially equivalent" to a legally marketed predicate device, rather than proving that it meets acceptance criteria for an AI/algorithm-based diagnostic device through a study with human readers.

    Therefore, the information requested in the prompt, such as acceptance criteria for AI algorithms, sample sizes for AI test sets, expert adjudication methods, MRMC studies, standalone AI performance, ground truth establishment for training sets, and the number and qualifications of experts for ground truth, is not applicable to this document.

    The document describes performance testing for a physical medical device (gloves) against established standards for physical properties and biocompatibility.

    Here's an analysis based on the information available in the document, framed to address the prompt's structure where possible, but highlighting the inapplicability of AI-specific criteria.


    Analysis of Acceptance Criteria and Device Performance (for Disposable Nitrile Gloves)

    The document, K211781, describes the submission for Disposable Nitrile Gloves. The "acceptance criteria" in this context refer to the performance specifications and safety standards for the physical product, not for an AI/algorithm. The study presented is a series of non-clinical bench tests and biocompatibility assessments demonstrating the gloves meet these standards and are substantially equivalent to a predicate device.

    1. Table of Acceptance Criteria and the Reported Device Performance

    The acceptance criteria are derived from recognized standards (ASTM and ISO). The reported device performance shows the proposed device meets or exceeds these criteria.

    Test Methodology / StandardPurposeAcceptance CriteriaReported Device Performance
    Biocompatibility Tests
    ISO 10993-10:2010 (Irritation & Skin Sensitization)Assess potential contact hazards (skin/mucosal irritation, eye irritation, skin sensitization).Skin Sensitization: Grades less than 1. Skin Irritation: Primary Irritation Index 0-0.4 (Negligible).All grades are 0. All animals survived with no abnormal signs. Primary irritation index is 0 (negligible response).
    ISO 10993-5:2009 (In Vitro Cytotoxicity)Assess in vitro cytotoxicity.Viability reduced to <70% of blank indicates cytotoxic potential.Viab.% of 100% test article extract is 13.7%. Indicates potential toxicity to L-929 in MTT method. (Note: This is a deviation from the "positive" acceptance criteria, but later mitigated by systemic toxicity test).
    ISO 10993-11:2017 (Systemic Toxicity)Evaluate potential for medical device materials to cause adverse systemic reactions.Within 72h, no greater toxicosis response in testing group than control group.No evidence of systemic toxicity from the extract. (This mitigates the cytotoxicity finding, suggesting it's not systemically toxic).
    Physical Properties Tests
    ASTM D6124-06 (Reapproved 2017), Residual Powder on Medical GlovesDetermine amount of residual powder.Powder residue limit of 2.0 mg.0.06 mg/glove. (Meets criterion).
    ASTM D5151-19, Detection of Holes in Medical GlovesCover detection of holes in medical gloves.Samples: 125 gloves. AQL: 2.5 (ISO 2859). Criterion ≤ 7 gloves for water leakage.No glove water leakage found. (Meets criterion).
    ASTM D6319-19, Nitrile Examination Gloves for Medical ApplicationCovers general requirements (sterility, freedom from holes, dimensions, thickness, physical properties).Sterility: No need (for this non-sterile device). Freedom from holes: AQL 2.5. Dimensions: (e.g., S: width 80±10mm, L ≥220mm; M: width 95±10mm, L ≥230mm; L: width 110±10mm, L ≥230mm; XL: width 120±10mm, L ≥230mm). Thickness: Finger ≥0.05 mm, Palm ≥0.05 mm. Physical Properties: (Before Aging) Tensile strength ≥14MPa, Ultimate Elongation ≥500%. (After Aging) Tensile strength ≥14MPa, Ultimate Elongation ≥400%.N.A. for sterility. Freedom from holes: Met (as per D5151). Dimensions: S: width 84-88 mm, L 251-261 mm; M: width 91-95 mm, L 252-256 mm; L: width 103-106 mm, L 250-256 mm; XL: width 118-121mm, L 253-261 mm. (All meet D6319-19 requirements, though different from predicate detailed in Table 2). Thickness: Finger 0.14-0.15 mm, Palm 0.08-0.09 mm. (All meet D6319-19 requirements). Physical Properties: (Before Aging) Tensile strength 14.67-26.48 MPa, Ultimate Elongation 585.48% - 846.81%. (After Aging) Tensile strength 14.40-22.23 MPa, Ultimate Elongation 527.73% - 775.00%. (All meet D6319-19 requirements).

    2. Sample size used for the test set and the data provenance

    • Sample Size:
      • For Freedom from Holes (ASTM D5151-19): 125 gloves.
      • For other physical tests (Dimensions, Thickness, Physical Properties, Powder Content): Not explicitly stated as a "test set" sample size but implied by compliance with ASTM D6319-19, which dictates sampling plans.
      • For Biocompatibility Tests (ISO 10993): Animal study sample sizes (for irritation, sensitization, systemic toxicity) are not explicitly stated, but are standard for these in vivo tests. Cell cultures are used for cytotoxicity.
    • Data Provenance: The testing was conducted by or on behalf of "Jiangsu Jinke Protective Equipment Co., Ltd" in China. The data would be considered prospective, as it was generated specifically for this submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not Applicable. This is a physical device. "Ground truth" in this context refers to the measured physical and biological properties of the gloves as determined by validated test methods and laboratory analysis, not by human expert interpretation of images or other data. No human experts were used to establish "ground truth" for diagnostic purposes.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable. As above, this is for a physical medical device. Not an AI diagnostic study.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. No AI component; therefore, no MRMC study.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable. No AI component; therefore, no standalone AI performance testing.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • For physical properties, the ground truth is established by objective measurements using standardized laboratory equipment and methods (e.g., tensile testers for strength, calipers for dimensions, water for hole detection, chemical analysis for powder content).
    • For biocompatibility, the ground truth is established by biological responses observed in validated in vitro (cytotoxicity) or in vivo (irritation, sensitization, systemic toxicity) animal models according to ISO standards.

    8. The sample size for the training set

    • Not Applicable. This is not an AI/machine learning device; there is no training set in the context of AI.

    9. How the ground truth for the training set was established

    • Not Applicable. As there is no training set for an AI model, the concept of establishing ground truth for it does not apply.

    Conclusion from the Document:

    The document successfully demonstrates that the Disposable Nitrile Gloves meet the performance specifications outlined in relevant ASTM and ISO standards for physical properties and biocompatibility, particularly when compared to the predicate device. The one noted 'potential cytotoxicity' was mitigated by a successful systemic toxicity test. The regulatory submission hinges on substantial equivalence rather than AI performance criteria.

    Ask a Question

    Ask a specific question about this device

    K Number
    K212009

    Validate with FDA (Live)

    Date Cleared
    2021-09-15

    (79 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Disposable nitrile gloves are disposable devices intended for medical purposes that are worn on the examiner's hands to prevent contamination between patient and examiner.

    Device Description

    The proposed device is Powder Free Disposable nitrile gloves. The proposed device is blue. The design of proposed device is addressing the standards as ASTM D6124,ASTM D5151, and ASTM D6319. The proposed device is non-sterile.

    AI/ML Overview

    The provided document describes the non-clinical performance testing for "Disposable Nitrile Gloves" (K212009) to demonstrate substantial equivalence to a predicate device (K171422).

    Here's the breakdown of the acceptance criteria and the study results:

    1. Table of Acceptance Criteria and Reported Device Performance

    | No. | Test Methodology / Standard | Purpose | Acceptance Criteria | Reported Device Performance (Results) |
    |-----|-------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-
    | 1 | ISO 10993-10:2010 (Skin Sensitization Test) | Assesses possible contact hazards from chemicals released from medical devices, that may produce skin sensitization. | Grades less than 1, otherwise sensitization. | All grades are 0. All animals survived and no abnormal signs were observed. |
    | 2 | ISO 10993-10:2010 (Skin Irritation Test) | Assesses possible contact hazards from chemicals released from medical devices, that may produce skin irritation. | Primary irritation index of 0-0.4 for negligence. | Primary irritation index is 0. The response of the proposed device was categorized as negligible. |
    | 3 | ISO 10993-5:2009 (Cytotoxicity) | Assesses the in vitro cytotoxicity of medical devices. | Viability percentage (Viab.%) of the 100% extract of the test article is the final result. If viability is reduced to <70% of the blank, it has cytotoxic potential. | Viab.% of 100% test article extract is 71.75%. This means the proposed device has no potential toxicity to L-929 in the MTT method. |
    | 4 | ASTM D6124-06 (Reapproved 2017) (Residual Powder on Medical Gloves) | Designed to determine the amount of residual powder (or filter-retained mass) found on medical gloves. | Powder residue limit of 2.0 mg/glove. | 0.06 mg/glove. |
    | 5 | ASTM D5151-06 (Reapproved 2015) (Detection of Holes) | Covers the detection of holes in medical gloves. | Sample number: 125 gloves. AQL: 2.5 (ISO 2859). Criterion ≤ 7 gloves for water leakage. | No glove water leakage found. |
    | 6 | ASTM D6319-10 (Reapproved 2015) (Nitrile Examination Gloves) | Covers certain requirements for nitrile rubber gloves used in medical examinations and diagnostic and therapeutic procedures. | Freedom from holes: Refer to No. 5. Dimensions: S: width 80±10mm, Length ≥220 mm; M: width 95±10mm, Length ≥230 mm; L: width 110±10mm, Length ≥230 mm; XL: width 120±10mm, Length ≥230 mm. Thickness: Finger ≥0.05 mm, Palm ≥0.05 mm. Physical properties (Before Aging): Tensile strength ≥14MPa, Ultimate Elongation ≥500%. Physical properties (After Accelerated Aging): Tensile strength ≥14MPa, Ultimate Elongation ≥400%. | Dimensions: S: width: 82-86 mm, Length 238-245 mm; M: width 97-99 mm, Length 243-249 mm; L: width 108-111 mm, Length 244-254 mm; XL: width 115-118 mm, Length 237-246 mm. Thickness: Finger 0.13-0.15 mm, Palm 0.08 mm. Physical properties (Before Aging): Tensile strength 14.12-22.10 MPa, Ultimate Elongation 537.05%-907.44%. Physical properties (After Accelerated Aging): Tensile strength 14.46-20.20 MPa, Ultimate Elongation 542.91%-856.03%. |

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Sizes:

      • Skin Sensitization, Skin Irritation, and Cytotoxicity: The document states "All grades are 0. All animals were survived and no abnormal signs were observed during the study." for sensitization, and "The primary irritation index is 0." for irritation, and "Viab.% of 100% test article extract is 71.75%" for cytotoxicity. While it indicates successful testing, the specific numerical sample size for these biological evaluation tests (e.g., number of animals or cell cultures) is not explicitly provided in the summary.
      • Freedom from Holes (ASTM D5151-06): 125 gloves.
      • Residual Powder (ASTM D6124-06): Not specified, but implied to be sufficient for a quantitative measurement (0.06 mg/glove).
      • Dimensions and Physical Properties (ASTM D6319-10): Specific sample sizes for each dimension, thickness, and physical property measurement are not provided in this summary, but the results for ranges are given.
    • Data Provenance: The document does not explicitly state the country of origin or whether the studies were retrospective or prospective. Given that this is a 510(k) summary for a Chinese manufacturer (JIANGSU SHENGLIJIE SAFETY PRODUCTS CO.,LTD), it is highly probable that the testing was conducted in China. These types of performance studies are typically prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This information is not applicable to this type of device and study. The ground truth for these tests (e.g., irritation index, viability percentage, presence of holes, tensile strength) is established by standardized laboratory procedures and measurements, not by expert consensus or interpretation of images.

    4. Adjudication Method for the Test Set

    This information is not applicable. As stated above, the "ground truth" for these performance tests is determined by objective, standardized measurements rather than human interpretation that would require adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This type of study is relevant for diagnostic imaging devices where human readers interpret and a comparative effectiveness with and without AI assistance is measured. This document is for disposable nitrile gloves, which do not involve such a scenario.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    No, a standalone (algorithm only) study was not done. This concept applies to AI/ML software as a medical device. Disposable nitrile gloves are a physical medical device, and their performance is evaluated through physical and biological testing, not through algorithms.

    7. The Type of Ground Truth Used

    The ground truth used for these tests is objective, standardized measurements and laboratory observations according to established international and national standards (ISO 10993, ASTM D6124, ASTM D5151, ASTM D6319). This includes:

    • Quantitative measurements (e.g., primary irritation index, viability percentage, powder content, dimensions, tensile strength, elongation).
    • Qualitative observations based on standardized protocols (e.g., presence/absence of abnormal signs for sensitization, water leakage for holes).

    8. The Sample Size for the Training Set

    Not applicable. This device (disposable nitrile gloves) does not utilize a training set in the context of machine learning or AI models. The testing described is for physical and biological performance, not for developing an algorithm.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. There is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K210276

    Validate with FDA (Live)

    Date Cleared
    2021-04-16

    (74 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Disposable Nitrile Gloves are intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.

    Device Description

    The subject device is powder free nitrile examination gloves. The subject device is blue. The subject device is non-sterile.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification summary for Disposable Nitrile Gloves (K210276). It is not for an AI/ML medical device, but rather for a physical medical device (gloves). Therefore, the information requested in your prompt (e.g., sample size for test set, number of experts for ground truth, MRMC study, training set details) is not applicable to this document.

    The document discusses the substantial equivalence of the Disposable Nitrile Gloves to a predicate device (K171422) based on performance testing and adherence to recognized standards.

    Here's an attempt to interpret the provided content in the context of your request, while acknowledging that it doesn't fit the typical AI/ML device evaluation structure:

    1. A table of acceptance criteria and the reported device performance:

    The document outlines comparisons between the subject device (K210276) and the predicate device (K171422), effectively serving as the acceptance criteria for substantial equivalence. The reported device performance is demonstrated by its compliance with the listed standards.

    Acceptance Criteria (Common to Subject & Predicate)Reported Device Performance (K210276)
    General Characteristics
    Product Code: LZALZA
    Regulation No: 21CFR880.625021CFR880.6250
    Class: II
    Intended Use: Medical purposes; worn on examiner's hands to prevent contamination between patient and examiner.Same
    Powdered/Powered free: Powdered freePowdered free
    Design Feature: AmbidextrousAmbidextrous
    Labeling Information: Single-use indication, powder free, device color, device name, glove size and quantitySingle-use indication, powder free, device color, device name, glove size and quantity, Nitrile Glove Powder Free Blue, Non-Sterile
    Dimensions (Compliance with ASTM D6319-19)
    Length, mm: 230 min (for predicate)230 mm (S, M, L, XL) with ±10 tolerance
    Width, mm: 75, 85, 95, 105, 115 (for predicate) with ±5 tolerance85 (S), 95 (M), 110 (L), 115 (XL) with ±10 tolerance
    Thickness, mm: Finger 0.05 min0.05 min
    Thickness, mm: Palm 0.05 min0.05 min
    Physical Properties
    Before Aging: Tensile Strength 14MPa, min14MPa, min
    Before Aging: Ultimate Elongation 500% min500% min
    After Aging: Tensile Strength 14MPa, min14MPa, min
    After Aging: Ultimate Elongation 400%min400%min
    Freedom from Holes: Comply with ASTM D6319, be free from holes when tested in accordance with ASTM D5151 AQL=2.5Comply with ASTM D6319, be free from holes when tested in accordance with ASTM D5151 AQL=2.5
    Powder Content: Meet requirements of ASTM D61240.01 mg per glove
    Safety/Biocompatibility (Compliance with ISO 10993)
    Material: NitrileNitrile
    Irritation: Not an irritant (comply with ISO10993-10)Not an irritant (under the conditions of the study)
    Sensitization: Not a sensitizer (comply with ISO10993-10)Not a sensitizer (under conditions of the study)
    Cytotoxicity: Device extract is not cytotoxic (comply with ISO10993-5)Device extract is not cytotoxic (under conditions of the study)

    2. Sample size used for the test set and the data provenance:

    • Sample Size: The document does not specify exact sample sizes for each test conducted (e.g., number of gloves tested for tensile strength, holes, etc.). It only states that "Non-clinical tests were conducted to verify that the proposed device met all design specifications." The standards themselves (e.g., ASTM D6319, ASTM D5151) would define the required sample sizes for their respective tests, but these details are not present in this summary.
    • Data Provenance: The tests were non-clinical performance tests, usually performed in a lab setting by the manufacturer or a contracted lab. The manufacturer is Fujian Ercon Medical Management Co., Ltd. in China. The provenance is internal testing, not patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not Applicable. This is not a study requiring expert readers or ground truth establishment in the context of interpreting medical images or data. The "ground truth" for these tests is defined by the technical specifications and standards (e.g., a hole is present or absent according to ASTM D5151 criteria).

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not Applicable. As this is not an interpretive/diagnostic study, no adjudication method for human readers is relevant. The tests are laboratory measurements.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not Applicable. This is a physical medical device (gloves), not an AI/ML diagnostic aid. There were no human readers or AI assistance involved.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not Applicable. This is a physical medical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The "ground truth" for this device's performance is objective measurements and compliance with industry standards (e.g., ASTM D6319-19 for glove specifications, ASTM D5151-19 for holes, ISO 10993 for biocompatibility). It is based on pre-defined engineering specifications and test protocols.

    8. The sample size for the training set:

    • Not Applicable. This is not an AI/ML device; there is no training set.

    9. How the ground truth for the training set was established:

    • Not Applicable. As there is no training set for an AI/ML model, this question is not relevant.

    In summary, the provided document details the regulatory submission for a physical medical device, not an AI/ML diagnostic tool. Therefore, many of the questions related to AI/ML study design are not applicable. The device's acceptance is based on non-clinical performance testing confirming its compliance with recognized standards and substantial equivalence to a legally marketed predicate device.