510(k) Data Aggregation

    K Number
    K241498

    Validate with FDA (Live)

    Date Cleared
    2024-11-18

    (174 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Diode laser Treatment System (Model: K18) is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.

    Device Description

    Diode laser Treatment System adopts Gold standard 808nm semi-conductor laser, based on selective photothermolysis principle, through specific wavelength, penetrate epidermis into dermis, optical energy was absorbed and translated into heat energy with restraining hair follicle tissues, and produce photothermal effect. It takes laser energy in hair follicle with rich in melanin, surrounding tissues absorb less even no energy, reach to restrained melanin of hair follicles and remove hair. The Diode laser Treatment System utilize a semiconductor diode with invisible infrared radiation as a laser source to emit 808nm wavelength laser which is absorbed by melanin. The laser power is delivered to the treatment area via laser handpiece. The emission laser is activated by a foot-switch.

    AI/ML Overview

    This document is a 510(k) Pre-Market Notification from the FDA, describing the substantial equivalence of a Diode Laser Treatment System (Model: K18) to a predicate device. It does not contain information about acceptance criteria and a study proving a device meets these criteria in the context of an Artificial Intelligence (AI) / Machine Learning (ML) enabled medical device.

    The entire document focuses on demonstrating that the proposed device, a Diode Laser Treatment System, is substantially equivalent to an existing predicate device based on non-clinical tests (electrical safety, EMC, laser safety, and biocompatibility).

    Here's why the requested information cannot be extracted from this document:

    • No AI/ML Component: The device described is a physical laser treatment system, not an AI/ML software or system. Therefore, there are no AI/ML specific acceptance criteria or studies to evaluate its performance in the way described in the request.
    • No Clinical Study: The document explicitly states: "No clinical study is included in this submission." This means there is no data from patient trials or diagnostic accuracy studies, which are typically used to establish performance for AI/ML devices.
    • Focus on Substantial Equivalence: The purpose of a 510(k) summary for a traditional medical device like this is to show that it is as safe and effective as a legally marketed predicate device. This is achieved through engineering tests and comparison of specifications, not through performance metrics against a human "ground truth" or expert consensus (as would be the case for AI/ML).

    Therefore, I cannot provide the requested information from the provided text for the following reasons:

    1. A table of acceptance criteria and the reported device performance: Not applicable. Performance is established through engineering and safety standards, not clinical accuracy metrics.
    2. Sample size used for the test set and the data provenance: Not applicable. No clinical test set.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. No clinical ground truth is established.
    4. Adjudication method: Not applicable.
    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done: Not applicable.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
    7. The type of ground truth used: Not applicable.
    8. The sample size for the training set: Not applicable.
    9. How the ground truth for the training set was established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K182924

    Validate with FDA (Live)

    Date Cleared
    2019-04-15

    (178 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    0.52876712328 - 120
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Diode Laser Treatment System is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.

    Device Description

    The Diode Laser Treatment System consists of the main unit and a hand piece. The system uses a diode laser as an active medium placed in an optical cavity to produce amplified beam at the wavelength of 808 nm. A microprocessor is used to control electronics for the front panel. A self-contained water cooling system is built into the power supply unit. The device provides 36 working modes, which are six modes for men and six modes for women. The men or women mode includes face, armpit, arm, body, bikini, leg mode respectively for different treatment part. The diode laser operates in a pulsed mode with a fixed pulse width and fixed pulse duration of the pulse train for each mode. The number of pulses can be adjusted within the preset range.

    AI/ML Overview

    This document describes a 510(k) premarket notification for the Weifang KM Electronics Co., Ltd. Diode Laser Treatment System. The notification aims to demonstrate substantial equivalence to legally marketed predicate devices for hair removal and permanent hair reduction.

    Based on the provided text, a clinical study proving the device meets specific performance-based acceptance criteria for hair reduction was not described. The document focuses on demonstrating substantial equivalence primarily through technological characteristics and non-clinical performance data (biocompatibility, electrical safety, EMC, and laser safety).

    Here's an analysis based on your requested information:

    1. A table of acceptance criteria and the reported device performance

    The provided text does not explicitly state performance-based acceptance criteria for hair reduction (e.g., a percentage of hair reduction required) or reported device performance in terms of clinical efficacy. Instead, it defines "Permanent hair reduction" as "the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime." This is a definition rather than an acceptance criterion with a specific threshold.

    The document does include acceptance criteria and reported performance for non-clinical testing:

    Acceptance CriteriaReported Device Performance
    Biocompatibility:
    ISO 10993-1:2009 for "Surface – Mucosal Membrane" with contact duration "< 24 hours"All evaluation acceptance criteria were met for Cytotoxicity, Irritation, and Sensitization.
    Electrical Safety:
    IEC 60601-1:2012The system has been tested to comply with this standard.
    IEC 60601-2-22:2007The system has been tested to comply with this standard.
    Laser Safety:
    IEC 60825-1: 2007The system has been tested to comply with this standard.
    Electromagnetic Compatibility (EMC):
    IEC 60601-1-2:2007The system has been tested to comply with this standard.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    There is no information provided in the document about a clinical test set, its sample size, or data provenance related to hair reduction efficacy. The non-clinical tests mentioned (biocompatibility, electrical safety, EMC, laser safety) do not typically involve a patient sample size or data provenance in the context of clinical efficacy.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    Since no clinical test set for assessing hair reduction efficacy is described, there is no information provided regarding experts used to establish ground truth.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Since no clinical test set for assessing hair reduction efficacy is described, there is no information provided regarding an adjudication method.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    The device is a Diode Laser Treatment System for hair removal and not an AI-assisted diagnostic or interpretation device that would typically involve a multi-reader multi-case (MRMC) study or human readers. Therefore, no such study was done or reported in this document.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This point is not applicable to a laser treatment system that performs a physical action (hair removal). The "standalone" performance for such a device would refer to its inherent operational capabilities, which are covered by the non-clinical tests (electrical safety, laser safety, etc.) but not in the context of algorithm-only performance.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the non-clinical tests, the "ground truth" is defined by the requirements of the specific international standards (ISO, IEC). For example, for biocompatibility, the ground truth is whether the device materials elicit a toxic, irritant, or sensitizing response according to the defined test methods. As no clinical efficacy study is described, no clinical ground truth (like pathology or outcomes data for hair reduction) is mentioned.

    8. The sample size for the training set

    The document describes a physical medical device (Diode Laser Treatment System), not an AI/ML algorithm that would require a "training set." Therefore, this is not applicable, and no information is provided.

    9. How the ground truth for the training set was established

    As the device is not an AI/ML algorithm requiring a training set, this is not applicable, and no information is provided.