510(k) Data Aggregation

    K Number
    K193262

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-03-09

    (104 days)

    Product Code
    Regulation Number
    892.1740
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DR 600 system is a General Radiography X-ray imaging system used in hospitals, clinics and medical practices by radiographers, radiologists and physicists to make, process and view static X-ray radiographic images of the skeleton (including skull, spinal column and extremities), chest, abdomen and other body parts on adult, pediatric or neonatal patients.

    In addition, the system provides the Agfa tomosynthesis option, which is intended to acquire tomographic slices of human anatomy and to be used with Agfa DR X-ray systems. Digital tomosynthesis is used to synthesize tomographic slices from a single tomographic sweep.

    Applications can be performed with the patient in the sitting, standing or lying position.

    This system is not intended for mammography applications.

    Device Description

    The DR 600 with Tomosynthesis is a tomographic and solid state x-ray system (product codes IZF and MQB) intended to capture tomographic slices and static images of the human body. The DR 600 with Tomosynthesis is a ceiling mounted tomographic and general radiographic system that consists of a tube and operator console with a motorized patient table and/or wall stand. The DR 600 with Tomosynthesis uses Agfa's NX workstation with MUSICA2 ™ image processing and flat-panel detectors of the scintillator-photodetector type (Cesium Iodide - CsI or Gadolinium Oxysulfide - GOS). It is capable of replacing other direct radiography, tomography, image intensifying tubes and TV cameras, including computed radiography systems with conventional or phosphorous film cassettes.

    AI/ML Overview

    The provided text describes the Agfa DR 600 with Tomosynthesis device and its K193262 510(k) submission. However, it does not contain specific acceptance criteria or a detailed clinical study demonstrating the device's meeting of these criteria. The document focuses on showing substantial equivalence to predicate devices primarily through technological characteristics and bench testing, not through comparative clinical effectiveness studies with explicit acceptance criteria for diagnostic performance.

    Therefore, many of the requested details about acceptance criteria, clinical study design, sample sizes, ground truth establishment, expert qualifications, and MRMC studies are not present in the provided text. The document explicitly states: "No clinical trials were performed in the device. No animal or clinical studies were performed in the development of the new device."

    Based on the available information, here's what can be extracted and what is missing:


    Acceptance Criteria and Device Performance (as inferred from the document's approach to substantial equivalence):

    Since no specific acceptance criteria for diagnostic performance (e.g., sensitivity, specificity, AUC) are presented, the "acceptance criteria" for this 510(k) appear to be primarily focused on demonstrating substantial equivalence to predicate devices through technical specifications, image quality evaluations (bench testing), and compliance with various electrical safety, EMC, and software standards.

    Acceptance Criteria (Inferred from Document)Reported Device Performance
    1. Technological Characteristics are Identical/Equivalent to Predicates: - Communications (DICOM) - Flat Panel Detectors (type, material, sizes, pixel size, dynamic range) - Operator Workstation (Agfa NX) - Image Processing (MUSICA DTS, MUSICA2, MUSICA3/3+) - Operating System (Windows 7, 8, 8.1, 10) - Display System (Separately cleared medical display) - Power Supply - Generators2. Indication for Use statement is consistent/identical to predicates.3. Performance/Functionality as Intended: - Confirmed functions and performs as intended. - Supports a tomographic workflow and Smart Dr visualization (including adult and pediatric patients).4. Image Quality Equivalent to Predicate: - For both adult and pediatric patients.5. Software Validation: - Verification and validation plans confirmed. - Risk assessment shows no unacceptable risks.6. Electrical Safety and EMC Compliance: - Adherence to specified IEC standards (60601-1, 60601-1-2, 60601-1-3, 60601-1-6, 60601-2-28, 60601-2-54). - Compliance with FDA Subchapter J mandated performance standard 21 CFR 1020.30 and 1020.31.7. Quality Management System Compliance: - Adherence to ISO 13485:2015, ISO 14971:2012, ACR/NEMA PS3.1-3.20 (DICOM).1. Technological Characteristics: - Communications: Same as both predicates (DICOM). - Flat Panel Detectors: Same as both predicates (Flat Panel Detectors, GOS/CsI Scintillator, various sizes 17x17, 14x17, 10x10 in., 148 µm pixel size in primary predicate, 139 µm in other, 16 bit dynamic range in primary predicate, 14 bit in other). The new device shares characteristics with both, indicating equivalence. - Operator Workstation: Same as both predicates (Agfa NX). - Image Processing: MUSICA DTS, MUSICA2, MUSICA3/3+. The addition of tomographic image processing is identical to the DR 800 (K183275) primary predicate device. - Operating System: Same as predicate K183275 (Windows 7, 8, 8.1, 10). - Display System: Same as both predicates (Separately cleared medical display K051901). - Power Supply: Same as predicate K152639 (50-60 Hz, 380/400/415/440/480V + 10%). - Generators: Same as predicate K183275 (Choice of three models: 50, 65KW, 80 KW). - Overall: "Principles of operation and technological characteristics of the new and predicate devices are the same."2. Indication for Use: "The DR 600 system is a General Radiography X-ray imaging system... In addition, the system provides the Agfa tomosynthesis option... Digital tomosynthesis is used to synthesize tomographic slices from a single tomographic sweep." This is stated to be "virtually identical" to K152639 with the tomosynthesis addition from K183275.3. Performance/Functionality: - "Technical and acceptance testing was completed on the DR 600 in order to confirm the medical device functions and performs as intended. All deviations or variances are documented... All design input requirements have been tested and passed." - "Functionality evaluations were conducted with three qualified radiographers. Usability and TBD. The results of these tests fell within the acceptance criteria for the DR 600; therefore, the DR 600 supports a tomographic workflow and Smart Dr visualization including adult and pediatric patients."4. Image Quality: "Image quality bench tests were conducted in support of this 510(k) submission in which anthropomorphic adult and pediatric images taken with the DR 600 and the primary predicate device, DR 800 (K183275) were compared to ensure substantial equivalency. The test results indicated the image processing of the DR 600 passed the acceptance criteria and was equal to the image processing for the primary predicate, DR 800 (K183275) device for both adult and pediatric patients."5. Software Validation: "Verification and validation testing confirmed the device meets performance, safety, usability and security requirements... For the NX22 (NX Nomad) software there are a total of 342 risks in the broadly acceptable region and 27 risks in the ALARP region with only one of these risks identified. Zero risks were identified in the Not Acceptable Region."6. Electrical Safety and EMC Compliance: Document states compliance with all listed IEC standards and FDA performance standards.7. Quality Management System Compliance: Document states adherence to all listed ISO and other standards.

    Study Details (Based on the provided text):

    1. Sample sizes used for the test set and the data provenance:

      • Test Set: No specific numerical sample size is mentioned for image quality evaluations beyond "anthropomorphic adult and pediatric images." The document mentions "functionality evaluations were conducted with three qualified radiographers," but this refers to human user testing of workflow and usability, not diagnostic image performance.
      • Data Provenance: Not explicitly stated, but likely retrospective as it refers to comparisons of images taken with the new device and a predicate device. The comparison of anthropomorphic phantom images suggests a controlled laboratory setting.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • The document states "Laboratory data and image quality evaluations conducted with internal and independent specialists confirm that performance is equivalent to the predicates." It also mentions "clinical image quality evaluations for adults and pediatric patients" and "functionality evaluations were conducted with three qualified radiographers."
      • Number of Experts: At least "three qualified radiographers" for functionality, and "internal and independent specialists" for image quality, but exact numbers or specific qualifications (e.g., years of experience, board certification) are not detailed.
      • Qualifications: "Qualified radiographers" and "internal and independent specialists."
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • No adjudication method is described for the image quality evaluations or other performance tests. The comparison to predicates implies direct visual or quantitative comparison by specialists.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No MRMC comparative effectiveness study was done. The document explicitly states: "No clinical trials were performed in the device. No animal or clinical studies were performed in the development of the new device."
      • Therefore, no effect size of human readers improving with AI assistance is provided as this type of study was not conducted. The device in question is an imaging system, not an AI-based diagnostic assistance tool.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • The "performance data including clinical image quality evaluations for adults and pediatric patients" involved the system's output. The "image processing of the DR 600 passed the acceptance criteria and was equal to the image processing for the primary predicate, DR 800 (K183275) device for both adult and pediatric patients." This implies an evaluation of the algorithm's output (image quality) without necessarily focusing on a human-in-the-loop diagnostic task. So, in essence, standalone image quality performance was evaluated against a predicate.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The "ground truth" for image quality evaluation appears to be comparison to a cleared predicate device's image quality, as judged by "internal and independent specialists," using anthropomorphic phantoms. There is no mention of pathological confirmation or patient outcomes for establishing ground truth, as it was not a clinical trial.
    7. The sample size for the training set:

      • This device is an X-ray system with image processing, not a machine learning/AI algorithm that requires a "training set" in the traditional sense of AI development. The software capabilities (MUSICA DTS, MUSICA2, MUSICA3/3+) are described as being identical to previously cleared versions in predicate devices. Therefore, a "training set" for a new AI model is not applicable here.
    8. How the ground truth for the training set was established:

      • Not applicable, as no new AI model training set is described. The image processing algorithms are identical to those previously cleared.
    Ask a Question

    Ask a specific question about this device

    K Number
    K152639

    Validate with FDA (Live)

    Device Name
    DR 600
    Date Cleared
    2015-12-11

    (87 days)

    Product Code
    Regulation Number
    892.1680
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DR 600 is a GenRad X-ray imaging system used in hospitals, clinics and medical practices by radiographers, radiologists and physicists to make, process and view static X-ray radiographic images of the skeleton (including skull, spinal column and extremities), chest, abdomen and other body parts on adults and pediatric patients. Applications can be performed with the patient in sitting, standing or lying position.

    The DR 600 is not intended for use in Mammography applications.

    Device Description

    Agfa's DR 600 is a solid state x-ray system, a direct radiography (DR) system (product code MOB) intended to capture images of the human body. The device is a combination of a conventional x-ray system with digital image capture. The DR 600 is a ceiling mounted tube and operator console with a motorized patient table and/or wall stand. The DR 600 uses Agfa's NX workstation with MUSICA2 ™ image processing and flat-panel detectors of the scintillatorphotodetector type (Cesium Iodide - CsI or Gadolinium Oxysulfide - GOS). It is compatible with Agfa's computed radoigraphy systems as well.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for the Agfa DR 600, a stationary X-ray system. The core of this submission is to demonstrate substantial equivalence to previously cleared devices, not to demonstrate clinical efficacy or a specific performance benchmark for an AI model.

    Therefore, the acceptance criteria and study proving an AI device meets those criteria, as typically understood in Machine Learning or AI product development, are not explicitly present in this document. The document focuses on demonstrating that the DR 600, as an imaging system, performs equivalently to its predicates based on physical and technological characteristics, and that its image quality is comparable.

    However, I can extract the closest analogous information to the requested points, interpreting "acceptance criteria" through the lens of demonstrating "substantial equivalence" and "device performance" in terms of physical and image quality comparisons.

    Here's how the document addresses the spirit of your request, adapted for a medical imaging device without a specific AI component being evaluated for a diagnostic task:


    Acceptance Criteria and Study for Agfa DR 600 (Interpreted for a Standard X-Ray System)

    The primary "acceptance criterion" for this 510(k) submission is demonstrating substantial equivalence to legally marketed predicate devices. This means proving that the DR 600 is as safe and effective as the predicate devices and does not raise different questions of safety and effectiveness.

    1. Table of Acceptance Criteria and Reported Device Performance

    Since this is a substantial equivalence claim for an X-ray system, the "acceptance criteria" are implied by comparing the new device's technical specifications and image quality to well-established predicates. No numerical "performance metrics" in the typical AI sense (e.g., AUC, sensitivity, specificity) for a diagnostic task are provided, as the device itself is the imaging system, not a diagnostic algorithm.

    Parameter/CharacteristicAcceptance Criteria (Implied by Predicates)Reported DR 600 Performance
    Image QualityEquivalent to or better than predicate devices (DR 400 & DX-M) based on visual assessment by qualified radiologists. No artifacts influencing image quality.Laboratory image quality comparison of DR 600 (with flat-panel & CR cassettes) and DR 400 (predicate K141192) and DX-M (CR System) anthropomorphic phantoms were performed. "The study confirmed that the Agfa DR 600 system... was equivalent to or better in performance than the DR 400 and DX-M." No artifacts detected using DR 600 that could influence image quality.
    Usability & FunctionalitySupports radiographic workflow, design, functionality, and usability within a hospital environment. FLFS workflow rated positive. Fulfilled intended use.Usability and functionality evaluations conducted. "The results of the usability test fell within the acceptance criteria for all components; therefore, the DR 600 supports a radiographic workflow. The usability and functionality of Full Leg Full Spine (FLFS) workflow for DR was rated positive as well. The intended use is fulfilled using different flat-panel detectors."
    Grid Tests ConsistencyConsistent with predicate DR 400 results. Varian (DX-D 10) and Vieworks (DX-D 40) detectors work well.Flat field, chest, and skull phantoms created using all grids for DR 600. "The results of the grid tests remained consistent with the DR 400 (predicate K141192) results. The Varian (DX-D 10) and Vieworks (DX-D 40) detectors worked well and had positive results. The intended use is fulfilled using different flat-panel detectors and/or CR cassettes and plates."
    Technological Characteristics"Same" or "similar" to predicates for key components (e.g., communication, detector material, pixel size, dynamic range, workstation, image processing)."Principles of operation and technological characteristics of the new and predicate devices are the same." (Detailed comparisons in pages 6-7). Minor differences like ceiling vs. floor mounted tube, and specific generator/tube models, are deemed not to alter intended diagnostic effect.
    Regulatory ComplianceMeets relevant safety and performance standards (e.g., IEC 60601 series, ISO 14971, ISO 13485).Compliance demonstrated through laboratory testing and software verification/validation against standards such as IEC 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-2-28, IEC 60601-2-54, ACR/NEMA PS3.1-3.20 (DICOM), ISO 14971, ISO 13485 (Page 8).
    Risk AssessmentRisks are broadly acceptable or ALARP (As Low As Reasonably Practicable), with zero in the "Not Acceptable Region.""For the DR 600 there are a total of 97 risks in the broadly acceptable region and eight risks in the ALARP region. Zero risks were identified in the Not Acceptable Region." (Page 8).

    2. Sample Size and Data Provenance for Test Set

    • Sample Size: The document mentions "anthropomorphic phantoms" for image quality comparison and "all grids" for grid tests. It does not specify a specific number of phantom images or actual patient data. For usability, "qualified independent radiographers" and "qualified internal radiographer" participated.
    • Data Provenance: The data appears to be prospective laboratory testing using phantoms rather than retrospective patient data. The country of origin for the data is not specified but would presumably be Agfa's testing facilities (likely in Belgium or the US, given the submission details).

    3. Number of Experts and Qualifications for Ground Truth

    • Number of Experts:
      • Image Quality: "Qualified independent radiologists" (plural, so at least two).
      • Usability & Functionality: "Qualified independent radiographers" (plural, so at least two).
      • Grid Tests: "Qualified internal radiographer" (singular, presumably one).
    • Qualifications: "Qualified" is a general term. Specific years of experience or board certifications are not provided in this document.

    4. Adjudication Method for the Test Set

    • Image Quality: "Performed in pairs by the qualified independent radiologists." This suggests a consensus or comparison method, possibly side-by-side. The specific adjudication rule (e.g., majority vote, forced consensus) is not detailed.
    • Usability & Functionality / Grid Tests: No specific adjudication method is mentioned beyond the conduct of the studies by qualified individuals.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No, an MRMC comparative effectiveness study was not done in the context of human readers improving with AI vs. without AI assistance. This document describes an X-ray imaging system, not an AI software intended for diagnostic assistance. The "image quality evaluations conducted with independent radiologists" were for demonstrating primary image quality equivalence, not for evaluating radiologist performance with or without an AI.

    6. Standalone Performance (Algorithm Only)

    • Not applicable in the context of an AI algorithm. This device is a complete X-ray imaging system. Its performance (image quality, technical specifications) is evaluated as a standalone product. The "laboratory image quality comparison" can be considered as the "algorithm only" type of evaluation in the sense that the system's output (images) were assessed directly.

    7. Type of Ground Truth Used

    • For image quality, the ground truth appears to be expert visual consensus/comparison on anthropomorphic phantom images.
    • For usability and functionality, the ground truth is based on user feedback and assessment against predefined usability criteria.
    • For grid tests, the ground truth is the consistency of the results with predicate device expectations and visual assessment by an expert.
    • No pathology or outcomes data were used as ground truth, as this is a technical equivalence submission for an imaging device, not a diagnostic AI.

    8. Sample Size for the Training Set

    • Not applicable. This document describes a medical device, not a machine learning model. There is no concept of a "training set" for the DR 600 system itself. Its internal image processing (MUSICA2™) is mentioned as being identical to that used in the predicate devices, implying it's a pre-existing, validated algorithm rather than one trained specifically for this submission.

    9. How Ground Truth for the Training Set Was Established

    • Not applicable. As there is no training set for the DR 600 as a whole, this question is not addressed. The MUSICA2™ image processing is a pre-existing component validated with prior devices.

    Summary: This 510(k) submission successfully demonstrates substantial equivalence of the DR 600 to its predicate devices through rigorous bench testing, including technical specifications comparison, image quality evaluation with phantoms by qualified radiologists, usability studies, and compliance with relevant safety and performance standards. The evaluations performed align with the requirements for showing that a new hardware medical imaging device is functionally and safely comparable to existing ones, rather than testing a novel AI diagnostic algorithm.