Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K190057

    Validate with FDA (Live)

    Date Cleared
    2019-06-07

    (147 days)

    Product Code
    Regulation Number
    888.3358
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Joint replacement is indicated for patients suffering from disability due to:

    • · noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis of the natural femoral head;
    • rheumatoid arthritis:
    • · correction of functional deformity;
    • · femoral fracture.

    This device may also be indicated in the salvage of previously failed surgical attempts.

    This device is indicated for cementless use.

    Device Description

    The DJO Acetabular System allows for the total replacement of the acetabulum. The system consists of porous coated titanium acetabular cups, bone screws for use with the cups, and acetabular liners manufactured from highly crossed linked polyethylene with Vitamin E. It is designed for compatibility with currently cleared DJO stems and femoral heads.

    AI/ML Overview

    This looks like a 510(k) summary for a medical device that does not involve AI. Therefore, most of the questions about acceptance criteria, study design, ground truth, and expert involvement are not applicable in the context of AI/ML devices.

    Here's a breakdown based on the provided document:

    1. A table of acceptance criteria and the reported device performance

    The provided document describes a traditional medical device (hip implant) and does not specify quantitative acceptance criteria in the way you would expect for an AI/ML device (e.g., sensitivity, specificity thresholds). Instead, the acceptance criteria are implicitly that the device performs equivalently to the predicates in non-clinical mechanical testing.

    Acceptance Criteria (Implicit)Reported Device Performance
    Ability to perform under expected conditions consistent with predicates in:Device demonstrated ability to perform under expected conditions.All testing determined that the device is substantially equivalent to the predicate devices.
    - Range of Motion Analysis(Passed - Implicitly equivalent to predicates)
    - Impingement Testing(Passed - Implicitly equivalent to predicates)
    - Push Out, Lever Out, and Torsional Strength(Passed - Implicitly equivalent to predicates)
    Pyrogen limit specifications met via Kinetic Chromogenic methodPyrogen limit specifications are met via the Kinetic Chromogenic method for bacterial endotoxin testing.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This device clearance did not involve clinical testing as stated: "Clinical Testing: Clinical testing was not required". Therefore, there is no "test set" of patient data in the context of an AI/ML study. The "testing" referred to is mechanical testing of the physical device components. The document does not specify sample sizes for these mechanical tests, nor is data provenance (country, retrospective/prospective) relevant here.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable, as no clinical testing or AI/ML ground truth was established by experts.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, as no clinical testing or AI/ML ground truth was established.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a physical hip implant device, not an AI/ML diagnostic or assistive tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. There is no algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Not applicable. For mechanical testing, the "ground truth" is typically defined by engineering specifications, material properties, and comparison to predicate device performance under controlled laboratory conditions, rather than clinical outcomes or expert consensus on a diagnosis.

    8. The sample size for the training set

    Not applicable. There is no AI/ML model to train.

    9. How the ground truth for the training set was established

    Not applicable. There is no AI/ML model to train.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1