510(k) Data Aggregation

    K Number
    K232475

    Validate with FDA (Live)

    Date Cleared
    2024-05-10

    (268 days)

    Product Code
    Regulation Number
    880.5440
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is intended to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into the vein under the action of gravity.

    Device Description

    The device is a Disposable infusion set with needle. Its configuration includes a Protective Cap of Closure-piercing Device, Closure-piercing Device, Clamp, Fluid channel (tubing), Drip chamber upper cover (including the dropper), Drip chamber, Drip chamber base (including Fluid Filter membrane), Flow regulator, Regulator wheel, Luer Lock Connector, Infusion needle hub, Needle tube, Plastic needle handle, Infusion needle protector, Integral air-inlet with air filter and closure, Tubing, Lubricant, and Adhesives. It has a Tubing Diameter of 4 mm and Filter Characteristics of 15µm. It is available in various Needle Gauges and Lengths.

    AI/ML Overview

    The provided document is a 510(k) summary for a "Disposable infusion set with needle," which is a Class II medical device. This type of device is a physical medical device, not an AI/software device. Therefore, the request for "acceptance criteria and the study that proves the device meets the acceptance criteria" in terms of AI/software performance metrics (like sensitivity, specificity, MRMC studies, ground truth establishment by experts, training/test set sample sizes for AI, etc.) is not applicable to this document.

    The document discusses physical, mechanical, chemical, and biocompatibility testing according to various ISO and ASTM standards to demonstrate substantial equivalence to a predicate device.

    However, to address the spirit of the request regarding "acceptance criteria and the study that proves the device meets the acceptance criteria" for a physical medical device as described in this document, I can extract the relevant performance criteria and testing methods from the provided text.

    Here's an interpretation based on the provided content for a physical medical device:


    Acceptance Criteria and Device Performance for a Disposable Infusion Set with Needle

    This document describes the regulatory submission for a Disposable Infusion Set with Needle (K232475) to demonstrate substantial equivalence to a predicate device (K163160). The "acceptance criteria" here refer to the performance requirements established by recognized standards, and the "study" refers to non-clinical performance testing conducted to prove the device meets these criteria.

    1. Table of Acceptance Criteria and the Reported Device Performance

    The document presents a comparison between the predicate device and the subject device, outlining various performance parameters. The "acceptance criteria" for the subject device are generally implied to be meeting the same performance standards as the predicate, or demonstrating that any differences do not adversely affect safety or performance, as evaluated against relevant ISO and ASRM standards.

    "Reported Device Performance" is indicated by the statement that the subject device "met all design specifications as Substantially Equivalent (SE) to the predicate device" and that "The requirements of the standards are met." Specific quantitative results are generally not provided in this summary, but rather a confirmation of compliance.

    ParameterAcceptance Criteria (Implied by Predicate & Standards)Reported Device Performance (Subject Device)
    Infusion Set PerformanceCompliance Confirmed
    Particulate contaminationContamination index limit is less than 90Contamination index limit is less than 90 (Met)
    LeakageNo leakageNo leakage (Met)
    Tensile strengthWithstand a static tensile force of not less than 15N for 15sCan withstand a static tensile force of 15 N for 15s (Met)
    Closure piercing deviceComply the dimension of ISO 8536-4Comply the dimension of ISO 8536-4 (Met)
    Air-inlet deviceFlow of fluid reduced less than 20% of that from a freely ventilated containerFlow of fluid reduced less than 20% of that from a freely ventilated container (Met)
    Fluid filterRetention of latex particles is not less than 80%Retention of latex particles is not less than 80% (Met)
    Drip chamber and drip tube20 drops of distilled water equivalent to (1±0.1) ml/g20 drops of distilled water equivalent to (1±0.1) ml/g (Met)
    Flow rateDeliver not less than 1000mlDeliver not less than 1000ml (Met)
    Construction (Clamps)Resist flow of fluid/air at 50 kPa when closedResist flow of fluid and air at an applied pressure of 50 kPa when closed (Met)
    Needle PerformanceCompliance Confirmed
    CleanlinessFree from particles and extraneous matterFree from particles and extraneous matter (Met)
    Needle lengthMeet ISO 7864:2016 Clause 4.10Met ISO 7864:2016 Clause 4.10 for respective gauges (e.g., 20.72mm for 23G) (Met)
    Bond of hub and needle tubeNot broken by the minimum forceNot broken by the minimum force (Met)
    StiffnessDeflection is less than 0.50mmDeflection is less than 0.50mm (Met)
    Dimensions of needle tubingMeet ISO 9626:2016 Clause 5.6Met ISO 9626:2016 Clause 5.6 for respective gauges (e.g., 0.627mm for 23G) (Met)
    Resistance to breakageThe tubing is not breakNo breakage (Met)
    Resistance to corrosionNo evidence of corrosionNo corrosion (Met)
    BiocompatibilityCompliance Confirmed
    CytotoxicityNo CytotoxicityNo Cytotoxicity (Met)
    Intracutaneous ReactivityNo Intracutaneous ReactivityNo Intracutaneous Reactivity (Met)
    Skin SensitizationNo Skin SensitizationNo Skin Sensitization (Met)
    Acute Systemic ToxicityNo Acute Systemic ToxicityNo Acute Systemic Toxicity (Met)
    HemolysisNo HemolysisNo Hemolysis (Met)
    PyrogenNo pyrogenNo pyrogen (Met)
    Other testsCompliance Confirmed
    Seal strength(Standards apply, e.g., ASTM F88/F88M-15)Met (Tests performed per ASTM F88/F88M-15)
    Internal pressure(Standards apply, e.g., ASTM F1140/F1140M-13)Met (Tests performed per ASTM F1140/F1140M-13)
    Dye Penetration(Standards apply, e.g., ASTM F1929-15)Met (Tests performed per ASTM F1929-15)
    EO residue(Standards apply, e.g., ISO 10993-7:2008)Met (Tests performed per ISO 10993-7:2008)
    ECH residue(Standards apply, e.g., ISO 10993-7:2008)Met (Tests performed per ISO 10993-7:2008)
    Bacterial Endotoxin Limit(Standards apply, e.g., USP NF <161>)Met (Tests performed per USP NF <161>)
    Shelf-Life EvaluationMaintain performance after accelerated agingMet (Physical, Mechanical, Chemical, Package & Sterility Tests on aged samples)
    Transportation TestingMaintain package integrityMet (Tests performed to verify package integrity)

    2. Sample Size Used for the Test Set and the Data Provenance

    The document does not explicitly state the sample sizes used for each specific test (e.g., number of infusion sets tested for leakage, tensile strength, etc.). It generally refers to compliance with recognized standards (e.g., ISO 8536-4, ISO 7864), which themselves specify required sample sizes for testing.

    • Data Provenance: The tests are non-clinical (laboratory/bench testing) performed by the manufacturer, Shandong Zhushi Pharmaceutical Group Co., Ltd. in China. The data would be derived from these internal tests. The term "retrospective or prospective" is not applicable to this type of device performance testing.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This question is not applicable. For a physical medical device like an infusion set, "ground truth" is established by direct physical, chemical, and biological testing against engineering and safety standards, not by expert consensus or interpretation of images. The tests evaluate measurable properties (e.g., tensile strength, flow rate, particulate count, presence of toxins) directly.

    4. Adjudication Method for the Test Set

    This question is not applicable. Adjudication methods (e.g., 2+1, 3+1) are common in studies involving human interpretation (e.g., radiologists reviewing images) where there might be inter-reader variability. For physical device performance testing, results are typically objective measurements or observations against predefined pass/fail criteria from a standard.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done

    This question is not applicable. MRMC studies are used to assess the performance of diagnostic devices or AI systems, particularly how human readers perform with and without AI assistance on cases. This device is an infusion set, not a diagnostic or AI-enabled device.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This question is not applicable. This device is a physical medical device, not an algorithm or AI.

    7. The type of ground truth used

    For this physical medical device, the "ground truth" is defined by:

    • Established Industry Standards: Such as ISO 8536-4, ISO 7864, ISO 9626, USP NF <788>, USP NF <161>, ISO 10993 series, ASTM F1929, ASTM F88, etc. These standards specify acceptable ranges, limits, or characteristics for device performance and material safety.
    • Direct Measurement and Observation: Performance is determined by physical tests (e.g., measuring tensile strength, observing leakage), chemical analyses (e.g., residual EO, particulate count), and biological assays (e.g., cytotoxicity, pyrogen testing).

    8. The sample size for the training set

    This question is not applicable. There is no "training set" in the context of a physical medical device. This term refers to data used to train an AI algorithm.

    9. How the ground truth for the training set was established

    This question is not applicable for the same reason as #8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K131646

    Validate with FDA (Live)

    Date Cleared
    2013-11-26

    (174 days)

    Product Code
    Regulation Number
    880.5440
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    JT series Disposable infusion set is used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into the vein.

    Device Description

    The applicant device-JT series disposable infusion set is plastic, single-use, sterile disposable infusion device, which is intended to be used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into the vein. The protective cap is intended to protect the needle; The puncturing needle is made of Polyethylene and used to pierce the container; the catheter is made of Polyvinyl chloride and used to connect various components; the drip is transparent so that the user can observe the dropping condition of the medical solution, and it has a filtration mesh which can prevent the micro particle with diameter larger than 200 um to from entering human vessel; Flow regulator is used to adjust the flow rate from zero to maximum; Male conical fitting is used for connecting with vein needle for medical solution transfusion. The proposed device is provided sterilized.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Disposable Infusion Set, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The provided document does not explicitly list specific acceptance criteria with numerical targets for the device's performance. Instead, it states that the device "comply[s] with voluntary standards as detailed in Section 9, 11 and 17 of this premarket submission" (which are not included in the provided text).

    The reported device performance is indirect and qualitative:

    Acceptance Criteria (Implied)Reported Device Performance
    Compliance with voluntary standardsDevice complies with voluntary standards as detailed in Sections 9, 11, and 17 of the premarket submission (details not provided here).
    SafetyConsidered "as safe" as the predicate device(s).
    EffectivenessConsidered "as effective" as the predicate device(s).
    PerformancePerformance is "substantially equivalent" to the predicate device(s).

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size: Not specified. The document states "Testing on unit level (Module verification)", "Integration testing (System verification)", "Final acceptance testing (Validation)", "Performance testing (Verification)", and "Safety testing (Verification)" were performed, but no sample sizes are given for these tests.
    • Data Provenance: Not specified. However, the manufacturer is Shanghai Jinta Medical Co., Ltd. in China, so it's likely the testing occurred in China. The study type is non-clinical testing, which implies lab-based testing rather than patient data.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    • Number of Experts: Not applicable. This was a non-clinical study, so there was no "ground truth" established by medical experts for a test set of patient data. The ground truth would be based on engineering specifications and compliance with standards.
    • Qualifications of Experts: Not applicable.

    4. Adjudication Method:

    • Adjudication Method: Not applicable, as this was a non-clinical study without human interpretation of clinical cases.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • MRMC Study: No, an MRMC comparative effectiveness study was not done. The document explicitly states: "The subject of this premarket submission, Disposable infusion set, did not require clinical studies to support substantial equivalence."

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

    • Standalone Study: Yes, this essentially describes the non-clinical performance and safety testing conducted. The device (Disposable Infusion Set) itself is a physical medical device, not an algorithm. The testing described ("Risk Analysis," "Requirements Reviews," "Design Reviews," "Testing on unit level," "Integration testing," "Final acceptance testing," "Performance testing," "Safety testing") demonstrates its standalone performance against design specifications and voluntary standards. There is no "human-in-the-loop" aspect to the function of this specific device to be evaluated.

    7. Type of Ground Truth Used (for the test set):

    • Ground Truth Type: For this non-clinical submission, the "ground truth" for the test set would be derived from engineering specifications, design requirements, and voluntary standards. Performance was measured against these predefined technical benchmarks and safety requirements for medical devices of this type.

    8. Sample Size for the Training Set:

    • Sample Size: Not applicable. This document describes the submission for a physical medical device (infusion set), not an AI/ML algorithm. Therefore, there is no "training set" in the machine learning sense. The device is designed and manufactured based on established engineering principles and materials.

    9. How Ground Truth for the Training Set Was Established:

    • Ground Truth Establishment: Not applicable, as there is no training set for this type of device. The device's design and manufacturing are based on established engineering and material science principles, guided by regulatory standards for medical devices.
    Ask a Question

    Ask a specific question about this device

    K Number
    K090012

    Validate with FDA (Live)

    Date Cleared
    2009-06-11

    (160 days)

    Product Code
    Regulation Number
    880.5440
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Disposable Infusion Set is used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into the vein.

    Device Description

    The applicant device is plastic, single-use, sterile disposable infusion device, which is intended to be used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into the vein. The protective cap is intended to protect the needle; The puncturing needle is made of Polyethylene and used to pierce the container; the catheter is made of Polyvinyl chloride and used to connect various components, there is no DEHP used in the proposed device; the drip is transparent so that the user can observe the dropping condition of the medical solution, and it has a filtration mesh which can prevent the micro particle with diameter larger than 200 um to from entering human vessel; Flow regulator is used to adjust the flow rate from zero to maximum; Infusion needle is inserted into human vessel for medical solution transfusion, it is made of stainless steel. The proposed device is provided sterilized.

    AI/ML Overview

    The provided text describes a 510(k) summary for a Disposable Infusion Set and does not contain information about acceptance criteria, device performance, sample sizes for test or training sets, ground truth establishment, expert qualifications, adjudication methods, or MRMC studies.

    The document discusses:

    • Device Name: Disposable Infusion Set
    • Intended Use: To administer fluids from a container to a patient's vascular system through a needle or catheter inserted into the vein.
    • Device Description: Plastic, single-use, sterile infusion device with components like a protective cap, puncturing needle, catheter (no DEHP), drip with filtration mesh, flow regulator, and infusion needle.
    • Test Conclusion: "Laboratory testing was conducted to validate and verify that disposable infusion set met all design specifications and was substantially equivalent to the predicate device."
    • Predicate Device: Tianjin Medis Disposable Infusion Set (K060082)
    • Substantial Equivalence Conclusion: The subject device is substantially equivalent to the predicate device.

    Therefore, I cannot provide the requested information as it is not present in the given text. The information provided focuses on the regulatory submission process and the declaration of substantial equivalence based on unspecified "laboratory testing" and "design specifications."