510(k) Data Aggregation

    K Number
    K172077

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-10-27

    (109 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The enlighten III laser system is intended for use in surgical and aesthetic applications in the medical specialties of dermatology and general and plastic surgery.

    1064 nm:
    The 1064 nm wavelength of the enlighten III laser system is indicated for:

    • treatment of benign pigmented lesions on patients with all skin types (Fitzpatrick I-VI)
    • tattoo removal for dark colored tattoo inks and for multicolored tattoos containing dark colored tattoo inks on patients with all skin types (Fitzpatrick I-VI)

    532 nm:

    • The 532 nm wavelength of the enlighten III laser system is indicated for:
    • treatment of benign pigmented lesions on patients with Fitzpatrick skin types I-IV
    • tattoo removal for lighter colored tattoo inks, including red and yellow inks, on patients with Fitzpatrick skin types I-III

    670 nm:

    • The 670 nm wavelength of the enlighten III laser system is indicated for:
    • treatment of benign pigmented lesions on patients with Fitzpatrick skin types I-III
    • tattoo removal for lighter colored tattoo inks, including green and blue inks, on patients with Fitzpatrick skin types II-IV
    Device Description

    The enlighten III laser system is a multi-wavelength, pulsed laser system. A key feature of the device is its ability to produce multiple laser wavelengths and pulse widths (1064 and 532 nm: 750 ps and 2 ns; 670 nm: 660 ps and 2 ns).

    The laser, power supplies, and control electronics are housed inside a console equipped with a touchscreen control panel. The laser treatment parameters are selected using the control panel. Laser emission is activated by depressing a footswitch. An articulated arm with rotational mirror knuckles delivers the laser beam from a laser head inside the console to a handpiece. The handpiece is equipped with an optical zoom assembly that determines the spot size of the laser beam on the treatment surface. The beam of a low-power diode laser is also transmitted through the arm to provide an aiming beam.

    AI/ML Overview

    The Cutera enlighten III Laser System was studied to assess the safety and efficacy of its 670 nm wavelength for clearing tattoos containing green and/or blue ink.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Primary Efficacy Endpoint)Reported Device Performance (670 nm wavelength for green/blue ink)
    ≥ 50% clearing of green and blue inks in ≥ 80% of patientsGreater than 50% clearing was seen in green and blue ink in the tattoos for 84% of subjects.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: 27 subjects were initially enrolled. Data analysis was conducted on 25 subjects (2 subjects voluntarily withdrew).
    • Data Provenance: The study was an IRB-approved, prospective, two-center clinical study. The country of origin is not explicitly stated in the provided text.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    • Number of Experts: Three blinded dermatologists.
    • Qualifications: "Blinded dermatologists" is the only qualification mentioned. Specific experience levels (e.g., years of experience) are not provided.

    4. Adjudication Method for the Test Set

    • The text states that "Before and after images of the twenty-five subject tattoos were randomized and sent to three blinded dermatologists for correct identification of the before from after image, and grading of the clearing of the green and/or blue ink."
    • It then mentions that "All blinded dermatologists correctly identified the before image for all subjects."
    • Regarding grading, the aggregated result is reported (84% of subjects achieved >50% clearing). The specific adjudication method for combining the dermatologists' individual grading scores (e.g., simple majority, average, or if discrepancies were resolved) is not explicitly stated. However, the fact that an overall performance metric is given implies a consensus or combined outcome.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No, a multi-reader multi-case (MRMC) comparative effectiveness study comparing AI-assisted vs. non-AI-assisted human readers was not conducted. This study focused on the device's standalone performance in a clinical setting.

    6. Standalone Performance Study

    • Yes, a standalone study was conducted. The clinical trial assessed the safety and efficacy of the enlighten III laser system (the algorithm/device) for tattoo removal when used by medical professionals. This is a direct measure of the device's performance without specific "human-in-the-loop" AI assistance considerations.

    7. Type of Ground Truth Used

    • Expert Consensus (Blinded Dermatologist Assessment): The ground truth for tattoo clearing was established through the photographic assessment and grading by three blinded dermatologists.
    • Investigator Assessment: There was also an investigator assessment of tattoo clearing at six-week post-treatment.

    8. Sample Size for the Training Set

    • The document does not provide information about a separate training set or its sample size. The clinical study described appears to be a validation/test set for the device's performance.

    9. How the Ground Truth for the Training Set Was Established

    • As no information about a separate training set is provided, the method for establishing its ground truth is not applicable from this document. The device's performance in the clinical study was evaluated against the ground truth established by the blinded dermatologists and investigators using the actual subjects' tattoos.
    Ask a Question

    Ask a specific question about this device

    K Number
    K170936

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-04-25

    (26 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The enlighten III laser system is intended for use in surgical and aesthetic applications in the medical specialties of dermatology and general and plastic surgery.

    1064 nm:
    The 1064 nm wavelength of the enlighten III laser system is indicated for:

    • treatment of benign pigmented lesions on patients with all skin types (Fitzpatrick I-VI)
    • tattoo removal for dark colored tattoo inks and for multicolored tattoos containing dark colored tattoo inks on patients with all skin types (Fitzpatrick I-VI)

    532 nm:
    The 532 nm wavelength of the enlighten III laser system is indicated for:

    • treatment of benign pigmented lesions on patients with Fitzpatrick skin types I-III
    • tattoo removal for lighter colored tattoo inks, including red and yellow inks, on patients with Fitzpatrick skin types I-III

    670 nm:
    The 670 nm wavelength of the enlighten III laser system is indicated for treatment of benign pigments with Fitzpatrick skin types I-III.

    Device Description

    The modified Cutera enlighten III laser system is a multi-wavelength, pulsed laser system. A key feature of the device is its ability to produce multiple laser wavelengths and pulse widths (1064 and 532 nm: 750 ps and 2 ns; 670 nm: 660 ps and 2 ns).

    The laser, power supplies, and control electronics are housed inside a console equipped with a touchscreen control panel. The laser treatment parameters are selected using the control panel. Laser emission is activated by depressing a footswitch. An articulated arm with rotational mirror knuckles delivers the laser beam from a laser head inside the console to a handpiece. The handpiece is equipped with an optical zoom assembly that determines the spot size of the laser beam on the treatment surface. The beam of a low-power diode laser is also transmitted through the arm to provide an aiming beam.

    AI/ML Overview

    This document is an FDA 510(k) premarket notification for a modification to the Cutera enlighten III Laser System. It focuses on demonstrating substantial equivalence to a previously cleared device. Therefore, it does not contain the kind of detailed study data (like sample sizes, expert qualifications, or comparative effectiveness with human readers) that would be present in a clinical trial report for a new device.

    Here's an analysis of what is and isn't provided based on your request:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" for clinical performance. Instead, it details modifications to the device's operational parameters and then asserts that these modifications do not introduce new risks or change the intended use, implying that the device, with these modifications, still meets the safety and effectiveness profile of its predicate.

    However, the key performance data discussed relates to the device's technical specifications and compliance with safety standards.

    Acceptance Criteria (Implied / Stated Compliance)Reported Device Performance (with modifications)
    Safety and Electrical Standards ComplianceComplies with: - IEC 60601-1 (Edition 3.1, 2012) - IEC 60601-1-6 (Edition 3.1, 2013) - IEC 62366 (Edition 1.1, 2014) - IEC 60601-2-22 (Edition 3.1, 2012) - IEC 60825-1 (Edition 2.0, 2007) - AAMI/ANSI ES60601-1:2005+A2 (R2012)+A1 - IEC 60601-1-2 (Edition 3, 2007)
    Maximum Delivered Energy1064 nm: Increased from 600 mJ to 800 mJ (maintaining 10 J/cm² max fluence) 532 nm: Increased from 300 mJ to 400 mJ (maintaining 2.5 J/cm² max fluence)
    Maximum Treatment Fluence1064 nm: Maintained at 10 J/cm² 532 nm: Maintained at 2.5 J/cm²
    Spot SizesAdded 9 mm and 10 mm incremental spot sizes for 1064 nm and 532 nm wavelengths.
    Pulse DurationAdded 2 ns pulse duration for the 670 nm wavelength.
    No New Risks IntroducedAffirmed based on unchanged maximum treatment fluence and no changes to intended use or indications for use.
    Equivalence in Intended Use/IndicationsIdentical to predicate device.
    Equivalence in Technical SpecificationsSimilar or within range of predicate and reference devices in spot sizes, wavelengths, pulse widths, laser types, and repetition rates.
    Equivalence in Operating PerformanceSimilar to predicate and reference devices.
    Equivalence in General DesignSimilar to predicate and reference devices.

    2. Sample size used for the test set and data provenance

    The document refers to a "Performance Data: enlighten III Software Verification and Validation Testing Report (V0140 rA1)". This report likely contains the details of the test set and data provenance for the software modifications. However, the FDA letter does not provide any specific sample sizes or data provenance (e.g., country of origin, retrospective/prospective) for clinical or performance data related to the device's efficacy in treating patients.

    The "testing" mentioned primarily concerns software verification and validation and compliance with electrical and safety standards, not a clinical study on human subjects with a "test set" in the context of diagnostic accuracy.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not provided in the document. As this is a 510(k) for a laser system modification, the "ground truth" would likely refer to the physical output parameters of the laser, verified by engineering and metrology, rather than expert clinical consensus on patient outcomes, which is typical for AI diagnostic devices.

    4. Adjudication method for the test set

    This information is not provided. Given the nature of the device (a laser system for treatment), "adjudication method" in the context of reconciling expert opinions (as described by "2+1, 3+1") is not relevant here. The "testing" pertains to engineering standards and device performance specifications.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done

    No, an MRMC comparative effectiveness study was not done or described in this document. This type of study is relevant for evaluating the impact of AI algorithms on human reader performance, typically in diagnostic imaging, which is not the function of this laser system.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This concept is not applicable to this device. The Cutera enlighten III is a physical laser system, not a diagnostic algorithm. The "algorithm" here refers to the control software that dictates the laser's operational parameters, and its "performance" is judged by its ability to control the hardware within specified tolerances and safety limits.

    7. The type of ground truth used

    The implicit "ground truth" for this device's performance is its technical specifications and adherence to international safety and electrical standards (e.g., IEC 60601-1, IEC 60825-1, etc.), as well as the accurate output of laser parameters (energy, fluence, spot size, pulse duration). This is established through engineering measurements and metrology, not pathology, outcomes data, or expert clinical consensus on patient diagnosis.

    8. The sample size for the training set

    This information is not applicable and therefore not provided. The device's "software modification" does not involve machine learning or AI models that require a "training set" in the typical sense of classifying data or predicting outcomes. The software is a set of control logic and algorithms to operate the laser system.

    9. How the ground truth for the training set was established

    This information is not applicable for the same reasons as #8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K160488

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-10-28

    (249 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The enlighten III laser system is intended for use in surgical and aesthetic applications in the medical specialties of dermatology and general and plastic surgery.

    1064 nm:
    The 1064 nm wavelength of the enlighten III laser system is indicated for:

    • treatment of benign pigmented lesions on patients with all skin types (Fitzpatrick I-VI)
    • tattoo removal for dark colored tattoo inks and for multicolored tattoos containing dark colored tattoo inks on patients with all skin types (Fitzpatrick I-VI)

    532 nm:
    The 532 nm wavelength of the enlighten III laser system is indicated for:

    • treatment of benign pigmented lesions on patients with Fitzpatrick skin types I-III
    • tattoo removal for lighter colored tattoo inks, including red and yellow inks, on patients with Fitzpatrick skin types I-III

    670 nm:
    The 670 nm wavelength of the enlighten III laser system is indicated for treatment of benign pigments with Fitzpatrick skin types I-III.

    Device Description

    The enlighten III laser system is a multi-wavelength, pulsed laser system. A key feature of the device is its ability to produce multiple laser wavelengths and pulse widths (1064 and 532 nm: 750 ps and 2 ns; 670 nm: 660 ps).

    The laser, power supplies, and control electronics are housed inside a console equipped with a touchscreen control panel. The laser treatment parameters are selected using the control panel. Laser emission is activated by depressing a footswitch. An articulated arm with rotational mirror knuckles delivers the laser beam from a laser head inside the console to a handpiece. The handpiece is equipped with an optical zoom assembly that determines the spot size of the laser beam on the treatment surface. The beam of a low-power diode laser is also transmitted through the arm to provide an aiming beam.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study proving the device's performance, based on the provided text:

    Device: enlighten III Laser System
    Indications for Use (670 nm wavelength): Treatment of benign pigmented lesions on patients with Fitzpatrick skin types I-III.


    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a formal "acceptance criteria" table with pre-defined numerical targets. Instead, the study aims to demonstrate safety and efficacy, implicitly establishing that a certain level of improvement and safety is acceptable for market clearance. Based on the study results, the implied acceptance criteria for efficacy appears to be achieving statistically significant lightening (>50% improvement) and high patient satisfaction for the treatment of benign pigmented lesions. Safety is assessed by the absence of serious adverse events and good tolerability.

    Feature / CriteriaImplied Acceptance Standard (from study results)Reported Device Performance (670 nm wavelength)
    Efficacy
    >50% BPL Lightening (Cohort 1)Achieve statistically significant improvement.Observed in 66% of treated BPL areas (median improvement score of 2.00, statistically significant: p<0.0001).
    >50% BPL Lightening (Cohort 2)Achieve statistically significant improvement.Observed in 74% of treated subjects (median improvement score of 2.33, statistically significant: p<0.001).
    Patient Satisfaction (Cohort 1)High percentage of subjects satisfied/extremely satisfied and willing to recommend/repeat treatment.96% of subjects were satisfied or extremely satisfied, and would repeat the treatment or recommend it to others.
    Patient Satisfaction (Cohort 2)High percentage of subjects satisfied/extremely satisfied and willing to recommend/repeat treatment.85% of subjects were satisfied or extremely satisfied, 90% would repeat the treatment, and 95% would recommend it to others.
    Safety
    Pain levelsTolerable pain levels.Mean pain scores: Cohort 1 (1.0 ± 0.2 [1-2]); Cohort 2 (0.96 ± 0.37 [0-2]) on a 10-point scale. Considered "tolerated well."
    Adverse EventsNo serious adverse events; expected minor effects resolve.All subjects experienced minor erythema, edema, and lesion frosting immediately. Most experienced mild crusting. All adverse effects resolved or improved during the study period with no required intervention. No serious adverse events were noted.

    2. Sample Size Used for the Test Set and Data Provenance

    The study was a prospective clinical study. The document does not specify the country of origin of the data, but it is a US FDA 510(k) submission, suggesting the study likely took place in the US or followed international standards acceptable to the FDA.

    • Total Enrolled: 48 subjects with Fitzpatrick skin types I-III and benign pigmented lesions.
    • Cohort 1: 25 subjects with 32 subject-treatment-areas. (For analysis, a dataset of 25 observations was created by randomly selecting one treatment area per subject, 100 times).
    • Inter-cohort subjects: 3 additional subjects with 5 subject-treatment-areas.
    • Cohort 2: 20 subjects with a single subject-treatment-area.
    • Completed Study: 46 subjects.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    • Number of Experts: Three independent dermatologists.
    • Qualifications: "Independent dermatologists" – further specific qualifications (e.g., years of experience) are not provided in the document.

    4. Adjudication Method for the Test Set

    The adjudication method involved a "blinded assessment of improvement (clearing)" completed by the three independent dermatologists. The description provides details on consistency:

    • For Cohort 1, reviewers were consistent (kappa of 0.874) and accurate (kappa of 0.833) in identifying baseline and post-treatment photographs. 78% of photographs were scored identically or within one grade difference.
    • For Cohort 2, reviewers were consistent (kappa of 1.0) and accurate (kappa of 1.0). 95% of photographs were scored identically or within one grade difference.
    • Overall (presumably for all data by the three reviewers), scores were an exact match in 16% of BPL photographs, and within one grade difference in 79% of photographs.

    This indicates an expert consensus approach, implicitly averaging or considering the predominant score for analysis given the high consistency. It isn't a strict "2+1" or "3+1" where a specific number of agreements is required; rather, the agreement level itself is reported as favorable.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. The study evaluated the standalone performance of the enlighten III laser system for lesion clearing and safety, not how human readers' performance improved with AI assistance.


    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    This study is for a laser system (medical device), not an AI algorithm. Therefore, the concept of "standalone algorithm performance" as it applies to AI/software is not relevant here. The laser system itself is the "device" being evaluated in its standalone clinical application.


    7. The Type of Ground Truth Used

    The ground truth for efficacy was established through expert consensus based on blinded review of standardized photographs. The experts assessed "improvement (clearing)" of benign pigmented lesions. Safety ground truth was based on reported pain levels and observed/recorded adverse events by investigators.


    8. The Sample Size for the Training Set

    This document describes a clinical study for a laser device, not a machine learning algorithm. Therefore, there is no "training set" in the context of AI/ML. The clinical study itself serves as the primary evidence for safety and efficacy for the device. If the question implies a "training set" for the clinical practice of using the device, that is not specified in this regulatory document.


    9. How the Ground Truth for the Training Set Was Established

    As stated above, there is no training set in the context of AI/ML for this device.