Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K193556

    Validate with FDA (Live)

    Device Name
    Cryocheck Hex LA
    Manufacturer
    Date Cleared
    2020-10-23

    (305 days)

    Product Code
    Regulation Number
    864.7925
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    CRYOcheck Hex LA is for clinical laboratory use as a qualitative test kit intended to aid in the detection of lupus anticoagulants (LA) in 3.2% citrated human plasma by the application of hexagonal phase phospholipids. CRYOcheck Hex LA should be used as an integrated test for lupus anticoagulant detection. For in vitro diagnostic use. The performance of this device has not been established in neonate and pediatric patient populations.

    Device Description

    CRYOcheck Hex LA is comprised of three reagents supplied in a frozen format as follows:

    LA Start: Pooled normal plasma with buffer and a heparin neutralizer.

    LA Correct: Pooled normal plasma with buffer, a heparin neutralizer, and inverted hexagonal phase phospholipid.

    LA APTT: Silica-based lupus sensitive APTT reagent with stabilizer.

    AI/ML Overview

    1. Acceptance Criteria and Device Performance:

    Acceptance Criteria (Internal Precision)Reported Device Performance
    Pooled precision of < 5% CV for LA StartPooled precision of < 5% CV for LA Start (e.g., 2.6%-4.2%)
    Pooled precision of < 8% CV for LA CorrectPooled precision of < 8% CV for LA Correct (e.g., 4.2%-7.6%)
    Acceptance Criteria (Reproducibility)Reported Device Performance
    Pooled reproducibility of < 5% CV for LA Start (across 3 sites)Pooled reproducibility of < 5% CV for LA Start (e.g., 3.0%-4.8%)
    Pooled reproducibility of ≤ 8% CV for LA Correct (across 3 sites)Pooled reproducibility of ≤ 8% CV for LA Correct (e.g., 4.8%-8.0%)
    Acceptance Criteria (Normal Range / Cut-off Establishment)Reported Device Performance
    Cut-off for assay delta correction determined as mean + 4 SD from normal range study.Cut-off established as mean + 4 SD, resulting in:
    < 6.0 seconds = LA Negative
    ≥ 6.0 seconds = LA Positive
    Acceptance Criteria (Shelf Life Stability)Reported Device Performance
    Supports a shelf-life stability claim of at least 12 months at -70°C or colder.Supported for at least 12 months at -70°C or colder.
    Acceptance Criteria (In-Use Stability)Reported Device Performance
    Supports a 4-hour in-use stability at room temperature.Supported for 4 hours at room temperature.
    Supports an 8-hour in-use stability when stored on-board the instrument.Supported for 8 hours when stored on-board the instrument.
    Acceptance Criteria (Method Comparison - Qualitative Agreement vs. Predicate)Reported Device Performance
    Positive Percent Agreement (PPA)95.6% (95% CI, 91%-98%)
    Negative Percent Agreement (NPA)95.2% (95% CI, 92%-97%)
    Overall Agreement95.3% (95% CI, 93%-97%)
    Acceptance Criteria (Sample Integrity - Fresh Sample Stability)Reported Device Performance
    Supports fresh sample stability of 4 hours at room temperature.Supported for 4 hours at room temperature.
    Acceptance Criteria (Sample Integrity - Frozen Storage & Freeze-Thaw)Reported Device Performance
    Supports frozen storage claim of 2 months at ≤ -70°C, including one freeze-thaw cycle.Supported for 2 months at ≤ -70°C, including one freeze-thaw cycle.
    Acceptance Criteria (Interferences - No Interference)Reported Device Performance (no interference up to indicated concentrations)
    Hemoglobin ≤ 500 mg/dLHemoglobin: ≤ 500 mg/dL
    Bilirubin (unconjugated) ≤ 20 mg/dLBilirubin (unconjugated): ≤ 20 mg/dL
    Bilirubin (conjugated) ≤ 2 mg/dLBilirubin (conjugated): ≤ 2 mg/dL
    Intralipid ≤ 500 mg/dLIntralipid: ≤ 500 mg/dL
    Unfractionated Heparin ≤ 2 IU/mLUnfractionated Heparin: ≤ 2 IU/mL
    Low Molecular Weight Heparin ≤ 2 IU/mLLow Molecular Weight Heparin: ≤ 2 IU/mL
    Dabigatran, rivaroxaban, fondaparinux (no interpretation interference)No interpretation interference observed
    Elevated Factor VIII activity (up to 180%) (no interference)No interference observed
    Elevated fibrinogen concentrations (no interpretation interference)No interpretation interference observed
    C-reactive protein (no interpretation interference)No interpretation interference observed
    Factor VIII inhibitor antibodies (no interpretation interference)No interpretation interference observed
    Plasma samples with elevated INR (up to 4.5) (no interpretation interference)No interpretation interference observed
    High platelet counts (>10,000 platelets/µL) should show interferenceShowed interference
    Abnormally low factor II activities (below 50%) (may interfere, potentially false negative)May interfere, potentially resulting in false negative results
    Factor VII and factor IX deficiencies (no interference)No interference observed
    Abnormally low factor X activities (below 50%) (no interpretation interference)No interpretation interference observed

    2. Sample Sizes and Data Provenance:

    • Precision Study:
      • 3 control plasmas and 5 plasmas with varying LA positivity.
      • Tested in duplicate, twice a day for 20 days per lot of reagent (3 lots used).
    • Reproducibility Study:
      • 3 control plasmas and 5 plasmas with varying LA positivity.
      • Each sample tested in triplicate, twice a day for 5 days for each of the 3 lots of reagent.
    • Normal Range and Assay Cut-off Study:
      • Normal samples: 137 on Analyzer A, 126 on Analyzer B.
      • Each sample tested using 3 lots of CRYOcheck Hex LA.
    • Shelf Life Stability Study:
      • 3 lots of CRYOcheck Hex LA.
      • 10 replicates of 3 controls tested at time = 0 and regular intervals up to 37 months.
      • One additional plasma sample close to the cut-off for one lot.
    • In-Use Stability Study:
      • 3 lots of CRYOcheck Hex LA.
      • 5 replicates of 3 control plasmas and 4 test plasmas with varying levels of LA.
      • Tested at 0, 2, 4, 6, 7, 8, and 9 hours.
    • Interference Studies:
      • Patient plasma samples spiked with possible interferents.
      • 20 replicates of spiked samples tested alongside 20 replicates of corresponding blank matrix control.
      • Single lot of CRYOcheck Hex LA used.
    • Method Comparison Studies (Test Set):
      • Total samples: 446
        • 124 known (previously characterized) LA positive samples.
        • 75 normal (presumed LA negative) samples from individuals with other medical conditions including autoimmune disorders.
        • 220 LA target screening population samples.
      • Data Provenance: The document does not explicitly state the country of origin for the patient samples. The study involved one internal site and three external sites, suggesting a multi-center study. It is a retrospective study since samples were "known (previously characterized) LA positive" or "presumed LA negative."
    • Sample Integrity Study:
      • 64 samples.

    3. Number of Experts and Qualifications for Ground Truth:

    • This device is an in vitro diagnostic (IVD) test, not an AI/imaging device requiring expert interpretation for ground truth.
    • The ground truth for the method comparison study was established by comparing the CRYOcheck Hex LA results against a legally marketed predicate device, Staclot® LA (K923731).
    • The determination of "known (previously characterized)" LA positive samples and "presumed LA negative" samples would implicitly rely on established clinical or laboratory diagnostic procedures, which are performed by qualified laboratory personnel, not typically "experts" in the sense of a radiology reader study.

    4. Adjudication Method for the Test Set:

    • Not applicable as this is a comparison study against a predicate device's results, not a human reader study needing adjudication. The "ground truth" for the test set was the result obtained from the predicate device (Staclot® LA).

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • No, an MRMC comparative effectiveness study was not done. This is an IVD device measuring a biomarker, not an imaging AI device assisting human readers. Therefore, the concept of "effect size" of how human readers improve with AI vs. without AI assistance is not relevant here.

    6. Standalone Performance:

    • Yes, the performance data presented (Precision, Reproducibility, Normal Range, Stability, Interferences) are essentially standalone (algorithm-only) performance characteristics of the CRYOcheck Hex LA device in a laboratory setting.
    • The method comparison study also evaluated the device's performance independently against a predicate, effectively demonstrating its standalone performance in identifying LA status.

    7. Type of Ground Truth Used:

    • For the method comparison study (test set), the ground truth was comparison to a legally marketed predicate device (Staclot® LA).
    • For the known LA positive/negative samples, the ground truth was "previously characterized" LA status, likely established through a combination of clinical diagnosis and existing laboratory methods.

    8. Sample Size for the Training Set:

    • This document describes a 510(k) submission for an in vitro diagnostic (IVD) device, not an AI/machine learning device that typically involves a "training set" in the same computational sense.
    • The development and optimization of such an IVD assay would involve internal development samples and studies, but these are not referred to as "training sets" in the context of this regulatory filing. The provided information focuses on analytical and clinical performance validation.

    9. How the Ground Truth for the Training Set Was Established:

    • Not applicable due to the nature of the device as explained in point 8. The device's "training" or development would involve laboratory optimization and calibration using reference materials and characterized samples, but not "ground truth" in the AI sense for a dedicated "training set."
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1