Search Filters

Search Results

Found 7 results

510(k) Data Aggregation

    K Number
    K241557

    Validate with FDA (Live)

    Device Name
    Cercon® yo ML
    Manufacturer
    Date Cleared
    2024-08-28

    (89 days)

    Product Code
    Regulation Number
    872.6660
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Cercon® yo ML is indicated in the anterior and posterior segments for:

    • Crowns
    • Multi-unit bridges (with a maximum of two pontics between abutment crowns)
    • Inlays, onlays and veneers
    Device Description

    The proposed Cercon® yo ML is a partially sintered ceramic blank composed of yttrium oxide-(yttria-) stabilized zirconium oxide (zirconia) (Y-TZP). Cercon® yo ML is a multilayer disk built out of different shade layers and yttrium oxide contents which create a natural gradient of the restoration after sintering. The proposed device is supplied to dental professionals as a blank and then processed by machining and subsequent sintering to fabricate dental ceramic restorations.

    AI/ML Overview

    The provided text describes acceptance criteria and the study conducted for the Cercon® yo ML device, specifically focusing on non-clinical performance data.

    Here's a breakdown of the requested information:

    1. A table of acceptance criteria and the reported device performance

    TestTest Method/Applicable StandardAcceptance CriteriaResults
    Material Performance (ISO 6872:2015 Amd. 1. 2018)
    Flexural StrengthISO 6872:2015 Amd 1. 2018 Dentistry Ceramic Materials$\geq$ 800 MPaPass
    CTE (Coefficient of Thermal Expansion)ISO 6872:2015 Amd 1. 2018 Dentistry Ceramic MaterialsThe CTE of the ceramics shall not deviate by more than 0.5 x 10 $^{-6}$ K $^{-1}$ from the value stated by the manufacturer.Pass
    SolubilityISO 6872:2015 Amd 1. 2018 Dentistry Ceramic Materials<100µg/cm² for each tested shadePass
    RadioactivityISO 6872:2015 Amd 1. 2018 Dentistry Ceramic Materials<1.0 Bq/g of $^{238}$ UPass
    Shrinkage factorClass 5 Type II materialAbsolute Accuracy ± 0.002Pass
    UniformityISO 6872:2015 Amd 1. 2018 Dentistry Ceramic MaterialsPigments shall be uniformly dispersedPass
    Freedom from extraneous materialsISO 6872:2015 Amd 1. 2018 Dentistry Ceramic MaterialsFree from extraneous materials when assessed by visual inspection and radioactivity measurementPass
    Compatibility Performance (ISO 9693:2019)
    De-bonding / crack-initiationISO 9693:2019 Dentistry - Compatibility testing>20 MPa zirconia material with at least one specified dental veneering ceramicPass
    Thermal shock resistanceISO 9693:2019 Dentistry - Compatibility testingMeet critical quenching temperature of at least 120 °CPass
    Biocompatibility Performance(Various ISO 10993 standards and USP <151>)Meets ISO 10993 requirements (implied by passing tests)Pass

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not explicitly state the sample sizes for the bench tests. It refers to general standards (ISO 6872, ISO 9693, ISO 10993 series) which typically define the number of samples required for each test. The data provenance is not mentioned, but these are non-clinical engineering and biological tests, not patient-specific data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This section is not applicable. The studies described are non-clinical bench tests and biocompatibility assessments, which rely on established scientific and engineering standards and laboratory measurements, not expert consensus for ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This section is not applicable. Adjudication methods are typically used in clinical studies involving interpretation of data by multiple experts. The described studies are non-clinical bench tests.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    There is no mention of an MRMC study or AI assistance in the provided text. The device, Cercon® yo ML, is a partially sintered ceramic blank for dental restorations, not an AI-powered diagnostic or assistive device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This section is not applicable. The device is a material for dental restorations, not a software algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the non-clinical performance data:

    • Material properties (Flexural Strength, CTE, Solubility, Radioactivity, Shrinkage factor, Uniformity, Freedom from extraneous materials): The ground truth is based on the specified physical, chemical, and mechanical properties as defined by the ISO 6872:2015 Amd 1. 2018 standard. The "ground truth" is the established limit or range defined by the standard.
    • Compatibility (De-bonding / crack-initiation, Thermal shock resistance): The ground truth is based on the specified performance criteria defined by the ISO 9693:2019 standard.
    • Biocompatibility: The ground truth is established by the requirements of the ISO 10993 series of standards and USP <151>, which define acceptable biological responses and toxicity levels.

    8. The sample size for the training set

    This section is not applicable as the studies described are non-clinical bench tests and biocompatibility assessments, not machine learning model training.

    9. How the ground truth for the training set was established

    This section is not applicable as there is no training set mentioned or implied for this type of device and testing.

    Ask a Question

    Ask a specific question about this device

    K Number
    K162888

    Validate with FDA (Live)

    Device Name
    Cercon
    Manufacturer
    Date Cleared
    2017-02-24

    (130 days)

    Product Code
    Regulation Number
    872.6660
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Cercon® is indicated for all ceramic restorations for anterior and posterior locations, including crowns and 3-unit bridges in anterior and posterior regions.

    Device Description

    The subject device, Cercon® is a partially sintered ceramic blank composed of yttria stabilized zirconium oxide, (yttria stabilized tetragonal zirconia particle, Y-TZP). The subject device. Cercon® is supplied to dental professionals as a blank and then processed by machining and subsequent sintering to fabricate all ceramic restorations. Cercon® blanks are zirconia based millable blanks (yttria-) stabilized zirconium oxide (zirconia) (Y-TZP)) that are used for all ceramic restorations for anterior and posterior locations. Specifically, Cercon® is indicated for all ceramic restorations for anterior and posterior locations, including crowns and 3-unit bridges in anterior and posterior regions.

    The subject device, Cercon® is available in various shades (colors) for esthetics. Specifically, the subject Cercon® material is offered in the following shades: A1, A2, A3, A3.5, A4, B1, B2, B3, B4, C1, C2, C3, C4, D2, D3, D4 and white.

    The sintered restorations which are fabricated using the subject Cercon® material may be veneered with compatible veneering porcelains.

    AI/ML Overview

    The provided text describes the 510(k) premarket notification for the Dentsply Sirona Cercon®, which is a ceramic blank for dental restorations. The document focuses on demonstrating substantial equivalence to predicate devices based on physical properties and biocompatibility, as clinical performance data was not submitted.

    Here's a breakdown of the requested information based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (Set by ISO 6872:2015 for Type II Class 4)Reported Device Performance (Cercon® K162888)
    Flexural strengthMeets requirements per ISO 6872:2015
    Chemical solubilityMeets requirements per ISO 6872:2015
    Coefficient of thermal expansion, CTEA4: 10.1 x 10^-6 K^-1 (25-500°C), 10.2 x 10^-6 K^-1 (25-600°C) C4: 10.0 x 10^-6 K^-1 (25-500°C), 10.1 x 10^-6 K^-1 (25-600°C)
    RadioactivityMeets requirements per ISO 6872:2015
    UniformityMeets requirements per ISO 6872:2015
    Shrinkage factorMeets requirements per ISO 6872:2015
    Freedom from extraneous materialsMeets requirements per ISO 6872:2015
    Fracture ToughnessMeets requirements per ISO 6872:2015
    Biocompatibility (Cytotoxicity)Meets requirements per ISO 10993-5
    Finished restoration quality (fit, margin, surface quality)Evaluations of finished restorations fabricated with Cercon® and predicate Cercon ht (K112152) met predetermined acceptance criteria, and all finished dental restorations (Cercon®) were found to be crack-free after milling and sintering.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not explicitly state the sample sizes for the physical property tests or the finished restoration evaluations. It mentions that testing was conducted "in accordance with ISO 6872:2015 (Dentistry - Ceramic materials)" and ISO 10993-5 for biocompatibility. The provenance of the data (country of origin, retrospective/prospective) is not specified, but it can be inferred that these are bench tests conducted in a laboratory setting for regulatory submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not provided. The "ground truth" for the physical properties is defined by the ISO standards themselves. For the "predetermined acceptance criteria" for finished restoration quality, the number and qualifications of experts involved in setting or evaluating against these criteria are not mentioned.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    The document does not describe any adjudication method. The testing appears to be based on physical measurements and evaluations against established standards, not on subjective interpretations requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. The device (Cercon®) is a zirconia-based ceramic material for dental restorations, not an AI-assisted diagnostic or treatment planning tool. No MRMC study was conducted or is relevant to this device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable. The device is a material, not an algorithm. The performance described is inherent to the material properties and the final restoration's physical characteristics, not an algorithmic output.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The ground truth for the physical properties is established by international standards (ISO 6872:2015 and ISO 10993-5). For the finished restoration evaluations, the ground truth was based on "predetermined acceptance criteria," which would likely have been derived from dental engineering and clinical best practices for material performance. No pathology or outcomes data is mentioned here for establishing ground truth.

    8. The sample size for the training set

    This is not applicable. Cercon® is a material, not a machine learning model, so there is no concept of a "training set" in this context.

    9. How the ground truth for the training set was established

    This is not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K112152

    Validate with FDA (Live)

    Device Name
    CERCON HT
    Date Cleared
    2011-11-09

    (105 days)

    Product Code
    Regulation Number
    872.6660
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Cercon® ht is indicated in the anterior and posterior segments for:

    • crowns
    • telescopic primary crowns
    • multi-unit bridges (with no more than two pontics between abutment crowns)
      Cercon® ht can be used as a substructure (framework) which is then veneered with a dental veneering ceramic or can be used for full-contour application (without veneering) as well. In the case of telescopic primary crowns the substructure is not veneered.
    Device Description

    Cercon ht is a ceramic composed of partially sintered yttria stabilized zirconia powder (Y-TZP). It is supplied to the dental laboratory as a blank and then processed in the dental-laboratory by machining and subsequent sintering to full density. It can be veneered with a dental veneering ceramic or can be used as anatomically shaped full-contour crown or bridge without veneering. It is designed for anterior and posterior locations as a substructure (framework) for single-tooth or bridge type restorations. It can also be used for the preparation of telescopic primary crowns.

    AI/ML Overview

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance CriteriaReported Device Performance
    Mechanical Strength (Flexural Strength)Similar to predicate device (Cercon Base)"material has the same mechanical (flexural strength)... as the predicate device."
    Chemical Stability (Solubility)Similar to predicate device (Cercon Base)"material has the same... chemical stability (solubility) as the predicate device."
    Wear (of unveneered material itself)Significantly lower abrasion compared to conventional veneering porcelain"significantly lower compared to conventional veneering porcelain."
    Wear (of antagonist teeth opposed to unveneered material)Significantly lower abrasion compared to conventional veneering porcelain"significantly lower compared to conventional veneering porcelain."
    Fit of Large BridgesComparable to conventional metal cast frameworks"comparable to conventional metal cast frameworks."
    Fitting Accuracy of Telescopic Primary CrownsGood fitting accuracy"a good fitting accuracy... is achieved."
    Surface Quality of Telescopic Primary CrownsGood surface quality"a good surface quality is achieved."

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample sizes used for each non-clinical test (flexural strength, solubility, wear tests, fit tests, telescopic primary crown tests). The data provenance is not specified regarding country of origin; however, the tests were "Non-Clinical Performance Data" and "bench tests and professional evaluations" performed to demonstrate substantial equivalence to a predicate device. This suggests lab-based, retrospective testing rather than prospective clinical trials.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not specify the number of experts or their qualifications for establishing ground truth. It mentions "professional evaluations" in the context of proving substantial equivalence but does not detail the methodology for these evaluations.

    4. Adjudication Method for the Test Set

    The document does not describe any specific adjudication method (e.g., 2+1, 3+1) for the test set. Given the non-clinical nature of the tests, adjudication for human-based assessment as seen in clinical trials would not typically apply. The evaluation seems to rely on direct measurement and comparison to established benchmarks or predicate device performance.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The submission relies solely on non-clinical performance data (mechanical and chemical property testing, wear tests, fit tests). The document explicitly states: "Clinical Performance Data. Not applicable."

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This question is not applicable to the device described. Cercon ht is a physical dental material (zirconia ceramic) and not a software algorithm or AI device that would have "standalone" performance in the context of human-in-the-loop assessments.

    7. The Type of Ground Truth Used

    The ground truth for the non-clinical tests appears to be based on:

    • Established physical and chemical properties: Standardized measurements for flexural strength and solubility.
    • Comparative data against predicate device: Evaluating if Cercon ht performs "the same as" the predicate Cercon Base for mechanical and chemical properties.
    • Benchmarks against conventional materials: Comparing wear to "conventional veneering porcelain" and fit to "conventional metal cast frameworks."
    • Qualitative assessments of fit and surface quality: "good fitting accuracy and a good surface quality."

    8. The Sample Size for the Training Set

    This question is not applicable. Cercon ht is a physical dental material, not an AI/ML model that requires a "training set."

    9. How the Ground Truth for the Training Set Was Established

    This question is not applicable as there is no training set for a physical dental material.

    Ask a Question

    Ask a specific question about this device

    K Number
    K041707

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2004-07-30

    (37 days)

    Product Code
    Regulation Number
    872.6660
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Designed for use on zirconia (zirconium oxide) in single tooth or bridge type restorations. Applications include tooth anterior and posterior locations.

    Device Description

    The CERCON® CERAM EXPRESS Ceramic System consists of a dental ceramic veneering material developed for veneering zirconium oxide substructures for fixed prosthodontics devices that include both anterior and posterior crowns and bridges. The modified device is a glass/glass-ceramic mixture used to veneer zirconia substrates in a medium-pressure. lost-wax injection molding (pressing) process. The modified device consists of a translucent material and an opaque material.

    AI/ML Overview

    This 510(k) K041707 submission focuses on a dental ceramic veneering material (CERCON® CERAM EXPRESS Ceramic System) and demonstrates substantial equivalence to a predicate device, Cercon® Ceram S Porcelains (K022796).

    The provided text does not contain information about specific acceptance criteria, a detailed study proving performance against those criteria, or the methodology (sample size, expert qualifications, adjudication, etc.) of such a study.

    Instead, the submission primarily focuses on:

    • Device Description and Intended Use: CERCON® CERAM EXPRESS Ceramic System is a glass/glass-ceramic mixture for veneering zirconia substructures in fixed prosthodontics.
    • Technological Characteristics: It's a modification of an existing device (Cercon® Ceram S Porcelains, K022796) with changes to formulation, material form, and processing.
    • Basis for Substantial Equivalence: The manufacturer asserts substantial equivalence based on the prior use of components in legally marketed devices, similarity in formulations, and data regarding modifications. This implies that the modifications did not introduce new questions of safety or effectiveness.

    Therefore, I cannot populate the table or answer the specific questions regarding acceptance criteria, study details, sample sizes, expert qualifications, or ground truth establishment based on the provided document.

    It's important to understand that for materials like this, "performance" is often demonstrated through mechanical and biocompatibility testing (e.g., flexural strength, bond strength, cytotoxicity) rather than clinical studies with human readers or AI algorithms as would be relevant for diagnostic imaging devices. Such tests would typically compare the modified device's properties to the predicate device or established standards. However, the details of these tests are not present in the provided summary.

    Here's what can be inferred or stated about what is NOT present:

    1. Table of Acceptance Criteria and Reported Device Performance: Not available in this document.
    2. Sample Size and Data Provenance (Test Set): Not applicable or not reported for the type of comparison made. There is no "test set" in the context of human readers or AI performance.
    3. Number of Experts and Qualifications (Ground Truth): Not applicable.
    4. Adjudication Method: Not applicable.
    5. MRMC Comparative Effectiveness Study: Not conducted or reported. This device is a material, not a diagnostic tool requiring human reader interpretation or AI assistance in that sense.
    6. Standalone (Algorithm Only) Performance: Not applicable. There is no algorithm.
    7. Type of Ground Truth Used: Not applicable in the context of expert consensus, pathology, or outcomes data for a diagnostic algorithm.
    8. Sample Size for Training Set: Not applicable. There's no AI model being trained.
    9. How Ground Truth for Training Set was Established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K022796

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2002-09-06

    (14 days)

    Product Code
    Regulation Number
    872.6660
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Designed for use on zirconia (zirconium oxide) in single tooth or bridge type restorations. Applications include both anterior and posterior locations.

    Device Description

    CERCON® CERAM S PORCELAIN is a dental ceramic veneering material developed for veneering Cercon zirconia or an equivalent zirconium oxide substructure for fixed prosthodontics devices that include both anterior and posterior crowns/bridges. The CERCON® CERAM S PORCELAIN System consists of Dentin/Transparent/Incisal, Liner/Opaque, Shoulder, and Correction/Glaze/Stain Porcelains. The final restoration matches more shades than the predicate device.

    AI/ML Overview

    This document is a 510(k) summary for a dental ceramic veneering material named CERCON® CERAM S PORCELAIN. It does not contain information about a study with acceptance criteria, sample sizes, expert ground truth, or adjudication methods in the context of device performance metrics typically associated with AI/ML devices or diagnostic tools.

    Instead, this submission focuses on demonstrating substantial equivalence to a predicate device (Cercon® Ceram K011333) based on:

    • Device Description: It describes the use of the new porcelain for veneering zirconia substructures in fixed prosthodontic devices.
    • Intended Use: It specifies the use on zirconia for single tooth or bridge type restorations in anterior and posterior locations.
    • Technological Characteristics: It highlights that CERCON® CERAM S PORCELAIN is a modification of the predicate device with minor formulation changes.
    • Biocompatibility: It states that the device was evaluated and passed biocompatibility testing for cytotoxicity.
    • Components: It mentions that all components have been used in legally marketed devices.

    The conclusion is that the prior use of components, similarity in formulation, and biocompatibility data support the safety and effectiveness of the new device.

    Therefore, I cannot provide the requested table and study details as they are not present in the provided text. The document is a regulatory submission for a material, not a diagnostic or AI-powered device that would typically involve such performance studies.

    Ask a Question

    Ask a specific question about this device

    K Number
    K013230

    Validate with FDA (Live)

    Device Name
    CERCON BASE
    Date Cleared
    2001-10-25

    (28 days)

    Product Code
    Regulation Number
    872.6660
    Panel
    Dental
    Reference & Predicate Devices
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Cercon Base is a partially sintered yttria (yttrium oxide) stabilized zirconia (zirconium oxide) powder that is machined and then sintered to full density. It is designed for use as a substructure (framework) for single tooth or bridge type restorations. Applications include both anterior and posterior locations.

    Device Description

    Cercon Base is a partially sintered yttria (yttrium oxide) stabilized zirconia (zirconium oxide) powder that is machined and then sintered to full density.

    AI/ML Overview

    I am sorry, but based on the provided text, I cannot extract any information about acceptance criteria or a study proving that a device meets such criteria. The document is a 510(k) premarket notification letter from the FDA regarding a dental ceramic device named "Cercon Base." It primarily discusses the regulatory approval process, classification, and general compliance requirements for marketing the device. There is no mention of specific acceptance criteria, performance studies, sample sizes, expert qualifications, or ground truth establishment.

    Ask a Question

    Ask a specific question about this device

    K Number
    K011333

    Validate with FDA (Live)

    Device Name
    CERCON CERAM
    Date Cleared
    2001-08-20

    (110 days)

    Product Code
    Regulation Number
    872.6660
    Panel
    Dental
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview