Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K191221

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-01-10

    (248 days)

    Product Code
    Regulation Number
    872.6070
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Cybird LED Curing Light is intended to polymerize resinous dental materials, restorative composite materials and orthodontic brackets, bonding and sealing materials that are photo-polymerized in the 400-480 nm waveband of visible light.

    Device Description

    Cybird is a LED curing light intended for polymerization of light-cured materials by dental professionals. This product is effective on various light-cured materials. Cybird's body is made from industrial-grade aluminum which is a durable material intended to ensure heat dissipation. The Cybird features multiple curing modes to ensure functionality.

    AI/ML Overview

    The provided text describes the 510(k) summary for the Cybird LED Curing Light (Model: Cybird WPT) and its comparison to a predicate device. It explicitly states that clinical testing was not required to demonstrate substantial equivalence, meaning a study proving the device meets specific acceptance criteria based on clinical outcomes or human reader performance was not performed or provided.

    Instead, the submission relies on non-clinical tests and a comparison of technical specifications to a previously cleared predicate device. Therefore, it's not possible to provide all the requested information, specifically points 2 through 9, as they pertain to clinical or human-reader studies.

    Here's a breakdown of what information can be extracted based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission does not present a formal "acceptance criteria" table with specific thresholds that the device had to meet to be deemed effective in a clinical context. Instead, it compares the performance attributes of the Cybird WPT to its predicate device, the Cybird LED Curing Light (Model: Gold & XD, K173876). The "acceptance criteria" here implicitly relates to demonstrating that the new device performs "similar or better" than the predicate in non-clinical tests and that the differences do not "represent significant issues to clinical evaluation."

    Performance AttributePredicate Device (Cybird Gold/XD - K173876)Subject Device (Cybird WPT)Comparison / "Acceptance" Status
    Intended UsePolymerize resinous dental materials, restorative composite materials, and orthodontic brackets, bonding and sealing materials that are photo-polymerized in the 400-480 nm waveband of visible light.Same as predicate.Same
    Device Design4.2VDC Lithium Ion battery chargerInductive Wireless ChargerDifferent (Same user convenience; spectral irradiance identical)
    Operational Modes (Plasma/Ortho)2, 3, 4 sec (Plasma); 55 sec (Ortho)3, 6, 9 sec (Plasma); 55 sec (Ortho)Different (Longer curing time to achieve equivalent polymerization with black light guides)
    Light SourceLED light, blue and violet wavelengths; 8 light head diameterSame as predicate.Same
    AccessoriesNot specified.Not specified.Not specified.
    Composition of Patient-Contacting MaterialsAluminum, anodized white; Barrier Sleeves; Light ShieldSame as predicate.Same
    Light GuideFiber optic glass Clear (11x8, 11x11)Fiber optic glass Black (11x8, 11x11)Different (Less light penetration through black guides)
    Light Intensity (mW/Cm²)Plasma E. Mode/Ortho Mode: 11x8 Clear Shor (2581); 11x11 Clear Short (1800); High Power Mode: 11x8 Clear Shor (1521); 11x11 Clear Short (1043)Plasma E. Mode/Ortho Mode: 11x8 Black Long (2034); 11x11 Black Long (1294); High Power Mode: 11x8 Black Long (1104); 11x11 Black Long (675)Different (WPT's irradiance is ~20-30% lower with black light guide)
    Peak WavelengthDual peak: Approximately 405nm, 460nm (GOLD), 460nm (XD)Dual peak: Approximately 405nm, 460nmSame
    Depth of Cure (mm) - Spident Escom 100Plasma mode (2,3,4 sec): 11x8 Clear: 3.214.12; 11x11 Clear: 2.893.45; High Power mode (5,10,15,20 sec): 11x8 Clear: 3.785.23; 11x11 Clear: 3.414.45Plasma mode (3,6,9 sec): 11x8 Black: 3.084.47; 11x11 Black: 2.242.88; High Power mode (5,10,15,20 sec): 11x8 Black: 3.54.99; 11x11 Black: 3.053.95Similar (Performances are similar using Spident ESCOM – K110428)
    Parameters of DisinfectionChemical disinfection with approved cleaning/sanitizing agents (Cavicide, Isopropyl alcohol (75%), Ethyl alcohol based, Lysol disinfectant (alcohol-based only))Same as predicate.Same
    Usability/Ergonomics2 buttons – 1 power, 1 mode selectSame as predicate.Same
    Power Source3.7V Li-ion Battery3.7V Li-ion Battery with inductive wireless chargingSame (Battery type)
    Curing HardnessTestedTestedSimilar or better (demonstrated that the Cybird WPT LED Curing light performed similar or better compared with the predicate device.)
    Spectral Irradiance OutputTestedTestedSimilar or better (demonstrated that the Cybird WPT LED Curing light performed similar or better compared with the predicate device.)

    Summary of Acceptance: The Cybird WPT LED Curing Light was deemed substantially equivalent based on:

    • Identical intended use.
    • Demonstration through non-clinical performance and material testing that differences in device design (wireless charger, adjusted curing times for black light guides) and resulting light intensity do not negatively impact the device's ability to achieve polymerization of dental materials, and that the depth of cure is similar to the predicate.
    • Compliance with voluntary standards for biocompatibility, electrical safety, EMC, and software verification/validation.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not applicable. No clinical test set data was used. Non-clinical tests were performed on the device itself and material samples. The text does not specify the sample sizes for these non-clinical tests (e.g., how many devices were tested for irradiance intensity or depth of cure) nor the country of origin of this testing data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable. No clinical "ground truth" established by experts was used, as clinical testing was not required for this submission.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. No clinical test set requiring adjudication was used.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This device is an LED curing light, not an AI-powered diagnostic or assistive tool. No MRMC study was performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This device is hardware for polymerization; it does not involve an algorithm with standalone performance in the context of diagnostic or interpretive tasks. Software verification and validation were done for its internal operation.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • For the non-clinical tests, the "ground truth" or reference for performance would be established by:
      • Validated laboratory methods and standards: Such as those for measuring irradiance intensity, spectral output, and depth of cure (e.g., ADA Specification No. 48).
      • Predicate device performance: The performance of the predicate device (Cybird Gold / XD) served as the benchmark for comparison for most attributes.
      • Material specifications: For biocompatibility, materials were tested against ISO standards (ISO 10993-5, 10993-10).

    8. The sample size for the training set

    • Not applicable. This device does not use machine learning or AI models that require a training set.

    9. How the ground truth for the training set was established

    • Not applicable. No training set was used.
    Ask a Question

    Ask a specific question about this device

    K Number
    K173876

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-01-24

    (34 days)

    Product Code
    Regulation Number
    872.6070
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Cybird LED Curing Light is intended to polymerize resinous dental materials, restorative composite materials and orthodontic brackets, bonding and sealing materials that are photo-polymerized in the 400-480 nm waveband of visible light.

    Device Description

    Cybird is a LED curing light intended for rapid polymerization of light-cured materials by dental professionals. This product effectively reduces polymerization time on various lightcured materials and provides excellent treatment results every time. Cybird's body is made from industrial-grade aluminum which ensures its durability and excellent heat dissipation. The Cybird features multiple curing modes for maximum functionality.

    The Cybird curing light produces visible blue light in the 405 nm (Gold) & 465 nm (Gold / XD) peak waveband of spectrum with a power density of 1,500mW/cm² (High Power Mode), 2,500 mW/cm2 (Ortho Mode, Gold), 2,700mW/cm2 (Ortho Mode, XD). These power densities are sufficient for the product intended uses.

    405 nm (Gold) & 465 nm (Gold / XD) peak waveband of visible light cured (VLC) orthodontic brackets and orthodontic bonding and sealing materials.

    Exposure times can be set for 1.5, 2, 2,5, 3 seconds for XD in Ortho Mode; 2, 3, 4 seconds for Gold in Ortho Mode; and 5,10,15, or 20 seconds in High Power Mode.

    AI/ML Overview

    The provided text describes the 510(k) premarket notification for the "Cybird LED Curing Light" and its substantial equivalence to a predicate device. However, it does not explicitly detail acceptance criteria and a study proving the device meets those criteria in the way a clinical trial or algorithm performance study would.

    Instead, the document focuses on demonstrating substantial equivalence to a legally marketed predicate device through non-clinical performance testing and compliance with various standards. The "acceptance criteria" here are implied to be the standards and measurements taken to show the device performs similarly or equivalently to the predicate and adheres to regulatory requirements.

    Here's an attempt to structure the information based on your request, acknowledging the limitations of the provided text:

    Acceptance Criteria and Device Performance

    The "acceptance criteria" for this device are primarily demonstrated through non-clinical performance tests and compliance with established standards, aiming to show substantial equivalence to a predicate device.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Test/StandardAcceptance Goal (Implied)Reported Device Performance (Cybird LED Curing Light)
    Curing PerformanceCuring hardnessSufficient polymerization for intended usePerformed; results not explicitly quantified as pass/fail against specific target, but implies successful curing.
    Curing PerformanceDepth of cure (per ADA Specification No. 48)Sufficient depth of cure for intended use (e.g., 2.5mm of predicate)Plasma mode (2 sec): 2.035mm - 2.380mm (for various materials); High Power mode (10 sec): 2.370mm - 3.795mm (for various materials).
    Light IntensityMaximum light intensity measurements at 2mmWithin acceptable range for intended use; comparable to predicate1,500mW/cm² (High Power Mode), 2,500 mW/cm2 (Ortho Mode, Gold), 2,700mW/cm2 (Ortho Mode, XD).
    Light SpectrumSpectral irradiation plots of all power modesAppropriate waveband for polymerization (e.g., 400-480 nm)Dual peak: 405nm, 465nm (GOLD); Single peak: 465nm (XD) - within 400-480 nm range.
    BiocompatibilityIn vitro cytotoxicity (AAMI / ANSI / ISO 10993-5)No cytotoxicityConformity with AAMI / ANSI / ISO 10993-5:2009(R) 2014.
    BiocompatibilityIrritation and skin sensitization (AAMI / ANSI / ISO 10993-10)No irritation or sensitizationConformity with AAMI / ANSI / ISO 10993-10:2010.
    Electrical SafetyEssential performance testing (AAMI ANSI ES 60601-1)Compliance with safety standardsConducted in accordance with AAMI ANSI ES 60601-1:2005/(R) 2012 and A1:2012.
    EMC TestingElectromagnetic Compatibility (IEC 60601-1-2)Compliance with EMC standardsConducted in accordance with IEC 60601-1-2 Ed.3: 2007-03.
    SoftwareVerification and Validation (IEC 62304)Compliance with software lifecycle processesConducted as recommended in IEC62304:2006 and FDA's Guidance Document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."
    Substantial EquivalenceComparison to predicate deviceNo new questions of safety or effectiveness compared to predicateAll differences (LED light source, light guard diameter, power mode, exposure time) found not to raise new questions of safety and effectiveness.

    2. Sample Size and Data Provenance

    • Test Set Sample Size: The document does not specify a "test set" in the context of discrete data points for a predictive model. The testing involved various materials (e.g., Handae Chemical, Hanfil; Spident Escom 100; Dentkist Charmfil Plus; Mediclus Anycom) for depth of cure and hardness. The number of samples for each material or measurement is not provided.
    • Data Provenance: The tests are non-clinical (laboratory-based performance tests) and were conducted by DXM Co., Ltd. (Republic of Korea). The data is prospective in the sense that it was generated specifically for this 510(k) submission.

    3. Number of Experts and Qualifications for Ground Truth

    • This information is not applicable as the document describes non-clinical performance testing of a physical device, not an AI/ML algorithm requiring expert-established ground truth for image interpretation or diagnosis. The "ground truth" for these tests would be the physical measurements and observations made by trained technicians and engineers in a laboratory setting.

    4. Adjudication Method for the Test Set

    • This information is not applicable for the reasons stated above. Adjudication methods like 2+1 or 3+1 are typically used for establishing ground truth in clinical data reviewed by multiple experts, not for physical device performance measurements.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not performed. This type of study is relevant for evaluating the impact of AI on human reader performance for tasks like diagnostic interpretations. The "Cybird LED Curing Light" is a physical dental device, not a diagnostic AI tool.

    6. Standalone (Algorithm Only) Performance Study

    • No, a standalone (algorithm only) performance study was not done. This is also not applicable as the device is not an algorithm but a medical device.

    7. Type of Ground Truth Used

    • The "ground truth" for the tests described is based on physical measurements and adherence to recognized standards. For example, Depth of Cure was measured per ADA Specification No. 48, biocompatibility per ISO 10993 standards, and electrical safety per IEC 60601 standards. These are objective, empirically verifiable metrics, not expert consensus or pathology in a clinical sense.

    8. Sample Size for the Training Set

    • This information is not applicable. There is no "training set" as this is not an AI/ML device.

    9. How Ground Truth for the Training Set Was Established

    • This information is not applicable for the reasons stated above.
    Ask a Question

    Ask a specific question about this device

    K Number
    K042703

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2004-10-12

    (12 days)

    Product Code
    Regulation Number
    872.6070
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Cybird LED Curing Light is intended to polymerize resinous dental materials, restorative composite materials, and orthodontic brackets, bonding and sealing materials that are photo-polymerized in the 430-490 nm waveband of visible light.

    Device Description

    Cybird by DXM Co., LTd is classified as an Ultraviolet Activator for polymerization(21 C.f.r.872.6070) because it is a device intended for the photo-polymerization light cured dental materials. The Cybird is a universal photo-polymerization light curing source working in cordless conditions and producing visible blue light in the 440 to 490nm waveband of spectrum with a power density comprised 1,000W/cm², 11-8mm light-guide).These power densities are sufficient for the product intended uses ,namely, Photopolymerization in the 440-490 nm waveband of visible light cured (VLC) dental materials , Photo – polymerization in the 440-490nm waveband of visible light cured(VLC) restorative composite materials, and Photo-polymerization in the 440-490 nm waveband of visible light cured (VLC) orthodontic brackets and orthodontic bonding and sealing materials. Exposure times can be set for 5,10,15, or 20 seconds.

    AI/ML Overview

    The DXM Co., Ltd. Cybird LED Curing Light is a device intended for the photo-polymerization of light-cured dental materials. The submission states that no clinical testing was conducted to establish its safety and effectiveness. Instead, the device's substantial equivalence to a predicate device, the Satelec Mini L.E.D (K032465), was demonstrated through non-clinical testing.

    Therefore, the following information cannot be extracted from the provided text: A table of acceptance criteria and the reported device performance, sample size used for the test set and data provenance, number of experts for ground truth and qualifications, adjudication method, MRMC comparative effectiveness study results, standalone performance, type of ground truth used, sample size for training set, and how ground truth for the training set was established.

    However, based on the provided text, the following details regarding the acceptance criteria and the study that "proves" the device meets them can be summarized:

    Non-Clinical Acceptance Criteria and Reported Performance (via equivalence to a predicate device):

    The acceptance criteria for the Cybird LED Curing Light were not explicitly stated as quantitative thresholds for performance metrics. Instead, the device's acceptability was based on demonstrating substantial equivalence to a legally marketed predicate device (Satelec Mini L.E.D, K032465) through non-clinical testing. This implies that the device's performance, as measured by these non-clinical tests, should be comparable to or within acceptable limits relative to the predicate device and established safety standards.

    Acceptance Criteria CategoryReported Device Performance (via Non-Clinical Testing)
    Intended UseDemonstrated substantial similarity to the predicate device in its intended use, which is the photo-polymerization of resinous dental materials, restorative composite materials, and orthodontic brackets, bonding, and sealing materials that are photo-polymerized in the 430-490 nm waveband of visible light.
    Operational CharacteristicsProduces visible blue light in the 440 to 490nm waveband of spectrum with a power density comprised of 1,000W/cm² (11-8mm light-guide). Exposure times can be set for 5, 10, 15, or 20 seconds. These power densities are sufficient for the intended uses.
    Safety - Electrical SafetyConforms to EN 60601-1: 1990 +A,+B. No design characteristics violated the requirements of this standard or resulted in safety hazards.
    Safety - Electromagnetic Compatibility (EMC)Conforms to EN 60601-1-2: 2001, EN 61000-3-2:2000, EN61000-3-3: 1998/A1:2001, and EN 55011: 1998/A1:1999, Class B. No design characteristics violated the requirements of these standards or resulted in safety hazards.
    FunctionDemonstrated substantial similarity to the predicate device without any design characteristics violating standard requirements or resulting in safety hazards.

    Study Information (Non-Clinical Equivalence Study):

    1. Sample size used for the test set and the data provenance: Not applicable in the traditional sense of a clinical trial. The "test set" consisted of the Cybird LED Curing Light itself, undergoing various engineering and electrical safety tests. The "data provenance" is from these specific non-clinical tests performed on the device by DXM Co., Ltd. There is no mention of country of origin for the data or whether it was retrospective or prospective in the context of human subjects.

    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth, in this context, was established by compliance with international and national standards for electrical safety and electromagnetic compatibility (e.g., EN 60601-1, EN 61000 series, EN 55011). These standards themselves are established by expert consensus in their respective fields, but no specific number or qualification of experts for this particular device's testing is mentioned beyond the presumption that the company performing the tests had qualified personnel.

    3. Adjudication method: Not applicable. The process was one of compliance testing against defined standards. Either the device met the standard (passed) or it did not (failed). There was no independent adjudication of test outcomes by multiple parties mentioned.

    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is an LED curing light, not an AI-powered diagnostic or assistive tool for human readers. No MRMC study was performed.

    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. The device is a standalone hardware product for polymerization, not an algorithm.

    6. The type of ground truth used: The ground truth for proving the device's safety and effectiveness was based on compliance with established international and national standards for electrical safety (EN 60601-1), electromagnetic compatibility (EN 60601-1-2, EN 61000 series, EN 55011), and demonstration of functional equivalence (power output, wavelength) to the predicate device for its intended use.

    7. The sample size for the training set: Not applicable. There was no machine learning model or algorithm to train.

    8. How the ground truth for the training set was established: Not applicable, as there was no training set.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1