Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K211034

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-05-06

    (29 days)

    Product Code
    Regulation Number
    892.5050
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CORVUS system is a radiation treatment planning package designed to allow medical physicists, dosimetrists, and radiation oncologists to create conformal treatment plans using photon (x-ray, gamma ray) external beam radiation therapy. The treatment plans generated by CORVUS are based upon treatment machine-specific data and are intended to provide a guide to delivering external beam radiation therapy which conforms to the target volume defined by the radiation oncologist.

    The CORVUS system is valid for use only with external beam photon therapy; calculations for electrons and intracavity sources (Brachytherapy) are NOT supported.

    Device Description

    CORVUS is a semi-automatic planning system: For forward planning, the system allows the user to design a treatment plan. For IMRT, rather than simply verifying a user-designed plan, the system itself suggests a plan. A clinician then reviews and approves the plan.

    CORVUS is designed to generate plans for treatment delivery systems that can create multiple radiation patterns composed of pencil beams on which the intensity can be individually controlled. The treatment beams are weighted so that when they are projected into the treatment space they superimpose to give the desired dose distribution.

    Each radiation field is generated using one of several optimization methods provided with the system, including simulated annealing and gradient descent.

    The treatment beams are set not only to deliver the prescribed dose to the identified target volume, but also to keep the dose to other sensitive volumes below user-defined limits. Planning is done volumetrically: the beam weights for treating the entire target volume are generated simultaneously. The dose matrix is volumetric. The dose to each point is calculated to be that received from all beams and from all gantry angles. Dosage is calculated using a finite size pencil beam (FSPB) algorithm based on the beam characterization of clinically measured data. The degree to which a treatment plan is optimized is determined in part by constraints placed on the planning algorithm. The user has direct control over these constraints, which include dose goals to the target structures, dose limits to the sensitive structures, and the specification of arcs or fixed gantry positions in the treatment plan.

    CORVUS treatment plans need not have the isocenter located within the target volume. An unlimited number of targets falling within the treatment volume can be planned for at the same time. Dose may be prescribed for up to 32 structures, 29 of them user-selectable, any number of which may be separate targets or radiation-sensitive structures. Each structure can have a separate dose prescription.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study details for the CORVUS device, based on the provided FDA 510(k) summary:

    The CORVUS device is a radiation treatment planning system. The supplied document focuses on demonstrating substantial equivalence to a predicate device (CORVUS 14) rather than a direct clinical performance study against specific acceptance criteria for diagnostic accuracy in the typical sense of AI/ML devices.

    Therefore, the acceptance criteria and study detailed below are primarily centered around dosimetric accuracy, software functionality, and compliance with relevant standards, which are the key performance indicators for a treatment planning system seeking 510(k) clearance.


    1. Table of Acceptance Criteria and Reported Device Performance

    Given that this is a 510(k) submission for a treatment planning system and not a diagnostic AI device, the "acceptance criteria" are generally related to dosimetric accuracy and software performance compared to established benchmarks or the predicate device.

    Acceptance Criteria CategorySpecific Acceptance Criteria (Inferred from documentation)Reported Device Performance
    Dosimetric Accuracy- Treatment plan calculations (dose distribution) align with medical physics measurements. - Calculations for Cobalt-60 based Forward Planning align with expected accuracy (implied by adding this new feature). - Pencil-beam algorithm and dose calculation options (homogeneous, EPL) maintain accuracy.- "The accuracy of treatment plans was evaluated through comparison with medical physics measurements 2D array and ion chamber measurements." - The submission states that the "verification and validation results demonstrate that the CORVUS 15 system met its design requirements and specifications." - The fundamental scientific technology and intended use/indications are unchanged from the predicate, implying similar dosimetric performance for existing features.
    Software Functionality- All device functionalities work as per its intended use. - Risk mitigation is ensured for software. - Conforms to required software standards (IEC 62304, IEC 62366-1, IEC 62083). - New features (Cobalt-60 Forward Planning) function correctly and safely.- "Design verification and validation testing was performed to ensure that the device functionality works as per its intended use, all risks are mitigated, is substantially equivalent, and the product conforms to the required standards." - "The verification activities included system tests, module tests, anomaly verification, code reviews, and run-through integration tests." - "The validation activities included clinical workflow, treatment planning software usability, dosimetric accuracy, and import-export."
    Substantial Equivalence- No new issues of safety or effectiveness are raised compared to the predicate device (CORVUS 14). - Intended use and indications for use remain substantially similar or unchanged. - Fundamental technical characteristics are the same.- "The fundamental scientific technology of the CORVUS 15 with respect to its predicate device (CORVUS 14) system has not changed. The intended use and indications for use of the device have not changed. Based upon the performance testing results for CORVUS 15 (as detailed in the submission), the system raises no new issues of safety or effectiveness."

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify a "test set" in the sense of a dataset of patient cases used for clinical efficacy as would be seen for a diagnostic AI. Instead, the testing focuses on system verification and dosimetric validation.

    • Sample Size for Test Set: Not explicitly stated in terms of patient cases. The testing involved "medical physics measurements 2D array and ion chamber measurements." This typically involves a set of phantoms or carefully controlled geometric setups, but the number of such measurements or scenarios is not provided.
    • Data Provenance: Not applicable in the context of patient data. The "measurements" would be generated in a lab setting, not from patients.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    • Number of Experts: Not explicitly stated. The "accuracy of treatment plans was evaluated through comparison with medical physics measurements." This implies evaluation by medical physicists.
    • Qualifications of Experts: The involvement of "medical physicists" is mentioned. Their specific years of experience or board certifications are not provided in this summary.

    4. Adjudication Method for the Test Set

    Not applicable. This is not a study requiring adjudication of expert interpretations of images or patient outcomes. The "ground truth" for dosimetric accuracy would be independent physical measurements.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. This type of study is typically performed for diagnostic or AI-assisted interpretation devices to compare human performance with and without AI assistance. CORVUS is a treatment planning system, and its evaluation focuses on the accuracy of its calculations and software functionality, not on improving human reader performance in interpreting images.


    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, in the context of a treatment planning system. The "dosimetric accuracy" testing evaluates the algorithm's output (calculated dose distributions) against physical measurements, which is an algorithm-only evaluation. While the system is used by humans (medical physicists, dosimetrists, radiation oncologists), the core dose calculation component performs its function algorithmically, and its output is verified independently.


    7. The Type of Ground Truth Used

    The ground truth used for evaluating dosimetric accuracy was medical physics measurements (2D array and ion chamber measurements). This represents a physical, objective standard for verifying dose distribution accuracy.


    8. The Sample Size for the Training Set

    Not applicable. CORVUS is a rule-based or physics-based treatment planning system, not a machine learning (ML) or artificial intelligence (AI) device that typically requires a large 'training set' of data in the common sense for algorithm development. Its algorithms are based on established physics models and mathematical optimization techniques.


    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there isn't a "training set" in the context of machine learning. The algorithms are built upon fundamental physics and mathematical principles, and characterized/calibrated using machine-specific data (e.g., beam data from a linear accelerator). This characterization data is not a "ground truth for training" in the AI/ML sense but rather input parameters that define the physical behavior of the treatment machine.

    Ask a Question

    Ask a specific question about this device

    K Number
    K151469

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2015-11-25

    (177 days)

    Product Code
    Regulation Number
    892.5050
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CORVUS system is a radiation treatment planning package designed to allow medical physicists, dosimetrists, and radiation oncologists to create conformal treatment plans using photon (x-ray, gamma ray) external beam radiation therapy. The treatment plans generated by CORVUS are based upon treatment machine-specific data and are intended to provide a guide to delivering external beam radiation therapy which conforms to the target volume defined by the radiation oncologist.

    The CORVUS system is valid for use only with external beam photon therapy; calculations for electrons and intracavity sources (Brachytherapy) are NOT supported.

    Device Description

    CORVUS is a semi-automatic planning system: rather than simply verifying a user-designed plan, the system itself suggests a plan. A clinician then reviews and approves the plan.

    CORVUS is designed to generate plans for treatment delivery systems that can create multiple radiation patterns composed of pencil beams on which the intensity can be individually controlled. The treatment beams are weighted so that when they are projected into the treatment space they superimpose to give the desired dose distribution.

    Each radiation field is generated using one of several optimization methods provided with the system, including simulated annealing and gradient descent.

    The treatment beams are set not only to deliver the prescribed dose to the identified target volume, but also to keep the dose to other sensitive volumes below user-defined limits. Planning is done volumetrically: the beam weights for treating the entire target volume are generated simultaneously. The dose matrix is volumetric. The dose to each point is calculated to be that received from all beams and from all gantry angles. Dosage is calculated using a finite size pencil beam (FSPB) algorithm based on the beam characterization of clinically measured data. The degree to which a treatment plan is optimized is determined in part by constraints placed on the planning algorithm. The user has direct control over these constraints, which include dose goals to the target structures, dose limits to the sensitive structures, and the specification of arcs or fixed gantry positions in the treatment plan.

    CORVUS treatment plans need not have the isocenter located within the target volume. An unlimited number of targets falling within the treatment volume can be planned for at the same time. Dose may be prescribed for up to 32 structures, 29 of them user-selectable, any number of which may be separate targets or radiation-sensitive structures. Each structure can have a separate dose prescription.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study information for the CORVUS Radiation Therapy Planning System, based on the provided document:

    This document describes a premarket notification (510(k)) for the CORVUS Radiation Therapy Planning System (K151469), which is an accessory to medical devices of Major Level of Concern. The submission focuses on the upgrade from CORVUS 09 to CORVUS 2011, primarily adding support for Cobalt-60 based external beam radiation treatment planning (Gamma Tomotherapy) and updated operating system/hardware.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria provided are mostly qualitative (e.g., "clinically acceptable," "similar") or defined by pass/fail thresholds for dosimetric accuracy.

    Sr. No.TestAcceptance CriteriaReported Device Performance
    1.Linac Dose Equivalence TestLinac treatment plans recomputed with CORVUS 2011 should have equivalent dose and sequences to those created on CORVUS 09.The results showed that the Linac treatment plans recomputed (resegmented and redosed) with CORVUS 2011 had equivalent dose and sequences to those created on CORVUS 09.
    2.Clinical comparison of CORVUS 09 treatment plans with Cobalt treatment plans on CORVUS 2011Treatment plans created for Cobalt treatment on CORVUS 2011 should be clinically acceptable and comparable to Linac based treatment plans on CORVUS 09.The results showed that the treatment plans created for Cobalt based treatment were clinically acceptable and were comparable to Linac based treatment plans on CORVUS 09.
    3.Dosimetric Accuracy>95% of points pass relative comparison for all film. Chamber reading within 1.5% (pass/fail criteria of 4% / 4mm).>95% of points pass relative comparison for all film. Chamber reading within 1.5% (pass/fail criteria of 4% / 4mm).
    4.Comparison of calculated versus measured valuesDosimetric parameters (e.g., percentage dose depth and output factors) for Cobalt-based plans should show calculated values similar to those measured using EDR2 films and CC01 ionization chamber. Output factors and percentage depth dose curves agreed within <4% (pass/fail criteria of 4% / 4mm).Multiple test results of dosimetric parameters (such as percentage dose depth and output factors for various pencil beam sizes) showed that the Cobalt plan based calculated values are similar to those measured using film and ion chambers. Output factors and percentage depth dose curves agreed within <4% (pass/fail criteria of 4% / 4mm).
    5.Dose Comparisons for CORVUS 2011 using Gamma TomotherapyConsistency in dose calculation with growth regions, between EPL and homogeneous calculations in water-equivalent phantom, for hybrid and patient plans, for single/multi-fraction plans, for dose scaling and decay calculations.The results indicated that the growth margin does not affect the dose distribution. Profiles matched exactly. EPL and homogeneous calculation matched. Hybrid plan matched original patient plan. Dose to chamber per fraction agreed. Scaling changed treatment time and mean chamber dose as expected. Decay calculation agreed as expected.
    6.Transfer to R&V SystemAll fields (Energy, Field Size, Gantry Start Angle, Gantry Stop Angle, Collimator Angle, Couch Angle, and MU of each arc) transferred to Mosaiq R&V system via DICOM RT should be correct.All fields including Energy, Field Size, Gantry Start Angle, Gantry Stop Angle, Collimator Angle, Couch Angle, and MU of each arc were verified to be correct.
    7.Comparison of Linac plan OptimizationOptimized Linac plans in CORVUS 2011 should be clinically similar to CORVUS 09 in terms of statistics, isodose distributions, and dose volume histograms.Comparative analysis of the optimized plans (statistics, isodose distributions and dose volume histograms) indicated that these plans were clinically similar.
    8.Final build – validating dose accuracy>98% of points pass for all Cobalt validation plans in the final build (pass/fail criteria of 4%/4mm).>98% of the points pass for all the plans (pass fail criteria of 4%/4mm).

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document does not explicitly state the numerical sample size for patient cases or plans used in the test sets for most clinical validation tests. It mentions "various pencil beam sizes" (Test 4), "multiple test results" (Test 4), and "all Cobalt validation plans" (Test 8), which are not specific numbers. For Test 1 (Linac Dose Equivalence) and Test 2 (Clinical Comparison), it refers to "the same patients" and "treatment plans created for Cobalt treatment," but no specific count of patients or plans is given.
    • Data Provenance: The document does not specify the country of origin of the data. It also does not explicitly state whether the data was retrospective or prospective. Given the context of clinical comparison and re-calculation, it is likely retrospective, using existing patient data or phantom measurements, but this is not definitively stated.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    • The document mentions that the CORVUS system is "designed to allow medical physicists, dosimetrists, and radiation oncologists to create conformal treatment plans." And that "It is the physician's responsibility to verify that the dose distributions... are appropriate for a particular patient."
    • However, for the specific validation studies on the test set, the document does not explicitly state the number of experts or their qualifications used to establish the "ground truth" or to review the clinical acceptability of plans. The "clinical acceptability" (Test 2) implies expert review, but details are not provided. The dosimetric accuracy tests (Tests 3, 4, 8) rely on direct physical measurements (film and ionization chambers) rather than expert consensus as the ground truth.

    4. Adjudication Method for the Test Set

    • The document does not describe any explicit adjudication method for expert review (e.g., 2+1, 3+1). For the dosimetric tests, the adjudication is based on direct measurement comparison against defined physical thresholds (e.g., 4% / 4mm).

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No, a MRMC comparative effectiveness study was not done as described in the document. The studies were primarily focused on comparing the CORVUS 2011 algorithm's output against:
      • Previous version (CORVUS 09) outputs (Tests 1, 2, 7)
      • Physical measurements (Tests 3, 4, 8)
      • Expected computational outcomes (Test 5)
    • There is no mention of human readers, their performance without AI (the planning system itself), or an effect size of improvement with AI assistance. The system is a planning tool, not an AI for interpretation or diagnosis that would augment human review in that manner.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    • Yes, the studies are largely standalone in terms of the algorithm's performance. The tests primarily evaluate the calculations and outputs of the CORVUS 2011 system itself (dose equivalence, dosimetric accuracy, consistency of calculations, transfer of data).
    • While the system is a "semi-automatic planning system" that "suggests a plan" which a "clinician then reviews and approves," the validation tests described focus on the accuracy and equivalence of the generated plans and calculations themselves, rather than the overall human-in-the-loop workflow's effectiveness. Test 2 (Clinical comparison) does allude to "clinical efficacy" and "clinical acceptability" but without specifying a human-in-the-loop scenario. The dosimetric tests are purely algorithmic output vs. physical measurement.

    7. Type of Ground Truth Used

    • Mixed:
      • For dosimetric accuracy (Tests 3, 4, 8): Physical measurements using EDR2 film and ionization chambers (0.125 cc chamber, CC01 ionization chamber). This is a objective, empirical ground truth.
      • For clinical comparisons and equivalence (Tests 1, 2, 7): The ground truth is established by comparison to the predicate device's output (CORVUS 09) and expert consensus on "clinical acceptability" or "clinical similarity," though the specifics of this expert consensus are not detailed.
      • For dose comparison consistency (Test 5) and system transfer (Test 6): Expected computational outcomes and predefined data formats (DICOM RT) serve as the ground truth.

    8. Sample Size for the Training Set

    • The document does not provide any information about the sample size for a training set. CORVUS is described as a "semi-automatic planning system" using "optimization methods including simulated annealing and gradient descent," and dosage is calculated "using a finite size pencil beam (FSPB) algorithm based on the beam characterization of clinically measured data." This suggests a model-based approach with measured beam data, rather than a machine learning model that would typically have a "training set" in the modern sense. The "clinically measured data" for beam characterization itself would be a foundational dataset, but its size is not specified as a "training set."

    9. How the Ground Truth for the Training Set Was Established

    • As a "training set" in the modern AI/ML sense is not explicitly mentioned, the concept of establishing ground truth for it is not addressed. The "beam characterization of clinically measured data" for the FSPB algorithm would have its ground truth established through direct physical measurements of radiation beam properties.
    Ask a Question

    Ask a specific question about this device

    K Number
    K100092

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2010-02-23

    (41 days)

    Product Code
    Regulation Number
    892.5050
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Intended Use: CORVUS is intended for use as a planning tool for conformal radiation therapy. Using operator-supplied input and patient scans, it creates a plan for treatment delivery systems and generates a set of beam weights that, when applied to a compatible system, facilitates delivery of an intensity-modulated 3D conformal radiation therapy treatment. CORVUS is intended only to suggest a delivery plan. It is the physician's responsibility to verify that the dose distributions which would result from plan implementation are appropriate for a particular patient. The CORVUS system is intended to be used as an integrated system with a modulating device for planning and delivery of conformal radiation therapy. The modulating device can be the NOMOS MIMIC, nomosSTAT MLC, or a supported MLC. CORVUS produces radiation fields which are modulated to conform to the projected tumor volume plus margins. The system tries to achieve target goals while sparing sensitive structures.

    Indications for Use: The CORVUS system is a radiation treatment planning package designed to allow medical physicists, dosimetrists, and radiation oncologists to create conformal treatment plans using photon (x-ray) external beam radiation therapy. The treatment plans generated by CORVUS are based upon treatment machine-specific data and are intended to provide a guide to delivering external beam radiation therapy which conforms to the target volume defined by the radiation oncologist. The CORVUS system is valid for use only with external beam photon therapy; calculations for electrons and intracavity sources (Brachytherapy) are NOT supported.

    Device Description

    CORVUS is a semi-automatic planning system: rather than simply verifying a userdesigned plan, the system itself suggests a plan. A clinician then reviews and approves the plan. CORVUS is designed to generate plans for treatment delivery systems that can create multiple radiation patterns composed of pencil beams on which the intensity can be individually controlled. The treatment beams are weighted so that when they are projected into the treatment space they superimpose to give the desired dose distribution. Each radiation field is generated using one of several optimization methods provided with the system, including simulated annealing and gradient descent. The treatment beams are set not only to deliver the prescribed dose to the identified target volume, but also to keep the dose to other sensitive volumes below user-defined limits. Planning is done volumetrically: the beam weights for treating the entire target volume are generated simultaneously. The dose matrix is volumetric. The dose to each point is calculated to be that received from all beams and from all gantry angles. Dosage is calculated using a finite size pencil beam (FSPB) algorithm based on the beam characterization of clinically measured data. The degree to which a treatment plan is optimized is determined in part by constraints placed on the planning algorithm. The user has direct control over these constraints, which include dose goals to the target structures, dose limits to the sensitive structures, and the specification of arcs or fixed gantry positions in the treatment plan. CORVUS treatment plans need not have the isocenter located within the target volume. An unlimited number of targets falling within the treatment volume can be planned for at the same time. Dose may be prescribed for up to 32 structures, 29 of them userselectable, any number of which may be separate targets or radiation-sensitive structures. Each structure can have a separate dose prescription.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and the study proving the device meets them, structured as requested:

    Device Name: CORVUS Radiation Therapy Treatment Planning System (Model: CORVUS 09)


    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document describes the CORVUS 09 as an update to an existing device (CORVUS 5.0M) and focuses on the equivalence of the updated device to its predicate, with specific improvements. Therefore, the "acceptance criteria" are implied to be equivalence to the predicate device in overall treatment planning quality and improved accuracy in specific areas.

    Acceptance Criteria (Implied)Reported Device Performance
    Overall Treatment Plan Quality (EPL Algorithm)"Treatment plan quality validation was accomplished by comparing treatment plans generated by CORVUS 09 using the EPL algorithm with those generated by a prior version (CORVUS 08) and determined to be substantially equivalent."
    Accuracy of Dose Calculation (EPL Algorithm)"Dosimetric validation of CORVUS' EPL algorithm was performed by comparing it with film measurements, MOSFETS and ion-chambers. The accuracy of the system was confirmed and found to be conformant with our requirements."
    Accuracy of Dose Calculation in Low-Density Regions (LDI Algorithm)"Dosimetric validation of CORVUS' LDI algorithm was performed similarly by comparing it with film measurements, Mosfets and ion-chambers. The new lateral disequilibrium inclusive pencil-beam algorithm provides improved performance in low density regions such as those found in the lung."
    Treatment Plan Quality (LDI Algorithm)"Treatment plan quality validation using the new LDI algorithm was done by qualified personnel who are familiar with using an inverse treatment planning system in a clinical setting. The plan quality was evaluated and found substantially equivalent."
    Accuracy of Leakage Calculation for Varian and Siemens MLCs"Performance testing for the improvements to leakage calculation for the Varian and Siemens MLCs were conducted and found to be conformant with our requirements."
    Overall System Functionality and Defect Resolution"A total of 62 system tests passed the criteria. Module tests were completed for new or modified code. Defect resolutions were verified by an independent evaluator. New or modified code was evaluated by an independent reviewer. Final high level run-through tests were performed to confirm the final functionality of CORVUS."
    Substantial Equivalence to Predicate (CORVUS 5.0M)"The fundamental scientific technology for the CORVUS 5.0M and CORVUS 09 systems has not changed. The intended use of the device has not changed. Based upon the performance testing results for CORVUS 09, the system raises no new issues of safety or effectiveness."

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: The document does not specify a distinct "test set sample size" in terms of patient data or specific clinical cases.
      • For dosimetric validation, it states: "A total of 53 comparisons with measurement or comparisons with Monte-Carlo were completed." This refers to specific measurements against known standards.
      • For system functionality, it states: "A total of 62 system tests passed the criteria."
    • Data Provenance: The document does not specify the country of origin of data or whether it was retrospective or prospective. The validation appears to be primarily focused on physical measurements and comparisons with Monte-Carlo simulations, rather than clinical patient data.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • Number of Experts: Not explicitly stated as a specific number.
    • Qualifications of Experts: For the LDI algorithm's treatment plan quality validation, it states: "qualified personnel who are familiar with using an inverse treatment planning system in a clinical setting." For other validations, it refers to "medical physicists skilled in the art of conformal radiation therapy." No specific years of experience are mentioned.

    4. Adjudication Method for the Test Set

    The document does not describe an adjudication method (e.g., 2+1, 3+1) for establishing ground truth, as the testing primarily involves comparing device outputs to physical measurements (film, MOSFETs, ion-chambers) and Monte-Carlo simulations, or through direct comparison of plan quality by qualified personnel.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. The document describes technical performance validation and comparison to a predicate device, not a human reader study in combination with AI assistance. Therefore, no effect size of human readers improving with/without AI assistance is provided. The device (CORVUS) is the planning tool, and ActiveRx (a feature of CORVUS 09) involves a user in the optimization process, but this is not described as an MRMC comparative effectiveness study where human readers interpret output to make diagnostic decisions.


    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, implicitly. The "performance testing data" section focuses on the algorithm's accuracy (dosimetric validation, leakage calculation, system tests) independent of a human's final clinical interpretation, although a physician approves the plan. The system "suggests a plan" and "generates a set of beam weights" which are then reviewed. The validation studies for dose calculation and plan quality, comparing against physical measurements and Monte-Carlo, are standalone assessments of the algorithm's output.


    7. The Type of Ground Truth Used

    The ground truth for the performance testing was established using:

    • Physical Measurements: Film measurements, MOSFETs (Metal-Oxide-Semiconductor Field-Effect Transistors), and ion-chambers.
    • Computational/Simulated Data: Monte-Carlo calculations.
    • Clinical/Expert Criteria: "Clinical criteria which are associated with conformal therapy" for evaluating treatment plan quality, as assessed by "qualified personnel" or "medical physicists."

    8. The Sample Size for the Training Set

    The document does not provide information on the sample size used for the training set. It describes optimization algorithms (simulated annealing, gradient descent) but not how a training set was used to develop or train the system. This suggests that the system's dose calculation and optimization methods are based on physics-based models and algorithms rather than a machine learning model that requires a distinct "training set" in the conventional sense.


    9. How the Ground Truth for the Training Set Was Established

    As no specific training set is mentioned in the context of machine learning, there is no information on how its ground truth was established. The system's underlying physics models and algorithms would have been developed based on established principles and characterized with machine-specific data mentioned in the "Indications for Use" section.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1