    K Number
    K020763

    Validate with FDA (Live)

    Date Cleared
    2002-05-10

    (64 days)

    Product Code
    Regulation Number
    862.3250
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Cocaine Metabolite Enzyme Immunoassay is a homogeneous enzyme immunoassay with a 300 ng/mL cutoff. The assay is intended for use in the qualitative and semiquantitative analyses of benzoylecgonine (cocaine metabolite) in human urine.

    The Cocaine Metabolite Enzyme Immunoassay provides only a preliminary analytical test result. A more specific alternative chemical method must be used to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Clinical consideration and professional judgement should be applied to any drug-of-abuse test result, particularly when preliminary positive results are used.

    Device Description

    LZI's Cocaine Metabolite Enzyme Immunoassay is a ready-to-use, liquid reagent, homogeneous enzyme immunoassay. The assay uses specific antibody that can detect benzovlecgonine (cocaine metabolite) in human urine with minimal cross-reactivity to various, common prescription drugs and abused drugs.

    The assay is based on competition between benzoylecgonine labeled with glucose-6phosphate dehydrogenase (G6PDH) enzyme, and free drug from the urine sample for a fixed amount of specific antibody. In the absence of free drug from the urine sample the specific antibody binds to the drug labeled with G6PDH enzyme causing a decrease in enzyme activity. The G6PDH enzyme activity is determined spectrophotometrically at 340 nm by measuring its ability to convert nicotinamide adenine dinucleotide (NAD) to NADH.

    AI/ML Overview

    Here's an analysis of the provided text regarding the Lin-Zhi International, Inc.'s Cocaine Metabolite Enzyme Immunoassay, broken down by your requested categories:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't explicitly state "acceptance criteria" as pass/fail thresholds against which the new device was measured. Instead, it presents performance characteristics of the new device and compares them to a legally marketed predicate device (DRI's Cocaine Metabolite EIA). The implicit acceptance criterion is that the new device's performance should be "comparable" or "acceptable" relative to the predicate, demonstrating substantial equivalence.

    Performance CharacteristicAcceptance Criteria (Implicit - based on predicate)Reported Device Performance (LZI's Cocaine Metabolite EIA)
    Within Run Precision (Qualitative)Comparable to DRI's (e.g., %CV < 1.0)Negative: 0.37 %CV
    225 ng/mL: 0.50 %CV
    300 ng/mL: 0.48 %CV
    375 ng/mL: 0.56 %CV
    3000 ng/mL: 0.36 %CV
    Within Run Precision (Semi-quantitative)N/A (DRI had no data)225 ng/mL: 1.45 %CV
    300 ng/mL: 1.53 %CV
    375 ng/mL: 1.37 %CV
    Run-To-Run Precision (Qualitative)Comparable to DRI's (e.g., %CV < 1.5)Negative: 0.36 %CV
    225 ng/mL: 0.67 %CV
    300 ng/mL: 0.93 %CV
    375 ng/mL: 0.53 %CV
    3000 ng/mL: 0.49 %CV
    Run-To-Run Precision (Semi-quantitative)N/A (DRI had no data)225 ng/mL: 1.41 %CV
    300 ng/mL: 1.65 %CV
    375 ng/mL: 1.83 %CV
    SensitivityComparable to DRI's (40 ng/mL)4 ng/mL
    Accuracy (vs. Commercial EIA)100% Sensitivity, 100% Specificity100% Sensitivity, 100% Specificity
    Analytical Recovery (Qualitative)N/A (DRI had no data)100% accuracy on positive vs. negative tests
    Analytical Recovery (Semi-quantitative)N/A (DRI had no data)Quantitate within ±10% of nominal concentration between 30 ng/mL and 2100 ng/mL. Average 100.6% recovery at 225 ng/mL, 97.3% recovery at 375 ng/mL.
    SpecificityComparable to predicate deviceComparable to the predicate device
    Cutoff Level300 ng/mL300 ng/mL
    Assay Range0 to 1000 ng/mL (DRI's current) / 0 to 3000 ng/mL (DRI's previous)0 to 3000 ng/mL

    Study Proving Acceptance Criteria (Substantial Equivalence):

    The submission functions as the study report. It aims to demonstrate substantial equivalence to the predicate device (DRI/Microgenics Corp.'s Cocaine Metabolite Enzyme Immunoassay, K960187) by comparing various performance characteristics. The conclusion states: "LZI's Cocaine Metabolite Enzyme Immunoassay was evaluated for several performance characteristics including precision, sensitivity, accuracy, analytical recovery, and specificity. [These] showed acceptable results when compared to the predicate device."

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: The sample sizes are not explicitly stated for individual tests beyond the "Mean Rate," "SD," and "%CV" values, which imply multiple runs or measurements. For precision studies, there are multiple data points for "Within Run" and "Run-To-Run" precision for different concentration levels (Negative, 225, 300, 375, 3000 ng/mL). The specific number of replicates or runs for each condition is not provided.
    • Data Provenance: Not specified. It's likely an internal study conducted by Lin-Zhi International, Inc. The document does not indicate country of origin for test data or whether it was retrospective or prospective. Given the nature of a 510(k) submission for a new in-vitro diagnostic, it would typically involve prospective testing to generate the performance data.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • This information is not provided in the document. For an enzyme immunoassay testing for a drug metabolite, the "ground truth" would typically be established by preparing samples with known concentrations of the analyte (benzoylecgonine) or by confirming the presence/absence and concentration using a highly accurate reference method like GC/MS. The document mentions GC/MS as the preferred confirmatory method for preliminary positive results, but it doesn't detail how known concentrations were prepared or verified for the internal test set.

    4. Adjudication Method for the Test Set

    • Not applicable/Not specified. This refers to consensus-building amongst experts for diagnostic ground truth. For this type of IVD, the ground truth is established through chemical analysis methods (e.g., preparing spiked samples with known concentrations, or using a reference method like GC/MS).

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No, an MRMC comparative effectiveness study was not done. This type of study is relevant for imaging devices or those requiring human interpretation. This device is an automated enzyme immunoassay, so human interpretation is not the primary measure of performance in the same way. The comparison is between the performance metrics of the new assay and the predicate assay.

    6. Standalone Performance (Algorithm Only without Human-in-the-Loop Performance)

    • Yes, the provided data represents standalone performance. The "device" is the enzyme immunoassay kit itself, which provides an analytical result (rate or concentration) without human intervention in the result generation process beyond operating the analyzer. The precision, sensitivity, accuracy, and recovery data are measures of the assay's performance on its own.

    7. Type of Ground Truth Used

    • The ground truth for evaluating the device's performance (precision, sensitivity, analytical recovery) appears to be based on known concentrations of benzoylecgonine. This is achieved by either:
      • Creating samples with precisely prepared, spiked concentrations of benzoylecgonine.
      • Using samples previously characterized by a reference method (like the mentioned GC/MS).
    • For accuracy, the "ground truth" was established by comparing the LZI assay's results against a commercial EIA (presumably the predicate device or another established method) and reporting 100% sensitivity and specificity, indicating agreement with a widely accepted method.

    8. Sample Size for the Training Set

    • This information is not provided. This device is an enzyme immunoassay, not a machine learning algorithm that typically requires a large "training set" in the same sense. The development of such an assay involves formulation and optimization, but the document doesn't detail the experimental scale of that development process. The term "training set" is more common in AI/ML contexts.

    9. How the Ground Truth for the Training Set Was Established

    • Since there's no explicitly mentioned "training set" in the AI/ML sense, this question isn't directly applicable. If interpreted as how the assay was developed and optimized, the ground truth would have been established through a series of experiments using known concentrations of benzoylecgonine in urine samples to optimize reaction conditions, antibody concentrations, and enzyme activity, ensuring accurate detection across the intended range. However, these specific details are not elaborated upon in the 510(k) summary.
    Ask a Question

    Ask a specific question about this device

    K Number
    K013004

    Validate with FDA (Live)

    Device Name
    COCAINE
    Date Cleared
    2002-03-13

    (188 days)

    Product Code
    Regulation Number
    862.3250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Cocaine assay is used for the qualitative analysis of benzolecgonine (cocainc metabolite) in human urine with a cutoff of 300 ng/mL for use in clinical laboratories. Measurements obtained by this device are used in the diagnosis and treatment of cocaine use or overdose.

    The Cocainc assay is calibrated with Benzoylcegonine and will detect cocaine and the major metabolites.

    The Cocaine assay provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.

    Device Description

    Cocaine is an in vitro diagnostic assay for the qualitative analysis of benzoylecgonine (cocaine metabolite) in human urine. The assay is a homogencous enzyme immunoassay with a 300 ng/mL cutoff. The assay is based on competition between drug in the specimen and drug labeled with the enzyme glucose-6-phosphate dehydrogenasc (G6PDH) for antibody binding sites. Enzyme activity decreases upon binding to the antibody, so the drug concentration in the specimen can be measured in terms of enzyme activity.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Abbott Laboratories Cocaine assay, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    FeatureAcceptance CriteriaReported Device Performance
    Agreement with Predicate Device (Emit® II Cocaine assay on SYVA® 30R Analyzer)Acceptable correlation (implied by "substantially equivalent" and "similar performance characteristics")100% agreement (concordance)
    Agreement with GC/MS (Gold Standard for Confirmation)Acceptable correlation (implied for substantial equivalence)87% agreement (concordance)
    Within-run Total Precision for Verifier INot explicitly stated (implied for acceptable precision)0.48% CV
    Within-run Total Precision for Cutoff CalibratorNot explicitly stated (implied for acceptable precision)0.64% CV
    Within-run Total Precision for Verifier IINot explicitly stated (implied for acceptable precision)0.48% CV
    Within-run Total Precision for -25% of Cutoff CalibratorNot explicitly stated (implied for acceptable precision)0.42% CV
    Within-run Total Precision for +25% of Cutoff CalibratorNot explicitly stated (implied for acceptable precision)0.47% CV
    Cutoff300 ng/mL300 ng/mL
    Limit of Detection (Sensitivity)Not explicitly stated (implied for acceptable sensitivity)35 ng/mL

    Note: The document explicitly states "acceptable correlation" as the general criteria for comparison studies. For precision, the low coefficient of variation (%CV) values implicitly define the acceptable range.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document mentions that the clinical specimens tested for the GC/MS comparison ranged from 56.8 to 70,040.0 ng/mL, implying a range of cocaine metabolite concentrations. However, the number of clinical specimens used as a test set for the comparison studies (both with the predicate and GC/MS) is not explicitly stated.
    • Data Provenance: Not specified (e.g., country of origin, retrospective or prospective). The term "clinical specimens tested" suggests real-world samples, but further details are not provided.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • Number of Experts: Not applicable, as this device uses laboratory assays as its comparative standards, not human expert interpretation of images or clinical data.
    • Qualifications of Experts: Not applicable.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not applicable. The comparisons are based on analytical results from other laboratory assays (predicate device and GC/MS), which have predefined analytical criteria for results.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • MRMC Study: No, a multi-reader multi-case (MRMC) comparative effectiveness study was not performed. This device is an in vitro diagnostic assay, meaning its performance is evaluated against other laboratory tests, not by human readers interpreting results.

    6. Standalone (Algorithm Only) Performance Study

    • Standalone Study: Yes, the performance characteristics (precision, cutoff, limit of detection) of the Cocaine assay were evaluated in a standalone manner. The comparison studies (with the predicate device and GC/MS) also represent the standalone performance of the new device against established methods.

    7. Type of Ground Truth Used

    • Type of Ground Truth: The ground truth for evaluating the Cocaine assay's performance was established using two methods:
      • Comparison to a legally marketed predicate device: The Emit® II Cocaine assay on the SYVA® 30R Analyzer (K902579).
      • Comparison to a confirmatory chemical method: Gas Chromatography/Mass Spectrometry (GC/MS). GC/MS is explicitly stated as the "preferred confirmatory method" and serves as a higher-level analytical ground truth for drug testing.

    8. Sample Size for the Training Set

    • Sample Size for Training Set: Not explicitly stated. The document describes performance and comparison studies but does not provide details about a specific "training set" in the context of machine learning, as this device is an immunoassay, not an AI/ML-based diagnostic. It's likely that internal development and validation samples were used during the assay's creation and optimization, but these are not specified as a "training set" in the provided text.

    9. How the Ground Truth for the Training Set Was Established

    • Ground Truth for Training Set: Not applicable in the context of machine learning. For an immunoassay, the "ground truth" for development would involve known concentrations of the analyte (cocaine metabolite) in control samples, and the assay is optimized to accurately detect these concentrations around the specified cutoff. This is established through standard laboratory practices for assay development and analytical validation using certified reference materials and known-concentration samples.
    Ask a Question

    Ask a specific question about this device

    K Number
    K010239

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2001-06-25

    (151 days)

    Product Code
    Regulation Number
    862.3250
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K001197

    Validate with FDA (Live)

    Date Cleared
    2000-05-11

    (30 days)

    Product Code
    Regulation Number
    862.3250
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K960187

    Validate with FDA (Live)

    Date Cleared
    1996-04-02

    (77 days)

    Product Code
    Regulation Number
    862.3250
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1