Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K981931

    Validate with FDA (Live)

    Date Cleared
    1998-08-25

    (84 days)

    Product Code
    Regulation Number
    862.1295
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Elecsys® Red Blood Cell Folate Lysing Reagent is to be used in combination with the Elecsys® Folate Assay for the quantitation of folate in human red blood cells. The electrochemiluminescence assay "ECLIA" is intended for use on the Boehringer Mannheim Elecsys 2010 immunoassay analyzer.

    Device Description

    The Elecsys® test principle is based on the competitive principle. Total duration of assay: 90 +- 15 minutes. Reconstitution Step: Add contents of the ascorbic acid packet to 100 mL distilled or deionized water. Pretreatment Step: By incubating the sample (15 μl) with the folate pretreatment 1 (15 μl) and pretreatment 2 (10 μl), bound folate is liberated into the serum. Proceed to conduct the Elecsys® Folate Assay (K973674) as per insert instructions.

    AI/ML Overview

    The provided text does not contain detailed acceptance criteria or a study that explicitly proves the device meets specific performance criteria beyond general statements about evaluation of assay precision, demonstration of linearity, and correlation with a predicate device. The document primarily focuses on regulatory approval and equivalence to a predicate device.

    However, based on the information provided, we can infer some details related to the study conducted.

    Here's an attempt to structure the information based on your request, with significant caveats that much of the detailed information you asked for is not present in the provided text.

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (Implied)Reported Device Performance (Implied from studies performed)
    Assay Precision acceptable per NCCLS recommendationsEvaluation of assay precision according to NCCLS recommendations (result not detailed)
    Demonstration of linearityDemonstration of linearity (result not detailed)
    Correlation with predicate deviceCorrelation with the predicate device (result not detailed)
    Suitability for various sample typesSample type studies (result not detailed)
    Performance of the Elecsys® Red Blood Cell Folate Lysing Reagent in combination with the Elecsys® Folate Assay for quantitation of folate in human red blood cells.The device is intended for this use on the Boehringer Mannheim Elecsys 2010 immunoassay analyzer. Studies were performed to support this.

    2. Sample sized used for the test set and the data provenance

    • Sample Size for Test Set: Not specified.
    • Data Provenance: Not specified (e.g., country of origin, retrospective or prospective).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not specified. The context suggests a laboratory assay, so "experts" in the clinical interpretation sense might not be directly applicable for establishing ground truth for the raw assay results themselves, but rather for clinical validation or interpretation of results.

    4. Adjudication method for the test set

    Not specified.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is an in-vitro diagnostic (IVD) assay, not an AI-assisted diagnostic imaging device requiring human reader comparison.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, in a sense. This is an IVD assay. Its performance is evaluated as a standalone system (reagent + analyzer) that produces a quantitative result without a "human-in-the-loop" step in generating the primary measurement. The "study" here refers to the analytical performance of the assay itself.

    7. The type of ground truth used

    For an IVD assay like this, the "ground truth" for assay performance evaluation is typically established through:

    • Reference Methods: Using established, highly accurate analytical methods to determine true folate concentrations in samples.
    • Known Concentration Samples: Using samples with accurately known concentrations of the analyte (e.g., spiked samples, certified reference materials).
    • Clinical Samples with Clinical Diagnosis: For correlation studies, comparing assay results to a clinical diagnosis (though the document only mentions correlation with the predicate device, not directly with patient outcomes).
    • Predicate Device Comparison: The document explicitly mentions "correlation with the predicate device," implying the predicate device's results serve as a comparative "ground truth" or standard.

    8. The sample size for the training set

    Not applicable in the context of traditional machine learning "training sets" as this is a chemical assay, not an AI algorithm. The assay itself doesn't have a "training set" in that sense. However, the development and optimization of the assay would involve various experiments and sample analyses, but these are not referred to as a "training set."

    9. How the ground truth for the training set was established

    Not applicable.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1