Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K162282

    Validate with FDA (Live)

    Date Cleared
    2016-12-22

    (129 days)

    Product Code
    Regulation Number
    862.1345
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CardioChek Home Test System (consisting of the CardioChek Home analyzer and CardioChek Home Chol+HDL+Glu test strips) is for the quantitative determination of total cholesterol, HDL (high density lipoprotein) cholesterol and glucose in capillary whole blood from the fingertip and is intended to be used by a single person and should not be shared. This system is for in vitro diagnostic use only.

    • Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders.
    • HDL (lipoprotein) measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases.
    • Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
      A Chol/HDL ratio is calculated by the CardioChek Home analyzer.

    The CardioChek Plus Test System (consisting of the CardioChek Plus analyzer and PTS Panels Chol+HDL+Glu test strips) is for the quantitative determination of total cholesterol, HDL (high density lipoprotein) cholesterol and glucose in venous whole blood and capillary whole blood from the fingertip and is intended for multiple patient use in professional healthcare settings. This system should only be used with single-use, auto-disabling lancing devices. This system is for in vitro diagnostic use only.

    • Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipoprotein metabolism disorders.
    • HDL (lipoprotein) measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases.
    • Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
      A Chol/HDL ratio is calculated by the CardioChek Plus analyzer.

    The CardioChek Home Test System consisting of the CardioChek Home analyzer and CardioChek Home Chol+HDL test strips) is for the quantitative determination of total cholesterol and HDL (high density lipoprotein) cholesterol in capillary whole blood from the fingertip and is intended to be used by a single person and should not be shared. This system is for in vitro diagnostic use only.

    • Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders.
    • HDL (lipoprotein) measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases.
      A Chol/HDL ratio is calculated by the CardioChek Home analyzer.

    The CardioChek Plus Test System (consisting of the CardioChek Plus analyzer and PTS Panels Chol+HDL test strips) is for the quantitative determination of total cholesterol and HDL (high density lipoprotein) cholesterol in venous whole blood and capillary whole blood from the fingertip and is intended for multiple patient use in professional healthcare settings. This system should only be used with single-use, auto-disabling lancing devices. This system is for in vitro diagnostic use only.

    • Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders.
    • HDL (lipoprotein) measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases.
      A Chol/HDL ratio is calculated by the CardioChek Plus analyzer.

    The CardioChek Home Test System (consisting of the CardioChek Home analyzer and CardioChek Home Chol+Glu test strips) is for the quantitative determination of total cholesterol and glucose in capillary whole blood from the fingertip and is intended to be used by a single person and should not be shared. This system is for in vitro diagnostic use only.

    • Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders.
    • Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

    The CardioChek Plus Test System (consisting of the CardioChek Plus analyzer and PTS Panels Chol+Glu test strips) is for the quantitative determination of total cholesterol and glucose in venous whole blood and capillary whole blood from the fingertip and is intended for multiple patient use in professional healthcare settings. This system should only be used with single-use, auto-disabling lancing devices. This system is for in vitro diagnostic use only.

    • Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders.
    • Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
    Device Description

    The PTS Panels Chol+HDL+Glu, Chol+HDL and Chol+Glu test strips and the CardioChek Home Chol+HDL+Glu, Chol+HDL and Chol+Glu test strips are used with the CardioChek Plus and CardioChek Home analyzers to measure total cholesterol, HDL cholesterol and glucose in whole blood. The test strips utilize enzymatic methods on a dry strip that is read by reflectance photometry. These test strips are for in vitro diagnostic use only.

    AI/ML Overview

    I am sorry, but the provided text does not contain the specific information required to answer your request.

    The document is a 510(k) premarket notification letter from the FDA regarding the "Cardiochek Plus Test System" and "Cardiochek Home Test System". It describes the devices, their intended use, and states that they were found to be substantially equivalent to previously marketed predicate devices.

    However, it does not include:

    • A table of acceptance criteria and reported device performance.
    • Details on sample sizes, data provenance, or ground truth establishment for a specific study proving the device meets acceptance criteria.
    • Information on expert involvement, adjudication methods, MRMC studies, or standalone algorithm performance.
    • Sample size or ground truth details for a training set.

    The document primarily focuses on the regulatory aspects of substantial equivalence based on modifications to existing devices, rather than presenting a detailed study report with performance metrics against acceptance criteria.

    Ask a Question

    Ask a specific question about this device

    K Number
    K151545

    Validate with FDA (Live)

    Date Cleared
    2016-12-22

    (562 days)

    Product Code
    Regulation Number
    862.1175
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CardioChek Home Test System (consisting of the CardioChek Home analyzer and CardioChek Home Lipid Panel test strips) is for the quantitative determination of total cholesterol. HDL (high density lipoprotein) cholesterol and triglycerides in capillary whole blood from the fingertip and is intended to be used by a single person and should not be shared. This system is for in vitro diagnostic use only.

    • o Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders.
    • HDL (lipoprotein) measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes ● mellitus), atherosclerosis, and various liver and renal diseases.
    • Triglycerides measurements are used in the diagnosis and treatment of patients with diabetes mellitus, nephrosis, liver obstruction, other diseases involving lipid metabolism or various endocrine disorders.
      A Chol/HDL ration and estimated values for LDL (low density lipoprotein) cholesterol are calculated by the CardioChek Home Analyzer.

    The CardioChek Plus Test System (consisting of the CardioChek Plus analyzer and PTS Panels Lipid Panel test strips) is for the quantitative determination of total cholesterol. HDL (high density lipoprotein) cholesterol and triglycerides in venous whole blood and capillary whole blood from the fingertip and is intended for multiple patient use in professional healthcare settings. This system should only be used with single-use, auto-disabling lancing devices. This system is for in vitro diagnostic use only.

    • o Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders.
    • HDL (lipoprotein) measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes ● mellitus), atherosclerosis, and various liver and renal diseases.
    • Triglycerides measurements are used in the diagnosis and treatment of patients with diabetes mellitus, ● nephrosis, liver obstruction, other diseases involving lipid metabolism or various endocrine disorders.
      A Chol/HDL ration and estimated values for LDL (low density lipoprotein) cholesterol are calculated by the CardioChek Plus Analyzer.

    The CardioChek PA Test System (consisting of the CardioChek PA analyzer and PTS Panels Lipid Panel test strips) is for the quantitative determination of total cholesterol, HDL (high density lipoprotein) cholesterol and triglycerides in venous whole blood and capillary whole blood from the fingertip and is intended for multiple patient use in professional healthcare settings. This system should only be used with single-use, auto-disabling lancing devices. This system is for in vitro diagnostic use only.

    • Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol . in the blood and lipid and lipoprotein metabolism disorders.
    • HDL (lipoprotein) measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases.
    • Triglycerides measurements are used in the diagnosis and treatment of patients with diabetes mellitus, nephrosis, liver obstruction, other diseases involving lipid metabolism or various endocrine disorders.
      A Chol/HDL ration and estimated values for LDL (low density lipoprotein) cholesterol are calculated by the CardioChek PA Analyzer.

    The CardioChek PA Home Test System (consisting of the CardioChek PA Home analyzer and CardioChek Home Livid Panel test strips) is for the quantitative determination of total cholesterol. HDL (high density lipoprotein) cholesterol and triglycerides in capillary whole blood from the fingertip and is intended to be used by a single person and should not be shared. This system is for in vitro diagnostic use only.

    • Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol o in the blood and lipid and lipoprotein metabolism disorders.
    • HDL (lipoprotein) measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases.
    • Triglycerides measurements are used in the diagnosis and treatment of patients with diabetes mellitus, nephrosis, liver obstruction, other diseases involving lipid metabolism or various endocrine disorders.
      A Chol/HDL ration and estimated values for LDL (low density lipoprotein) cholesterol are calculated by the CardioChek PA Home Analyzer.
    Device Description

    The PTS Panels Lipid Panel test strips and the CardioChek Home Lipid Panel test strips are used with the CardioChek Plus, CardioChek PA, CardioChek Home and CardioChek PA Home analyzers to measure total cholesterol and triglycerides in whole blood. The test strips utilize enzymatic methods on a dry strip that is read by reflectance photometry. These test strips are for in vitro diagnostic use only.

    AI/ML Overview

    1. Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the CardioChek Test Systems (CardioChek Plus, CardioChek Home, CardioChek PA, CardioChek PA Home) involve demonstrating accuracy and precision for HDL cholesterol measurements within a modified dynamic range of 20-120 mg/dL. The reported performance is based on comparison to a reference laboratory method (Roche Cobas Integra 400 plus) and precision studies.

    Performance MetricAcceptance Criteria (Implied by equivalence claim and historical device performance)Reported Device Performance
    Accuracy (Correlation to Reference Method)High correlation (r ≥ 0.90 typically expected for clinical agreement, though not explicitly stated as a statistical threshold)CardioChek Plus Home: y = 0.99x + 0.55, r = 0.98
    CardioChek Plus: y = 0.99x + 0.55, r = 0.98
    CardioChek PA: y = 0.93x + 0.98, r = 0.98
    CardioChek PA Home: y = 0.93x + 0.98, r = 0.98
    Precision (Coefficient of Variation, CV%)Generally, CV% values for clinical chemistry analytes are expected to be low, typically < 5-10% depending on the analyte and concentration level. No explicit numerical criteria are given in the document.CardioChek PA HDL:
    Level 1 (38.3 mg/dL): CV = 4.3%
    Level 2 (62.4 mg/dL): CV = 3.6%
    Level 3 (106.0 mg/dL): CV = 4.0%
    CardioChek PA Home HDL:
    Level 1 (38.3 mg/dL): CV = 4.3%
    Level 2 (62.4 mg/dL): CV = 3.6%
    Level 3 (106.0 mg/dL): CV = 4.0%
    CardioChek Plus HDL:
    Level 1 (39.5 mg/dL): CV = 4.1%
    Level 2 (63.3 mg/dL): CV = 4.2%
    Level 3 (108.3 mg/dL): CV = 4.8%
    CardioChek Home HDL:
    Level 1 (39.5 mg/dL): CV = 4.1%
    Level 2 (63.3 mg/dL): CV = 4.2%
    Level 3 (108.3 mg/dL): CV = 4.8%
    Measuring Range (Dynamic Range)20-120 mg/dL HDL cholesterolDemonstrated with samples in the range of 21 to 112 mg/dL HDL cholesterol (for accuracy study) and various levels within the range for precision.

    2. Sample Size and Data Provenance for the Test Set

    • Accuracy Test Set:
      • Sample Size: 80 samples for each analyzer type (CardioChek Plus Home, CardioChek Plus, CardioChek PA, CardioChek PA Home).
      • Data Provenance: Not explicitly stated, but based on the nature of medical device studies for FDA submission, it would typically be prospective data collected for the purpose of the study. The country of origin is not specified, but the submission is to the U.S. FDA.
    • Precision Test Set:
      • Sample Size: 80 observations for Level 1 and Level 2 for all analyzers, and 80 observations (CardioChek PA/PA Home) or 78 observations (CardioChek Plus/Home) for Level 3. These observations were obtained by "three operators tested three levels of whole blood samples on five analyzers... over three time periods."
      • Data Provenance: Not explicitly stated, but likely prospective data collected in a controlled laboratory setting.

    3. Number of Experts and Qualifications for Ground Truth for the Test Set

    • This relates to a diagnostic device for quantitative measurement, not an AI classification system requiring expert interpretation for ground truth.
    • For Accuracy: The ground truth for HDL cholesterol measurements was established using a Roche Cobas Integra 400 plus HDL Cholesterol instrument, which is a recognized reference method in clinical chemistry. This is a laboratory analyzer, not an expert panel.
    • For Precision: The "ground truth" is the mean value obtained from repeated measurements of the same whole blood samples.

    4. Adjudication Method for the Test Set

    • Not applicable as this is a quantitative measurement device study, not an AI diagnostic study requiring human expert adjudication of interpretations. The "ground truth" is derived from a reference laboratory instrument.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not conducted as this is a study for a quantitative measurement device (lipid panel test system), not an AI algorithm intended for interpretation by human readers. The context is not one of human-in-the-loop performance improvement with AI assistance.

    6. Standalone (Algorithm Only) Performance Study

    • Yes, the accuracy and precision studies performed can be considered standalone performance of the device (analyzer with test strips), as it evaluates the device's ability to accurately and precisely measure HDL cholesterol compared to a reference method, without human interpretation as part of the primary output.

    7. Type of Ground Truth Used

    • Accuracy: The ground truth for the accuracy study was established using a reference laboratory instrument (Roche Cobas Integra 400 plus HDL Cholesterol). This is considered a highly reliable, objective, and standardized method for clinical chemistry analytes.
    • Precision: The ground truth for precision is the central tendency (mean) of repeated measurements of controlled samples.

    8. Sample Size for the Training Set

    • This document describes performance testing for a medical device (CardioChek Test System) that determines quantitative values using enzymatic methods on dry strips read by reflectance photometry. It is a traditional in vitro diagnostic device, not an AI/machine learning algorithm that requires a "training set" in the conventional sense for model development. The device's calibration parameters are programmed into a lot-specific memory chip. Therefore, there is no "training set" in the context of machine learning model development. The calibration and manufacturing processes would involve internal quality control and standardization.

    9. How the Ground Truth for the Training Set Was Established

    • As there is no "training set" for an AI/machine learning model, this question is not applicable. The device's operational parameters and calibration are established through manufacturing processes and validated through studies like those described (accuracy and precision).
    Ask a Question

    Ask a specific question about this device

    K Number
    K140068

    Validate with FDA (Live)

    Date Cleared
    2015-05-22

    (494 days)

    Product Code
    Regulation Number
    862.1345
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CardioChek Home Test System is a small portable analyzer and test strip system for self-testing by lay users. It is intended to be used by a single person and should not be shared. This system is for in vitro diagnostic use only. The tests strips are for the quantitative determination of glucose, total cholesterol, HDL (high density lipoprotein) cholesterol and triglycerides in capillary whole blood from the fingertip. A Chol/HDL ratio and estimated values for LDL (low density lipoprotein) cholesterol and non-HDL cholesterol are calculated by the CardioChek Plus analyzer.

    • Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess ● cholesterol in the blood and lipoprotein metabolism disorders.
    • HDL (lipoprotein) measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases.
    • Triglycerides measurements are used in the diagnosis and treatment of patients with diabetes mellitus, ● nephrosis, liver obstruction, other diseases involving lipid metabolism or various endocrine disorders.
    • Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

    The CardioChek Plus Test System is a small portable analyzer and test strip system intended for multiple patient use in professional healthcare settings. This system should only be used with single-use, auto-disabling lancing devices. This system is for in vitro diagnostic use only. The tests strips are for the quantitative determination of glucose, total cholesterol, HDL (high density lipoprotein) cholesterol and triglycerides in venous whole blood and capillary whole blood from the fingertip. A Chol/HDL ratio and estimated values for LDL (low density lipoprotein) cholesterol and non-HDL cholesterol are calculated by the CardioChek Plus analyzer.

    • Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipoprotein metabolism disorders.
    • HDL (lipoprotein) measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases.
    • Triglycerides measurements are used in the diagnosis and treatment of patients with diabetes mellitus, nephrosis, liver obstruction, other diseases involving lipid metabolism or various endocrine disorders.
    • Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders ● including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
    Device Description

    The CardioChek Plus professional test system consists of the CardioChek Plus professional analyzer and analyte specific test strips and is intended for professional use. The CardioChek Plus professional analyzer is an in vitro diagnostic device consisting of both a reflectance photometer and amperometer. This device measures various analytes in blood once the blood is applied to dry phase test strips that are specifically designed for reflectance or amperometric (electrochemical) analysis.

    The CardioChek Home test system consists of the CardioChek Home analyzer and analyte specific test strips. The CardioChek Home analyzer is an in vitro diagnostic device consisting of both a reflectance photometer and amperometer. This device measures various analytes in blood once the blood is applied to dry phase test strips that are specifically designed for reflectance or amperometric (electrochemical) analysis.

    The CardioChek Plus professional and CardioChek Home test systems are a modification of the original device (cleared as BioScanner Plus, K014099) to include the following modifications:

    • · Wireless communications capability (professional system)
    • · Software solutions capability
    • · Printer connectivity capability
    • · MEMo chip appearance
    • · Battery type
    • · Multiple language software capability
    • · Wired PC Communication
    • · Analyzer dimensions
    AI/ML Overview

    The provided text is a 510(k) summary for the CardioChek Plus and CardioChek Home Test Systems. It describes the devices, their intended use, and states their substantial equivalence to a predicate device (BioScanner Plus System, K014099). However, it does not contain specific acceptance criteria, study data, or details about patient sample sizes, training sets, or ground truth establishment methods.

    Therefore, I cannot fulfill your request for:

    1. A table of acceptance criteria and the reported device performance.
    2. Sample sizes used for the test set and the data provenance.
    3. Number of experts used to establish the ground truth for the test set and their qualifications.
    4. Adjudication method.
    5. Multi-reader multi-case (MRMC) comparative effectiveness study results or effect size.
    6. Standalone (i.e., algorithm only without human-in-the-loop performance) study results.
    7. The type of ground truth used.
    8. The sample size for the training set.
    9. How the ground truth for the training set was established.

    The document primarily focuses on the device modifications and
    the argument for substantial equivalence based on similarities and minor
    differences from the predicate device (BioScanner Plus System K014099).
    It states that "The CardioChek Plus and CardioChek Home test systems are as safe and effective and perform as well as our predicate device, the BioScanner Plus."
    However, it does not provide the detailed study results that would
    substantiate this claim in terms of specific performance metrics or
    acceptance criteria.

    Key information present in the document:

    • Device Names: CardioChek Plus Test System, CardioChek Home Test System
    • Intended Use: Quantitative determination of glucose, total cholesterol, HDL, and triglycerides in capillary whole blood (CardioChek Home) and venous/capillary whole blood (CardioChek Plus).
    • Predicate Device: BioScanner Plus System (K014099)
    • Modifications: Wireless communications, software solutions, printer connectivity, MEMo chip appearance, battery type, multi-language software, wired PC communication, analyzer dimensions. These are cosmetic, connectivity, and usability changes, not fundamental changes to the core measurement technology.
    • Conclusion: The new devices are "as safe and effective and perform as well as our predicate device, the BioScanner Plus." This conclusion implies that the performance of the predicate device serves as the implicit "acceptance criteria" and that the modifications did not degrade this performance.

    To answer your questions thoroughly, detailed study reports (e.g., analytical and clinical validation studies) would be required, which are not included in this 510(k) summary document.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1