510(k) Data Aggregation

    K Number
    K223470

    Validate with FDA (Live)

    Device Name
    C50
    Manufacturer
    Date Cleared
    2023-12-14

    (392 days)

    Product Code
    Regulation Number
    872.1745
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    C50 is intended for clinical practice of general dentistry, as an aid in the diagnosis of pit and fissure caries, as an aid to highlight dental plaque and gingival inflammations (restricted to gingival inflammations which lead to bleeding upon probing) and as intra-oral camera to visualize anatomical details invisible to the naked eye or with a mirror (thanks to magnification).

    Device Description

    The C50 is an intraoral video camera equipped with LEDs to illuminate the inspection site. The C50 optics and its complementary metal oxide semiconductor (CMOS) sensor capture the natural fluorescence of the site under observation and convert the images into a video signal that is sent to the computer via a USB interface. The dental practitioner can use the images displayed on the computer screen as an aid in diagnosis. A USB Type-C (camera) to USB-A cable is provided to connect the C50 to a computer. The C50 provides the following functions: Aid in the detection of pit and fissure caries Information about patient dental hygiene Highlighting of dental plaque Highlighting of gingival inflammations (restricted to gingival inflammation that leads to bleeding upon probing) Showing the difference between "before" and "after" care (follow-up). In CARIO mode (blue mode), the camera assists the dental practitioner by highlighting potential carious areas in pits and fissures on the occlusal surface of the teeth. In DAYLIGHT mode (white mode), the camera enables visualization of anatomical details invisible to the naked eye or with a mirror (thanks to its magnification). DAYLIGHT+ mode (white mode) is a shortcut version of DAYLIGHT mode where the contrast has been pushed from 20 to 70. In PERIO mode (yellow mode), the camera helps the dental practitioner to visualize dental plaque, in addition to highlighting areas of gingival inflammation (restricted to gingival inflammation that leads to bleeding upon probing).

    AI/ML Overview

    The C50 is an intraoral camera intended for use in general dentistry. Its indications for use include aid in the diagnosis of pit and fissure caries, highlighting dental plaque and gingival inflammations (restricted to those leading to bleeding upon probing), and functioning as an intra-oral camera for magnified visualization of anatomical details.

    Here's the breakdown of acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document describes comparison testing and other verification and validation tests rather than specific quantitative acceptance criteria for clinical performance metrics (e.g., sensitivity, specificity for caries detection). Instead, the performance testing focuses on demonstrating substantial equivalence to the predicate device and compliance with relevant safety and technical standards.

    Acceptance Criteria CategorySpecific Test/Performance AspectReported Device Performance
    Functional EquivalenceClinical Modes (CARIO, DAYLIGHT, PERIO)Demonstrated to be identical to the predicate device.
    New Mode (DAYLIGHT+)Functions similarly to DAYLIGHT mode but offers higher contrast.
    Image Quality (sharpness, size, resolution, focus position)Comparison testing showed substantial equivalence to the predicate device, even with the Full HD resolution enhancement of the C50.
    Safety and PerformanceTemperature elevation in target tissueComparison testing showed substantial equivalence to the predicate device.
    Mechanical performance of reprocessed C50TIPS (maintains position, blocks external light)Testing ensured the device maintains its position and completely blocks external light.
    Fit of third-party protective sheath with C50 and C50TIPSMechanical testing confirmed appropriate fit. -----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
    Electrical Safety (IEC 60601-1 and 60601-1-2)Testing was conducted, and the device was found to be consistent with appropriate sections of the standards. -----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
    EMC (IEC 60601-1-2, IEC/TR 60601-4-2)Testing was conducted, and the device was found to be consistent with appropriate sections of the standards. -------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
    Photobiological Safety (IEC 62471)Testing was conducted, and the device was found to be consistent with the appropriate section of the standard. ---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
    Software Validation (IEC 62304)Validation was performed on the C50's firmware, consistent with the appropriate section of the standard. ---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
    Reprocessing Validation (FDA Guidance Document "Reprocessing Medical Devices in Health Care Settings")Validation was conducted on the C50, C50TIPS, and USB cable in accordance with FDA guidance. ---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
    Biocompatibility (ISO 10993-1, FDA Guidance Document "Use of International Standard ISO 10993-1") (C50TIPS)Chemical characterization, cytotoxicity, and acute systemic toxicity tests were considered, and acceptance criteria were met for all. The patient contact materials are identical in formulation, manufacturing, processing, and geometry to the predicate device.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state a test set in the traditional sense of a clinical diagnostic study with patient data used for accuracy evaluation. Instead, it refers to "comparison testing of image quality" and "mechanical performance testing" conducted on the device and its accessories (C50, C50TIPS, protective sheath, USB cable).

    • Sample Size: Not specified in terms of number of patients or cases. The testing appears to have been performed on the devices themselves and associated accessories.
    • Data Provenance: Not applicable in the context of device performance testing. The "data" generated comes from direct laboratory and safety testing of the device hardware and software. No patient or human subject data is mentioned for performance evaluation.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. As described above, the testing performed was focused on device characteristics and substantial equivalence to a predicate, not on a diagnostic accuracy study requiring expert ground truth establishment for patient cases.

    4. Adjudication Method for the Test Set

    Not applicable, as no clinical diagnostic test set with multiple reader interpretations requiring adjudication is described.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs without AI Assistance

    No MRMC comparative effectiveness study is described. The C50 is an intraoral camera and not an AI-powered diagnostic algorithm that assists human readers in interpretation. It provides visual information to the dental practitioner.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    The C50 is a hardware device (intraoral camera with different illumination modes) that provides visual information. It is not an algorithm, and therefore, a standalone algorithm performance study is not applicable. The device's "diagnosis aid" function relies on the visual information it provides to the human practitioner.

    7. The Type of Ground Truth Used

    The ground truth used for relevant testing was based on:

    • Comparison to a predicate device (SOPROCARE): Many aspects of the C50's performance (image quality, temperature elevation, functional modes) were evaluated against the known performance of the predicate device to establish substantial equivalence.
    • Compliance with established standards: Electrical safety, EMC, photobiological safety, software validation, reprocessing validation, and biocompatibility were evaluated against recognized international and FDA standards.
    • Engineering specifications/functional requirements: Mechanical tests ensured specific physical functionalities like the fit of accessories and light blocking.

    8. The Sample Size for the Training Set

    Not applicable. The C50 is a medical device, not an AI/machine learning model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K191106

    Validate with FDA (Live)

    Date Cleared
    2019-12-13

    (232 days)

    Regulation Number
    870.1025
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The C50 and C80 patient monitors are intended to be used for monitoring, displaying, reviewing, alarming and storing of multiple physiological parameters as following: ECG (3-lead or 12-lead selectable, arrhythmia detection, heart rate (HR)), Respiration rate (RR), SpO2, pulse rate (PR), non-invasive blood pressure (NIBP), temperature (Temp), invasive blood pressure (IBP), carbon dioxide (CO2), anesthetic gas (AG), cardiac output (C.O.) for single patient. The C80 can also monitor the ICG (impedance cardiography) and BIS (bispectral index).

    All the parameters can be monitored on single adult, pediatric, and neonatal patients with the exception of the following:

    · The ICG monitoring is applicable to the adult patients of 122229cm in height and 30159Kg (67~341 pounds) in weight only;

    · NIBP measurement continual mode is not applicable to neonates;

    · Anesthetic depth (BIS) of the multi-parameter patient monitor is not intended for neonatal patients;

    The monitors are to be used in general healthcare facilities by clinical physicians or appropriate medical staff under the direction of physicians. The monitors are not intended for home use.

    Device Description

    The C50 and C80 patient monitors are intended to be used for monitoring, displaying, reviewing, alarming and storing of multiple physiological parameters as following: ECG (3-lead, 5-lead or 12-lead selectable, arrhythmia detection, heart rate (HR)), Respiration rate (RR), SpO2, pulse rate (PR), non-invasive blood pressure (NIBP), temperature (Temp), invasive blood pressure (IBP), carbon dioxide (CO2), anesthetic gas (AG), cardiac output (C.O.) for single patient. The C80 can also monitor the ICG (impedance cardiography) and BIS (bispectral index).

    All the parameters can be monitored on single adult, pediatric, and neonatal patients with the exception of the following:
    • The ICG monitoring is applicable to the adult patients of 122229cm in height and 30159Kg (67~341 pounds) in weight only;
    • NIBP measurement continual mode is not applicable to neonates;
    • Anesthetic depth (BIS) of the multi-parameter patient monitor is not intended for neonatal patients;
    Those monitors provide patient monitoring capabilities by using corresponding accessories.
    The multi-parameter Patient monitor, model: C50 and C80 have same design principle and technical characteristics:
    But the software in the host and modules and components are different.
    The differences between C50 and C80 are ICG, BIS and size of monitor.

    AI/ML Overview

    The provided document describes the acceptance criteria and a study proving the C50 and C80 Multi-parameter Patient Monitor meets these criteria.


    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally established through adherence to recognized international standards and clinical performance metrics. The document details a comparison with a predicate device (K170876, Passport Series Patient Monitors). While explicit "acceptance criteria" for each parameter are not always separately listed with a single performance value from the new device, the "Comparison" column in the table below infers if the new device meets or exceeds the predicate's performance or if sufficient justification is provided for differences.

    FeaturePredicate Device (K170876) PerformanceSubject Device (C50 & C80) PerformanceComparison to Acceptance Criteria (Predicate) & Justification
    Monitor SizePassport 17m: 400x370x193mm; Passport 12m: 297x336x187mmC80: Approx. 344x291x165mm; C50: Approx. 291.7x250x187mmThe monitor size of the subject device is smaller than the predicate device, but it "won't affect the safety and effectiveness of the subject device." Implied acceptance: smaller size is acceptable if safety/effectiveness are not compromised.
    Integrated display & touch screenColor TFT LCD (17-inch, 1280x1024 pixels for 17m; 12-inch, 800x600 pixels for 12m)Color TFT LCD (C80: 12.1-inch, 800x600 pixels; C50: 10.4-inch, 800x600 pixels)Screen size is smaller, but resolution for C80/C50 is the same as Passport 12m. "Same. The screen of Subject device is smaller than predicate device. Complying with IEC 60601-1 and IEC 62366-1 also indicates the clinical use is safe and effective." Implied acceptance: smaller screen size is acceptable due to compliance with relevant IEC standards for safety and usability.
    Power supplyTwo rechargeable Lithium-ion batteries or AC for Passport 17m; One rechargeable Lithium-ion battery or AC for Passport 12mC50 and C80: Powered either by built-in battery or external AC."Same, both powered by battery and AC." Implied acceptance: functional equivalence.
    Battery11.1V, 4500mAh11.1V, 2200mAh/4400mAh"The battery capacity is different. The batteries of C50/C80 have complied with IEC 62133. C50 and C80 have conformed to IEC 60601-1." Implied acceptance: different capacity is acceptable as long as safety (IEC 62133, IEC 60601-1) is maintained.
    ECG3-lead, 5-lead and 12-lead selectable, heart rate (HR)3-lead, 5-lead and 12-lead selectable, heart rate (HR)"Same." Implied acceptance: functional equivalence.
    ECG (Arrhythmia Analysis)Asystole, VFib/VTac, Vtac, Vent. Brady, Extreme Tachy, Extreme Brady, PVC, Couplet, Bigeminy, Trigeminy, Ron T, Run PVCs, PVCs, Tachy, Brady, Missed Beats, Vent. Rhythm, PNP, PNC, Multif. PVC, Nonsus. Vtac, Pause, Irr. Rhythm, AFib (24 types)Asystole, ventricular fibrillation, R ON T, VT >2, Couplet, PVC, Bigeminy, Trigeminy, Brady (Bradycardia), PNC (Pacer Not Capture), PNP (Pacer Not Pace), Missed Beats, IHB (Irregular Heart Beat), VTAC (Ventricular Tachycardia), Tachy (Tachycardia), PVC Too High, Extreme Tachycardia, Extreme Bradycardia, Ventricular Rhythm, Heart Pause (20 types)"C50/C80 has 20 types of arrhythmias in total, 18 of which is same as the predicate device. The other two (underline) is supported by compliance with EC57." Implied acceptance: fewer types are acceptable as long as key arrhythmia detections are present and supported by EC57 compliance.
    RespirationMethod: Trans-thoracic impedance. Range: adult:0-120 rpm; pediatrics:0-150rpm; neonate:0-150rpm. Accuracy: 7 to 150rpm: ± 2rpm or ± 2%, whichever is greater. 0 to 6rpm: not specified.Method: Trans-thoracic impedance. Range: adult:0-120 rpm; pediatrics:0-150rpm; neonate:0-150rpm. Accuracy: 7 to 150rpm: ± 2rpm or ± 2%, whichever is greater. 0 to 6rpm: not specified."Same." Implied acceptance: functional and performance equivalence.
    Pulse oxygen saturation (SpO2)Method: red and infrared light method. Masimo SpO2: Range:1100%. Accuracy: No motion: 70-100% ±2%(adult/pediatric), ±3%(neonate); Motion: 70-100% ±3%. Nellcor SpO2: Range:0100%. Accuracy: 70-100% ±2%(adult/pediatric), ±3%(neonate). Mindray SpO2: Range:0~100%. Accuracy: 70-100% ±2%(adult/pediatric), ±3%(neonate).Method: same. Masimo SpO2: same range & accuracy. Nellcor SpO2: same range & accuracy. Comen SpO2: Range: 0%-100%. Accuracy: 70-100% ±2%(adult/pediatric, non-motion), ±3%(neonate, non-motion)."Same." (for Masimo and Nellcor modules). For their own Comen SpO2 module, it shows equivalent accuracy. Implied acceptance: performance equivalence with predicate's different OEM modules and new Comen module meets similar accuracy.
    Pulse rate (PR)From Mindray SpO2 (Range 20-254 bpm, Acc ±3 bpm), Masimo SpO2 (Range 25-240 bpm, Acc ±3 bpm no motion, ±5 bpm motion), Nellcor SpO2 (Range 20-300 bpm, Acc ±3 bpm for 20-250 bpm), IBP sensor (Range 25-350 bpm, Acc ±1 bpm or ±1%).From Masimo SpO2 (Range 25-240 bpm, Acc ±3 bpm no motion, ±5 bpm motion), Nellcor SpO2 (Range 20-300 bpm, Acc ±3 bpm for 20-250 bpm), Comen SpO2 (Range 20-254 bpm, Acc ±2 bpm), IBP sensor (Range 25-350 bpm, Acc ±1 bpm or ±1%), NIBP sensor (Range 40-240 bpm, Acc ±3 bpm or ±3%)."The PR from Comen SpO2 is more accuracy than the PR form Mindray SpO2. The PR form Masimo SpO2, Nellcor SpO2 and IBP sensor of C50 and C80 are the same with the predicate device. In addition, C50 and C80 have one more PR data source (NIBP) than predicate device. C50 and C80 have complied with ISO 80601-2-61." Acceptance criteria met or exceeded; added NIBP source (improvement). Adherence to ISO 80601-2-61.
    Non-invasive blood pressure (NIBP)Method: Oscillometry. Range: Adult (systolic 25-290mmHg, diastolic 10-250mmHg), pediatric (systolic 25-240mmHg, diastolic 10-200mmHg), neonate (systolic 25-140mmHg, diastolic 10-115mmHg). Accuracy: Max mean error: ±5 mmHg; Max standard deviation: 8 mmHg.Method: same. Range: Adult (systolic 40-270mmHg, diastolic 10-215mmHg), pediatric (systolic 40-200mmHg, diastolic 10-150mmHg), neonate (systolic 40-135mmHg, diastolic 10-100mmHg). Accuracy: 0-300mmHg: ±3mmHg."The NIBP measure range for adult/pediatric/neonate in C50 and C80 is smaller than that in predicate device. The C50 and C80 is more accuracy than predicate device. C50 and C80 have conformed to IEC 80601-2-30." Acceptance criteria met or exceeded in accuracy, and justification for smaller range implicitly accepted by meeting IEC 80601-2-30.
    Temperature (Temp)Method: Thermal resistance. Range: 0 ~ 50°C. Accuracy: ±0.1°C.Method: Thermal resistance. Range: 0 ~ 50°C. Accuracy: ±0.2°C."The predicate device is more accuracy than C50 and C80. C50 and C80 have complied with ISO 80601-2-56." Implied acceptance: slightly lower accuracy is acceptable as long as it complies with ISO 80601-2-56.
    Carbon dioxide (CO2)Method: Infrared absorption. Masimo CO2: Sidestream 0-99mmHg, AwRR 0-120rpm. Microstream 0-99mmHg, AwRR 0-150rpm. Accuracy: Sidestream 0-40mmHg ±2mmHg, 41-76mmHg ±5%, 77-99mmHg ±10%, AwRR ±2rpm. Microstream 0-38mmHg ±2mmHg, 39-99mmHg ±5% +0.08%, AwRR 0-70rpm ±1rpm, 71-120rpm ±2rpm, 121-150rpm ±3rpm.Method: same. Masimo CO2: Sidestream 0-190mmHg, 0-25% (760mmHg), AwRR 0-150rpm. Mainstream 0-190mmHg, 0-25% (760mmHg), AwRR 0-150rpm. Respironics CO2: Sidestream 0-150mmHg, 0-19.7%, AwRR 0, 2-150rpm. Mainstream 0-150mmHg, 0-19.7%, AwRR 0, 2-150rpm. Accuracy (Masimo): ±(2.25mmHg +reading x 4%), AwRR ±1rpm. Accuracy (Respironics): 0-40mmHg ±2mmHg, etc., AwRR ±1rpm."The C50 and C80 are more accuracy and measured wider than predicate device. For C50 and C80, Masimo CO2 module (K123043 and K103604) and Respironics CO2 module (K040183 and K053174) have been cleared. Both modules have complied with ISO 80601-2-55." Acceptance criteria met or exceeded; wider range and higher accuracy; existing modules previously cleared and comply with ISO 80601-2-55.
    Invasive blood pressure (IBP)Method: Direct invasive measurement. Range: -50 to 300 mmHg. Accuracy: ±2% or ±1 mmHg, whichever is greater (without sensor).Method: Direct invasive measurement. Range: -50 to 300 mmHg. Accuracy: ±2% or ±1 mmHg, whichever is greater (without sensor)."Same." Implied acceptance: functional and performance equivalence.
    Cardiac output (C.O.)Method: Thermodilution method. Range: 0.1 to 20 L/min. Accuracy: ±5% or ±0.1 L/min, whichever is greater.Method: Thermodilution method. Range: 0.1 to 20 L/min. Accuracy: ±5% or ±0.1 L/min, whichever is greater."Same." Implied acceptance: functional and performance equivalence.
    Anesthetic gas (AG)Method: Infrared absorption. Range (CO2, N2O, Hal, Enf, Iso, Sev, O2, AwRR). Accuracy (CO2 ±0.3% ABS, N2O ±(8%REL+2%ABS), Other AG 8%REL).Method: same. Range (CO2, N2O, Hal, Enf, Iso, Sev, O2, AwRR). Accuracy (CO2 0-15%: ±(0.2kPa+readingx2%), N2O ±(2 kPa+readingx2%), Hal/Enf/Iso 0-8%: ±(0.15%+readingx5%), Sev 0-10%: ±(0.15%+readingx5%), Des 0-22%: ±(0.15%+readingx5%), O2 ±(1%+readingx2%), Masimo AG AwRR ±1rpm)."For CO2/Enf/Hal/Iso/Sev/Des, The measurement range of C50 and C80 is smaller than the predicate device. For awRR, the measurement range of C80 is better than the predicate device. The C50 and C80 are more accuracy than predicate device. The measurement range of C50 and C80 is enough for most environments; it won't affect the safety and effectiveness. The C50 and C80 supports two AG modules: MASIMO ISA AX+ Sidestream module and Masimo IRMA AX+ Mainstream module, both of them have been cleared in K103604." Acceptance criteria: different ranges are acceptable with justification of sufficient range for most environments and high accuracy. Utilizes previously cleared and compliant modules.
    BIS (C80 only)Range: BIS, BIS L, BIS R: 0-100; SQI, SQI L, SQI R: 0-100%; EMG, EMG L, EMG R: 0-100 Db; SR, SR L, SR R: 0-100%; SEF, SEF L, SEF R: 0.5-30.0 Hz; TP, TP L, TP R: 40-100 Db; BC, BC L, BC R: 0-30; sBIS L, sBIS R: 0-10.0; Semg L, Semg R: 0-10.0; ASYM: 0-100%. Accuracy: 1% for BIS, SQI, EMG, ESR (implied).Range and Accuracy: BIS: same; accuracy: 1%. SQI: same; accuracy: 1%. EMG: same; accuracy: 1%. ESR: 0-100%; accuracy: 1%."The underline parameters are what C50 and C80 doesn't have. The BIS module and sensor have been cleared by FDA and its 510k numbers are K040183 and K002734." Acceptance criteria for essential parameters are met by using previously cleared FDA modules.
    ICG (C80 only)Method: Indirect impedance cardiograph measurement. Range: SV: 5-250 ml; HR: 44-2m; C.O. 1.4-15 L/min. Accuracy: SV: Not specified; HR: ±2 bpm; C.O. Not specified.Method: Indirect impedance cardiograph measurement. Range: HR: 40-250bpm; SV: 0-250mL; C.O.:0-30L/min; TFC: 5-150 /KΩ; SVR:0-3500 dyn・s ・ cm-5. Accuracy: SV: Not specified; HR: ±2 bpm; C.O. Not specified."The underline parameter is what C50 and C80 doesn't have. The ICG electrode cable and sensors are all cleared by FDA, 510(k) number of which is K110645." Acceptance criteria for essential parameters are met by using previously cleared FDA components.

    Study that Proves the Device Meets Acceptance Criteria

    The essential studies to prove the device meets acceptance criteria are divided into Non-Clinical Performance Data and Clinical Studies.

    Non-Clinical Performance Data:

    • Software Verification and Validation Testing: Conducted in accordance with FDA guidance "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." Risk analysis was performed, and the software was deemed a "major" level of concern.
    • Electrical safety and Electromagnetic Compatibility (EMC): Testing conducted according to:
      • ANSI AAMI ES60601-1:2005/(R)2012 And A1:2012 (IEC 60601-1:2005, MOD)
      • IEC 60601-1-2 Edition 4: 2014-02
    • Bench Testing: Functional and system-level tests were performed, showing the devices meet specifications and perform equivalently to the predicate.
    • Biocompatibility Testing: Performed for patient-contacting components (ECG cable, SpO2 probes, temperature probes) that were not previously cleared. Cytotoxicity, sensitization, and irritation testing were conducted based on ISO 10993-1 and FDA's 2016 biocompatibility guidance.
    • Conformance with Applicable Standards: The device demonstrated compliance with numerous recognized consensus standards, in addition to those listed above for specific modules (e.g., ISO 80601-2-55 for Respiratory Gas Monitors, ISO 80601-2-56 for Thermometers, ISO 80601-2-61 for Pulse Oximeter Equipment, ANSI AAMI EC57:2012 for Cardiac Rhythm and ST-Segment measurement algorithms).

    Clinical Studies:

    1. NIBP Clinical Study - for adults and children

    • Device Parameter: Non-invasive Blood Pressure (NIBP)
    • Acceptance Criteria/Standard: Complies with ISO 81060-2 Second Edition 2013-05-01 "Non-Invasive Sphygmomanometers - Part 2: Clinical Validation Of Automated Measurement Type."
    • Sample Size: 25 patients (19 adults, 6 children).
      • Demographics: 10 men, 15 women. 6 patients aged 3-12, 2 aged 12-40, 10 aged 40-60, 7 above 60.
    • Data Provenance: Retrospective or Prospective not explicitly stated, but implies prospective data collection from "The Second Affiliated Hospital of Guangzhou Medical University and The First Affiliated Hospital of Guangzhou Traditional Chinese Medical University" from 2012 to 2013 (China).
    • Ground Truth: Implicitly referent method (e.g., auscultation by trained observers) as required by ISO 81060-2.
    • Adjudication Method: Not specified, but standard practice for NIBP clinical validation involves multiple observers for reference measurements.
    • Results: "The results are accurate and reliable, and the repeated measurement consistency is in good condition, within the measurement range. Moreover, No adverse events and side effects were found in clinical trials."

    2. NIBP Clinical Study - for neonate and infants

    • Device Parameter: Non-invasive Blood Pressure (NIBP)
    • Acceptance Criteria/Standard: Complies with ISO 81060-2 Second Edition 2013-05-01 "Non-Invasive Sphygmomanometers - Part 2: Clinical Validation Of Automated Measurement Type."
    • Sample Size: 20 subjects.
      • Demographics: 10 males, 10 females. 13 subjects <29 days, 4 subjects 29 days to 1 year, 3 subjects 1 to 3 years old.
      • Data Points: 10 groups of data collected for each subject, totaling 200 groups.
    • Data Provenance: Retrospective or Prospective not explicitly stated, but implies prospective data collection from "The First Affiliated Hospital of Guangxi Medical University and The First Affiliated Hospital of Xinxiang Medical University" from 2012 to 2013 (China).
    • Ground Truth: Implicitly referent method (e.g., auscultation by trained observers) as required by ISO 81060-2.
    • Adjudication Method: Not specified.
    • Results: "The results are accurate and reliable, and the repeated measurement consistency is in good condition, within the measurement range. Moreover, No adverse events and side effects were found in clinical trials."

    3. SpO2 Clinical Study

    • Device Parameter: Oxygen saturation (SpO2)
    • Acceptance Criteria/Standard: Complies with FDA Certification Guidance Document of Photoelectric Oximeter and ISO 80601-2-61 "Medical Electrical Equipment - Part 2-61: Particular Requirements For Basic Safety And Essential Performance Of Pulse Oximeter Equipment."
    • Sample Size: 30 subjects (24 adult subjects, 6 neonates).
      • Demographics (Adults): 24 subjects (7 males, 17 females) aged 24-44 years old (20 yellows, 4 blacks).
      • Demographics (Neonates): 6 neonates (5 males, 1 female) aged 1-24 days.
    • Data Provenance: Implies prospective data collection from "the Affiliated Hospital of Guilin Medical University" from January 20, 2016 to June 10, 2016 (China).
    • Ground Truth: Implicitly a controlled desaturation study where arterial blood gas samples are taken and analyzed for SaO2, which is compared to the pulse oximeter readings.
    • Adjudication Method: Not specified.
    • Results: "Per the clinical test analysis result, there is no adverse events, the oxygen saturation module and probe are safe and effective for the intended use, as well as demonstrating accurate pulse measurement."

    Additional Requested Information:

    • 2. Sample size for the test set and the data provenance:

      • NIBP (Adult/Children): 25 patients from China (2012-2013).
      • NIBP (Neonate/Infants): 20 subjects (200 data groups) from China (2012-2013).
      • SpO2: 30 subjects (24 adults, 6 neonates) from China (Jan-June 2016).
      • All clinical data appears to be prospective as it describes clinical trials "conducted in" specific hospitals during specific periods.
    • 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • The document does not specify the number or qualifications of experts (e.g., radiologists) for establishing ground truth in clinical studies. For NIBP validation against ISO 81060-2, ground truth typically involves at least two trained observers performing auscultatory measurements. For SpO2, ground truth is established by laboratory CO-oximetry measurements of arterial blood, usually managed by laboratory technicians in a controlled setting, not interpretation by "experts" in the sense of image analysis.
    • 4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • The document does not specify the adjudication method for any of the clinical or non-clinical tests. For NIBP, ISO 81060-2 dictates specific statistical analysis methods (e.g., Bland-Altman) for comparing device readings to reference measurements, often requiring two or more observers.
    • 5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • This device is a multi-parameter patient monitor measuring physiological parameters (ECG, SpO2, NIBP, Temp, etc.). It does not describe an AI-powered diagnostic imaging device or an AI-assistance feature for human readers. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance is not applicable and was not performed.
    • 6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • The device itself is a standalone patient monitor that provides automated measurements and alarms. The performance of its modules (e.g., NIBP, SpO2 algorithms) was evaluated independently against reference standards (clinical validation) and through bench testing. So, yes, standalone performance of the underlying algorithms and hardware was demonstrated through the clinical and non-clinical tests.
    • 7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • NIBP: Likely comparative measurements against a reference standard method (e.g., auscultatory measurements by trained observers) as specified by ISO 81060-2.
      • SpO2: Involves laboratory arterial blood gas analysis (CO-oximetry) during controlled desaturation studies, comparing the device's SpO2 readings to the directly measured SaO2 values.
      • Other Parameters (ECG, Temp, IBP, CO2, AG, BIS, ICG): Performance is primarily verified through bench testing against calibrated simulators and reference equipment and adherence to respective ISO/IEC standards.
    • 8. The sample size for the training set:

      • The document describes pre-market notification (510k) for a patient monitor, not an AI/machine learning device that typically involves a training set. The clinical studies performed are for validation of the device's performance against human subjects, not for training internal algorithms. Therefore, there is no explicit "training set" mentioned in the context of machine learning.
    • 9. How the ground truth for the training set was established:

      • As there is no explicit mention of an AI/machine learning training set, this question is not applicable. The ground truth for validation was established by reference standard methods in clinical trials and bench testing as described in point 7.