Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K162146

    Validate with FDA (Live)

    Date Cleared
    2017-01-05

    (157 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BRIGHTWAY BRAND NITRILE EXAMINATION GLOVES, POWDER FREE, [LAVENDER] is a disposable device intended for medical purpose, to be worn on the examiners hand or finger to prevent contamination between patient and examiner.

    Device Description

    The subject device in this 510(k) Notification is a Lavender Examination Glove. The subject device is a patient examination glove made from a Nitrile compound, Lavender in colour, powder free and non sterile (as per 21 CFR 880.6250, class I). The principle operation of the medical device to provide single use barrier protection for the wearer and the device that meets all the requirement specifications for Barrier Protection, tensile properties as defined in ASTM D6319-10 ; Standard specification for Nitrile Examination Gloves.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Brightway Brand Nitrile Examination Gloves, Powder Free (Lavender), based on the provided document:

    Acceptance Criteria and Reported Device Performance

    ParameterAcceptance Criteria (Standard)Reported Device PerformanceResult
    Dimension and Thickness
    Weight (for each size XS-XL)$Range around target (e.g., XS: $2.7 \pm 0.3g$)All samples within specificationPass
    Length (Min-Max)Min - 230mm, Tar - 240mm, Max - 258mmAll samples within specificationPass
    Width (Min-Max for each size XS-XL)Range around targetAll samples within specificationPass
    Finger Tip Thickness (Min-Max)Min - 0.06mm, Tar - 0.08mm, Max - 0.11mmAll samples within specificationPass
    Palm Thickness (Min-Max)Min - 0.05mm, Tar - 0.07mm, Max - 0.09mmAll samples within specificationPass
    Cuff Thickness (Min-Max)Min - 0.05mm, Tar - 0.06mm, Max - 0.08mmAll samples within specificationPass
    Freedom from HolesASTM D5151; AQL 1.5, Ac=7 Rej=8 for 200 pcs1 defective foundPass
    Physical Properties (Before Aging)
    Tensile Strength (MPa)16 (Min) [ASTM D 6319-10]27.805 (Avg.)Pass
    Ultimate Elongation (%)500 (Min) [ASTM D 6319-10]588.8 (Avg.)Pass
    Physical Properties (After Aging)
    Tensile Strength (MPa)15 (Min) [ASTM D 6319-10]26.075 (Avg.)Pass
    Ultimate Elongation (%)450 (Min) [ASTM D 6319-10]523.5 (Avg.)Pass
    Biocompatibility
    Primary Skin Irritation testISO 10993-10: Not an irritantNot an irritantPass
    Dermal Sensitization assayISO 10993-10: Not a contact sensitizerNot a contact sensitizerPass
    Systemic ToxicityISO 10993-11: Non-systemically toxicNon-systemically toxicPass
    Residual PowderASTM D6124: Meets requirementsMeets requirementsPass

    Study Information

    1. Sample size used for the test set and the data provenance:

      • Dimension Measurement: 20 pieces per batch. Random Sampling based on ISO2859-1:1999; S2 AQL 2.5, Ac=1 Rej=2.
      • Water-tight Test: 200 pieces per batch. Random Sampling based on ISO2859-1:1999; G2 AQL 1.5, Ac=7 Rej=8.
      • Physical Properties Analysis: Not explicitly stated but "Single Normal S2 AQL: 2.5 Acc/Rej : 1/2" for sampling plan. Given that 20 samples were tested and individual results are shown, it seems 20 samples were used for the physical properties test.
      • Biocompatibility Studies (Primary Skin Irritation, Dermal Sensitization, Systemic Toxicity): Sample sizes for these tests are not provided in the document.
      • Data Provenance: The tests were conducted by Brightway Holdings Sdn Bhd (the manufacturer) in Malaysia. The document does not specify if the data is retrospective or prospective, but as it's part of a 510(k) submission for a new device, it would typically involve prospective testing of newly manufactured devices.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

      • This information is not applicable to the device described. The "ground truth" for examination gloves is established by standardized physical and chemical testing methods (ASTM, ISO), not by expert consensus or interpretations.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not applicable. Performance is measured against objective standards, not subjective interpretations requiring adjudication.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not applicable. This device is an examination glove, not an AI-powered diagnostic or assistive tool.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Not applicable. This device is an examination glove, not an algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

      • The "ground truth" for this device is based on objective, standardized measurements and tests outlined in recognized standards such as ASTM D6319-10 (Standard Specification for Nitrile Examination Gloves), ASTM D5151 (Water Leak Test), ASTM D6124 (Residual Powder), and ISO 10993 (Biocompatibility).
    7. The sample size for the training set:

      • Not applicable. This device is not an AI/machine learning model that requires a training set.
    8. How the ground truth for the training set was established:

      • Not applicable. This device is not an AI/machine learning model.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1