510(k) Data Aggregation

    K Number
    K253578

    Validate with FDA (Live)

    Date Cleared
    2026-02-26

    (101 days)

    Product Code
    Regulation Number
    892.2080
    Age Range
    18 - 120
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BriefCase-Triage: CARE (Clinical AI Reasoning Engine) Multi-Triage CT for Pneumothorax; Pericardial effusion; Large aortic aneurysm; Shoulder fracture or dislocation device is a radiological computer aided triage and notification software indicated for use in the analysis of contrast and non-contrast CT images of the chest, abdomen, or chest/abdomen, in adults or transitional adolescents aged 18 and older. The device is intended to assist hospital networks and appropriately trained medical specialists in workflow triage by flagging and communicating suspected positive findings, per study, of:

    • Pneumothorax;
    • Pericardial effusion;
    • Large aortic aneurysm
    • Shoulder Fracture or Dislocation

    The device flags cases with at least one suspected finding to assist with triage/prioritization of medical images. The device will provide a flag for each suspected finding within this study. A preview image will be provided for each distinct suspected finding.

    BriefCase-Triage uses a foundation model-based artificial intelligence (AI) system to analyze images and highlight cases with detected findings in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected findings. Notifications include compressed preview images for each suspected finding that are meant for informational purposes only and not intended for diagnostic use beyond notification. The device does not alter the original medical images and is not intended to be used as a diagnostic device.

    The results of BriefCase-Triage are intended to be used in conjunction with other patient information and based on their professional judgment to assist with triage/prioritization of medical images. Notified clinicians are responsible for viewing full images per the standard of care.

    Device Description

    BriefCase-Triage: CARE Multi-Triage CT for Pneumothorax; Pericardial effusion; Large aortic aneurysm; Shoulder fracture or dislocation device is a radiological computer-assisted triage and notification software device. The software is based on an algorithm programmed component and is intended to run on a linux-based server in a cloud environment.

    The BriefCase-Triage device receives images that match meta-data criteria according to the BriefCase-Triage predefined set of parameters. Then, the BriefCase-Triage processes the series chronologically, identifying cases with suspected positive finding(s) and selecting key slice(s) for preview. BriefCase-Triage output consists of suspected positive flag/notification regarding the existence of each finding in the analyzed study. Each finding includes a Representative Key Slice. The Key Slice(s) may be presented to the users as compressed, low-quality, grayscale, preview images with the date and time imprinted. The previews are not annotated and are captioned with the disclaimer "Not for diagnostic use, for prioritization only" according to the device requirement from the Image Communication Platform (ICP).

    Presenting the users with worklist prioritization facilitates efficient triage by prompting the user to assess the relevant original images in the PACS. Thus, the suspect case receives attention earlier than would have been the case in the standard of care practice alone.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study proving device performance, based on the provided FDA 510(k) clearance letter:


    1. Acceptance Criteria and Reported Device Performance

    The core acceptance criteria are based on standalone performance metrics for each of the four clinical indications.

    IndicationAcceptance Criteria (Default Operating Point)Reported Device Performance (Default Operating Point)
    PneumothoraxAUC > 0.95 (lower bound 95% CI); Sensitivity > 80%; Specificity > 80%AUC: 98.9 (95% CI: 97.8-99.7) Sensitivity: 94.8% (95% CI: 89.5%-97.9%) Specificity: 95.9% (95% CI: 91.3%-98.5%)
    Pericardial effusionAUC > 0.95 (lower bound 95% CI); Sensitivity > 80%; Specificity > 80%AUC: 99.1 (95% CI: 98.0-99.8) Sensitivity: 96.4% (95% CI: 91.7%-98.8%) Specificity: 96.5% (95% CI: 92.0%-98.8%)
    Large aortic aneurysmAUC > 0.95 (lower bound 95% CI); Sensitivity > 80%; Specificity > 80%AUC: 99.5 (95% CI: 98.9-99.9) Sensitivity: 97.1% (95% CI: 92.7%-99.2%) Specificity: 97.2% (95% CI: 92.9%-99.2%)
    Shoulder fracture or dislocationAUC > 0.95 (lower bound 95% CI); Sensitivity > 80%; Specificity > 80%AUC: 99.9 (95% CI: 99.7-100) Sensitivity: 97.8% (95% CI: 93.7%-99.5%) Specificity: 99.3% (95% CI: 96.2%-100.0%)
    Time-to-notificationComparability with predicate device in time savings to standard of care.Subject Device Mean: 49.9 seconds (95% CI: 46.4-53.5) Predicate Device Mean: 10.7 seconds (95% CI: 10.5-10.9) Note: While the subject device's time is longer, the conclusion states comparability regarding time savings to standard of care review, implying it still offers significant benefit.

    Study Proving Device Meets Acceptance Criteria

    The study conducted was a retrospective, blinded, multicenter standalone performance analysis.

    2. Sample size used for the test set and the data provenance:
    * Sample Size: N = 280 for each of the 4 clinical indications, totaling 772 unique scans across all indications.
    * Data Provenance: The cases were collected from 6 US-based clinical sites, representing diverse geographic locations and site types. The data was "distinct in time or center from the cases used to train the algorithm," and "sequestered from algorithm development activities." This indicates a high level of independence for the test set. The data is retrospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
    * Number of Experts: Three (3)
    * Qualifications: Senior board-certified radiologists.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
    * The document states ground truth was "determined by three senior board-certified radiologists." It does not explicitly mention an adjudication method like 2+1 or 3+1, but the plural "radiologists" and the method of "determined by" suggests a consensus or majority opinion among these three, rather than individual opinions without interaction.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
    * No MRMC comparative effectiveness study was explicitly described. The study was a "standalone performance analysis" of the software itself. The comparison of "time-to-notification" with the predicate device implies a comparison of software performance characteristics related to triage, not a study of human readers with and without AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
    * Yes, a standalone performance study was done. The document explicitly refers to it as a "standalone performance analysis" to "evaluate the software's performance."

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
    * Expert Consensus: The ground truth was established by "three senior board-certified radiologists."

    8. The sample size for the training set:
    * The document does not specify the exact sample size for the training set. It only mentions that the "algorithm was trained during software development on images of the pathology."

    9. How the ground truth for the training set was established:
    * The ground truth for the training set was established by "labeled ('tagged') images. In that process, each image in the training dataset was tagged based on the presence of the critical finding." The method or type of tagging (e.g., by experts, automated, etc.) is not detailed, but it's implied that there was a process of assigning labels/tags to the images to indicate the presence or absence of the target pathologies.

    Ask a Question

    Ask a specific question about this device

    K Number
    K251195

    Validate with FDA (Live)

    Device Name
    BriefCase-Triage
    Date Cleared
    2026-01-27

    (285 days)

    Product Code
    Regulation Number
    892.2080
    Age Range
    18 - 120
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BriefCase-Triage is a radiological computer aided triage and notification software indicated for use in the analysis of contrast-enhanced CT images that include the brain, in adults or transitional adolescents aged 18 and older. The device is intended to assist hospital networks and appropriately trained medical specialists in workflow triage by flagging and communication of suspected positive cases of Brain Aneurysm (BA) findings that are 3.0 mm or larger.

    BriefCase-Triage uses an artificial intelligence algorithm to analyze images and flag suspect cases in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for suspect cases. Notifications include compressed preview images that are meant for informational purposes only and not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device.

    The results of BriefCase-Triage are intended to be used in conjunction with other patient information and based on professional judgment, to assist with triage/prioritization of medical images. Notified clinicians are responsible for viewing full images per the standard of care.

    Device Description

    BriefCase-Triage is a radiological computer-assisted triage and notification software device.

    The software is based on an algorithm programmed component and is intended to run on a linux-based server in a cloud environment.

    The BriefCase-Triage receives filtered DICOM Images, and processes them chronologically by running the algorithms on each series to detect suspected cases. Following the AI processing, the output of the algorithm analysis is transferred to an image review software (desktop application). When a suspected case is detected, the user receives a pop-up notification and is presented with a compressed, low-quality, grayscale image that is captioned "not for diagnostic use, for prioritization only" which is displayed as a preview function. This preview is meant for informational purposes only, does not contain any marking of the findings, and is not intended for primary diagnosis beyond notification.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study details for the BriefCase-Triage device, based on the provided FDA 510(k) clearance letter:

    Acceptance Criteria and Reported Device Performance

    MetricAcceptance Criteria (Performance Goal)Reported Device Performance
    Primary Endpoints
    Sensitivity80%87.8% (95% CI: 83.1%-91.6%)
    Specificity80%91.6% (95% CI: 87.9%-94.5%)
    Secondary Endpoints
    Time-to-Notification (mean)Comparable to predicate device44.8 seconds (95% CI: 41.4-48.2)
    Negative Predictive Value (NPV)N/A98.9% (95% CI: 98.4%-99.2%)
    Positive Predictive Value (PPV)N/A47.6% (95% CI: 38.4%-57.1%)
    Positive Likelihood Ratio (PLR)N/A10.5 (95% CI: 7.2-15.3)
    Negative Likelihood Ratio (NLR)N/A0.13 (95% CI: 0.1-0.19)

    Note on Additional Operating Points (AOPs): The device also met performance goals (80% sensitivity and specificity) for three additional operating points (AOP1, AOP2, AOP3) with slightly varying sensitivity/specificity trade-offs (e.g., AOP3: Sensitivity 86.2%, Specificity 93.6%).

    Study Details

    1. Sample size used for the test set and the data provenance:

    • Sample Size: 544 cases
    • Data Provenance: Retrospective, blinded, multicenter study from 6 US-based clinical sites. The cases were distinct in time or center from those used for algorithm training.

    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Number of Experts: Three (3) senior board-certified radiologists.
    • Qualifications: "Senior board-certified radiologists." (Specific number of years of experience not detailed in the provided text).

    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • The text states the ground truth was "determined by three senior board-certified radiologists." It doesn't explicitly describe an adjudication method like "2+1" or "3+1." This implies a consensus approach where all three radiologists agreed, or a majority rule, but the exact mechanism for resolving discrepancies (if any) is not specified.

    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, an MRMC comparative effectiveness study was NOT done. The study's primary objective was to evaluate the standalone performance of the BriefCase-Triage software. The secondary endpoint compared the device's time-to-notification to that of the predicate device, but not its impact on human reader performance.

    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Yes, a standalone performance study was done. The primary endpoints (sensitivity and specificity) measure the algorithm's performance in identifying Brain Aneurysm (BA) findings.

    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Expert Consensus: The ground truth was "determined by three senior board-certified radiologists."

    7. The sample size for the training set:

    • Not explicitly stated. The document mentions the algorithm was "trained during software development on images of the pathology" and that "critical findings were tagged in all CTs in the training data set." However, the specific sample size for this training data is not provided.

    8. How the ground truth for the training set was established:

    • Manually labeled ("tagged") images: The text states, "As is customary in the field of machine learning, deep learning algorithm development consisted of training on manually labeled ('tagged') images. In that process, critical findings were tagged in all CTs in the training data set." It does not specify who performed the tagging or their qualifications, nor the method of consensus if multiple taggers were involved.
    Ask a Question

    Ask a specific question about this device

    K Number
    K252970

    Validate with FDA (Live)

    Date Cleared
    2026-01-07

    (112 days)

    Product Code
    Regulation Number
    892.2080
    Age Range
    18 - 120
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BriefCase-Triage: CARE (Clinical AI Reasoning Engine) Multi-Triage CT Body is a radiological computer aided triage and notification software indicated for use in the analysis of contrast and non-contrast CT images of the chest, abdomen, and/or pelvis, in adults or transitional adolescents aged 18 and older. The device is intended to assist hospital networks and appropriately trained medical specialists in workflow triage by flagging and communicating suspected positive findings, per study, of:

    1. Diverticulitis;
    2. Abdominal-pelvic abscess;
    3. Appendicitis;
    4. Intestinal ischemia and/or pneumatosis;
    5. Obstructive renal stone;
    6. Small bowel obstruction;
    7. Large bowel obstruction;
    8. Spleen injury;
    9. Liver injury;
    10. Kidney injury;
    11. Pelvic fracture.

    The device flags cases with at least one suspected finding to assist with triage/prioritization of medical images. The device will provide a flag for each suspected finding within this study. A preview image will be provided for each distinct suspected finding.

    BriefCase-Triage uses a foundation model-based artificial intelligence (AI) system to analyze images and highlight cases with detected findings in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected findings. Notifications include compressed preview images for each suspected finding that are meant for informational purposes only and not intended for diagnostic use beyond notification. The device does not alter the original medical images and is not intended to be used as a diagnostic device.

    The results of BriefCase-Triage are intended to be used in conjunction with other patient information and based on their professional judgment to assist with triage/prioritization of medical images. Notified clinicians are responsible for viewing full images per the standard of care.

    Device Description

    Briefcase-Triage is a radiological computer-assisted triage and notification software device. The software is based on an algorithm programmed component and is intended to run on a linux-based server in a cloud environment.

    The BriefCase-Triage receives images that match meta-data criteria according to BriefCase-Triage: CARE Multi-Triage CT Body's predefined set of parameters. Then, the BriefCase-Triage processes the series chronologically, identifying cases with suspected positive finding(s) and selecting key slice(s) for preview. BriefCase-Triage output consists of suspected positive flag/notification regarding the existence of each finding in the analyzed study. Each finding includes a Representative Key Slice. The Key Slice(s) may be presented to the users as compressed, low-quality, grayscale, preview images with the date and time imprinted. The previews are not annotated and are captioned with the disclaimer "Not for diagnostic use, for prioritization only" according to the device requirement from the Image Communication Platform (ICP).

    AI/ML Overview

    Acceptance Criteria and Study Details for BriefCase-Triage: CARE Multi-triage CT Body

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the BriefCase-Triage: CARE Multi-triage CT Body device were primarily defined by performance goals for Area Under the Curve (AUC), Sensitivity (Se), and Specificity (Sp). The study demonstrated that the device met and exceeded these criteria for all 11 indications.

    IndicationPerformance Goal (Acceptance Criteria)Reported Device Performance (Mean)95% Confidence Interval (Reported)
    Primary Endpoints
    Finding-level AUC> 0.950.974 - 1.000.952 - 1.00
    Sensitivity (Se)> 80%94.0% - 99.3%88.9% - 100%
    Specificity (Sp)> 80%95.7% - 99.3%91% - 100%
    Secondary Endpoints (Comparable to Predicate)
    BriefCase time-to-notificationComparable to predicate45 seconds43.4 - 46.5 seconds

    Note: The reported device performance for AUC, Sensitivity, and Specificity are ranges covering the minimum and maximum values observed across the 11 indications in the pivotal study. Detailed values for each indication are provided in the source text.

    2. Sample Size and Data Provenance for the Test Set

    • Sample Size: N = 280 for each of the 11 clinical indications, resulting in 1769 unique scans included across all device indications.
    • Data Provenance: The data was collected from 6 US-based clinical sites. It was retrospective and the cases were distinct in time or center from the cases used to train the algorithm.

    3. Number of Experts and Qualifications for Ground Truth

    • Number of Experts: Three senior board-certified radiologists.
    • Qualifications: The document specifically states "senior board-certified radiologists." No further details on years of experience were provided.

    4. Adjudication Method for the Test Set

    The adjudication method used to establish ground truth was based on the "consensus" of the three senior board-certified radiologists ("as determined by three senior board-certified radiologists"). This implies a consensus-based adjudication, but the specific mechanics (e.g., majority vote like 2+1, or requiring all three to agree) are not explicitly detailed.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC study comparing human readers with AI assistance versus without AI assistance was reported in this document. The study described is a standalone performance analysis of the algorithm.

    6. Standalone Performance Study

    Yes, a standalone performance study was done. The document states: "Aidoc conducted a retrospective, blinded, multicenter study with the Briefcase-Triage software to evaluate the standalone performance analysis individually for each of the 11 clinical indications supported by BriefCase-Triage: CARE Multi-triage CT Body device."

    7. Type of Ground Truth Used

    The ground truth was established by expert consensus of three senior board-certified radiologists.

    8. Sample Size for the Training Set

    The sample size for the training set is not explicitly provided in the given text. It is only mentioned that "the algorithm was trained during software development on images of the pathology."

    9. How the Ground Truth for the Training Set was Established

    The ground truth for the training set was established through labeled ("tagged") images. The document states: "As is customary in the field of machine learning, deep learning algorithm development consisted of training on labeled ("tagged") images. In that process, each image in the training dataset was tagged based on the presence of the critical finding." The specific method or expert involvement in this tagging process is not detailed, but it implies human expert labeling.

    Ask a Question

    Ask a specific question about this device

    K Number
    K253265

    Validate with FDA (Live)

    Device Name
    BriefCase-Triage
    Date Cleared
    2025-11-06

    (38 days)

    Product Code
    Regulation Number
    892.2080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BriefCase-Triage is a radiological computer aided triage and notification software indicated for use in the analysis of abdominal CT images in adults or transitional adolescents aged 18 and older. The device is intended to assist hospital networks and appropriately trained medical specialists in workflow triage by flagging and communication of suspected positive findings of Intra-abdominal free gas (IFG) pathologies.

    BriefCase-Triage uses an artificial intelligence algorithm to analyze images and highlight cases with the detected findings in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected findings. Notifications include compressed preview images that are meant for informational purposes only and not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device.

    The results of BriefCase-Triage are intended to be used in conjunction with other patient information and based on their professional judgment, to assist with triage/prioritization of medical images. Notified clinicians are responsible for viewing full images per the standard of care.

    Device Description

    Briefcase-Triage is a radiological computer-assisted triage and notification software device.

    The software is based on an algorithm programmed component and is intended to run on a linux-based server in a cloud environment.

    The Briefcase-Triage receives filtered DICOM Images, and processes them chronologically by running the algorithms on each series to detect suspected cases. Following the AI processing, the output of the algorithm analysis is transferred to an image review software (desktop application). When a suspected case is detected, the user receives a pop-up notification and is presented with a compressed, low-quality, grayscale image that is captioned "not for diagnostic use, for prioritization only" which is displayed as a preview function. This preview is meant for informational purposes only, does not contain any marking of the findings, and is not intended for primary diagnosis beyond notification.

    Presenting the users with worklist prioritization facilitates efficient triage by prompting the user to assess the relevant original images in the PACS. Thus, the suspect case receives attention earlier than would have been the case in the standard of care practice alone.

    The algorithm was trained during software development on images of the pathology. As is customary in the field of machine learning, deep learning algorithm development consisted of training on labeled ("tagged") images. In that process, each image in the training dataset was tagged based on the presence of the critical finding.

    AI/ML Overview

    Here's a detailed breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided FDA 510(k) clearance letter for BriefCase-Triage:

    Acceptance Criteria and Reported Device Performance

    ParameterAcceptance Criteria (Performance Goal)Reported Device Performance
    Primary Endpoints
    Sensitivity80%94.2% (95% CI: 89.6%, 97.2%)
    Specificity80%94.6% (95% CI: 90.7%, 97.2%)
    Secondary Endpoint
    Time-to-notification (Subject Device)Comparability with predicate (time savings to standard of care review)10.4 seconds (95% CI: 10.1-10.8)
    Time-to-notification (Predicate Device)(for comparison)264.4 seconds (95% CI: 222-300)

    Note: The document explicitly states that the primary endpoints were "sensitivity and specificity with an 80% performance goal." The reported performance for both sensitivity and specificity (94.2% and 94.6% respectively) significantly exceeds this 80% goal. The time-to-notification for the subject device is significantly faster than the predicate, demonstrating improved "time savings to the standard of care review."

    Study Information

    1. Sample Size Used for the Test Set and Data Provenance:
    * Sample Size: 394 cases
    * Data Provenance:
    * Country of Origin: US (6 clinical sites)
    * Retrospective/Prospective: Retrospective
    * Additional Detail: Cases were distinct in time or center from the training data.

    2. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:
    * Number of Experts: 3
    * Qualifications: Senior board-certified radiologists

    3. Adjudication Method for the Test Set:
    * The document states "as determined by three senior board-certified radiologists." While it doesn't explicitly state "2+1" or "3+1," this implies a consensus-based approach among the three experts. Without further detail, it's reasonable to infer a consensus was reached, or a specific rule for disagreement (e.g., majority) was applied.

    4. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
    * No, a multi-reader multi-case (MRMC) comparative effectiveness study was not conducted to assess how much human readers improve with AI vs. without AI assistance. The study focuses purely on the standalone performance of the AI algorithm.

    5. Standalone Performance Study (Algorithm Only):
    * Yes, a standalone study was performed. The "Pivotal Study Summary" describes evaluating "the software's performance to the ground truth," indicating a standalone performance assessment of the algorithm without human-in-the-loop performance measurement.

    6. Type of Ground Truth Used:
    * Expert consensus (as determined by three senior board-certified radiologists).

    7. Sample Size for the Training Set:
    * The document states, "The algorithm was trained during software development on images of the pathology." However, it does not provide a specific sample size for the training set.

    8. How the Ground Truth for the Training Set Was Established:
    * "each image in the training dataset was tagged based on the presence of the critical finding." This indicates that human experts (or a similar method to the test set ground truth) labeled the images in the training set for the presence of the pathology. However, the specific number and qualifications of these experts are not explicitly stated for the training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K251406

    Validate with FDA (Live)

    Device Name
    BriefCase-Triage
    Date Cleared
    2025-05-30

    (24 days)

    Product Code
    Regulation Number
    892.2080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BriefCase-Triage is a radiological computer aided triage and notification software indicated for use in the analysis of CT chest, abdomen, or chest/abdomen exams with contrast (CTA and CT with contrast) in adults or transitional adolescents aged 18 and older. The device is intended to assist hospital networks and appropriately trained medical specialists in workflow triage by flagging and communication of suspected positive findings of Aortic Dissection (AD) pathology.

    BriefCase-Triage uses an artificial intelligence algorithm to analyze images and highlight cases with detected findings on a standalone desktop application in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected findings. Notifications include compressed preview images that are meant for informational purposes only and not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device.

    The results of BriefCase-Triage are intended to be used in conjunction with other patient information and based on their professional judgment, to assist with triage/ prioritization.

    Device Description

    Briefcase-Triage is a radiological computer-assisted triage and notification software device. The software is based on an algorithm programmed component and is intended to run on a linux-based server in a cloud environment.

    The Briefcase-Triage receives filtered DICOM Images, and processes them chronologically by running the algorithms on each series to detect suspected cases. Following the AI processing, the output of the algorithm analysis is transferred to an image review software (desktop application). When a suspected case is detected, the user receives a pop-up notification and is presented with a compressed, low-quality, grayscale image that is captioned "not for diagnostic use, for prioritization only" which is displayed as a preview function. This preview is meant for informational purposes only, does not contain any marking of the findings, and is not intended for primary diagnosis beyond notification.

    Presenting the users with worklist prioritization facilitates efficient triage by prompting the user to assess the relevant original images in the PACS. Thus, the suspect case receives attention earlier than would have been the case in the standard of care practice alone.

    AI/ML Overview

    Here's a detailed breakdown of the acceptance criteria and study findings for BriefCase-Triage, based on the provided FDA 510(k) clearance letter:


    1. Table of Acceptance Criteria and Reported Device Performance

    ParameterAcceptance CriteriaReported Device Performance
    Primary EndpointsA lower bound 95% Confidence Interval (CI) of 80% for Sensitivity and Specificity at the default operating point.Default Operating Point: - Sensitivity: 92.7% (95% CI: 88.2%, 95.8%). The lower bound (88.2%) is > 80%. - Specificity: 92.8% (95% CI: 89.2%, 95.4%). The lower bound (89.2%) is > 80%. Additional Operating Points (AOPs) meeting criteria: - AOP1: Sensitivity 95.6% (95% CI: 91.8%-98.0%), Specificity 88.2% (95% CI: 84.0%-91.6%) - AOP2: Sensitivity 94.1% (95% CI: 90.0%-96.9%), Specificity 89.8% (95% CI: 85.8%-93.0%) - AOP3: Sensitivity 89.3% (95% CI: 84.2%-93.2%), Specificity 94.7% (95% CI: 91.6%-97.0%) - AOP4: Sensitivity 86.3% (95% CI: 80.9%-90.7%), Specificity 97.7% (95% CI: 95.3%-99.1%)
    Secondary Endpoints (Comparability with Predicate)Time-to-notification metric for the Briefcase-Triage software should demonstrate comparability with the predicate device.Briefcase-Triage (Subject Device): Mean time-to-notification = 10.7 seconds (95% CI: 10.5-10.9) Predicate AD: Mean time-to-notification = 38.0 seconds (95% CI: 35.5-40.4) The subject device's time-to-notification is faster than the predicate, demonstrating comparability and improvement in time savings.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: 509 cases.
    • Data Provenance:
      • Country of origin: 5 US-based clinical sites.
      • Retrospective or Prospective: Retrospective.
      • Data Sequestration: Cases collected for the pivotal dataset were "all distinct in time or center from the cases used to train the algorithm," and "Test pivotal study data was sequestered from algorithm development activities."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • Number of Experts: Three (3) senior board-certified radiologists.
    • Qualifications: "Senior board-certified radiologists." (Specific years of experience are not provided.)

    4. Adjudication Method for the Test Set

    • The text states "the ground truth, as determined by three senior board-certified radiologists." This implies a consensus-based adjudication, likely 3-0 or 2-1 (majority vote), but the exact method (e.g., 2+1, 3+1) is not explicitly detailed.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • Was it done? No.
    • The study primarily focused on the standalone performance of the AI algorithm compared to ground truth and a secondary comparison of time-to-notification with a predicate device. It did not evaluate human reader performance with and without AI assistance.

    6. Standalone Performance Study

    • Was it done? Yes.
    • The study evaluated the algorithm's performance (sensitivity, specificity, PPV, NPV, PLR, NLR) in identifying AD pathology without human intervention as a primary and secondary endpoint. The device's output is "flagging and communication of suspected positive findings" and "notifications include compressed preview images that are meant for informational purposes only and not intended for diagnostic use beyond notification," confirming a standalone function.

    7. Type of Ground Truth Used

    • Ground Truth: Expert Consensus, specifically "as determined by three senior board-certified radiologists."

    8. Sample Size for the Training Set

    • The document states, "The algorithm was trained during software development on images of the pathology." However, it does not specify the sample size for the training set. It only mentions that the pivotal test data was "distinct in time or center" from the training data.

    9. How the Ground Truth for the Training Set Was Established

    • "As is customary in the field of machine learning, deep learning algorithm development consisted of training on labeled ("tagged") images. In that process, each image in the training dataset was tagged based on the presence of the critical finding."
    • While it indicates images were "labeled ("tagged")" based on the "presence of the critical finding," it does not explicitly state who established this ground truth for the training set (e.g., experts, pathology, etc.). It's implied that medical professionals were involved in the labeling process, but no specific number or qualification is provided for the training set.
    Ask a Question

    Ask a specific question about this device

    K Number
    K250248

    Validate with FDA (Live)

    Device Name
    BriefCase-Triage
    Date Cleared
    2025-02-14

    (18 days)

    Product Code
    Regulation Number
    892.2080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BriefCase-Triage is a radiological computer aided triage and notification software indicated for use in the analysis of contrast-enhanced images that include the lungs in adults or transitional adolescents age 18 and older. The device is intended to assist hospital networks and appropriately trained medical specialists in workflow triage by flagging and communication of suspect cases of incidental Pulmonary Embolism (iPE) pathologies.

    BriefCase-Triage uses an artificial intelligence algorithm to analyze images and highlight cases with detected findings in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for suspect cases. Notifications include compressed preview images that are meant for informational purposes only and not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device.

    The results of BriefCase-Triage are intended to be used in conjunction with other patient information and based on their professional judgment, to assist with triage/prioritization of medical images. Notified clinicians are responsible for viewing full images per the standard of care.

    Device Description

    Briefcase-Triage is a radiological computer-assisted triage and notification software device. The software is based on an algorithm programmed component and is intended to run on a linux-based server in a cloud environment.

    The Briefcase-Triage receives filtered DICOM Images, and processes them chronologically by running the algorithms on each series to detect suspected cases. Following the AI processing, the output of the algorithm analysis is transferred to an image review software (desktop application). When a suspected case is detected, the user receives a pop-up notification and is presented with a compressed, low-quality, grayscale image that is captioned "not for diagnostic use, for prioritization only" which is displayed as a preview function. This preview is meant for informational purposes only, does not contain any marking of the findings, and is not intended for primary diagnosis beyond notification.

    Presenting the users with worklist prioritization facilitates efficient triage by prompting the user to assess the relevant original images in the PACS. Thus, the suspect case receives attention earlier than would have been the case in the standard of care practice alone.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study details for the BriefCase-Triage device, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    MetricAcceptance Criteria (Performance Goal)Reported Device Performance (Default Operating Point)95% Confidence Interval
    Sensitivity≥ 80%91.7%[87.5%, 94.9%]
    Specificity≥ 80%91.4%[87.3%, 94.6%]
    NPVNot specified99.8%[99.6%, 99.8%]
    PPVNot specified22.2%[16.1%, 29.9%]
    PLRNot specified10.7[7.2, 16.0]
    NLRNot specified0.09[0.06, 0.14]
    Time-to-NotificationComparability to predicate (282.63s)40.2 seconds[36.9, 43.5] (for subject device)

    Note on Additional Operating Points (AOPs): The document also describes four additional operating points (AOPs 1-4) designed to maximize specificity while maintaining a lower bound 95% CI of 80% for sensitivity. These AOPs achieved similar performance metrics, further demonstrating the device's flexibility.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: 498 cases
    • Data Provenance: Retrospective, blinded, multicenter study. Cases were collected from 6 US-based clinical sites. The test data was distinct in time or center from the cases used for algorithm training and was sequestered from algorithm development activities.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    • Number of Experts: Three (3)
    • Qualifications of Experts: Senior board-certified radiologists.

    4. Adjudication Method for the Test Set

    The specific adjudication method (e.g., 2+1, 3+1) is not explicitly stated in the provided text. It only mentions that the ground truth was "determined by three senior board-certified radiologists."

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    A Multi-Reader Multi-Case (MRMC) comparative effectiveness study comparing AI assistance versus without AI assistance for human readers was not described for this device. The study primarily evaluated the standalone performance of the AI algorithm and its time-to-notification compared to the predicate device, not the improvement in human reader performance with AI assistance.

    6. If a Standalone (Algorithm Only Without Human-in-the-loop Performance) was done

    Yes, a standalone performance study was done. The "Pivotal Study Summary" sections describe the device's sensitivity, specificity, and other metrics based on its algorithm's performance in identifying iPE cases compared to the ground truth established by radiologists. The device is intended to provide notifications and not for diagnostic use, suggesting its performance is evaluated in its autonomous triage function.

    7. The Type of Ground Truth Used

    The ground truth used was expert consensus, specifically "determined by three senior board-certified radiologists."

    8. The Sample Size for the Training Set

    The sample size for the training set is not explicitly stated in the provided text. It only mentions that the algorithm was "trained during software development on images of the pathology."

    9. How the Ground Truth for the Training Set Was Established

    The ground truth for the training set was established through manual labeling ("tagging") by experts. The text states: "deep learning algorithm development consisted of training on manually labeled ("tagged") images. In that process, critical findings were tagged in all CTs in the training data set."

    Ask a Question

    Ask a specific question about this device

    K Number
    K243548

    Validate with FDA (Live)

    Device Name
    BriefCase-Triage
    Date Cleared
    2024-12-11

    (26 days)

    Product Code
    Regulation Number
    892.2080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BriefCase-Triage is a radiological computer aided triage and notification software indicated for use in the analysis of CT images with or without contrast that include the ribs, in adults or transitional adolescents aged 18 and older. The device is intended to assist hospital networks and appropriately trained medical specialists in workflow triage by flagging and communicating suspect cases of three or more acute Rib fracture (RibFx) pathologies.

    BriefCase-Triage uses an artificial intelligence algorithm to analyze images and highlight cases with detected findings in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected RibFx findings. Notifications include compressed preview images that are meant for informational purposes only, and not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device.

    The results of BriefCase-Triage are intended to be used in conjunction with other patient information and based on their professional judgment, to assist with triage/prioritization of medical images. Notified clinicians are responsible for viewing full images per the standard of care.

    Device Description

    BriefCase-Triage is a radiological computer-assisted triage and notification software device. The software is based on an algorithm programmed component and is intended to run on a linuxbased server in a cloud environment.

    The BriefCase-Triage receives filtered DICOM Images, and processes them chronologically by running the algorithms on each series to detect suspected cases. Following the Al processing, the output of the algorithm analysis is transferred to an image review software (desktop application). When a suspected case is detected, the user receives a pop-up notification and is presented with a compressed, low-quality, grayscale image that is captioned "not for diagnostic use, for prioritization only" which is displayed as a preview function. This preview is meant for informational purposes only, does not contain any marking of the findings, and is not intended for primary diagnosis beyond notification.

    Presenting the users with worklist prioritization facilitates efficient triage by prompting the user to assess the relevant original images in the PACS. Thus, the suspect case receives attention earlier than would have been the case in the standard of care practice alone.

    AI/ML Overview

    Acceptance Criteria and Device Performance for BriefCase-Triage

    1. Table of Acceptance Criteria and Reported Device Performance

    MetricAcceptance Criteria (Performance Goal)Reported Device Performance
    AUC> 0.9597.2% (95% Cl: 95.5%-99.0%)
    Sensitivity> 80%95.2% (95% Cl: 89.1%-98.4%)
    Specificity> 80%95.1% (95% Cl: 91.2%-97.6%)
    Time-to-notification (Mean)Comparability with predicate (70.1 seconds)41.4 seconds (95% Cl: 40.4-42.5)

    Note: The acceptance criteria for sensitivity and specificity are extrapolated from the statement "As the AUC exceeded 0.95 and sensitivity and specificity both exceeded 80%, the study's primary endpoints were met."

    2. Sample Size and Data Provenance for Test Set

    • Sample Size for Test Set: 308 cases
    • Data Provenance: Retrospective, multicenter study from 5 US-based clinical sites. The cases collected for the pivotal dataset were distinct in time or center from the cases used to train the algorithm.

    3. Number and Qualifications of Experts for Ground Truth

    • Number of Experts: Three senior board-certified radiologists.
    • Qualifications: Senior board-certified radiologists. (Specific years of experience are not provided in the document).

    4. Adjudication Method for Test Set

    The adjudication method is not explicitly stated. The document mentions "ground truth, as determined by three senior board-certified radiologists," but does not detail how disagreements among these radiologists were resolved (e.g., 2+1, 3+1, or simple consensus).

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study was done to assess the effect of AI assistance on human readers' performance. The study focused on the standalone performance of the AI algorithm and its time-to-notification compared to a predicate device.

    6. Standalone Performance Study

    Yes, a standalone performance study was done. The "Pivotal Study Summary" section explicitly details the evaluation of the software's performance (AUC, sensitivity, specificity, PPV, NPV, PLR, NLR) in identifying RibFx without human intervention, comparing it to the established ground truth.

    7. Type of Ground Truth Used

    The ground truth used was expert consensus, determined by three senior board-certified radiologists.

    8. Sample Size for Training Set

    The sample size for the training set is not explicitly provided in the document. The document states, "The algorithm was trained during software development on images of the pathology. As is customary in the field of machine learning, deep learning algorithm development consisted of training on labeled ("tagged") images." It also mentions, "The cases collected for the pivotal dataset were all distinct in time or center from the cases used to train the algorithm, as was used for the most recent clearance (K230020)."

    9. How Ground Truth for Training Set Was Established

    The ground truth for the training set was established by labeling ("tagging") images based on the presence of the critical finding (three or more acute Rib fractures). This process is described as "each image in the training dataset was tagged based on the presence of the critical finding." The document does not specify who performed this tagging or the exact methodology for establishing the ground truth for the training set (e.g., expert consensus, pathology).

    Ask a Question

    Ask a specific question about this device

    K Number
    K242837

    Validate with FDA (Live)

    Device Name
    BriefCase-Triage
    Date Cleared
    2024-10-18

    (29 days)

    Product Code
    Regulation Number
    892.2080
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BriefCase-Triage is a radiological computer aided triage and notification software indicated for use in the analysis of CT scans that include the cervical spine, in adults or transitional adolescents aged 18 and older. The device is intended to assist hospital networks and appropriately trained medical specialists in workflow triage by flagging and communication of linear lucencies in the cervical spine bone in patterns compatible with fractures.

    BriefCase-Triage uses an artificial intelligence algorithm to analyze images and highlight cases with detected findings in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected findings. Notifications include compressed preview images that are meant for informational purposes only and not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device.

    The results of BriefCase-Triage are intended to be used in conjunction with other patient information and based on their professional judgment, to assist with triage/prioritization of medical images. Notified clinicians are responsible for viewing full images per the standard of care.

    Device Description

    BriefCase-Triage is a radiological computer-assisted triage and notification software device.

    The software is based on an algorithm programmed component and is intended to run on a linuxbased server in a cloud environment.

    The BriefCase-Triage receives filtered DICOM Images, and processes them chronologically by running the algorithms on each series to detect suspected cases. Following the Al processing, the output of the algorithm analysis is transferred to an image review software (desktop application). When a suspected case is detected, the user receives a pop-up notification and is presented with a compressed, low-quality, grayscale image that is captioned "not for diagnostic use, for prioritization only" which is displayed as a preview function. This preview is meant for informational purposes only, does not contain any marking of the findings, and is not intended for primary diagnosis beyond notification.

    Presenting the users with worklist prioritization facilitates efficient triage by prompting the user to assess the relevant original images in the PACS. Thus, the suspect case receives attention earlier than would have been the case in the standard of care practice alone.

    The algorithm was trained during software development on images of the pathology. As is customary in the field of machine learning algorithm development consisted of training on manually labeled ("tagged") images. In that process, critical findings were tagged in all CTs in the training data set.

    AI/ML Overview

    The Aidoc BriefCase-Triage device, intended for triaging cervical spine CT scans for fractures, underwent a retrospective, blinded, multicenter study to evaluate its performance.

    Here's a breakdown of the acceptance criteria and study details:

    1. A table of acceptance criteria and the reported device performance

    Acceptance CriteriaReported Device Performance
    Sensitivity (performance goal ≥ 80%)92.1% (95% CI: 87.5%, 95.4%)
    Specificity (performance goal ≥ 80%)92.6% (95% CI: 89.0%, 95.4%)
    Time-to-Notification (comparable to predicate)15.1 seconds (95% CI: 14.1-16.2)

    2. Sample size used for the test set and the data provenance

    • Sample Size: 487 cases
    • Data Provenance: Retrospective, multicenter study from 6 US-based clinical sites. The cases collected for the pivotal dataset were all distinct in time or center from the cases used to train the algorithm.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Number of Experts: Three
    • Qualifications: Senior board-certified radiologists

    4. Adjudication method for the test set

    The document does not explicitly state the adjudication method (e.g., 2+1, 3+1). However, it implies that the ground truth was "as determined by three senior board-certified radiologists," suggesting a consensus-based approach among these experts.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    There is no MRMC comparative effectiveness study presented in this document that evaluates human readers' improvement with AI assistance versus without AI assistance. The study focuses on the standalone performance of the AI device and its time-to-notification compared to a predicate device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, a standalone algorithm-only performance study was done. The primary endpoints (sensitivity and specificity) and secondary endpoints (time-to-notification, PPV, NPV, PLR, NLR) directly measure the performance of the BriefCase-Triage software itself.

    7. The type of ground truth used

    The ground truth was established by expert consensus (determined by three senior board-certified radiologists).

    8. The sample size for the training set

    The document states, "The algorithm was trained during software development on images of the pathology." However, it does not specify the sample size for the training set. It only mentions that the test pivotal study data was sequestered from algorithm development activities.

    9. How the ground truth for the training set was established

    The ground truth for the training set was established by manually labeled ("tagged") images. The document states, "critical findings were tagged in all CTs in the training data set." This implies expert annotation of the training data.

    Ask a Question

    Ask a specific question about this device

    K Number
    K241727

    Validate with FDA (Live)

    Device Name
    BriefCase-Triage
    Date Cleared
    2024-07-12

    (28 days)

    Product Code
    Regulation Number
    892.2080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Briefcase-Triage is a radiological computer- aided triage and notification software indicated for use in the analysis of CTPA images, in adults or transitional adolescents aged 18 and older. The device is intended to assist hospital networks and appropriately trained medical specialists in workflow triage by flagging and communicating suspected positive cases of Pulmonary Embolism (PE) pathologies.

    Briefcase-Triage uses an artificial intelligence algorithm to analyze images and highlight cases with detected findings in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected PE findings. Notifications include compressed preview images that are meant for informational purposes only and not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device.

    The results of Briefcase-Triage are intended to be used in conjunction with other patient information and based on their professional judgment, to assist with triage/prioritization of medical images. Notified clinicians are responsible for viewing full images per the standard of care.

    Device Description

    Briefcase-Triage is a radiological computer-assisted triage and notification software device. The software is based on an algorithm programmed component and is intended to run on a linux-based server in a cloud environment.

    The Briefcase-Triage receives filtered DICOM Images, and processes them chronologically by running the algorithms on each series to detect suspected cases. Following the Al processing, the output of the algorithm analysis is transferred to an image review software (desktop application). When a suspected case is detected, the user receives a pop-up notification and is presented with a compressed, low-quality, grayscale image that is captioned "not for diagnostic use, for prioritization only" which is displayed as a preview function. This preview is meant for informational purposes only, does not contain any marking of the findings, and is not intended for primary diagnosis beyond notification.

    Presenting the users with worklist prioritization facilitates efficient triage by prompting the user to assess the relevant original images in the PACS. Thus, the suspect case receives attention earlier than would have been the case in the standard of care practice alone.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study details for the Aidoc BriefCase-Triage device, based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    Device Name: BriefCase-Triage (for Pulmonary Embolism - PE)

    Acceptance CriteriaPerformance GoalReported Device Performance (Default Operating Point)
    Sensitivity≥ 80%94.39% (95% CI: 90.41%, 97.07%)
    Specificity≥ 80%94.39% (95% CI: 91.04%, 96.67%)
    Time-to-notification (compared to predicate)Comparable benefit in time savingMean 26.42 seconds (95% CI: 25.3-27.54) vs Predicate's 78.0 seconds (95% CI: 73.6-82.3)

    Additional Operating Points (AOPs) Performance:

    Operating PointSensitivity (95% CI)Specificity (95% CI)
    AOP199.53% (97.42%-99.99%)86.67% (82.16%-90.39%)
    AOP297.66% (94.63%-99.24%)91.93% (88.14%-94.82%)
    AOP391.59% (87.03%-94.94%)96.49% (93.64%-98.3%)
    AOP485.98% (80.6%-90.34%)98.25% (95.95%-99.43%)

    Other reported secondary endpoints for the default operating point:

    • NPV: 98.96% (95% CI: 98.21%- 99.4%)
    • PPV: 74.79% (95% CI: 64.80%- 82.7%)
    • PLR: 16.81 (95% CI: 10.43- 27.09)
    • NLR: 0.059 (95% CI: 0.034- 0.103)

    2. Sample Size and Data Provenance for the Test Set

    • Sample Size: 499 cases
    • Data Provenance: Retrospective, multicenter study from 6 US-based clinical sites. The cases were distinct in time or center from the cases used to train the algorithm.

    3. Number of Experts and Qualifications for Ground Truth

    • Number of Experts: Three (3)
    • Qualifications: Senior board-certified radiologists.

    4. Adjudication Method

    • The document states "the ground truth as determined by three senior board-certified radiologists." It does not explicitly state the adjudication method (e.g., 2+1, consensus, majority vote). However, "determined by" implies that their collective judgment established the ground truth for each case.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not explicitly described. The study focuses on the standalone performance of the AI algorithm against a ground truth established by experts, and a comparison of its notification time with a predicate device. It does not evaluate human reader performance with and without AI assistance.

    6. Standalone Performance Study

    • Yes, a standalone study (algorithm only without human-in-the-loop performance) was conducted. The primary endpoints (sensitivity and specificity) evaluated the device's ability to identify PE cases independently.

    7. Type of Ground Truth Used

    • Expert Consensus: The ground truth was "determined by three senior board-certified radiologists"

    8. Sample Size for the Training Set

    • The document states that the algorithm was "trained during software development on images of the pathology" and that the subject device was trained on a "larger data set" compared to the predicate. However, it does not specify the exact sample size of the training set.

    9. How Ground Truth for the Training Set was Established

    • "critical findings were tagged in all CTs in the training data set." This implies manual labeling (annotation) of findings by experts on the training images. While not explicitly stated, it is common practice that these tags are done by medical professionals.
    Ask a Question

    Ask a specific question about this device

    K Number
    K232751

    Validate with FDA (Live)

    Device Name
    BriefCase-Triage
    Date Cleared
    2023-10-30

    (52 days)

    Product Code
    Regulation Number
    892.2080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BriefCase-Triage is a radiological computer-aided triaqe and notification software indicated for use in the analysis of CTPA images in adults or transitional adolescents aged 18 and older. The device is intended to assist hospital networks and appropriately trained medical specialists in workflow triage by flagging and communicating suspected positive cases of Central Pulmonary Embolism (Central PE).

    BriefCase-Triage uses an artificial intelligence algorithm to analyze images and highlight cases with detected findings in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected Central PE findings. Notifications include compressed preview images that are meant for informational purposes only and not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device.

    The results of BriefCase-Triage are intended to be used in conjunction with other patient information and based on their professional judgment, to assist with triage/prioritization of medical images. Notified clinicians are responsible for viewing full images per the standard of care.

    Device Description

    BriefCase-Triage is a radiological computer-assisted triage and notification software device. The software is based on an algorithm programmed component and is intended to run on a linux-based server in a cloud environment.

    The BriefCase-Triage receives filtered DICOM images, and processes them chronologically by running the algorithms on each series to detect suspected cases. Following the Al processing, the output of the algorithm analysis is transferred to an image review software (desktop application). When a suspected case is detected, the user receives a pop-up notification and is presented with a compressed, low-quality, grayscale image that is captioned "not for diagnostic use, for prioritization only" which is displayed as a preview function. This preview is meant for informational purposes only, does not contain any marking of the findings, and is not intended for primary diagnosis beyond notification.

    Presenting the users with worklist prioritization facilitates efficient triage by prompting the user to assess the relevant original images in the PACS. Thus, the suspect case receives attention earlier than would have been the case in the standard of care practice alone.

    The algorithm was trained during software development on images of the pathology. As is customary in the field of machine learning, deep learning algorithm development consisted of training on manually labeled ("tagged") images. In that process, critical findings were tagged in all CTs in the training data set.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study details for the Aidoc BriefCase-Triage device, based on the provided FDA 510(k) summary:

    1. Table of Acceptance Criteria and Reported Device Performance

    The FDA 510(k) summary explicitly states the primary performance goals and the achieved results.

    Acceptance Criteria (Performance Goal)Reported Device Performance
    Sensitivity ≥ 80%89.2% (95% CI: 82.5%, 93.9%)
    Specificity ≥ 80%94.5% (95% CI: 90.3%, 97.2%)

    Note on Secondary Endpoints: While time-to-notification, PPV, NPV, PLR, and NLR were assessed as secondary endpoints, the document does not state explicit acceptance criteria for them, but rather presents them as comparative data or additional performance metrics.

    2. Sample Size and Data Provenance

    • Test Set Sample Size: 328 cases from unique patients.
    • Data Provenance: Retrospective, multi-center study with data from 6 US-based clinical sites. The cases collected for the pivotal dataset were distinct in time or center from the cases used to train the algorithm.

    3. Number of Experts and Qualifications for Ground Truth

    • Number of Experts: 3 senior board-certified radiologists.
    • Qualifications of Experts: Senior board-certified radiologists. (No specific years of experience are detailed, but "senior" implies extensive experience).

    4. Adjudication Method for the Test Set

    • Adjudication Method: Majority voting among the three senior board-certified radiologists was used to establish the ground truth.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • Was an MRMC study done? No, a comparative effectiveness study involving human readers with and without AI assistance (MRMC) was not performed as the primary evaluation for this device. The study compared the algorithm's performance to ground truth, and the device is intended for workflow triage/notification, not as a diagnostic tool replacing human interpretation. The time-to-notification comparison was done between the device and a predicate device (PETN), not human readers.

    6. Standalone (Algorithm Only) Performance Study

    • Was a standalone study done? Yes. The pivotal study directly evaluated the BriefCase-Triage software's performance (sensitivity and specificity) in identifying Central PE by comparing its output against the established ground truth. This is a standalone performance evaluation.

    7. Type of Ground Truth Used

    • Type of Ground Truth: Expert consensus (majority voting by three senior board-certified radiologists).

    8. Sample Size for the Training Set

    • The document states that the algorithm was "trained during software development on images of the pathology." However, the exact sample size for the training set is not specified in the provided text.

    9. How Ground Truth for the Training Set Was Established

    • Method: "Critical findings were tagged in all CTs in the training data set." This implies manual labeling/tagging of findings by experts. The specific number or qualifications of these "tagging" experts are not detailed, but it's consistent with a machine learning development process.
    Ask a Question