    Intended Use

    BriefCase-Quantification is a radiological image management and processing system software indicated for use in the analysis of contrast-enhanced CT exams that include the aorta in adults or transitional adolescents aged 18 and older.

    BriefCase-Quantification of Aortic Measurement (M-Aorta) is intended to assist hospital networks and appropriately trained medical specialists by providing the user with aortic diameter measurements across the aorta. BriefCase-Quantification is indicated to evaluate normal and aneurysmal aortas and is not intended to evaluate post-operative aortas.

    The device provides the following assessments:

    • Aortic measurements at 10 anatomical landmarks;
    • Maximum aortic diameter of the abdominal aorta, descending aorta and ascending aorta.

    The BriefCase-Quantification results are not intended to be used on a stand-alone basis for clinical decision-making or otherwise preclude clinical assessment of cases. These measurements are unofficial, are not final, and are subject to change after review by a radiologist. For final clinically approved measurements, please refer to the official radiology report. Clinicians are responsible for viewing full images per the standard of care.

    Device Description

    BriefCase-Quantification is a radiological medical image management and processing device. The software consists of a single module based on an algorithm programmed component and is intended to run on a linux-based server in a cloud environment.

    The BriefCase-Quantification receives filtered DICOM Images, and processes them chronologically by running the algorithm on relevant series to measure the maximum abdominal aortic diameter. Following the AI processing, the output of the algorithm analysis is transferred to an image review software (desktop application), and forwarded to user review in the PACS.

    The BriefCase-Quantification produces images and tabular views of the landmarks and segments selected per installation to be displayed in the image review software, provided by the Image Communication Platform integrated with Briefcase-Quantification. The diameter marking is not intended to be a final output, but serves the purpose of visualization and measurement. The original, unmarked series remains available in the PACS as well. The preview image presents an unofficial and not final measurement, and the user is instructed to review the full image and any other clinical information before making a clinical decision. The image includes a disclaimer: "Not for diagnostic use. The measurement is unofficial, not final, and must be reviewed by a radiologist."

    BriefCase-Quantification is not intended to evaluate post-operative aortas.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study proving device performance, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Metric / Acceptance CriteriaReported Device Performance
    Primary Endpoint: Mean Absolute Error (MAE) between ground truth measurement and algorithm across all landmarks and studies.1.88 mm (95% CI: 1.78 mm, 1.99 mm). (This was "below the prespecified performance goal," thus achieving the primary endpoint. The reported MAE of the subject device (1.88 mm) was comparable to the predicate device [1.95 mm (95% CI: 1.59 mm, 2.32 mm)].)
    Secondary Endpoint: Bias between ground truth and algorithm output (Mean Difference in Bland-Altman analysis).0.1 mm (This indicates "little to no bias between the two measurements," demonstrating the study's secondary endpoint was achieved.)

    2. Sample Size and Data Provenance

    • Test Set Sample Size: 212 cases
    • Data Provenance: Retrospective, multicenter study. Cases were collected from 6 US-based clinical sites, including both academic and community centers. The cases were distinct in time and/or center from those used to train the algorithm.

    3. Number of Experts and Qualifications for Ground Truth

    • Number of Experts: Three (3)
    • Qualifications: US board-certified radiologists.

    4. Adjudication Method for the Test Set

    The document states the ground truth was "determined by three US board-certified radiologists." While it doesn't explicitly detail a 2+1 or 3+1 method, the implication of "determined by" three experts suggests a consensus-based approach, likely one where all three agreed or a majority agreement was required for the final ground truth. It does not mention any specific adjudication process explicitly (e.g., if disagreements were resolved by a fourth reader).

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • Was an MRMC study done? No. The study described is a standalone (algorithm-only) performance evaluation against expert-established ground truth. It does not compare human readers with and without AI assistance.
    • Effect size of improvement: Not applicable, as no MRMC study was performed.

    6. Standalone (Algorithm-Only) Performance

    • Was a standalone performance study done? Yes. The "Pivotal Study Summary" describes the evaluation of "the software's performance ... compared to the ground truth," focusing on the algorithm's accuracy in measurement against expert consensus.

    7. Type of Ground Truth Used

    • Type of Ground Truth: Expert consensus. Specifically, the ground truth was "determined by three US board-certified radiologists."

    8. Sample Size for the Training Set

    The document explicitly states: "The cases collected for the pivotal dataset were all distinct in time and/or center from the cases used to train the algorithm." However, the exact sample size for the training set is not provided in the given text.

    9. How the Ground Truth for the Training Set was Established

    The document does not detail how the ground truth for the training set was established. It only mentions that the training cases were distinct from the test set cases.

    Ask a Question

    Ask a specific question about this device

    K Number
    K241112

    Validate with FDA (Live)

    Date Cleared
    2024-05-15

    (23 days)

    Product Code
    Regulation Number
    892.2050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BriefCase-Quantification is a radiological image management and processing system software indicated for use in the analysis of CT exams with contrast, that include the abdominal aorta, in adults or transitional adolescents aged 18 and older.

    The device is intended to assist appropriately trained medical specialists by providing the user with the maximum abdominal aortic diameter measurement of cases that include the abdominal aorta (M-AbdAo). BriefCase-Quantification is indicated to evaluate normal and aneurysmal abdominal aortas and is not intended to evaluate post-operative aortas.

    The BriefCase-Quantification results are not intended to be used on a stand-alone basis for clinical decision-making or otherwise preclude clinical assessment of cases. These measurements are unofficial, are not final, and are subject to change after review by a radiologist. For final clinically approved measurements, please refer to the official radiology report. Clinicians are responsible for viewing full images per the standard of care.

    Device Description

    BriefCase-Quantification is a radiological medical image management and processing device. The software consists of a single module based on an algorithm programmed component and is intended to run on a linux-based server in a cloud environment.

    The BriefCase-Quantification receives filtered DICOM Images, and processes them chronologically by running the algorithm on relevant series to measure the maximum abdominal aortic diameter. Following the AI processing, the output of the algorithm analysis is transferred to an image review software (desktop application), and forwarded to user review in the PACS.

    BriefCase-Quantification produces a preview image annotated with the maximum diameter measurement. The diameter marking is not intended to be a final output, but serves the purpose of visualization and measurement. The original, unmarked series remains available in the PACS as well. The preview image presents an unofficial measurement which is not final, and the user is instructed to review the full image and any other clinical information before making a clinical decision. The image includes a disclaimer: "Not for diagnostic use. The measurement is unofficial, not final, and must be reviewed by a radiologist."

    BriefCase-Quantification is not intended to evaluate post-operative aortas.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    Acceptance Criteria and Device Performance

    Acceptance Criteria (Performance Goal)Reported Device Performance
    Mean Absolute Error (MAE) between ground truth and algorithm measurement below a prespecified goal.1.52 mm (95% Cl: 1.20 mm, 1.83 mm)
    Little to no bias between ground truth and algorithm output.Mean difference of 0.58 mm

    Study Details

    1. A table of acceptance criteria and the reported device performance:
    See table above.

    2. Sample size used for the test set and the data provenance:

    • Sample Size (Test Set): 162 cases
    • Data Provenance: Retrospective, from 6 US-based clinical sites (both academic and community centers). The cases were distinct in time and/or center from the training data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Number of Experts: Three (3)
    • Qualifications: US board-certified radiologists

    4. Adjudication method for the test set:
    The text states the ground truth was "as determined by three US board-certified radiologists." It doesn't explicitly specify an adjudication method like 2+1 or 3+1, but implies a consensus or agreement among the three experts formed the ground truth.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
    No, a multi-reader multi-case (MRMC) comparative effectiveness study of human readers with and without AI assistance was not conducted. The study focused on the standalone performance of the algorithm against a human-established ground truth.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
    Yes, a standalone study of the algorithm's performance was done. The study evaluated the BriefCase-Quantification software's performance in providing maximum diameter measurements compared to a ground truth established by radiologists.

    7. The type of ground truth used:
    Expert consensus among three US board-certified radiologists.

    8. The sample size for the training set:
    The sample size for the training set is not explicitly stated. The text only mentions that the subject device's improved performance is "due to its training on a larger data set" compared to the predicate device, but does not provide a specific number for this larger data set. It also states the test set cases were "distinct in time and/or center from the cases used to train the algorithm."

    9. How the ground truth for the training set was established:
    The text does not explicitly describe how the ground truth for the training set was established. It only refers to a "larger data set" used for training.

    Ask a Question

    Ask a specific question about this device

    K Number
    K231631

    Validate with FDA (Live)

    Date Cleared
    2023-11-28

    (176 days)

    Product Code
    Regulation Number
    892.1750
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BriefCase-Quantification is a software intended for use in the analysis of non-cardiac-gated non-contrast CT (NCCT) images that include the heart in adult patients aged 30 and older.

    The device is intended to assist physicians by providing the user with a four-category Coronary Artery Calcification (CAC) of plaques, which present a risk for coronary artery disease, together with preview axial images of the detected calcium meant for informational purposes only.

    The BriefCase-Quantification results are not intended to be used on a stand-alone basis for clinical decision-making or otherwise preclude clinical assessment of cases. Clinicians are responsible for viewing full images per the standard of care.

    Device Description

    BriefCase-Quantification is a standalone software as a medical device intended for use in the analysis of non-cardiac-gated non-contrast CT (NCCT) images that include the heart to assist hospital networks and appropriately trained medical specialists. The software consists of a single module based on an algorithm programmed component and is intended to run on a linux-based server in a cloud environment.

    The BriefCase-Quantification receives filtered routine, non-gated computed tomography (CT) scans, and processes them chronologically by running the algorithm on relevant series to evaluate calcified plaques in the coronary arteries. Following the AI processing, the output of the algorithm analysis is transferred to an image review software (the PACS or a desktop application).

    The device generates a four-category output corresponding with the estimated quantity of calcium detected: very low, low, medium, and high. The categories composing the output of the device correspond with a validated visual assessment categorization of none, mild, moderate, and severe [1] in agreement with categorized Agatston scores indicated in the literature (very low: 0; low: 1-100; medium: 101-400; high: ≥400). In addition, the categories accord with the 2016 SCCT/STR guidelines for coronary artery calcium scoring of non-contrast non-cardiac chest CT scans and are used as standard of care in clinical practice during CAC assessment in NCCT scans.

    The BriefCase-Quantification software generates a preliminary summary report that is provided in the desktop application that includes applicable user warnings, the CAC detection category and number of slices that include CAC. The report presents preliminary results only and instructs the user to review the full image and any other clinical information before making a clinical decision. For all analyzed scans, the user will be presented in the PACS with all the slices containing the measured coronary calcifications. On these images, the calcified areas will be represented to provide the user visibility on the areas which supported the category output. These slices will be presented along with the original slices. Preview images of the represented calcium are non-diagnostic and are available in the PACS for informational purposes only.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and the study for the BriefCase-Quantification device, based on the provided text:


    1. Table of Acceptance Criteria and Reported Device Performance

    CategoryAcceptance Criteria (Performance Goal)Reported Device Performance
    Overall Agreement (Primary Endpoint)"up to the prespecified performance goal"87.1%
    Agreement for 'Very Low' CAC (Secondary Endpoint)"up to the prespecified performance goal"95.1%
    Agreement for 'Low' CAC (Secondary Endpoint)"up to the prespecified performance goal"81.3%
    Agreement for 'Medium' CAC (Secondary Endpoint)"up to the prespecified performance goal"81.5%
    Agreement for 'High' CAC (Secondary Endpoint)"up to the prespecified performance goal"89.3%

    Note: The document states that the primary and secondary endpoints were achieved because the reported performance was "up to the prespecified performance goal," but it does not explicitly state the numerical value of the prespecified performance goal for each criterion.


    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: 433 cases
    • Data Provenance: Retrospective, multicenter study from 6 US-based clinical sites (both academic and community centers). The cases were distinct in time or center from the cases used to train the algorithm.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    • Number of Experts: At least one, potentially more. The text states "In cases where the reviewers disagree on the level of CAC, the senior US board-certified radiologist provided a final opinion which has established the ground truth." This implies primary reviewers and a senior radiologist for adjudication.
    • Qualifications of Experts: Senior US board-certified radiologist (for final opinion/adjudication). The qualifications of initial reviewers are not specified.

    4. Adjudication Method for the Test Set

    • Adjudication Method: 2+1 (or similar) where a "senior US board-certified radiologist provided a final opinion" in cases of disagreement among initial reviewers.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and its Effect Size

    • A standalone performance study of the algorithm against ground truth was performed.
    • The document does not mention a multi-reader multi-case (MRMC) comparative effectiveness study to assess how much human readers improve with AI vs without AI assistance. The study focuses on the algorithm's agreement with expert-established ground truth.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • Yes, a standalone study was done. The pivotal study evaluated "the software's performance in providing estimated coronary artery calcification detection category ... compared to the ground truth." The performance metrics (overall agreement and agreement per category) are for the algorithm's output directly compared to ground truth.

    7. The Type of Ground Truth Used

    • Type of Ground Truth: Expert consensus/adjudication. The text states, "In cases where the reviewers disagree on the level of CAC, the senior US board-certified radiologist provided a final opinion which has established the ground truth." The ground truth categories ([very low, low, medium, high] corresponding to Agatston scores) are also aligned with "validated visual assessment categorization of none, mild, moderate, and severe" and "2016 SCCT/STR guidelines."

    8. The Sample Size for the Training Set

    • The sample size for the training set is not provided in the document. The text only mentions that "The cases collected for the pivotal dataset were all distinct in time or center from the cases used to train the algorithm."

    9. How the Ground Truth for the Training Set was Established

    • The document does not explicitly describe how the ground truth for the training set was established. It only ensures that the training and test sets were distinct. While it's implied that similar expert-based ground truth would be used, the specifics are not detailed.
    Ask a Question

    Ask a specific question about this device

    K Number
    K232083

    Validate with FDA (Live)

    Date Cleared
    2023-11-13

    (123 days)

    Product Code
    Regulation Number
    892.2050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BriefCase-Quantification of Midline Shift (MLS) is a radiological image management and processing system software intended for automatic measurement of brain midline shift in non-contrast head CT (NCCT) images, in adults or transitional adolescents aged 18 years and older.

    The device is intended to assist appropriately trained medical specialists by providing the user with an automated current manual process of measuring midline shift.

    The device provides midline shift measurement from NCCT images acquired at a single time point, and can additionally provide an output with comparative analysis of two or more images that were acquired in the same individual at multiple time points.

    The device does not alter the original medical image and is not intended to be used as a diagnostic device. The BriefCase-Quantification results are not intended to be used on a stand-alone basis for clinical decision-making or otherwise preclude clinical assessment of cases. Clinicians are responsible for viewing full images per the standard of care.

    Device Description

    BriefCase-Quantification is a radiological image management and processing device. The software consists of a single module based on an algorithm programmed component and is intended to run on a linux-based server in a cloud environment.

    The BriefCase-Quantification receives filtered DICOM Images, and processes them chronologically by running the algorithm on relevant series to quantify the extent of midline shift. Following the Al processing, the output of the algorithm analysis is transferred to an image review software (the PACS or a desktop application).

    The device generates a summary report that includes a preview image of the slice with the largest midline shift. The preview image includes the measured shift, the annotation of the midline, and the annotation of the largest perpendicular distance between the midline and septum pellucidum. Also, the summary report includes a table and a graph showing the measured midline shift over time for patients with multiple scans.

    AI/ML Overview

    Here's a detailed breakdown of the acceptance criteria and study information for the BriefCase-Quantification device, based on the provided document:

    Acceptance Criteria and Device Performance

    CriteriaAcceptance CriteriaReported Device Performance
    Primary Endpoint: Mean Absolute Error (MAE)Mean absolute error estimate must be lower than prespecified performance goal.0.94 mm (95% CI: 0.74 mm, 1.14 mm) (Lower than prespecified goal)
    Secondary Endpoint: Bias (Bland-Altman plot)Little to no bias between ground truth and algorithm output.Mean difference of -0.15 mm (Little to no bias)
    Secondary Endpoint: MAE for multiple time points (First Case)Mean absolute error estimate must be lower than prespecified performance goal.1.16 mm (95% CI: 0.61 mm, 1.71 mm) (Lower than prespecified goal)
    Secondary Endpoint: MAE for multiple time points (Follow-up Cases)Mean absolute error estimate must be lower than prespecified performance goal.1.28 mm (95% CI: 0.68 mm, 1.88 mm) (Lower than prespecified goal)

    Study Details

    1. Sample size for the test set and data provenance:

      • Sample Size: 284 cases from 228 unique patients.
      • Data Provenance: Retrospective, multi-center study from 6 US-based clinical sites (both academic and community centers). The cases were distinct in time or center from the cases used to train the algorithm, indicating independent test data.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Number of Experts: Three neuroradiologists.
      • Qualifications: The document states "appropriately trained medical specialists" and specifically "three neuroradiologists," implying they are qualified experts in the field. Specific experience (e.g., "10 years of experience") is not provided.
    3. Adjudication method for the test set:

      • Method: The reference standard (ground truth) was created as the mean of all three independent measurements by the neuroradiologists. This implies a "consensus by average" approach rather than a specific 2+1 or 3+1 voting method.
    4. If a multi-reader, multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • MRMC Study: No, an MRMC comparative effectiveness study was not explicitly stated as performed with human readers and AI assistance. The study described focuses on the standalone performance of the AI algorithm against a neuroradiologist-established ground truth.
      • Effect Size of Human Improvement with AI: This information is not provided because an MRMC study comparing human readers with and without AI assistance was not detailed.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • Standalone Performance: Yes, a standalone performance study was done. The reported performance metrics (MAE, Bland-Altman) directly compare the algorithm's output to the ground truth established by experts, without human intervention in the device's measurement process. The device is intended to assist specialists by providing an automated process, but its performance evaluation here is purely algorithmic.
    6. The type of ground truth used:

      • Ground Truth Type: Expert consensus. Specifically, the "mean of all three [neuroradiologist] measurements."
    7. The sample size for the training set:

      • Training Set Sample Size: Not explicitly stated. The document only mentions that the "cases collected for the pivotal dataset were all distinct in time or center from the cases used to train the algorithm."
    8. How the ground truth for the training set was established:

      • Training Set Ground Truth: Not explicitly stated. Given the nature of a supervised learning algorithm, it is implied that the training data also had established ground truth measurements, likely derived through expert review, but the specific method or number of experts for the training data is not detailed in this summary.
    Ask a Question

    Ask a specific question about this device

    K Number
    K230534

    Validate with FDA (Live)

    Date Cleared
    2023-11-08

    (254 days)

    Product Code
    Regulation Number
    892.2050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BriefCase-Quantification is a radiological image management and processing system software indicated for use in the analysis of CT exams with contrast, that include the abdominal aorta, in adults or transitional adolescents aged 18 and older.

    The device is intended to assist appropriately trained medical specialists by providing the user with the maximum abdominal aortic axial diameter measurement of cases that include the abdominal aorta (M-AbdAo), BriefCase-Quantification is indicated to evaluate normal and aneurysmal abdominal aortas and is not intended to evaluate postoperative aortas.

    The BriefCase-Quantification results are not intended to be used on a stand-alone basis for clinical decision-making or otherwise preclude clinical assessment of cases. These measurements are unofficial, are not final, and are subject to change after review by a radiologist. For final clinically approved measurements, please refer to the official radiology report. Clinicians are responsible for viewing full images per the standard of care.

    Device Description

    BriefCase-Quantification is a radiological medical image management and processing device. The software consists of a single module based on an algorithm programmed component and is intended to run on a linux-based server in a cloud environment.

    The BriefCase-Quantification receives filtered DICOM Images, and processes them chronologically by running the algorithm on relevant series to measure the maximum abdominal aortic diameter. Following the Al processing, the output of the algorithm analysis is transferred to an image review software (desktop application), and forwarded to user review in the PACS.

    The BriefCase-Quantification produces a preview image annotated with the maximum axial diameter measurement. The diameter marking is not intended to be a final output, but serves the purpose of visualization and measurement. The original, unmarked series remains available in the PACS as well. The preview image presents an unofficial and not final measurement, and the user is instructed to review the full image and any other clinical information before making a clinical decision. The image includes a disclaimer: "Not for diagnostic use. The measurement is unofficial, not final, and must be reviewed by a radiologist."

    BriefCase-Quantification is not intended to evaluate post-operative aortas.

    AI/ML Overview

    1. A table of acceptance criteria and the reported device performance

    Acceptance CriteriaReported Device Performance
    Mean absolute error between ground truth measurement and algorithm1.95 mm (95% Cl: 1.59 mm, 2.32 mm)
    Performance GoalMean absolute error estimate below prespecified performance goal (specific numerical goal not explicitly stated, but was met)

    2. Sample size used for the test set and the data provenance

    • Test set sample size: 160 cases
    • Data provenance: Retrospective, from 6 US-based clinical sites (both academic and community centers). The cases were distinct in time and/or center from the training data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Number of experts: 3
    • Qualifications of experts: US board-certified radiologists.

    4. Adjudication method for the test set

    The document does not explicitly state the adjudication method (e.g., 2+1, 3+1). It only mentions that the ground truth was "determined by three US board-certified radiologists." This implies a consensus-based approach, but the specific decision rule is not detailed.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, a multi-reader multi-case (MRMC) comparative effectiveness study evaluating human reader improvement with AI assistance versus without AI assistance was not done. This study focused on the standalone performance of the algorithm against a ground truth established by experts.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, a standalone performance study of the algorithm was done. The "Primary Endpoint" section details the algorithm's performance (mean absolute error) compared to the ground truth.

    7. The type of ground truth used

    Expert consensus was used as the ground truth. The ground truth measurements were "determined by three US board-certified radiologists."

    8. The sample size for the training set

    The document does not specify the sample size for the training set. It only states that the cases collected for the pivotal dataset (test set) were "distinct in time and/or center from the cases used to train the algorithm."

    9. How the ground truth for the training set was established

    The document does not explicitly describe how the ground truth for the training set was established. It only mentions that the cases were "distinct in time and/or center from the cases used to train the algorithm," implying that training data also had a ground truth, likely established by similar expert review, but this is not detailed.

    Ask a Question