Device Name
Biomoneta Avata RxManufacturer
Date Cleared
2025-11-14
(149 days)
Product Code
Regulation Number
880.5045Type
TraditionalPanel
General HospitalAge Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use
The Biomoneta Avata Rx is an over-the-counter (OTC) medical recirculating air cleaner intended to remove particles for medical purposes. The device is intended for indoor use only.
The Biomoneta Avata Rx has been demonstrated to reduce Staphylococcus epidermidis, Klebsiella aerogenes, and MS2 bacteriophage by > 5 log when operated at its highest fan speed for 60 minutes.
Note: The designation "Rx" in the product name is part of the trade name and does not indicate prescription status.
Performance was validated in controlled test chambers measuring 9.1 ft × 9.1 ft × 7 ft (bioaerosol) and 12 ft × 12.5 ft × 15 ft (particle reduction).