Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K973932

    Validate with FDA (Live)

    Date Cleared
    1997-12-15

    (61 days)

    Product Code
    Regulation Number
    862.1150
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    CAL 2 (Calibrator 2), when used in conjunction with Beckman alpha,-antitrypsin (AAT), ceruloplasmin (CER), properdin factor B (PFB), beta-2-microglobulin (B2M), alphamacroglobulin (AMG) and antithrombin III (AT3) reagents, is intended for used on Array®, Array® 360, and IMMAGE™ Systems for the calibration of these reagents.

    Device Description

    Beckman Calibrator 2 is derived from fresh frozen human plasma that has been defibrinated and processed. Assay of CAL 2 provides a response value that is utilized for the adjustment of pre-programmed calibration curves from which AAT, CER, PFB, B2M, AMG, and AT3 concentration values are determined for test specimens.

    AI/ML Overview

    The provided document is a 510(k) summary for the Beckman Calibrator 2 (CAL2), a device used for calibrating diagnostic instruments. It describes the device's intended use and compares it to predicate devices. However, this document does not contain the detailed study results, acceptance criteria, or performance data typically associated with studies proving a device meets specific acceptance criteria for diagnostic accuracy metrics (e.g., sensitivity, specificity, AUC) or human reader performance with AI assistance.

    The document states: "The data in the Premarket Notification on safety and effectiveness supports a finding of substantial equivalence to chemistry test systems already in commercial distribution. The value assignment process for each analyte is correlated to a known standard via the anchor method. The Beckman Calibrator 2 value assignment and verification processes yield acceptable calibrator assigned values for calibration on the Array® and IMMAGE™ Systems."

    This indicates that internal testing was conducted to ensure the calibrator's values were acceptable and correlated to known standards, consistent with establishing substantial equivalence to existing predicate devices. However, it does not provide the specific numerical acceptance criteria or the raw performance data from such a study.

    Therefore, much of the requested information cannot be extracted from this specific document.

    Here's an attempt to answer the questions based only on the provided text, highlighting what is missing:


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Not explicitly stated in the document. The document implies "acceptable calibrator assigned values" and "correlation to a known standard" as criteria for "substantial equivalence."- "The value assignment process for each analyte is correlated to a known standard via the anchor method."- "The Beckman Calibrator 2 value assignment and verification processes yield acceptable calibrator assigned values for calibration on the Array® and IMMAGE™ Systems."- The device was found substantially equivalent to predicate devices (K771603, K780913, K791339, K791340, K901977, K940353).

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not specified in the document.
    • Data Provenance: Not specified in the document (e.g., country of origin, retrospective or prospective). The document only mentions "The data in the Premarket Notification."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • Not applicable/Not specified. This device is a calibrator for in-vitro diagnostic tests, not an interpretive AI device for medical images or clinical decision support that would typically involve expert ground truth establishment for a test set in the same way. The ground truth for a calibrator would be an established reference standard or a method traceable to an international standard.

    4. Adjudication Method for the Test Set

    • Not applicable/Not specified. Adjudication methods like 2+1 or 3+1 are typically used for expert consensus on image interpretation or clinical diagnosis, which is not the primary function of a calibrator.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    • No, an MRMC comparative effectiveness study was not done. This type of study involves human readers interpreting cases, often with and without AI assistance, to measure performance improvement. The Beckman Calibrator 2 is a diagnostic calibrator, not an AI-powered diagnostic tool used by human readers for interpretation.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    • Yes, in spirit. The testing described (though lacking detail) focuses on the performance of the calibrator itself in assigning values and its compatibility with the instruments. This is analogous to a "standalone" performance evaluation of the calibrator's function, independent of human operators, beyond its intended use in calibrating the instrument. The "value assignment and verification processes" would be an assessment of the calibrator's intrinsic performance.

    7. The Type of Ground Truth Used

    • Known Standard / Anchor Method: The document states, "The value assignment process for each analyte is correlated to a known standard via the anchor method." This implies that the ground truth for the calibrator's values is established by comparing them to pre-defined, validated reference materials or methods.

    8. The Sample Size for the Training Set

    • Not applicable/Not specified. This is a calibrator, not an AI algorithm that undergoes "training" in the machine learning sense. Its values are assigned using a chemical process and verified against standards.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable/Not specified, as this device does not have a "training set" in the context of AI. The "ground truth" for the calibrator's values is established through correlation to "known standards via the anchor method."
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1