510(k) Data Aggregation

    K Number
    K132441

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2013-12-13

    (129 days)

    Product Code
    Regulation Number
    878.3300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Bard® Ventralex™ Hernia Patch is intended for use in all forms of hernia repair requiring reinforcement with a nonabsorbable support material. The small Bard® Ventralex™ Hernia Patch (4.3 cm / 1.7") is also intended to repair soft tissue deficiencies, including deficiencies caused by trocars.

    Device Description

    The proposed Ventralex Hernia Patch is a self-expanding, sterile prosthesis, containing two primary layers of polypropylene mesh stitched with polytetrafluoroethylene (PTFE) monofilament to an expanded polytetrafluoroethylene (ePTFE) sheet. The mesh component is non-absorbable; however, the device contains a fully absorbable recoil ring using SorbaFlex™ Memory Technology, an absorbable polydioxanone (PDO) monofilament.

    The mesh is constructed from knitted polypropylene monofilament and forms a strong, porous, support material. The top layer of mesh has a mesh strap and forms a positioning pocket with the second mesh layer. The propylene mesh side promotes tissue in-growth and repairs the defect. The strap and pocket facilitate placement, positioning and fixation of the device. For a reference point, the small Ventralex Hernia Patch is constructed with a depth marker on the positioning strap. The depth marker is polypropylene monofilament dyed with a blue colorant to indicate the position of the device relative to the end of the laparoscopic trocar sleeve. After fixation to the abdominal wall, excess positioning strap material above the fixation line and at the level of the fascia is cut off and discarded.

    A layer of ePTFE is stitched with PTFE monofilament thread and edge heat sealed to the polypropylene mesh. The ePTFE layer minimizes tissue attachment to the device. The SorbaFlex Memory Technology is comprised of an absorbable PDO monofilament welded to form a ring. The PDO monofilament recoil ring provides memory and stability to the device, facilitating ease of initial insertion, proper placement, and fixation of the device. The PDO monofilament fully degrades by means of hydrolysis in vivo in 6 - 8 months. The PDO monofilament ring is placed within a mesh tube constructed from a knitted polypropylene monofilament. The purpose of the mesh tube is to contain the PDO monofilament recoil ring during the degradation process.

    AI/ML Overview

    This document is a 510(k) Premarket Notification for a medical device, specifically a surgical mesh for hernia repair. It is a submission to the FDA to demonstrate substantial equivalence to legally marketed predicate devices, not a study proving the device meets specific acceptance criteria in the sense of clinical performance metrics.

    Therefore, many of the requested categories (acceptance criteria, device performance, sample sizes for test/training sets, expert qualifications, adjudication methods, MRMC studies, standalone performance, ground truth types) are not applicable to this type of regulatory submission, as it focuses on material and design equivalence, not clinical efficacy or diagnostic accuracy.

    Here's an breakdown based on the provided text, addressing the applicable points:

    1. A table of acceptance criteria and the reported device performance

    The document does not present a table of specific acceptance criteria in the way one might for a diagnostic AI or imaging device's performance metrics (e.g., sensitivity, specificity thresholds). Instead, the "acceptance criteria" for this 510(k) submission are implicitly met by demonstrating substantial equivalence to predicate devices through various tests.

    The reported device performance is described qualitatively rather than with quantitative metrics against predefined numerical thresholds.

    Acceptance Criteria Category (Implicit)Reported Device Performance/Evidence
    Biocompatibility"All testing conducted to date (as per the requirements of ISO 10993) indicate that the device is biocompatible as per its intended use." (Relies on prior testing for predicate Ventrio Hernia Patch K081777).
    Device Integrity/Functionality (Recoil Ring & Mesh)"Bench testing results and in vivo simulated use experiments demonstrate that the proposed device design meets product specifications and intended uses." "The proposed Ventralex Hernia Patch was deployed in a laparoscopic deployment model to evaluate the integrity of the PDO monofilament recoil ring and mesh tube when in a folded position." "Additionally, the proposed Ventralex Hernia Patch underwent bench testing to ensure that the device recoiled to a flat position after folding."
    Material Degradation/Resorption"In vivo and in vitro resorption studies were performed and provided in support of the Ventrio Hernia Patch via K081777. These resorption studies were adopted for the proposed Ventralex Hernia Patch."
    Substantial Equivalence"All test results provided in this submission support the safety and effectiveness of the device for its intended use and demonstrate that the proposed device is substantially equivalent to its predicates."

    2. Sample size used for the test set and the data provenance

    • Sample Size: Not explicitly stated with numerical values. The text refers to "bench testing results," "in vivo simulated use experiments," and "whole system simulated use testing in an animal model." These are typically qualitative or small-scale engineering/animal studies rather than large clinical test sets with defined sample sizes for statistical inference.
    • Data Provenance: The document does not specify country of origin for any data. The studies are described as "bench testing" and "in vivo simulated use experiments" (animal model), implying laboratory-based or preclinical studies. They are prospective in the sense that they were performed specifically for this 510(k) submission or referenced from previous predicate device submissions.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This is a medical device 510(k) for a physical implant, not a diagnostic or AI algorithm requiring expert ground truth for interpretation. The "ground truth" here is the physical and biological performance of the device itself (e.g., does it recoil, is it biocompatible, does the material degrade as expected).

    4. Adjudication method for the test set

    Not applicable. This type of submission does not involve human readers interpreting data that would require an adjudication method.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done

    Not applicable. This is a medical device 510(k) for a physical implant, not a diagnostic or AI algorithm.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a medical device 510(k) for a physical implant, not an algorithm.

    7. The type of ground truth used

    The "ground truth" implicitly used for this device involves:

    • Bench Testing Standards: Acceptance against predefined engineering specifications for material properties, mechanical performance (e.g., recoil to a flat position), and physical integrity.
    • Biocompatibility Standards: Compliance with ISO 10993 standards for biological evaluation of medical devices.
    • In Vivo Observations (Animal Model): Direct observation of device behavior in a biological system (e.g., "whole system simulated use testing in an animal model") to ensure it meets user needs.
    • Resorption Studies: Chemical and biological analyses to confirm the degradation profile of the absorbable component.

    8. The sample size for the training set

    Not applicable. This is a medical device 510(k) for a physical implant, not an AI algorithm that requires a training set. The device manufacturing processes would have internal validation data, but this is not typically disclosed as a "training set" in a 510(k) summary.

    9. How the ground truth for the training set was established

    Not applicable for the same reasons as #8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K021736

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2002-07-16

    (49 days)

    Product Code
    Regulation Number
    878.3300
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Intended for the use in all forms of hernia repair requiring reinforcement with nonabsorbable support material.

    Device Description

    The proposed Ventralex Patch is a self-expanding, three laver device. Two layers consist of polypropylene mesh. The top layer of polypropylene mesh forms a positioning strap and pocket. The purpose of the strap and pocket is to facilitate placement, positioning and fixation of the device. After the device has been properly placed and attached, the positioning strap must be removed and properly discarded. The monofilament PET polymer "ring" is captured between the two layers of polypropylene mesh and adds stability to the device enabling greater simplicity and assurance in the proper placement. The third layer of the device is a single layer of expanded polytetrafluoroethylene (ePTFE) that is attached to the polypropylene mesh with an interlocking PTFE stitch pattern. The peripheral edge of the polypropylene mesh is heat sealed to the ePTFE layer.

    AI/ML Overview

    This is a 510(k) premarket notification for a medical device called the Ventralex Patch, a surgical mesh for hernia repair. The request asks for details about acceptance criteria and a study proving the device meets those criteria.

    However, the provided text does not describe a study with quantitative acceptance criteria for performance metrics like sensitivity, specificity, or accuracy, which are typically associated with diagnostic or AI-driven devices. This document is for a physical surgical implant, and the performance data focuses on materials, manufacturing, and biological compatibility rather than statistical performance metrics from a clinical study comparing the device's efficacy in patients against specific thresholds.

    Therefore, many of the requested details (sample size for test set, data provenance, number of experts, adjudication method, MRMC study, standalone performance, training set details) are not applicable or not present in this 510(k) summary.

    Here's an analysis based on the information available in the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria CategoryReported Device Performance (from text)
    Biocompatibility"Biocompatibility test results show that the material used in the design and manufacture of the device are non-toxic and non-sensitizing to biological tissues consistent with their intended use."
    Bench Testing"Laboratory test results demonstrate that the materials chosen and the design utilized in manufacturing the Ventralex Patch will meet the established specifications necessary for consistent performance during their intended use."
    Material Equivalence"The Ventralex Patch and the predicate Composix Kugel Mesh have a single layer of expanded polytetrafluoroethylene (ePTFE) and is attached to the polypropylene mesh with an interlocking stitch using polytetrafluoroethylene (PTFE) monofilament. The peripheral edge of the polypropylene mesh is heat sealed to the ePTFE layer." "The proposed device has the same the same materials, and similar physical attributes and manufacturing methods."
    Manufacturing Standards"The Ventralex Patch and the predicate Composix Kugel Mesh are both intended for the use in all forms of hernia repair requiring reinforcement with a nonabsorbable support material... The proposed device has the same the same materials, and similar physical attributes and manufacturing methods."

    2. Sample size used for the test set and the data provenance:

    • Not applicable / Not provided. The summary refers to "biocompatibility and bench testing" but does not detail specific sample sizes or data provenance (e.g., country of origin, retrospective/prospective clinical data) for these tests in the context of a "test set" as one would for an AI algorithm or diagnostic device. These tests would involve laboratory animals or materials testing according to established standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable / Not provided. Ground truth, in the context of this device, refers to the scientific and regulatory standards for biocompatibility and material performance. It's not established by a panel of medical experts reviewing individual cases in the way it would be for an AI diagnostic. Certification and standard compliance are the "ground truth" here.

    4. Adjudication method for the test set:

    • Not applicable / Not provided. Adjudication methods like 2+1 or 3+1 are used for reconciling expert disagreements in clinical studies, particularly for diagnostic devices. This 510(k) summary does not mention such a process.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, not applicable. This is not an AI device or a diagnostic imaging device that would involve human readers interpreting output. Therefore, an MRMC study is not relevant here.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • No, not applicable. This is a physical surgical implant, not an algorithm.

    7. The type of ground truth used:

    • For biocompatibility: Established biological testing standards and regulatory guidelines.
    • For bench testing: Validated engineering specifications and material science principles.
    • For substantial equivalence: The performance characteristics and material composition of the legally marketed predicate device (Bard Composix Kugel Mesh).

    8. The sample size for the training set:

    • Not applicable / Not provided. The concept of a "training set" is relevant for machine learning algorithms, not for the regulatory approval of a physical medical device like this surgical mesh. The "training" here would be the prior research, development, and manufacturing experience of Davol, Inc.

    9. How the ground truth for the training set was established:

    • Not applicable / Not provided. As above, the concept of a "training set" and its associated ground truth isn't applicable to this type of device and submission.