Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K243938

    Validate with FDA (Live)

    Date Cleared
    2025-04-01

    (102 days)

    Product Code
    Regulation Number
    870.1330
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Aristotle 14 Guidewire is intended for general vascular use within the neuro and peripheral vasculatures to introduce and position catheters and other interventional devices. The guidewire is not intended for use in the coronary vasculature.

    The Aristotle 18 Guidewire is intended for general vascular use within the neuro and peripheral vasculatures to introduce and position catheters and other interventional devices. The guidewire is not intended for use in the coronary vasculature.

    The Aristotle 24 Guidewire is intended for general vascular use within the neuro and peripheral vasculatures to introduce and position catheters and other interventional devices. The guidewire is not intended for use in the coronary vasculature.

    The Zoom Wire 14 Guidewire is intended for general vascular use within the neuro and peripheral vasculatures to introduce and position catheters and other interventional devices. The guidewire is not intended for use in the coronary vasculature.

    The Aristotle Colossus Guidewire is intended for general vascular use within the neuro and peripheral vasculatures to introduce and position catheters and other interventional devices. The guidewire is not intended for use in the coronary vasculature.

    Device Description

    The Scientia Vascular's Aristotle 14, Aristotle 18, Aristotle 24, Zoom Wire 14, and Aristotle Colossus Guidewires ("the guidewires") are steerable guidewires with a shapeable tip to aid in accessing the neuro and peripheral vasculatures. The guidewires are supplied sterile, for single use only.

    The distal portion of each guidewire's tip includes a radiopaque platinum wire marker coil to facilitate fluoroscopic visualization. All guidewires have a hydrophilic polymer coating on the distal portion and a polytetrafluoroethylene (PTFE) coating on the proximal portion to reduce friction during manipulation in vessels.

    The guidewires are provided with an accessory kit consisting of an introducer (to aid with the insertion of the guidewire into a catheter hub and/or a hemostasis valve), a torque device (to attach to the proximal portion of the guidewire to facilitate gripping and manipulation of the guidewire during use), and a shaping mandrel (to aid in shaping the flexible tip of the guidewire). These accessory devices are included to facilitate use of the guidewires and are not intended to contact the patient's body.

    AI/ML Overview

    This document describes a 510(k) clearance for several guidewire devices, not an AI/ML medical device. Therefore, the information requested about acceptance criteria and studies related to AI/ML device performance (such as sample sizes for test/training sets, expert ground truth establishment, MRMC studies, etc.) is not applicable and cannot be extracted from this document.

    The document primarily focuses on establishing substantial equivalence for physical medical devices—guidewires—by comparing their technological characteristics and performing non-clinical performance tests (biocompatibility, sterilization, bench performance).

    Here's a breakdown of why this document does not contain the requested information about an AI/ML device:

    • Device Type: The clearance is for "Aristotle 14 Guidewire; Aristotle 18 Guidewire; Aristotle 24 Guidewire; Zoom Wire 14 Guidewire; Aristotle Colossus Guidewire" which are physical catheter guide wires. These are traditional medical devices, not software or AI/ML devices.
    • Study Types: The studies described are "Biocompatibility," "Sterilization," and "Bench Performance Testing" (Visual Inspection, Coating Integrity, Corrosion Resistance, Agent Compatibility, Simulated Use, Particulate). These are mechanical and biological tests relevant to physical devices, not algorithmic performance.
    • Missing Information: All points you requested regarding AI/ML performance are absent because they are not relevant to this type of device clearance:
      • No acceptance criteria for AI performance metrics (e.g., sensitivity, specificity, AUC).
      • No test or training set sample sizes for AI data.
      • No information about experts establishing ground truth for images or data used by an AI.
      • No adjudication methods.
      • No MRMC comparative effectiveness studies.
      • No standalone algorithm performance.
      • No ground truth type related to AI (e.g., imaging labels, pathology).

    The document explicitly states:

    • "No animal testing was deemed necessary to support the substantial equivalence of the subject devices."
    • "No clinical testing was deemed necessary to support the substantial equivalence of the subject devices."

    This indicates that the clearance relies on non-clinical data and comparison to predicate devices, which is common for guidewires, especially when the changes are minor (like a different PTFE coating).

    Ask a Question

    Ask a specific question about this device

    K Number
    K220398

    Validate with FDA (Live)

    Date Cleared
    2023-03-10

    (392 days)

    Product Code
    Regulation Number
    870.1330
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Aristotle 14 Guidewire is intended for general vascular use within the neuro and peripheral vasculatures to introduce and position catheters and other interventional devices. The guidewire is not intended for use in the coronary vasculature.

    The Aristotle 18 Guidewire is intended for general vascular use within the neuro and peripheral vasculatures to introduce and position catheters and other interventional devices. The guidewire is not intended for use in the coronary vasculature.

    The Aristotle 24 Guidewire is intended for general vascular use within the neuro and peripheral vasculatures to introduce and position catheters and other interventional devices. The guidewire is not intended for use in the coronary vasculature.

    The Volo 14 Guidewire is intended for general vascular use within the neuro and peripheral vasculatures to introduce and position catheters and other interventional devices. The guidewire is not intended for use in the coronary vasculature.

    The Zoom Wire 14 Guidewire is intended for general vascular use within the neuro and peripheral vasculatures to introduce and position catheters and other interventional devices. The guidewire is not intended for use in the coronary vasculature.

    Device Description

    The Scientia Vascular's Aristotle 14, Aristotle 24, Volo 14, and Zoom Wire 14 Guidewires ("the guidewires") are steerable guidewires with a shapeable tip to aid in accessing the neuro and peripheral vasculatures. The guidewires are supplied sterile, for single use only, in the following diameters, stiffness profiles, and lengths:

    Aristotle 14 Guidewire: Diameter 0.014", Stiffness Profiles Soft, Standard, Lengths 200 cm, 300 cm
    Aristotle 18 Guidewire: Diameter 0.018", Stiffness Profiles Soft, Standard, Support, Length 200 cm
    Aristotle 24 Guidewire: Diameter 0.024", Stiffness Profiles Soft, Standard, Support, Length 200 cm
    Volo 14 Guidewire: Diameter 0.014", Stiffness Profiles Soft, Standard, Lengths 200 cm, 300 cm
    Zoom Wire 14 Guidewire: Diameter 0.014", Stiffness Profiles Support, Extra Support, Lengths 200 cm, 300 cm

    The distal portion of each guidewire's tip includes a radiopaque platinum wire marker coil to facilitate fluoroscopic visualization. All guidewires have a hydrophilic polymer coating on the distal portion and a polytetrafluoroethylene (PTFE) coating on the proximal portion to reduce friction during manipulation in vessels.

    The guidewires are provided with an accessory kit consisting of an introducer (to aid with the insertion of the guidewire into a catheter hub and/or a hemostasis valve), a torque device (to attach to the proximal portion of the guidewire to facilitate gripping and manipulation of the guidewire during use), and a shaping mandrel (to aid in shaping the flexible tip of the guidewire). These accessory devices are included to facilitate use of the guidewires and are not intended to contact the patient's body.

    AI/ML Overview

    This document concerns the 510(k) premarket notification for five guidewires: Aristotle 14 Guidewire, Aristotle 18 Guidewire, Aristotle 24 Guidewire, Volo 14 Guidewire, and Zoom Wire 14 Guidewire. The submission aims to demonstrate substantial equivalence to previously cleared predicate devices.

    1. Acceptance Criteria and Reported Device Performance:

    The acceptance criteria are not explicitly stated as numerical thresholds for performance metrics. Instead, the document indicates that the devices "met acceptance criteria" or "passed" various tests, implying successful adherence to pre-defined standards for each test. The study's focus is on demonstrating similar performance and safety to predicate devices, particularly given a change in the PTFE coating.

    Test CategoryTestAcceptance Criteria (Implied)Reported Device Performance
    BiocompatibilityCytotoxicityNon-cytotoxicPass: Non-cytotoxic
    PyrogenicityNon-pyrogenicPass: Non-pyrogenic
    SensitizationNon-sensitizingPass: Non-sensitizing
    IrritationNon-irritantPass: Non-irritant
    Acute Systemic ToxicityNo evidence of toxicityPass: No evidence of acute systemic toxicity
    HemocompatibilityNon-hemolyticPass: Non-hemolytic
    Complement Activation of SC5b-9Similar to comparator devicePass: The test article complement activation was similar to the comparator device.
    Partial Thromboplastin TimeSimilar to comparator devicePass: The test article clotting time was similar to the comparator device.
    In Vitro ThrombogenicityThromboresistantPass: Thromboresistant
    Latex DetectionLatex not detectedPass: Latex was not detected
    SterilizationSterilization Assurance Level (SAL), EO/ECH residuals, Bacterial Endotoxin LevelsSAL of 10^-6, acceptable residual levelsAchieved SAL of 10^-6, appropriate levels of residuals and endotoxins confirmed (implied from "existing validated sterilization cycle uses 100% EO to achieve a sterilization assurance level (SAL) of 10-").
    Bench PerformanceVisual Inspection and Dimensional VerificationMet ISO 11070 requirementsAcceptance criteria met
    Coating Integrity AssessmentMet ISO 11070 and FDA guidanceAcceptance criteria met
    Corrosion ResistanceMet ISO 11070 requirementsAcceptance criteria met
    Agent CompatibilityMet ISO 11070 requirementsAcceptance criteria met
    Simulated UseMet performance in anatomical modelAcceptance criteria met
    ParticulateAcceptable particulate countAcceptance criteria met

    2. Sample size used for the test set and the data provenance:

    • Test Set Sample Size: The document does not explicitly state the numerical sample sizes for each specific test (e.g., number of guidewires tested for coating integrity). It only mentions that "performance testing was performed on the subject devices" and "testing and evaluation... of the subject devices has been performed".
    • Data Provenance: The data provenance is internal to the manufacturer (Scientia Vascular, Inc.) and is prospective testing conducted specifically for this 510(k) submission. There is no mention of country of origin for the data other than it being part of an FDA submission from a U.S.-based company.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This submission pertains to a medical device's physical and biological performance, not an AI/software device that requires expert-established ground truth for diagnostic or interpretative tasks. Therefore, these details are not applicable. The "ground truth" for the tests mentioned (biocompatibility, sterilization, bench performance) is based on established, standardized test methods (e.g., ISO standards, FDA guidance documents).

    4. Adjudication method for the test set:

    • Not applicable, as this is not a diagnostic AI/software study involving human interpretation that requires adjudication of expert opinions. The performance is assessed against objective technical specifications and standards.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This is not an AI/software device that involves human readers or an AI assistance component.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is a hardware medical device (guidewire) and does not involve an algorithm or AI.

    7. The type of ground truth used:

    • The "ground truth" for the various tests conducted is based on:
      • Standardized Test Methods: Adherence to defined protocols from ISO 10993 series, ISO 14971:2019, ISO 11070:2014, and FDA guidance documents.
      • Objective Measurements: E.g., cell culture observations for cytotoxicity, temperature rise for pyrogenicity, measurement of physical dimensions, observation of material integrity.
      • "Pass/Fail" Criteria: Compliance with established thresholds or qualitative assessments as defined by the standard test methods (e.g., "Non-cytotoxic," "Non-hemolytic," "Acceptance criteria met").

    8. The sample size for the training set:

    • Not applicable. As this is not an AI/software device, there is no "training set" in the context of machine learning.

    9. How the ground truth for the training set was established:

    • Not applicable, as there is no training set for an AI/software algorithm.
    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Aristotle 14 Guidewire is intended for general vascular use within the neuro and peripheral vasculatures to introduce and position catheters and other interventional devices. The guidewire is not intended for use in the coronary vasculature.

    Device Description

    The Scientia Vascular Aristotle 14 Guidewire is a 0.014" diameter steerable guidewire with a shapeable tip to aid in accessing vasculature. The guidewire is supplied sterile and is for single use only. It is provided in a range of stiffness profiles, from soft to support, The product is provided in 200cm and 300cm lengths. The distal portion of the guidewire tip includes a radiopaque platinum wire marker coil to facilitate fluoroscopic visualization. The guidewire has a hydrophilic polymer coating on the distal portion and a polytetrafluoroethylene (PTFE) coating on the proximal portion to reduce friction during manipulation in vessels. The guidewire is provided with an introducer (to aid with the insertion of the guidewire into a catheter hub and/or a hemostasis valve) and a torque device (to attach to the proximal portion to facilitate gripping and manipulation of the guidewire during use). The introducer and torque accessory devices are included to facilitate use of the guidewire and are not intended to contact the patient's body.

    AI/ML Overview

    The provided text is a 510(k) summary for the "Aristotle 14 Guidewire". This document summarizes non-clinical performance tests to demonstrate substantial equivalence to a predicate device, rather than a study proving the device meets specific performance acceptance criteria for a new clinical claim or an AI algorithm.

    Therefore, many of the requested elements for an AI algorithm study (like sample size for test/training sets, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, or standalone algorithm performance) are not applicable to this submission.

    However, I can extract the acceptance criteria and reported device performance from the Non-Clinical Performance Tests section, focusing on the functional and biocompatibility tests.


    Acceptance Criteria and Reported Device Performance for the Aristotle 14 Guidewire

    The provided document details non-clinical performance tests for the Aristotle 14 Guidewire to demonstrate substantial equivalence to a predicate device (PVS 1300 Synchro® 0.014" Guidewire, K032146). The study is based on a series of laboratory and simulated use tests rather than a clinical trial with human subjects or an AI algorithm performance evaluation.

    Here's a summary of the acceptance criteria (implied by the test objective and relevant standards) and the reported device performance.

    1. Table of Acceptance Criteria and Reported Device Performance

    Biocompatibility Tests

    TestAcceptance Criteria (Implied)Reported Device Performance
    Cytotoxicity (L-929 MEM Elution)No reactivity (Grade 0)Cell culture treated with test sample exhibited no reactivity (Grade 0). Conclusion: Non-cytotoxic.
    Sensitization (Magnusson-Kligman)No erythema or edema (Grade 0) at challenge sitesChallenge sites treated with test sample exhibited no erythema or edema (Grade 0). Conclusion: Negative for dermal sensitization.
    Irritation (Intracutaneous Toxicity)Low mean test score (e.g., ≤1 for 0.9% Normal Saline and Sesame Oil extracts)The mean test score in 0.9% Normal Saline extract was 0, and in Sesame Oil was 0.1. Conclusion: Non-irritating.
    Systemic Toxicity (Acute)No abnormal clinical signs indicative of toxicity during the 72-hour test periodNo study animals were observed with abnormal clinical signs indicative of toxicity during the 72-hour test period. Conclusion: Non-toxic.
    Material Mediated Pyrogenicity (Rabbit)Temperature increases for all test animals not exceeding acceptable test limit for maximum individual temperature riseTemperature increases for all test animals did not exceed the acceptable test limit for maximum individual temperature rise. Conclusion: Non-pyrogenic.
    Hemolysis (Direct Contact and Extract)Difference between hemolytic indices of test article and negative control within acceptable limits (e.g., ≤2% for non-hemolytic)The difference between the hemolytic indices of the test article and the negative control was 0.00% for both direct and indirect hemolysis. Conclusion: Non-hemolytic.
    Partial Thromboplastin Time (PTT)Same thrombogenicity category as predicate device (minimal activator)Both the subject device and predicate device fell within the same thrombogenicity category (minimal activator). Conclusion: Subject and predicate devices showed similar thrombogenicity.
    Complement Activation (SC5b-9)Amounts of SC5b-9 generated comparable to predicate deviceAmounts of SC5b-9 generated by the test article and predicate device after exposure times of 30 and 60 minutes were not statistically different. The amount of SC5b-9 generated by the test article was statistically lower than that released by the predicate after 90 minutes exposure. Conclusion: Subject and predicate devices showed similar complement system activation.
    Thrombogenicity (Dog Model)Average thrombus score comparable to predicate deviceThe average thrombus score was the same for both the subject and predicate device. Conclusion: Subject and predicate devices showed similar thrombogenic potential.
    Latex Antigenic Protein Content (LEAP)No detectable Latex Antigenic ProteinsNo Latex Antigenic Proteins were detected. Conclusion: Contains no detectable traces of latex.

    Functional Testing

    TestTest Method & Acceptance Criteria (Implied / Explicit)Reported Device Performance
    Visual InspectionISO 11070: Absence of extraneous matter, process/surface defects, or defects that may cause trauma to vessels.No extraneous matter, surface defects, or visible droplets of coating were present on the Aristotle 14 Guidewires.
    Dimensional VerificationISO 11070: All guidewires meet dimensional specifications per engineering drawings.All guidewires met dimensional specifications.
    Flexing TestISO 11070: No defects or damage / flaking of the coating after flexing.No defects or damage / flaking of the coating were observed after flexing.
    Tensile StrengthISO 11070: All guidewires meet minimum force breakage requirements.All guidewires met minimum force breakage requirements specified in ISO 11070.
    Tip Shape, RetentionGuidewires must be shapeable and must retain shaped angle after simulated use.All tips met shaping and shape retention requirements after simulated use.
    TorqueabilityMeasurement of torque response (average input to output lag) in an anatomical model; acceptable torque responses, comparable to predicate device.All guidewires demonstrated acceptable torque responses. The torque response of the subject device was comparable to that of the predicate device.
    Torque StrengthTorque turns to failure in an anatomical model; acceptable torque strength, comparable to predicate device.All guidewires demonstrated acceptable torque strength. The torque strength of the subject device was comparable to that of the predicate device.
    Tip FlexibilityMeasure force to deflect guidewire tips to 45 and 90 degrees at 5mm, 10mm, and 20mm test lengths; acceptable forces & comparable flexibility to predicate.The forces required to deflect the guidewire tips were acceptable. The flexibility of the tips of all subject devices was comparable to the tip flexibility of the predicate guidewire.
    FractureISO 11070: No signs of fracture, loosening, or failure after wrapping around mandrel.No guidewires showed signs of fracture, loosening, or failure after wrapping them 8 times around a mandrel.
    Coating Lubricity and DurabilityFrictional force of coated guidewires determined after simulated use in a tortuous path; meet specified frictional force requirements.All guidewires met specified frictional force requirements.
    Coating IntegrityCoating uniformity and integrity visually examined on dyed samples after simulated use in a tortuous path; acceptable coating coverage.All samples showed acceptable coating coverage after simulated use.
    ParticulatesParticulates of various size ranges counted after simulated use in a tortuous path; comparable number of particulates to predicate.A comparable number of particulates was recovered from subject and predicate devices following simulated use.
    Torque Device - Introducer TestingISO 11070: Acceptance criteria for biocompatibility, visual inspection, corrosion resistance, tensile testing, luer taper dimensions, pouch seal strength.Acceptance criteria of all tests were met.
    Simulated Use Model Testing & Product CompatibilityAnatomical model to simulate neurovasculature; acceptable performance in evaluations of: Torqueability, Lubricity, Microcatheter Support & Tracking, Compatibility with accessories.Guidewires and predicate devices were found to perform acceptably in evaluations of: Torqueability in tortuous vasculature, Lubricity, Microcatheter Support & Tracking, Compatibility with Introducer, Compatibility with Torque Device, and Compatibility with Microcatheter.
    Cadaver TestingPhysician evaluation of subject and predicate guidewires for performance characteristics; acceptable performance.Subject and predicate guidewires both exhibited acceptable performance.
    RadiopacityPhysician evaluation in human cadaver; acceptable radiopacity.Both subject and predicate guidewires exhibited acceptable radiopacity.
    Corrosion ResistanceISO 11070: No signs of corrosion on guidewires after soaking in typical end-use solutions.There were no signs of corrosion on guidewires after soaking in typical end-use solutions.
    Chemical CompatibilityNo degradation, corrosion, or physical decomposition after exposure to saline and contrast agent/saline solutions.All guidewires showed no signs of degradation, corrosion or physical decomposition after exposure.
    MRI Compatibility(N/A, indicated as "MRI Unsafe")No testing performed. Aristotle Guidewires are labeled “MRI Unsafe.”
    LatexASTM D6499-07: No detectable traces of latex proteins.No detectable traces of latex were found.
    Package IntegrityASTM D 4169:16, ASTM F 88-15, ASTM F 2096-11: Maintenance of integrity, affixation/legibility of labeling after transport simulation.Following exposure to typical storage and transportation conditions, all sterile barrier pouches maintained their integrity and labeling remained affixed and legible.
    Shelf LifeASTM F1980: All device performance acceptance criteria met after accelerated aging conditions simulating 6 months real-time storage.After exposure to accelerated aging conditions simulating real-time storage under ambient conditions for 6 months, all device performance acceptance criteria were met, justifying labeling the devices with a 6-month shelf life.
    Sterilization ValidationAAMI TIR 28:2009, ISO11135:2007 Annex B: Justification for adoption into EO sterilization processing group, PCDs more difficult to sterilize than devices, bioburden enumeration performed, bacteriostasis/fungistasis tests demonstrate no inhibition of growth.Results justified adoption into the EO sterilization processing group: Comparative and Bioburden Resistance study results demonstrated that PCDs are more difficult to sterilize than devices; Bioburden Enumeration and Extraction Efficiency tests were used to enumerate the CFUs present on devices; and Bacteriostasis/Fungistasis test results demonstrated that the product does not inhibit the growth of organisms.
    Sterilization Validation: EO and ECH ResidualsAAMI/ANSI/ISO 10993-7: Residual traces of EO and ECH well below limits.The residual traces of EO and ECH remaining in the Aristotle 14 Guidewire after exposure to the EO sterilization process are well below the limits specified in ISO 10993-7.
    Sterilization Validation: Bacterial Endotoxin LevelsANSI/AAMI ST72:2011/(R)2016, USP <161>, and USP <85>: LAL testing; Endotoxin levels below specified limit (e.g., < 2.15 EU/device).< 2.15 EU/device.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not explicitly state a single "test set" sample size in the context of an AI algorithm, or detailed provenance for each test (e.g., country of origin, retrospective/prospective). Instead, it describes various non-clinical tests.

    • Sample Size: For each functional and biocompatibility test, "All guidewires" or "All samples" or "All test animals" met the criteria, suggesting that a sufficient number of devices/samples were tested according to relevant standards (e.g., ISO 11070, ISO 10993-1). Specific sample numbers for each test are not provided but are implied to be adequate for conformity to the referenced standards.
    • Data Provenance: This is not applicable in the context of an AI evaluation. The tests are laboratory-based, simulating use conditions or evaluating material properties. The cadaver testing involved "human cadaver" generally, without specifying origin.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Number of Experts: For the "Cadaver Testing" and "Radiopacity" tests, "Physicians" evaluated the guidewires. The specific number of physicians is not stated.
    • Qualifications of Experts: The specific qualifications (e.g., specialty, years of experience) of the "Physicians" involved in cadaver testing are not provided.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Adjudication methods like 2+1 or 3+1 are typically used in clinical studies or expert-driven evaluations where there's a need to resolve discrepancies in independent assessments. This is not applicable to the non-clinical, objective laboratory tests described in this 510(k) summary. For the cadaver studies, the method of aggregation or adjudication among physicians, if more than one, is not described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This document describes non-clinical performance and substantial equivalence tests for a medical device (guidewire), not an AI algorithm. Therefore, an MRMC study related to AI assistance was not performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. As stated above, this document concerns a medical device, not an AI algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • For the non-clinical tests, "ground truth" is established by:
      • Referenced Standards: Adherence to international and FDA-recognized standards (e.g., ISO 10993-1, ISO 11070) for material properties and functional performance.
      • Laboratory Measurements: Objective measurements (e.g., dimensions, force, torque, particulate counts, chemical analysis).
      • Qualitative Assessments by Trained Personnel: Visual inspections, qualitative assessments in simulated use.
      • Physician Evaluation: For cadaver testing and radiopacity, physician opinion served as the "ground truth" for acceptable performance.
      • Comparison to Predicate Device: Many tests establish equivalence by showing comparable performance to the legally marketed predicate device.

    8. The sample size for the training set

    • Not applicable. This document is for a physical medical device and does not involve an AI algorithm training set.

    9. How the ground truth for the training set was established

    • Not applicable. This document is for a physical medical device and does not involve an AI algorithm training set.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1