Found 2 results

510(k) Data Aggregation

    K Number
    K252181

    Validate with FDA (Live)

    Device Name
    AquaA
    Date Cleared
    2025-09-09

    (60 days)

    Product Code
    Regulation Number
    876.5665
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AquaA water purification system is a modular reverse osmosis unit intended for use with hemodialysis systems to remove organic and inorganic substances and microbial contaminants from the water used for treating hemodialysis patients or related therapies. This device is intended to be a component in a complete water purification system, and is not a complete water treatment system. The reverse osmosis unit must be preceded by pre-treatment devices, and may need to be followed by post-treatment devices as well, to meet current AAMI/ANSI/ISO and Federal (U.S.) standards.

    The Visual LED Indicator is used to remotely monitor the operating status of a connected device (e.g., the reverse osmosis water treatment system).

    The AquaDETECTOR leakage monitoring system provides full leakage monitoring of a dialysis center, typically during non-treatment hours. The central control unit can be programed to switch off dialysis water supply systems, dialysis concentrate preparation systems, dialysis concentrate supply systems, booster pump assemblies, and pre-filtration devices when a leak is detected.

    Device Description

    The AquaA system is a software controlled electromechanical device. It is a reverse osmosis (RO) water purification system that uses pretreated soft water to produce dialysis water for hemodialysis (HD) devices and for the preparation of dialysis concentrates.

    The AquaA system is a modular system consisting of a base module that can be used on its own or in conjunction with other modules. AquaA is the base module. AquaA2 is a second RO unit that can be added to increase the quality of dialysis water. AquaHT is a flow heater unit that can be added to provide heat disinfection for the RO system including membranes and ring main. The AquaHT module can also supply hot product water to connected HD devices. AquaUF is an ultrafiltration module for the AquaA system intended to improve dialysis water quality by retaining constituents such as endotoxins and bacteria.

    The AquaA system is configured by combining the different modules with the AquaA base module. The system configurations are:

    • AquaA (Single stage RO system)
    • AquaA + AquaUF (Single stage RO system with ultrafiltration)
    • AquaA + AquaHT (Single stage RO system with heat disinfection)
    • AquaA + AquaHT + AquaUF (Single stage RO system with heat disinfection and ultrafiltration)
    • AquaA + AquaA2 (Double stage RO system)
    • AquaA + AquaA2 +AquaUF (Double stage RO system with ultrafiltration)
    • AquaA + AquaA2 + AquaHT (Double stage RO system with heat disinfection)
    • AquaA + AquaA2 + AquaHT + AquaUF (Double stage RO system with heat disinfection and ultrafiltration)

    The Visual LED Indicator, Aqua DETECTOR, and Connecting Tube PVDF are accessories to the AquaA system.

    The Visual LED Indicator is a required accessory installed in the dialysis clinic. It reproduces the AquaA's visual indicator color signals in the clinic and communicates the AquaA system's operating status (including visual and audible alarm states) to the user.

    The AquaDETECTOR system monitors leakage throughout a dialysis center. The AquaDETECTOR system consists of a monitoring control center and up to 40 leakage sensors connected to 1–3 BUS lines.

    The Connecting Tube PVDF is used to connect the fluid path of the installed modules while maintaining placeholders for future modular expansion.

    The AquaA water purification system uses pretreated soft water to produce dialysis water for HD devices and for the preparation of dialysis concentrates. The system features the following main operating modes:

    • STANDBY – The system is on and dialysis water is not being produced
    • SUPPLY – The system produces dialysis water, controls the programmed yield, and monitors all relevant parameters
    • RINSE – The system is cleaned with water by rinsing all tubing sections and by replacing the specified dialysis water volume
    • CLEANING – The cleaning mode is used to decalcify or alkaline clean the RO system including membranes. (NOTE: The water distribution system is not included.) An acidic or alkaline solution is circulated in the system for a programmed time period. A Disinfection phase is initiated and followed by a Rinse phase.
    • DISINFECTION – The system, including the ring main, is chemically disinfected. The disinfectant is circulated throughout the system for a programmed time period and then a Rinse phase is initiated.
    • HEAT DISINFECTION – The AquaA system with the AquaHT module allows for heat disinfection of the AquaA system including RO membranes, ring main, and HD device interfaces
    • EMERGENCY OPERATION – In the event of a system malfunction if emergency mode is initiated, AquaA or AquaA2, depending on which RO module failed, begins emergency operation as a single stage RO system. Dialysis water is still produced to complete any HD therapy that is in progress. Dialysis water conductivity (AquaA and AquaA2) and temperature (AquaA) are monitored.
    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K213507

    Validate with FDA (Live)

    Device Name
    AquaA
    Date Cleared
    2022-07-27

    (268 days)

    Product Code
    Regulation Number
    876.5665
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AquaA water purification system is a modular reverse osmosis unit intended for use with hemodialysis systems to remove organic and inorganic substances and microbial contaminants from the water used for treating hemodialysis patients or related therapies. This device is intended to be a component in a complete water purification system, and is not a complete water treatment system. The reverse osmosis unit must be preceded by pre-treatment devices, and may need to be followed by post-treatment devices as well, to meet current AAMI/ANSI/ISO and Federal (U.S.) standards.

    Device Description

    The AquaA system is a reverse osmosis (RO) water purification system that uses pretreated soft water to produce dialysis water for hemodialysis (HD) devices and for the preparation of dialysis concentrates. The AquaA system is a modular system consisting of a base module that can be used on its own or in conjunction with other modules. AquaA is the base module. AquaA2 is a second RO unit that can be added to increase the quality of dialysis water. AquaHT is a flow heater unit that can be added to provide heat disinfection for the RO system including membranes and ring main. The AquaHT module can also supply hot product water to connected HD devices. AquaUF is an ultrafiltration module for the AquaA system intended to improve dialysis water quality by retaining constituents such as endotoxins and bacteria.

    AI/ML Overview

    This document describes the regulatory submission for the AquaA water purification system, not a study proving it meets specific acceptance criteria based on performance metrics such as accuracy, sensitivity, or specificity. The submission focuses on demonstrating substantial equivalence to a predicate device, AquaBplus (K133829), rather than presenting a performance study with detailed acceptance criteria and reported device performance in the typical sense for an AI/CADe device.

    Therefore, many of the requested sections about specific acceptance criteria, sample sizes, expert adjudication, MRMC studies, or standalone performance are not applicable based on the provided text.

    Here's an attempt to answer the questions based only on the provided text, while also noting when the information is not present:

    1. A table of acceptance criteria and the reported device performance

    The document does not provide a table of quantitative acceptance criteria (e.g., accuracy, sensitivity, specificity) with corresponding reported device performance values. Instead, the "acceptance criteria" are implied by the conformity to various standards and the successful verification of functional requirements and safety.

    Test ConductedImplied Acceptance Criteria (Conformity to Standards/Requirements)Reported Device Performance (as stated in document)
    Essential PerformanceConformity to:Not explicitly quantified performance metrics. The text states: "Testing to demonstrate that the device meets the following standards: ISO 13959, ISO 26722." This suggests the device did meet these standards.
    - ISO 13959 Third Edition 2014-04-01 (Water for haemodialysis and related therapies)
    - ISO 26722:2014 (Water treatment equipment for haemodialysis applications and related therapies)
    Disinfection ValidationValidation of chemical and heat disinfection labeling in accordance with FDA guidance document.Not explicitly quantified, but implied successful validation per guidance.
    Functional VerificationVerification of performance (e.g., conductivity and temperature) and functional (e.g., operating modes and generated alarms) requirements of the device.Not explicitly quantified performance metrics. The text states: "Complete system testing to verify the performance... and functional... requirements of the device." This implies successful verification.
    PackagingPackaging and transport verification according to ASTM D4169-16.Implied successful verification per ASTM D4169-16.
    Biocompatibility TestingConformity to ISO 10993-1:2018 and FDA guidance document. Endpoints evaluated: Chemical Characterization, Cytotoxicity, Sensitization, Irritation, Material-Mediated Pyrogenicity, Hemocompatibility.Implied successful evaluation/meeting of endpoints. The text states: "The following endpoints were evaluated to support the biological safety of the AquaA system."
    Electrical Safety (AquaA/Visual LED Indicator)Conformity to ANSI/AAMI ES 60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012.Implied successful testing and conformity.
    EMC (AquaA/Visual LED Indicator)Conformity to IEC 60601-1-2:2014.Implied successful testing and conformity.
    Electrical Safety (AquaDETECTOR)Conformity to IEC 61010-1:2010 (3rd Edition).Implied successful testing and conformity.
    EMC (AquaDETECTOR)Conformity to IEC 61326-1:2012, FCC, Part 15, Subpart B, Class A.Implied successful testing and conformity; "intended use and essential performance are maintained."
    Software Verification and Validation TestingEffectiveness of software and confirmation of device operation per FDA guidance documents.Implied successful verification; "performed to demonstrate the effectiveness of the software and to confirm operation of the device."

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided in the document. The testing described (functional verification, electrical safety, biocompatibility, etc.) appears to be device-centric testing rather than testing on a dataset with "test sets" in the context of AI/CADe. There are no mentions of specific patient data or provenance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable and not provided. The AquaA is a water purification system, not a diagnostic device that relies on expert interpretation for ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable and not provided.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No such study was mentioned or conducted. This is not an AI/CADe device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not an AI/CADe device, so standalone algorithm performance is not applicable. The device's "performance" is its ability to purify water according to standards, which is inherent in its operation, not an algorithm's output.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The concept of "ground truth" as typically used in AI/CADe evaluations (e.g., for diagnostic accuracy) is not applicable here. The "truth" or verification for this device is based on:

    • Conformity to established international and national standards (e.g., ISO, AAMI, ASTM).
    • Results from various engineering and safety tests (functional, electrical safety, EMC, biocompatibility).
    • Risk management evaluations.

    8. The sample size for the training set

    This is not an AI/CADe device, so the concept of a "training set" is not applicable.

    9. How the ground truth for the training set was established

    Not applicable.