Search Results
Found 49 results
510(k) Data Aggregation
(224 days)
The Affinity® Pediatric Arterial Blood Filter is indicated for use in cardiopulmonary bypass procedures up to 6 hours in duration for the removal of particulate and gaseous microemboli.
The Affinity Pixie™ Pediatric Arterial Filter is indicated for use in cardiopulmonary bypass procedures up to 6 hours in duration for the removal of particulate and gaseous microemboli.
The Affinity Pixie™ Arterial Filter with Balance™ Biosurface and Affinity® Pediatric Arterial Blood Filter are single-use, sterile, nonpyrogenic devices designed to filter microemboli greater than the specified micron size, from the circuit for periods up to six (6) hours, during cardiopulmonary bypass surgery. The Affinity Pixie™ Arterial Filter with Balance™ Biosurface device is coated with a non-leaching biocompatible surface to enhance blood compatibility and provide thromboresistant blood-contacting surfaces. The Affinity® Pediatric Arterial Blood Filter is uncoated. The models listed above will be referred to as Affinity Pixie™ Arterial Filter throughout the submission.
N/A
Ask a specific question about this device
(124 days)
Model BB811 and Model BB811-NS: The Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass procedures up to 6 hours in duration. The Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface is designed to filter from the circuit microemboli larger than the specified micron size for periods up to six hours during cardiopulmonary bypass surgery.
Model BB841: Oxygenator with Integrated Arterial Filter: The Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass procedures up to 6 hours in duration. The Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface is designed to filter from the circuit microemboli larger than the specified micron size for periods up to six hours during cardiopulmonary bypass surgery. Cardiotomy/Venous Reservoir: The Affinity Fusion Cardiotomy/Venous Reservoir with Balance Biosurface is intended to be used in an extracorporeal perfusion circuit to collect venous and cardiotomy suctioned blood during routine cardiopulmonary procedures up to 6 hours in duration. The CVR is also intended for use during vacuum assisted venous drainage (VAVD) procedures. The Affinity Fusion Cardiotomy/Venous Reservoir with Balance Biosurface is also intended for use after open heart surgery to collect autologous blood from the chest and to aseptically return the blood to the patient for blood volume replacement.
The Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface (BB811), Affinity Fusion Oxygenator with Integrated Arterial Filter and Cardiotomy/Venous Reservoir with Balance Biosurface (BB841), and Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface (Non-Sterile) (B811-NS) are collectively referred to as the Affinity Fusion Oxygenator in the summary. The Affinity Fusion Oxygenator is intended to be used in an extracorporeal perfusion blood circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass procedures up to 6 hours in duration. The Affinity Fusion Oxygenator contains both an integrated arterial filter and integrated heat exchanger. The Affinity Fusion Oxygenator is a microporous, hollow-fiber, gas-exchange devices available with Balance Biosurface bonded to the blood contacting surface of the device. The integrated arterial filter is designed to filter from the circuit microemboli larger than the specified micron size from the circuit for periods up to six hours during cardiopulmonary bypass surgery. Some models of the Affinity Fusion Oxygenator are packaged with an Affinity Fusion Cardiotomy/Venous Reservoir (CVR) with Balance Biosurface which is designed to be an integral part of a cardiopulmonary bypass circuit for use during cardiac surgery. The Affinity Fusion CVR is designed to collect venous and cardiotomy suctioned blood during routine cardiopulmonary procedures up to six (6) hours in duration. Additionally, the Affinity Fusion CVR may be used during vacuum assisted venous drainage (VAVD) procedures and collect autologous blood from the chest and to aseptically return the blood to the patient for blood volume replacement during open heart surgery.
The provided text discusses the Affinity Fusion Oxygenator and its acceptance criteria, specifically focusing on the addition of an alternate hollow fiber supplier.
Here's an analysis of the acceptance criteria and the study that proves the device meets them:
1. A table of acceptance criteria and the reported device performance:
| Test Item | Acceptance Criteria (Implied: PASS) | Reported Device Performance |
|---|---|---|
| O2 Transfer | PASS | PASS |
| CO2 Transfer | PASS | PASS |
| Blood Side Pressure Drop | PASS | PASS |
| 6 Hour O2 Transfer | PASS | PASS |
| 6 Hour CO2 Transfer | PASS | PASS |
| Pressure Integrity | PASS | PASS* |
| Burst | PASS | PASS |
| Gas Pathway Integrity | PASS | PASS |
| Plasma Breakthrough | PASS | PASS |
| Filtration Efficiency | PASS | PASS |
| Hemolysis (Max Flow) | PASS | PASS |
| White Blood Cell Retention | PASS | PASS |
| Platelet (PLT) retention | PASS | PASS |
| Functional platelet retention | PASS | NA (not specified as tested at this time point) |
| Platelet PLT Function (min flow) | PASS | NA (not specified as tested at this time point) |
| Prime Volume | PASS | PASS |
| Coverage | PASS | PASS |
| Leaching | PASS | PASS |
Biocompatibility Testing:
| Biocompatibility Test Method | Testing Category | Reported Device Performance |
|---|---|---|
| ISO MEM Elution Cytotoxicity, L929 Mouse Fibroblast Cells | Cytotoxicity | PASS |
| ISO Maximization Sensitization Study, 0.9%Sodium Chloride & Sesame Oil, NF extract | Sensitization | PASS |
| ISO Intracutaneous Study – 0.9% Sodium Chloride & Sesame Oil, NF extract | Intracutaneous Reactivity | PASS |
| ISO Systemic Toxicity Study - 0.9% Sodium Chloride & Sesame Oil, NF extract | Acute Systemic Toxicity | PASS |
| Pyrogen Study – Material Mediated -0.9%Sodium Chloride | Material Mediated Pyrogen Study | PASS |
| ASTM In-vitro Hemolysis, Rabbit Blood | Hemocompatibility | PASS |
| Complement Activation, SC5b-9, Human Serum | Hemocompatibility | PASS |
| ASTM Partial Thromboplastin Time, Direct Contact, Human Plasma | Hemocompatibility | PASS |
| Genotoxicity, Bacterial Reverse Mutation Study | Genotoxicity | PASS |
| Genotoxicity: Mouse Lymphoma Assay | Genotoxicity | PASS |
| Platelet and Leukocyte Count- with comparison article | Hemocompatibility | PASS |
*Note on "Pressure Integrity": Devices conditioned only to factors significant for leaks (per Two Proportion statistical test).
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
The document does not explicitly state the sample sizes for the test sets (number of oxygenator units tested) for design verification or biocompatibility. It also does not specify the country of origin of the data or whether the studies were retrospective or prospective. The nature of the tests (engineering performance and biocompatibility) suggests these were primarily laboratory-based prospective tests.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
This information is not applicable to the type of device and testing described. The "ground truth" for a medical device like an oxygenator is established through objective performance measurements and standardized biocompatibility tests, not subjective expert assessment as would be the case for image interpretation or diagnosis.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
This information is not applicable. Adjudication methods like "2+1" or "3+1" are typically used in clinical studies or studies involving subjective human interpretation (e.g., radiology reads) to resolve discrepancies. The tests described here are objective performance and safety tests.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
This is not applicable. The device is an oxygenator, not an AI-powered diagnostic tool requiring human reader studies.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
This is not applicable. The device is an oxygenator, not an algorithm.
7. The type of ground truth used (expert concensus, pathology, outcomes data, etc):
The ground truth for this device is based on:
- Engineering specifications and standards: For performance characteristics like O2 Transfer, CO2 Transfer, Pressure Drop, Burst, Filtration Efficiency, etc. "PASS" implies meeting predefined, objective measurable criteria.
- International standards for biocompatibility (ISO 10993-1:2018): For tests like cytotoxicity, sensitization, genotoxicity, hemocompatibility, etc. "PASS" means meeting the established safety thresholds for these biological endpoints.
8. The sample size for the training set:
This is not applicable. This is not an AI/machine learning device that requires a training set. The "study" described is a series of laboratory-based design verification and biocompatibility tests.
9. How the ground truth for the training set was established:
This is not applicable, as there is no training set for this type of device.
Ask a specific question about this device
(30 days)
Affinity NT Oxygenator and Uncoated Cardiotomy/ Venous Reservoir (541);
The AFFINITY NT Integrated CVR Membrane Oxygenator with Plasma Resistant Fiber is intended to be used in an extracorporeal perfusion circuit to collect venous and cardiotomy suctioned blood, cool or warm the blood and oxygenate and remove carbon dioxide from the blood during routine cardiopulmonary bypass procedures up to 6 hours in duration.
Indications for Use(541B)
The Affinity NT oxygenator with Balance biosurface is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass (CPB) procedures up to 6 hours in duration. The Affinity NT cardiotomy/venous reservoir (CVR) is intended to be used in an extracorporeal perfusion circuit to collect venous and cardiotomy suctioned blood during routine CPB procedures up to 6 hours in duration.
The Affinity NT Oxygenator with /without Balance Biosurface is a sterile membrane Oxygenator with plasma resistant fiber. It is a single-use gas exchange device with integral stainless-steel heat exchanger. Models 541 and 541B include a cardiotomy/venous reservoir.
The Luer cap 1140008-503 that is the 'subject' of this special 510(k) serve as protection of the Luer Access Port. This port can be used to prime the circuit or connect to another support device with a luer connector.
This is a 510(k) summary for a medical device change, not a study proving device meets acceptance criteria. Here's a breakdown of why and what information is provided:
This document describes a "Special 510(k) submission" for a change in a small component (a luer cap) within an already cleared medical device (Affinity NT Oxygenator). It is not a standalone study proving the device meets general acceptance criteria for a new device. The purpose of this submission is to demonstrate that the change to the luer cap does not adversely affect the safety and effectiveness of the previously cleared device, maintaining substantial equivalence to the predicate.
Therefore, most of the requested information (sample sizes, expert consensus, MRMC studies, etc.) is not applicable or not provided in the context of this specific regulatory submission.
Here's how to address the prompt based on the provided text:
1. A table of acceptance criteria and the reported device performance
The document doesn't explicitly define acceptance criteria in a quantitative table form for the overall device's performance. Instead, it focuses on demonstrating that the change to the luer cap does not negatively impact the device's functionality. The "acceptance criteria" for this specific change are implicitly tied to maintaining the original device's performance and safety.
| Acceptance Criterion (implicitly for the luer cap change) | Reported Device Performance |
|---|---|
| Biocompatibility | Pass (leveraged from K240534, a higher-risk classification) |
| Functionality (e.g., impact on final product functionality or performance) | Pass (dimensional comparison, torque removal test, pressure integrity test concluded no impact) |
| Performance of the modified device after the change | Substantially Equivalent to legally marketed predicate devices |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Test Set Sample Size: Not explicitly stated for specific tests related to the luer cap change. The document mentions "Risk-based testing and evaluations," suggesting a sample size appropriate for these specific tests, but the number is not provided.
- Data Provenance: Not specified. Given it's a regulatory submission by Medtronic, it's presumably internal testing conducted to support the change.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
Not applicable. This is a technical validation of a component change, not an evaluation requiring expert clinical interpretation for ground truth.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. This is not a study requiring adjudication of interpretations.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is not an AI-enabled device or an MRMC study.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is not an algorithm-only device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
For the luer cap change, the "ground truth" is established through engineering and biological testing:
- Biocompatibility: Conformance to ISO 10993-1:2018 standards, which involves specific chemical and biological tests. The "ground truth" here is the pass/fail result based on laboratory analysis against these standards, leveraged from a previous clearance (K240534) for a higher-risk product.
- Functionality: Direct measurements and observations (dimensional comparison, torque removal, pressure integrity) against established engineering specifications.
8. The sample size for the training set
Not applicable. This is not a machine learning or AI device that requires a training set.
9. How the ground truth for the training set was established
Not applicable.
Ask a specific question about this device
(29 days)
The Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass procedures up to 6 hours in duration. The Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface is designed to filter from the circuit microemboli larger than the specified micron size for periods up to six hours during cardiopulmonary bypass surgery.
The Affinity Fusion Cardiotomy/Venous Reservoir with Balance Biosurface is intended to be used in an extracorporeal perfusion circuit to collect venous and cardiotomy suctioned blood during routine cardiopulmonary procedures up to 6 hours in duration. The CVR is also intended for use during vacuum assisted venous drainage (VAVD) procedures. The Affinity Fusion Cardiotomy/Venous Reservoir with Balance Biosurface is also intended for use after open heart surgery to collect autologous blood from the chest and to aseptically return the blood to the patient for blood volume replacement.
The Affinity Fusion oxygenator with integrated arterial filter and Cortiva bioactive surface is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass (CPB) procedures up to 6 hours in duration. The Affinity Fusion oxygenator with integrated arterial filter and Cortiva bioactive surface is designed to filter from the circuit microemboli larger than the specified micron size for periods up to 6 hours during CPB surgery.
The Affinity Fusion Oxygenator is intended to be used in an extracorporeal perfusion blood circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass procedures up to 6 hours in duration. The Affinity Fusion Oxygenator contains both an integrated arterial filter and integrated heat exchanger. The Affinity Fusion Oxygenator is a microporous, hollow-fiber, gas-exchange devices available with Cortiva BioActive Surface or Balance Biosurface bonded to the blood contacting surface of the integrated arterial filter is designed to filter from the circuit microemboli larger than the specified micron size from the circuit for periods up to six hours during cardiopulmonary bypass surgery. Some models of the Affinity Fusion Oxygenator are packaged with an Affinity Fusion Cardiotomy/Venous Reservoir (CVR) with Balance Biosurface which is designed to be an integral part of a cardiopulmonary bypass circuit for use during cardiac surgery. The Affinity Fusion CVR is designed to collect venous and cardiotomy suctioned blood during routine cardiopulmonary procedures up to six (6) hours in duration. Additionally, the Affinity Fusion CVR may be used during vacuum assisted venous drainage (VAVD) procedures and collect autologous blood from the chest and to aseptically return the blood to the patient for blood volume replacement during open heart surgery.
The provided text is a 510(k) Summary for the Medtronic Affinity Fusion Oxygenator System. This document focuses on demonstrating substantial equivalence to a predicate device rather than presenting a standalone study with detailed acceptance criteria and performance data for a novel device. Therefore, much of the requested information about a specific study proving the device meets acceptance criteria is not explicitly provided in this type of regulatory submission.
However, I can extract the relevant information that is available and clarify what is not present.
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state quantitative acceptance criteria in a table format for performance metrics relevant to an "AI system" or "medical device performance" in the general sense, as typically sought in such a request. Instead, it refers to "predetermined requirements" and "improved bond performance" for specific components.
The "performance" referred to in this document primarily relates to the device meeting its intended function as an oxygenator and arterial filter during cardiopulmonary bypass. The summary states that "Operation and performance qualification... were completed" and that "the manufacturing process consistently produces product that meets predetermined requirements."
| Aspect of Performance/Criteria | Reported Device Performance / Evaluation Outcome |
|---|---|
| Overall Performance | - "No newly emerging hazards or risks were identified." - "Modifications... result in a substantially equivalent device because the fundamental scientific principle, operating principle, design features and intended use are unchanged from the predicate device." |
| Manufacturing Process | "Consistently produces product that meets predetermined requirements under challenge and normal operating conditions." |
| Pressure Integrity | Evaluated as part of OQ/PQ. (Specific criteria/results not detailed) |
| Burst (Strength) | Evaluated as part of OQ/PQ. (Specific criteria/results not detailed) |
| Torque (TMA bond) | "Improved bond performance relative to the current material." |
| Insertion Depth (TMA) | Evaluated as part of OQ/PQ. (Specific criteria/results not detailed) |
| Biocompatibility | New material is "biocompatible per the ISO 10993 standard and does not require new biocompatibility testing." |
| TMA Seal (non-patient contacting) | "Verified through the part specifications, along with random sampling and statistical modeling." |
2. Sample Size Used for the Test Set and Data Provenance
The document does not specify a "test set" in the context of an AI/algorithm evaluation. For the physical device modifications, it mentions "random sampling and statistical modeling" for the TMA seal, but no specific sample sizes or data provenance (country, retrospective/prospective) are provided for these manufacturing/design evaluations.
3. Number of Experts Used to Establish Ground Truth and Qualifications
This information is not applicable to the provided document. The submission pertains to modifications of a physical medical device (oxygenator) and its manufacturing process, not to an AI/algorithm that requires expert-established ground truth from a dataset.
4. Adjudication Method
This information is not applicable to the provided document, as it does not describe an AI/algorithm evaluation requiring adjudication of ground truth.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
This information is not applicable to the provided document. The submission is for a physical medical device and does not involve AI assistance to human readers.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
This information is not applicable to the provided document. The submission is for a physical medical device and does not involve an AI algorithm.
7. Type of Ground Truth Used
This information is not applicable in the context of an AI/algorithm. For the physical device, the "ground truth" or verification is based on engineering specifications, performance testing (pressure integrity, burst, torque, insertion depth), and biocompatibility standards (ISO 10993).
8. Sample Size for the Training Set
This information is not applicable to the provided document, as it does not describe an AI/algorithm that would require a training set.
9. How the Ground Truth for the Training Set Was Established
This information is not applicable to the provided document, as it does not describe an AI/algorithm.
Summary of what the document does provide:
The document describes a 510(k) premarket notification for the Affinity Fusion Oxygenator System, which involves modifications to a legally marketed predicate device (K203111). The core of the submission is to demonstrate substantial equivalence to this predicate device.
The study presented is not a classic "clinical trial" or "AI validation study" but rather a series of engineering and qualification tests to ensure that the modifications (specifically, changes to the Temperature Monitoring Adapter (TMA) insert component material and insertion depth, and different bioactive surfaces) do not alter the fundamental performance, safety, or efficacy of the device.
Key points from the document regarding "proof" and "acceptance":
- Proof of Meeting Acceptance Criteria: The proof is based on "Operation and performance qualification (OQ/PQ)" and "Design Characterization" activities. These activities evaluate various physical properties and manufacturing consistency.
- Acceptance Criteria (Implied): The implied acceptance criteria are that the modified device's performance (e.g., pressure integrity, burst, torque, insertion depth, biocompatibility) meets internal predetermined requirements and does not introduce new hazards compared to the predicate device. For example, the TMA bond evaluation "indicated improved bond performance relative to the current material," suggesting a performance target was met or exceeded. Biocompatibility was accepted based on the new material already being used in another cleared device and conforming to ISO 10993.
- Study Design: The study is a non-clinical engineering and bench testing evaluation rather than a clinical trial or AI performance study. It's focused on validating the impact of specific component changes.
- Conclusion: The manufacturer concluded, and the FDA agreed by clearing the device, that these modifications maintain substantial equivalence to the predicate device, meaning it is as safe and effective as a device already on the market.
Ask a specific question about this device
(56 days)
AFFINITY Proximal Tibia System is intended to treat fractures, nonunions of the proximal tibia including simple, comminuted, lateral wedge, depression, medial wedge, bicondylar combination of lateral wedge and depression, periprosthetic, and fractures with associated shaft fractures.
- · Simple metaphyseal fractures (Classification AO 41-A2)
- · Multifragmentary metaphyseal fractures (Classification AO 41-A3)
- · Simple bicondylar fractures (Classification AO41-C1, 41-C2)
- Multifragmentary bicondylar fractures (Classification AO 41-C3)
- · Simple joint, simple metaphyseal fractures (Classification AO 41-C1)
- · Diaphisary fractures (Classification AO 42A and 42B)
AFFINITY Proximal Tibia System consists of anatomical plates and screws for the placement of the proximal tibial condyles, improving the restoration of the original structure. Similarly, in combination with the variable angle technique, it allows for the placement of screws in different configurations providing appropriate support for the correct healing of fractures.
AFFINITY Proximal Tibia System consists of pre-contoured bone fixation plates and screws. The plates are made from biocompatible commercially pure titanium grade 4 according to ISO 5832-2 and ASTM F67 standard. The screws are made from biocompatible titanium alloy (Ti6Al4V) according to ISO 5832-3 and ASTM F136 standard.
The AFFINITY Proximal Tibia System plates can be fixed with variable angle technique.
The provided text is an FDA 510(k) summary for the AFFINITY Proximal Tibia System. This document describes a medical device (a bone plate and screw system for tibial fractures) and its claim of substantial equivalence to existing predicate devices.
Crucially, the document states that "Clinical testing was not necessary for the substantial equivalence determination." This means that the device's acceptance criteria and the proof it meets them do NOT involve human clinical studies or AI-driven performance metrics such as accuracy, sensitivity, specificity, or human-in-the-loop improvements.
Therefore, I cannot fulfill the request to describe the acceptance criteria and study that proves the device meets the acceptance criteria in the context of AI/machine learning performance. The provided text details non-clinical performance testing for a medical implant, focusing on mechanical, biocompatibility, and sterilization aspects.
Here's why I cannot provide the requested information based on the given text:
- No AI or algorithm present: There is no mention of any AI component, algorithm, or software in the device description or its testing.
- No clinical performance data (e.g., accuracy, sensitivity): The document explicitly states "Clinical testing was not necessary." The performance testing described is mechanical (bend, torsion, pullout strength), biocompatibility, and sterilization, which are relevant for physical implants, not AI diagnostic or assistive devices.
- No human reader studies (MRMC): Since no AI is involved, there are no studies assessing how human readers improve with AI assistance.
- No ground truth establishment for diagnostic performance: The "ground truth" mentioned in your prompt (expert consensus, pathology, outcomes data) is typically for validating diagnostic or classification algorithms. For a bone plate, ground truth relates to its mechanical integrity and biological safety, which were assessed through laboratory tests.
In summary, the provided document describes the regulatory approval of a physical medical device (a bone plate and screw system) through established non-clinical testing and comparison to predicates, not the validation of an AI/ML medical device.
If you have a document describing an AI/ML medical device's performance study, I would be happy to analyze it according to your detailed criteria.
Ask a specific question about this device
(336 days)
The Affinity Fusion oxygenator with integrated arterial filter and Cortiva bioactive surface is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass (CPB) procedures up to 6 hours in duration. The Affinity Fusion oxygenator with integrated arterial filter and Cortiva bioactive surface is designed to filter from the circuit microemboli larger than the specified micron size for periods up to 6 hours during CPB surgery.
The Affinity Fusion cardiotomy/venous reservoir with Balance biosurface is intended to be used in an extracorporeal perfusion circuit to collect venous and cardiotomy suctioned blood during routine cardiopulmonary procedures up to 6 hours in duration. The CVR is also intended for use during vacuum-assisted venous drainage (VAVD) procedures. The Affinity Fusion cardiotomy/venous reservoir with Balance biosurface is also intended for use after open heart surgery to collect autologous blood from the chest and to aseptically return the blood to the patient for blood volume replacement.
The Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass procedures up to 6 hours in duration. The Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface is designed to filter from the circuit microemboli larger than the specified micron size for periods up to six hours during cardiopulmonary bypass surgery.
The Affinity Fusion Cardiotomy/Venous Reservoir with Balance Biosurface is intended to be used in an extracorporeal perfusion circuit to collect venous and cardiotomy suctioned blood during routine cardiopulmonary procedures up to 6 hours in duration. The CVR is also intended for use during vacuum assisted venous drainage (VAVD) procedures. The Affinity Fusion Cardiotomy/Venous Reservoir with Balance Biosurface is also intended for use after open heart surgery to collect autologous blood from the chest and to aseptically return the blood to the patient for blood volume replacement.
The Affinity Pixie hollow fiber oxygenator with Cortiva bioactive surface are indicated for use for neonate, infant, and small pediatric patients undergoing cardiopulmonary bypass (CPB) procedures requiring a blood flow rate up to 2.0 L/min. The Affinity Pixie hollow fiber oxygenator is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine CPB procedures up to 6 hours in duration.
The Affinity Pixie hollow fiber oxygenator and cardiotomy/venous reservoir with Cortiva bioactive surface are indicated for use for neonate, infant, and small pediatric patients undergoing cardiopulmonary bypass (CPB) procedures requiring a blood flow rate up to 2.0 L/min. The Affinity Pixie hollow fiber oxygenator is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine CPB procedures up to 6 hours in duration. The Affinity Pixie cardiotomy/venous reservoir is intended to be used in an extracorporeal perfusion circuit to collect venous and cardiotomy suctioned blood during routine cardiopulmonary procedures up to 6 hours in duration. The CVR is also intended for use during vacuum-assisted venous drainage (VAVD) procedures.
The Affinity Pixie Hollow Fiber Oxygenator with Balance Biosurface are indicated for use for neonate, infant, and small pediatric patients undergoing cardiopulmonary bypass procedures requiring a blood flow rate up to 2.0 L/min. The Affinity Pixie Hollow Fiber Oxygenator is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass procedures up to 6 hours in duration.
The Affinity Pixie Hollow Fiber Oxygenator and Cardiotomy/Venous Reservoir with Balance Biosurface are indicated for use for neonate, infant, and small pediatric patients undergoing cardiopulmonary bypass procedures requiring a blood flow rate up to 2.0 L/ min. The Affinity Pixie Hollow Fiber Oxygenator is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass procedures up to 6 hours in duration. The Affinity Pixie Cardiotomy/Venous Reservoir is intended to be used in an extracorporeal perfusion circuit to collect venous and cardiotomy suctioned blood during routine cardiopulmonary procedures up to 6 hours in duration. The CVR is also intended for use during vacuum assisted venous drainage (VAVD) procedures.
The Affinity Fusion Oxygenator is intended to be used in an extracorporeal perfusion blood circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass procedures up to 6 hours in duration. The Affinity Fusion Oxygenator contains both an integrated arterial filter and integrated heat exchanger. The Affinity Fusion Oxygenator is a microporous, hollow-fiber, gas-exchange devices available with Cortiva BioActive Surface or Balance Biosurface bonded to the blood contacting surface of the device. The integrated arterial filter is designed to filter from the circuit microemboli larger than the specified micron size from the circuit for periods up to six hours during cardiopulmonary bypass surgery. Some models of the Affinity Fusion Oxygenator are packaged with an Affinity Fusion Cardiotomy/Venous Reservoir (CVR) with Balance Biosurface which is designed to be an integral part of a cardiopulmonary bypass circuit for use during cardiac surgery. The Affinity Fusion CVR is designed to collect venous and cardiotomy suctioned blood during routine cardiopulmonary procedures up to six (6) hours in duration. Additionally, the Affinity Fusion CVR may be used during vacuum assisted venous drainage (VAVD) procedures and collect autologous blood from the chest and to aseptically return the blood to the patient for blood volume replacement during open heart surgery.
The Affinity Pixie Oxygenator is intended to be used in an extracorporeal perfusion blood circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass (CPB) procedures up to 6 hours in duration. The Affinity Pixie hollow fiber oxygenator is a single-use, microporous, hollow-fiber, gas exchange device with plasma-resistant fiber and integrated heat exchanger. The primary blood-contacting surfaces of the oxygenator are coated with either Balance Biosurface or Cortiva BioActive Surface. Some models are packaged with an Affinity Pixie Cardiotomy/Venous Reservoir (CVR) which are designed to be an integral part of a cardiopulmonary bypass circuit for use during cardiac surgery. The Pixie CVR associated with the Balance or Cortiva coated Oxygenators are also Balance or Cortiva coated respectively. The Affinity Pixie CVR is designed to collect venous and cardiotomy suctioned blood during routine cardiopulmonary procedures up to six (6) hours in duration. Additionally, the Affinity Pixie CVR may be used during vacuum assisted venous drainage (VAVD) procedures. All are provided sterile, nonpyrogenic, and for single use.
I am sorry, but the provided text does not contain information about acceptance criteria or a study proving device performance in the context of an AI/ML device. The document is a 510(k) premarket notification for a medical device (cardiopulmonary bypass oxygenators) and focuses on the device description, indications for use, and comparison to predicate devices, primarily regarding a material modification. It discusses the device's function (oxygenation, CO2 removal, temperature control, filtering emboli, blood collection) but does not include any performance metrics for AI or ML algorithms, sample sizes for test sets, expert ground truth adjudication, or comparative effectiveness studies involving AI. Therefore, I cannot generate the requested information based on the given input.
Ask a specific question about this device
(86 days)
AFFINITY is a software application used to process, display, and manage nuclear medicine and other medical imaging data transferred from other workstations, PACS or acquisition stations. The information acquired from viewing the images is used, in conjunction with other patient related data, for diagnosis and monitoring of disease.
Warning! This application is not intended to replace visual assessment of tumors. The application is to provide pre-selection of lesions for visual confirmation and to provide consistency and reproducibility when assessing tumor response to treatment
Affinity is a software application used to process, display, analyse and manage nuclear medicine and other medical imaging data transferred from other workstations, PACS or acquisition stations. The information acquired from viewing the images is used, in conjunction with other patient related data, for diagnosis and monitoring of disease.
This application is not intended to replace visual assessment of tumors. The primary purpose of the application is to provide pre-selection of lesions for visual confirmation and to provide consistency and reproducibility when assessing tumor response to treatment.
Affinity can process data as whole body or constrained field of view (e.g. abdomen, brain) where scanning was performed with any of the following modalities PET/CT, MR or tomographic reconstructed SPECT/CT. The studies are read in DICOM format and if there are studies with different modalities, on the same patient, a co-registration and alignment is performed.
Software output - After the studies have been processed, these are presented visually in 2D or 3D for the user who can use tools to quantify the significant parameters SUV, SUVR, SUVbsa, SUVIbm, SUVbw, SUV Peak, SUV Mean, TLG and MTV.
Here's a breakdown of the acceptance criteria and study information for the Affinity device, based on the provided document:
Acceptance Criteria and Device Performance
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Split tool intra/inter-variability (quantitative values for tumor assessment) | Variability was 0, except for in the 14-17th decimal place (indicating high reproducibility). |
| Hottest Connected SUV peak value accuracy (comparison to manual calculation) | Identical results across all 13 tested scenarios. |
Study Details
1. Sample Sizes and Data Provenance
- Test Set (for Split tool intra/inter-variability): 13 subjects (7 males, 6 females) who received F18FDG-PET/CT studies.
- Data Provenance: Not explicitly stated (e.g., country of origin). The study states "13 subjects... that received a F18FDG-PET/CT study," implying prospective data collection for this specific validation, but this is not explicitly confirmed.
- Test Set (for Hottest Connected SUV peak value accuracy): A single synthetic software-generated study containing 8 voxels.
- Data Provenance: Synthetic (software-generated).
2. Number of Experts and their Qualifications for Ground Truth
- For Split tool intra/inter-variability: No experts were used to establish ground truth in the traditional sense. The validation focused on the reproducibility of the tool itself when used by validators.
- For Hottest Connected SUV peak value accuracy: The "ground truth" was established by manual calculation. No specific number or qualifications of human experts were mentioned for this manual calculation in the provided text.
3. Adjudication Method for the Test Set
- For Split tool intra/inter-variability: Two validators repeated the process 5 times each. The variability between these repeat measurements and between the two validators was assessed. This is not a classic adjudication method for ground truth, but rather a reproducibility assessment.
- For Hottest Connected SUV peak value accuracy: Direct comparison to manual calculation. No adjudication method involving multiple human readers was described.
4. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- No, an MRMC comparative effectiveness study involving human readers improving with AI vs. without AI assistance was not conducted or reported in the provided text. The studies focused on the performance characteristics of specific software features.
5. Standalone Performance Study
- Split tool intra/inter-variability: Yes, the study assesses the standalone reproducibility of the "threshold/split tool" by having two validators use the tool on patient data.
- Hottest Connected SUV peak value accuracy: Yes, the study assesses the standalone accuracy of the "Hottest Connected SUV peak value" by comparing its output to manual calculations.
6. Type of Ground Truth Used
- For Split tool intra/inter-variability: The "ground truth" for this test was the assumption that the tool should produce consistent quantitative values. The study measured the reproducibility of these measurements rather than comparing them to a separate, external ground truth like pathology. The process involved defining regions based on a predefined threshold, splitting them, and then deleting specific anatomical regions (brain, heart, urinary bladder) and ensuring pathological areas connected to non-pathological ones were deleted based on clinical importance. This implies an implicit clinical understanding, but not a formally established "ground truth" for each specific lesion.
- For Hottest Connected SUV peak value accuracy: Manual calculation on a synthetic dataset. This serves as the "truth" against which the algorithm's output is compared.
7. Sample Size for the Training Set
- The document does not specify a sample size for the training set. The descriptions provided relate to verification and validation testing, not model training.
8. How the Ground Truth for the Training Set Was Established
- Since no information on a training set was provided, the method for establishing its ground truth is also not available in the document.
Ask a specific question about this device
(29 days)
The Affinity CP centrifugal blood pump is used to pump blood through the extracorporeal bypass circuit for extracorporeal circulatory support for periods appropriate to cardiopulmonary bypass (up to 6 hours). It is also indicated for use in extracorporeal support systems (for periods up to 6 hours) not requiring complete cardiopulmonary bypass (e.g., valvuloplasty, circulatory support during mitral valve reoperation, surgery of the vena cava or aorta, liver transplants). The Affinity CP centrifugal blood pump is driven by the external drive motor or the emergency handcrank. The Affinity CP centrifugal blood pump is intended for use with pump speed controllers approved by Medtronic or may be used with the Stöckert and Sorin centrifygal pump systems or the Sams and Terumo centrifygal systems by attaching the Affinity CP adapter.
Not Found
This document, K201068, is an FDA 510(k) premarket notification for the Medtronic Affinity CP Centrifugal Blood Pump, including models with Balance Biosurface and Cortiva BioActive Surface. It does not contain information about acceptance criteria or a study proving the device meets the criteria in the context of an AI/ML medical device.
The information provided describes:
- The FDA's substantial equivalence determination for the pump.
- The general controls and additional regulations applicable to the device.
- The Indications for Use for the various models of the blood pump.
Therefore, I cannot provide details on the specific points requested (acceptance criteria table, sample sizes, expert ground truth, MRMC study, standalone performance, ground truth types, training set details) because this document does not pertain to the performance evaluation of an AI/ML-driven device. It is a clearance letter for a physical medical device (blood pump).
Ask a specific question about this device
(93 days)
AFFINITY is a software application used to process, display, and manage nuclear medicine and other medical imaging data transferred from other workstations, PACS or acquisition stations. The information acquired from viewing the images is used, in conjunction with other patient related data, for diagnosis and monitoring of disease.
The Affinity v1.0 is a Viewer that will be the first Hermes application released on the new development platform Affinity.
The application provides 2D and 3D visualization and processing of medical images in Digital Imaging and Communications in Medicine (DICOM) format from different modalities, such as PET/CT, MR and tomographic reconstructed SPECT from SPECT/CT. Affinity supports coregistration, with the exception of 2D images, and fusion of multiple time points with studies in the same frame of reference, different tracers, and modalities.
Affinity is developed with Microsoft Visual Studio on the .NET framework environment and designed for high throughput clinical scenarios with fast image loading and configurable workflows and layouts. In the design, emphasis has been placed on ease of use, where the user can easily access tools for 3D ROI and uptake analysis. The application supports pre-selection and automatic detection of all uptake areas within the body above a certain threshold level, by using threshold region tool. Where the user can define a threshold for any modality and unit within that modality to include all pixels of the study in a region or islands of regions.
Based on selected regions, quantification of the following parameters can be done SUV, SUVR, SUVbsa, SUVlbm, SUVbw, SUV Peak, SUV Mean, TLG and MTV.
Here's a summary of the acceptance criteria and the study conducted for the Affinity device based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance
Ask a specific question about this device
(119 days)
The Affinity NT oxygenator with Balance biosurface is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass (CPB) procedures up to 6 hours in duration.
The uncoated Affinity NT cardiotomy/venous reservoir (CVR) is intended to be used in an extracorporeal perfusion circuit to collect venous and cardiotomy suctioned blood during routine CPB procedures up to 6 hours in duration.
Not Found
I am sorry, but the provided text only contains an FDA clearance letter and an "Indications for Use" section for the Affinity NT Oxygenator with Balance Biosurface. It does not include any information about acceptance criteria for device performance or studies proving that the device meets such criteria. Therefore, I cannot generate the requested table and study details based on the given input.
Ask a specific question about this device
Page 1 of 5