510(k) Data Aggregation

    K Number
    K242190

    Validate with FDA (Live)

    Date Cleared
    2025-03-05

    (223 days)

    Product Code
    Regulation Number
    862.1205
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Access Cortisol assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of cortisol levels in human serum, plasma (heparin, EDTA) and urine using the Access Immunoassay Systems. A cortisol (hydrocortisone and hydroxycorticosterone) test system is a device intended to measure the cortisol hormones secreted by the adrenal gland in serum, plasma and urine. Measurements of cortisol are used in the diagnosis and treatment of disorders of the adrenal gland.

    The DxC 500i Clinical Analyzer combines the DxC 500 AU Clinical Chemistry Analyzer and the Access 2 Immunoassay System into a single instrument presentation. The system is for in vitro diagnostic use only.

    The chemistry module of the DxC 500i Clinical Analyzer is an automated chemistry analyzer that measures analytes in samples, in combination with appropriate reagents, calibrators, quality control (OC) material and other accessories. The immunoassay module of the DxC 500i Clinical Analyzer is an in-vitro diagnostic device used for the quantitative, semiquantitative, or qualitative determination of various analyte concentrations found in human body fluids.

    Device Description

    The Access Cortisol assay is a competitive binding immuno-enzymatic assay designed for use on Beckman Coulter's Access immunoassay analyzers in a clinical laboratory setting.

    The DxC 500i Clinical Analyzer is an integrated chemistry-immunoassay work cell that combines Beckman Coulter's DxC 500 AU Clinical Chemistry Analyzer and the Access 2 Immunoassay System into a single instrument presentation. The DxC 500i instrument has a single user interface and common point of entry for sample racks; the sample handling unit operates as a parallel processor and sample manager for both sides of the instrument. The DxC 500i operates in conjunction with the existing reagents, calibrators, controls, and system solutions for the AU and Access instrument families.

    AI/ML Overview

    The provided text describes the Beckman Coulter Access Cortisol assay on the DxC 500i Clinical Analyzer and its comparison to a predicate device. Here's a breakdown of the acceptance criteria and study information:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria CategorySpecific Acceptance CriteriaReported Device Performance
    Method ComparisonSlope criteria of 1.00 ± 0.12 (using Weighted Deming regression analysis when compared to predicate device)Serum: Slope = 0.974 (95% CI: 0.952 - 0.996) Urine: Slope = 1.002 (95% CI: 0.976 - 1.029)
    LinearityLinear throughout the analytical measuring range.Determined to be linear throughout the analytical measuring range (2.3 - 60.0 µg/dL).
    Imprecision (Repeatability & Total)Allowable imprecision of <12% at approximately 5 µg/dL and <10% for higher concentrations.Control L1 (approx 3 µg/dL): Instrument 1: Repeatability CV = 4.8%, Total CV = 6.4% Instrument 2: Repeatability CV = 5.6%, Total CV = 7.2% Control L2 (approx 17.7 µg/dL): Instrument 1: Repeatability CV = 3.2%, Total CV = 3.6% Instrument 2: Repeatability CV = 2.8%, Total CV = 4.1% Control L3 (approx 27 µg/dL): Instrument 1: Repeatability CV = 3.0%, Total CV = 3.7% Instrument 2: Repeatability CV = 2.9%, Total CV = 3.6% Serum Pool (approx 47-49 µg/dL): Instrument 1: Repeatability CV = 2.7%, Total CV = 3.8% Instrument 2: Repeatability CV = 3.0%, Total CV = 5.0%
    Detection CapabilityNot explicitly stated as acceptance criteria, but results are provided for LoB, LoD, and LoQ.Limit of Blank (LoB): ≤ 0.4 µg/dL Limit of Detection (LoD): ≤ 0.4 µg/dL Limit of Quantitation (LoQ) ≤ 20% within-lab CV: ≤ 0.8 µg/dL

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Method Comparison:
      • Serum: N = 104
      • Urine: N = 115
    • Imprecision: For each of the two instruments and four control levels/serum pools, N = 80 measurements were performed (this implies 20 days x 2 replicates x 2 runs, or similar, over the course of the study).
    • Provenance: The document does not specify the country of origin of the data or whether the studies were retrospective or prospective. It does refer to "patient samples" for the method comparison, suggesting real-world samples.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable as the device is an in-vitro diagnostic test for cortisol levels, not an imaging device requiring expert interpretation of results for ground truth. The "ground truth" for comparison and performance evaluation is generally established by the predicate device's measurements and reference methods.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable for this type of in-vitro diagnostic device. Adjudication methods like 2+1 or 3+1 are typically used in studies involving subjective human interpretation (e.g., radiology reads) where discrepancies between assessors need to be resolved. For quantitative laboratory tests, "ground truth" is typically established by comparing against established reference methods or predicate devices.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable as the device is an in-vitro diagnostic test for cortisol levels, not an AI-assisted diagnostic tool for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This concept is not directly applicable in the same way it would be for an AI algorithm. The Access Cortisol assay on the DxC 500i Clinical Analyzer is an automated in-vitro diagnostic system. The "standalone" performance is essentially what is being evaluated in the method comparison, linearity, imprecision, and detection capability studies, where the system itself generates the quantitative results without subjective human interpretation as part of the core measurement. Human operators are involved in running the instrument and interpreting the numerical results, but the measurement itself is automated.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" in these studies is established by:

    • Predicate Device Measurements: For the method comparison study, the Access Cortisol assay run on the predicate Access 2 Immunoassay System serves as the comparator or "reference" for evaluating the candidate device (Access Cortisol assay on the DxC 500i Clinical Analyzer).
    • Reference Materials/Established Protocols: For linearity, imprecision, and detection capability, the ground truth is often tied to known concentrations in control materials, spiked samples, and established analytical performance specifications derived from CLSI guidelines. The calibrators use "Human serum containing cortisol (purified chemical compound) at levels of 0 and approximately 2, 5, 10, 25, and 60 µg/dL," with traceability to "USP Reference Material."

    8. The sample size for the training set

    This information is not provided and is likely not explicitly documented in the submission for an IVD device of this nature. The instrument doesn't explicitly 'train' in the machine learning sense on a dataset. Its operational parameters are set during its development and validated through studies like those described.

    9. How the ground truth for the training set was established

    As there's no explicitly mentioned "training set" in the context of machine learning, this question is not directly applicable. The device's operational characteristics and performance specifications are established through rigorous analytical verification and validation using known standards, controls, and patient samples, rather than a "training set" in the AI sense.

    Ask a Question

    Ask a specific question about this device

    K Number
    K223038

    Validate with FDA (Live)

    Device Name
    Access Cortisol
    Date Cleared
    2023-02-08

    (132 days)

    Product Code
    Regulation Number
    862.1205
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Access Cortisol assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of cortisol levels in human serum, plasma (heparin, EDTA) and urine using the Access Immunoassay Systems.

    A cortisol (hydrocortisone and hydroxycorticosterone) test system is a device intended to measure the cortisol hormones secreted by the adrenal gland in serum, plasma and urine. Measurements of cortisol are used in the diagnosis and treatment of disorders of the adrenal gland.

    Device Description

    The Access Cortisol assay is a competitive binding immunoenzymatic assay. The Access Cortisol reagent kit is in a liquid ready-to-use format designed for optimal performance on Beckman Coulter's immunoassay analyzers. Each reagent kit contains two reagent packs. Other items needed to run the assay include Calibrators, substrate, and wash buffer. The Access Cortisol assay reagent pack, Access Cortisol assay calibrators, along with the UniCel Dxl wash buffer II are designed for use with the Dxl 9000 Access Immunoassay Analyzer in a clinical laboratory setting.

    AI/ML Overview

    The provided text is a 510(k) Summary for the Beckman Coulter Access Cortisol Assay, intended for in vitro diagnostic use. It describes the device, its intended use, and comparative studies to demonstrate substantial equivalence to a predicate device.

    Key points regarding acceptance criteria and study data fulfillment:

    This document is for an in vitro diagnostic (IVD) device, not an AI/ML-based device for image analysis. Therefore, many of the requested criteria (e.g., number of experts for ground truth, MRMC study, human-in-the-loop performance, training set details) are not applicable to this type of medical device submission. The acceptance criteria and studies for IVD devices focus on analytical performance parameters.

    Here's an analysis of the existing information based on the request:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Performance ParameterAcceptance Criteria (Implied/Stated)Reported Device Performance
    Method ComparisonR2 ≥ 0.90, Slope 1.00 ± 0.12Correlation Coefficient (R2): 1.00 (meets criteria) Slope: 1.01 (falls within 0.88 - 1.12 range, meets criteria) Slope 95% CI: 0.99 - 1.03 Intercept: -0.20 Intercept 95% CI: -0.41 - 0.056 (N=116 samples, concentration range 1.6 - 59 ug/dL)
    ImprecisionWithin-laboratory (total) %CV: ≤ 9.3% for levels > 5.0 ug/dL; SD ≤ 0.1 for levels ≤ 5.0 ug/dL (based on reported performance meeting target, specific criteria not explicitly stated but implied by acceptable results)Within-laboratory (total) %CV: Sample 1 (0.90 ug/dL): 13.1% (SD 0.1) - Note: This sample has a higher %CV than 9.3% for a low concentration, but the SD of 0.1 meets the low concentration criteria. Sample 2 (5.7 ug/dL): 7.1% Sample 3 (19 ug/dL): 8.6% Sample 4 (29 ug/dL): 9.3% Sample 5 (49 ug/dL): 3.9% All other samples meet the stated criteria for %CV or SD. (80 replicates per sample across 20 days, 3 analyzers, 3 reagent lots, 3 calibrator lots)
    LinearityNon-linearity within ± 1 ug/dL for values ≤ 5.0 ug/dL and ± 20% for values > 5.0 ug/dLMeets acceptance criteria, linear across 0.8 - 60 ug/dL.
    Limit of Blank (LoB)0.4 ug/dLAssay designed to meet claimed LoB of 0.4 ug/dL.
    Limit of Detection (LoD)0.4 ug/dLAssay designed to meet claimed LoD of 0.4 ug/dL.
    Limit of Quantitation (LoQ)0.8 ug/dL (based on 20% CV)LoQ designed to meet claimed LoQ of 0.8 ug/dL.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Method Comparison: N=116 samples. Data provenance is not specified (e.g., country of origin, retrospective/prospective), but implied to be from an internal site.
    • Imprecision: 5 serum samples were used, with 80 replicates per sample (total 400 measurements). Data provenance is from "one internal site."
    • Linearity, LoB/LoD, LoQ: Sample sizes are not explicitly stated for these studies, but they are performed as verification studies following CLSI guidelines. Data provenance is implied to be from an internal site.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    • Not Applicable. This is an IVD device for quantitative determination of a biomarker (cortisol) in patient samples. The "ground truth" is established by the analytical method itself and its traceability to a reference material (USP Reference Material for cortisol). There are no human experts "establishing ground truth" in the way it would be for an AI-based imaging device interpreting images.

    4. Adjudication Method for the Test Set:

    • Not Applicable. As per point 3, there is no expert adjudication for an IVD assay's analytical performance.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

    • No. This is an IVD device measuring a biomarker, not an AI/ML-based device assisting human readers in interpreting medical images. MRMC studies are not relevant for this type of device.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Yes, implicitly. The analytical performance studies (Method Comparison, Imprecision, Linearity, LoB/LoD, LoQ) evaluate the device (Access Cortisol Assay on Dxl 9000 Access Immunoassay Analyzer) in a standalone capacity, i.e., its ability to accurately and precisely measure cortisol levels. There is no "human-in-the-loop" aspect to the primary function of this diagnostic assay.

    7. The Type of Ground Truth Used:

    • The ground truth is established by:
      • Reference Method/Predicate Device: For method comparison, the predicate Access Cortisol assay on the Access Immunoassay System serves as the comparative "reference."
      • Traceability to Reference Material: The calibrators and the assay itself are traceable to USP Reference Material for cortisol. This is the ultimate "ground truth" for the accuracy of the cortisol measurements.
      • Defined Concentrations/Spiked Samples: For linearity and LoB/LoD/LoQ studies, samples with known or spiked concentrations are used.

    8. The Sample Size for the Training Set:

    • Not Applicable. This is not an AI/ML device that requires a "training set" in the context of machine learning. The device is a chemiluminescent immunoassay; its analytical characteristics are determined by its chemical and biological components and the instrument's engineering.

    9. How the Ground Truth for the Training Set was Established:

    • Not Applicable. As per point 8, there is no "training set" or corresponding "ground truth" establishment in the AI/ML sense for this device.