510(k) Data Aggregation

    K Number
    K980839

    Validate with FDA (Live)

    Device Name
    AVOCET PT
    Date Cleared
    1998-09-30

    (210 days)

    Product Code
    Regulation Number
    864.5425
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AvocetpT is an in vitro diagnostic system that provides a quantitative prothrombin time result, expressed as an International Normalized Ratio (INR). It is intended for use by health care professionals in monitoring patients who are on warfarin-type (coumarin) anticoagulation therapy.

    Device Description

    The Avocetpr is a membrane-based, dry-reagent system for use with fresh capillary or venous whole blood, and citrated venous whole blood or citrated plasma. The system uses an membrane to separate plasma from red cells. The membrane contains calcium and thromboplastin, and permits the reactions of the complete extrinsic pathway to occur with minimal distortion from membrane surface interactions. Thrombin generation is monitored optically using a rhodamine-110-based fluorescent thrombin substrate. Fluorescence kinetics are analyzed to produce a prothrombin-time-equivalent parameter that is converted to an international normalized ratio (INR) value.

    AI/ML Overview

    The provided text describes the AvocetpT System, a device for measuring Prothrombin Time (INR). However, it does not contain explicit acceptance criteria in the format of a table with specific thresholds or targets (e.g., sensitivity > X%, specificity > Y%).

    Instead, the acceptance criteria are implicitly stated through the claim of "substantial equivalence" to a predicate device and the outcomes of a performance study.

    Here's an attempt to extract and infer the requested information based on the provided text:


    AvocetpT System Acceptance Criteria and Study Details

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategoryReported Device Performance (as stated in document)
    AccuracyEquivalent to CoaguChek™ and a reference method (r > 0.95)
    PrecisionAcceptable
    LinearityAcceptable
    Interfering SubstancesResults are reflected in the product labeling
    HematocritResults are reflected in the product labeling
    RBC AbnormalitiesResults are reflected in the product labeling
    Factor DeficienciesResults are reflected in the product labeling

    Note: The document does not provide specific numerical thresholds for "acceptable" beyond the r > 0.95 for accuracy.

    2. Sample Size Used for the Test Set and Data Provenance

    The document mentions "field studies" for accuracy comparison.

    • Sample Size: Not explicitly stated.
    • Data Provenance: Not explicitly stated (e.g., country of origin, retrospective/prospective). However, "field studies" generally imply prospective data collection in a clinical setting.

    3. Number of Experts and Qualifications for Ground Truth Establishment

    • Number of Experts: Not explicitly stated.
    • Qualifications of Experts: Not explicitly stated. The comparison is made against a "reference method" and the predicate device (CoaguChek™), implying that the ground truth is established by these existing, presumably validated, methodologies, not necessarily by human expert interpretation of device raw output.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not applicable or not explicitly stated. The study focuses on comparing the AvocetpT System's measurements against a reference method and a predicate device. There is no indication of human interpretation requiring adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • MRMC Study: No. This device is an in vitro diagnostic system that provides quantitative numerical results (INR). It does not involve human readers interpreting images or other data that would necessitate an MRMC study or AI-assistance in interpretation.

    6. Standalone Performance Study

    • Standalone Performance: Yes, implicitly. The "accuracy of the AvocetpT was compared to the CoaguChek and a reference method in field studies." This indicates that the AvocetpT System's performance was evaluated as a standalone device without human-in-the-loop assistance in obtaining the INR output itself.

    7. Type of Ground Truth Used

    • Type of Ground Truth: The ground truth for the performance evaluation appears to be established by:
      • Measurements from a "reference method": This typically implies a well-established, highly accurate laboratory method for prothrombin time/INR.
      • Measurements from the predicate device (CoaguChek™): Used for concordance and equivalence comparison.

    8. Sample Size for the Training Set

    • Sample Size: Not applicable / Not stated. As an IVD device producing a direct quantitative measurement, it's unlikely to have a "training set" in the machine learning sense for its primary function. If there were internal calibration or algorithm development, details are not provided.

    9. How the Ground Truth for the Training Set Was Established

    • Ground Truth Establishment: Not applicable / Not stated for a "training set." The device itself provides a direct measurement based on a biochemical reaction and optical detection, rather than a learned prediction from a dataset.