510(k) Data Aggregation

    K Number
    K182420

    Validate with FDA (Live)

    Date Cleared
    2018-12-20

    (106 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This product is intended to provide support to facilitate the placement of guide wires in the coronary and peripheral vasculatures, and can be used to exchange one guide wire for another.

    This product is also intended to assist in the delivery of contrast media into the coronary, peripheral and abdominal vasculatures, and to assist in crossing de novo coronary total occlusions (CTO).

    This device should not be used in neurovasculature.

    Device Description

    The ASAHI Corsair Pro® XS consists of a distal tip and a shaft tube that are inserted into a vascular connector for catheter control and infusion of contrast media. The device has a hydrophilic coating on the outer surface of the shaft tube to provide a smooth transition in blood vessels. The distal tip of the ASAHI Corsair Pro® XS has a tapered shape and is designed to have increased flexibility towards the distal end. The inner lumen of the catheter is PTFE for the purposes of a smooth transition and exchange of guidewires.

    The microcatheter also contains wires to reinforce the distal tip and shaft tube to allow the physician greater control of the device during interventional procedures.

    As compared to the predicate device, the primary change presented in this 510(k) involves a minor change in the dimensions of the tip and catheter shaft and the structure of inner layer (rope coil) of the catheter shaft.

    AI/ML Overview

    I am sorry, but the provided text does not contain the acceptance criteria and study details for an AI-powered medical device.

    The document is a 510(k) Pre-market Notification for a medical device called "ASAHI Corsair Pro® XS", which is described as a percutaneous catheter used to facilitate the placement of guide wires and assist in the delivery of contrast media.

    The document discusses non-clinical testing and biocompatibility testing for this physical device to demonstrate its substantial equivalence to a predicate device. It does not mention any AI component, software, or algorithm that would require the kind of acceptance criteria, study design, and performance metrics typically associated with AI/ML medical devices.

    Therefore, I cannot provide the requested information based on the given context.

    Ask a Question

    Ask a specific question about this device

    K Number
    K161126

    Validate with FDA (Live)

    Date Cleared
    2016-08-25

    (126 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ASAHI Corsair Pro is intended to provide support to facilitate the placement of guide wires in the coronary and peripheral vasculature, and can be used to exchange one guide wire for another. The ASAHI Corsair Pro is also intended to assist in the delivery of contrast media into the coronary, peripheral, and abdominal vasculature. This device should not be used in neurovasculature.

    Device Description

    The ASAHI Corsair Pro consists of a distal tip and a shaft tube that are inserted into a vascular connector for catheter control and infusion of contrast media. The device has a hydrophilic coating on the outer surface of the shaft tube to provide a smooth transition in blood vessels. The distal tip of the Corsair Pro has a tapered shape and is designed to have increased flexibility towards the distal end. The inner lumen of the catheter is PTFE for the purposes of a smooth transition and exchange of guidewires. The microcatheter also contains wires to reinforce the distal tip and shaft tube to allow the physician greater control of the device during interventional procedures.

    AI/ML Overview

    The provided text describes the ASAHI Corsair Pro, a percutaneous catheter, and its substantial equivalence to predicate devices (ASAHI Corsair Microcatheter K151103, K083127). The document outlines the testing performed to demonstrate that the device meets acceptance criteria.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them:

    1. A table of acceptance criteria and the reported device performance

    The document states that "The in vitro bench tests demonstrated that the ASAHI Corsair Pro met all acceptance criteria." However, specific numerical acceptance criteria for each test (e.g., "Force at Break > X N") are not explicitly provided in this document. The table below lists the tests performed and the general statement of performance.

    Test PerformedReported Device Performance
    Appearance/DimensionsMet all acceptance criteria and performed similarly to predicate devices.
    Radio-DetectabilityMet all acceptance criteria and performed similarly to predicate devices.
    Slide DurabilityMet all acceptance criteria and performed similarly to predicate devices.
    Kink ResistanceMet all acceptance criteria and performed similarly to predicate devices.
    Force at BreakMet all acceptance criteria and performed similarly to predicate devices.
    FlexibilityMet all acceptance criteria and performed similarly to predicate devices.
    Torque TransmissionMet all acceptance criteria and performed similarly to predicate devices.
    Liquid Leakage under PressureMet all acceptance criteria and performed similarly to predicate devices.
    Leak and Damage under High Static PressureMet all acceptance criteria and performed similarly to predicate devices.
    Air LeakageMet all acceptance criteria and performed similarly to predicate devices.
    Corrosion ResistanceMet all acceptance criteria and performed similarly to predicate devices.
    Torque DurabilityMet all acceptance criteria and performed similarly to predicate devices.
    BiocompatibilityAll materials used are identical to the predicate device, demonstrating comparable biocompatibility.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not specify the sample sizes used for any of the non-clinical tests.
    The data provenance is from in vitro bench tests conducted by ASAHI Intecc Co., Ltd. The company is based in Nagoya, Aichi, Japan, with various branch offices and research facilities globally. The specific location where these tests were performed is not detailed. The tests are prospective in nature, as they are conducted on the new device to demonstrate its performance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This section is not applicable as the study described is a series of non-clinical, in vitro bench tests. These tests assess physical and mechanical properties of the device against predefined engineering specifications, not clinical outcomes requiring expert interpretation or ground truth establishment in a medical context.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This section is not applicable as the described study involves non-clinical bench testing. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies or studies involving expert review of medical images or data to establish a ground truth or resolve discrepancies.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This section is not applicable. The provided document pertains to the regulatory submission for a medical device (a microcatheter), and the studies described are non-clinical bench tests. It is not an AI-based diagnostic device, and therefore, an MRMC comparative effectiveness study involving human readers and AI assistance is not relevant to this content.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This section is not applicable. The device is a physical medical instrument (microcatheter), not an algorithm or AI system. Therefore, standalone algorithm performance is not relevant.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the non-clinical bench tests, the "ground truth" refers to predefined engineering specifications and performance standards. The device's performance in each test is measured and compared against these established criteria, which are derived from industry standards, regulatory requirements, and the performance characteristics of the predicate device. There is no biological or clinical "ground truth" derived from expert consensus, pathology, or outcomes data in this context.

    8. The sample size for the training set

    This section is not applicable. The document describes non-clinical testing of a physical medical device, not a machine learning model or algorithm that typically requires a training set.

    9. How the ground truth for the training set was established

    This section is not applicable for the same reasons as #8.