510(k) Data Aggregation

    K Number
    K234115

    Validate with FDA (Live)

    Device Name
    APRO 55 Catheter
    Manufacturer
    Date Cleared
    2024-03-15

    (79 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The APRO 55 Catheter is indicated for use in facilitating the insertion and guidance of appropriately sized interventional devices into a selected blood vessel in the peripheral and neurovascular systems. The APRO 55 Catheter is also indicated for use as a conduit for retrieval devices.

    Device Description

    The APRO 55 Catheter is a single-lumen, braid and coil reinforced catheter. The APRO 55 Catheter is designed to facilitate the insertion and guidance of interventional devices into peripheral and neuro vasculature. Using standard catheterization techniques under fluoroscopic guidance, the APRO 55 Catheter is introduced through a guide catheter or guide sheath and over a guidewire into the target vasculature. The distal segment of the catheter shaft has a hydrophilic coating to aid navigation through the vasculature. A radiopaque marker is located at the distal tip of the catheter for visualization under fluoroscopy. The APRO 55 Catheter is provided with an Introducer Sheath.

    AI/ML Overview

    The provided document is a 510(k) summary for a medical device (APRO 55 Catheter) seeking FDA clearance, not a study report detailing acceptance criteria for an AI/ML device. Therefore, it does not contain information typically found in such a study, such as acceptance criteria for machine learning performance metrics (e.g., sensitivity, specificity, AUC), sample sizes for test sets in an AI context, data provenance, expert ground truth establishment, adjudication methods, MRMC studies, standalone algorithm performance, or training set details.

    The document describes non-clinical bench testing for a physical medical catheter, including mechanical, material, and biocompatibility tests. The "acceptance criteria" mentioned are for these engineering and biological tests, not for an AI/ML diagnostic or predictive model.

    Therefore, I cannot fulfill your request for information related to AI/ML acceptance criteria and study details based on the provided text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K232971

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-02-29

    (161 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The APRO 55 Catheter with an aspiration pump and the Alembic Aspiration Tubing is intended for use in the revascularization of patients with acute stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.

    The Alembic Aspiration Tubing is intended to connect the APRO 55 Catheter to the aspiration pump.

    Device Description

    The APRO 55 Catheter is a single-lumen, braid and coil reinforced catheter. The APRO 55 Catheter is designed to remove thrombus from the vasculature using aspiration. The APRO 55 Catheter targets aspiration from a suction pump directly to the thrombus to remove thrombus from an occluded vessel. The APRO 55 Catheter is introduced through a guide catheter or long femoral sheath and into the intracranial vasculature and guided over a neurovascular guidewire under fluoroscopic visualization to the site of the primary occlusion. The distal shaft has a hydrophilic coating to aid navigation through the vasculature. A radiopaque marker is located at the distal end of the catheter for visualization under fluoroscopy. For the aspiration source, the APRO 55 Catheter is used in conjunction with an aspiration pump with pre-specified performance parameters that is connected using the Alembic Aspiration Tubing, along with a legally marketed canister and accessories kit. The APRO 55 Catheter is available in lengths of 125 cm and 137 cm and is provided with an introducer sheath.

    The Alembic Aspiration Tubing connects the APRO 55 Catheter to the aspiration pump. The flow control valve allows control of the aspiration flow using an ON/OFF switch. It is available in one size.

    AI/ML Overview

    Here's a summary based on the provided FDA 510(k) document, focusing on acceptance criteria and study details:

    Device: APRO 55 Catheter and Alembic Aspiration Tubing

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria listed are primarily for non-clinical bench tests. The reported device performance uniformly states that the APRO 55 Catheter (and Alembic Aspiration Tubing where applicable) met the acceptance criteria.

    TestAcceptance CriteriaReported Device Performance
    Visual and Dimensional CharacteristicsCatheter meets the visual and dimensional specifications. Introducer Sheath meets the visual and dimensional specifications.The APRO 55 Catheter met the acceptance criteria. The Introducer Sheath met the acceptance criteria.
    ParticulateCatheter meets the acceptance criteria. Subject device was evaluated with a predicate device under the same test conditions.The APRO 55 Catheter met the acceptance criteria.
    Vacuum IntegrityCatheter is free from collapse and loss of vacuum between aspiration source and catheter tip.The APRO 55 Catheter met the acceptance criteria.
    Kink ResistanceCatheter shaft shall not kink at clinically relevant radii.The APRO 55 Catheter met the acceptance criteria.
    Catheter Hub LeakageCatheter does not leak into hub assembly during aspiration, with methods specified in ISO 10555-1, Annex D.The APRO 55 Catheter met the acceptance criteria.
    Hub CompatibilityHub meets the requirements per EN ISO 80369-7.The APRO 55 Catheter met the acceptance criteria.
    Catheter Torque StrengthCatheter must withstand the minimum required number of rotations without breakage and without kinking compared to legally marketed devices.The APRO 55 Catheter met the acceptance criteria.
    Dynamic Burst PressureNo damage to catheter with dynamic pressure.The APRO 55 Catheter met the acceptance criteria.
    Fluid LeakageCatheter must withstand pressure with methods per ISO 10555-1 Annex C.The APRO 55 Catheter met the acceptance criteria.
    Static BurstCatheter must withstand pressures anticipated for clinical use.The APRO 55 Catheter met the acceptance criteria.
    Tensile Strength of Catheter Hub and ShaftCatheter hub and shaft must meet tensile strength specification.The APRO 55 Catheter met the acceptance criteria.
    Tensile Strength of Catheter TipCatheter tip must meet tip tensile strength specification.The APRO 55 Catheter met the acceptance criteria.
    Delivery and Retrieval ForceCatheter delivery and retrieval force must be acceptable. Forces were compared to a predicate.The APRO 55 Catheter met the acceptance criteria.
    Tip Buckling ForceCatheter tip buckling force must be acceptable. Forces were compared to a predicate.The APRO 55 Catheter met the acceptance criteria.
    Simulated Use including Clot RetrievalWhen used per the Instructions for Use with accessory devices in an anatomical neurovascular model, the Catheter and Aspiration Tubing must meet functionality specifications. Clot retrieval from various locations of the anatomical model was performed using the subject device system.The APRO 55 Catheter and Alembic Aspiration Tubing met the acceptance criteria.

    Biocompatibility Tests:

    TestResultsConclusions
    Sensitization (Guinea Pig Maximization)The APRO 55 Catheter did not elicit a sensitization response.Non-sensitizing
    Irritation/Intracutaneous ReactivityThe APRO 55 Catheter demonstrated no evidence of irritation.Non-Irritant
    Cytotoxicity (MEM Elution, L929 cells)The APRO 55 Catheter did not elicit a cytotoxic response at 24 hours and 48 hours.Non-cytotoxic
    Hemolysis – IndirectThere were no significant differences between the test article extract and negative control article results.Non-hemolytic
    Hemolysis – DirectThere were no differences between the hemolytic index of the test article and the negative control.Non-hemolytic
    Partial Thromboplastin Time (PTT)The average clotting time of the test article was greater than vehicle control and negative control.Acceptable clotting times
    SC5b9 Complement ActivationThe Sc5b9 concentration of the test article was statistically less than the positive control and was not statistically higher than the negative control.Acceptable
    Acute Systemic ToxicityNo weight loss, mortality, or evidence of systemic toxicity from the extract exposure to the mice.Non-toxic
    Material-Mediated PyrogenicityAll individual rabbits for both the test article and negative control showed a total rise in temperature of < 0.5 °C and were determined to be nonpyrogenic.Non-pyrogenic

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state a sample size for the "test set" in terms of patient data. All testing described is non-clinical bench testing and biocompatibility testing.

    • Data Provenance: Not applicable as no human patient data was used for effectiveness evaluation. The studies were laboratory-based bench tests and animal biocompatibility studies.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • This information is not provided as the studies were non-clinical bench tests and biocompatibility testing, which do not typically involve human expert ground truth for performance evaluation in the way imaging AI or diagnostic devices do. Standardized test methods and acceptance criteria are used.

    4. Adjudication Method for the Test Set

    • This information is not provided and is not applicable, as there was no test set requiring human adjudication for performance evaluation.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No, an MRMC comparative effectiveness study was not done. The document explicitly states: "Substantial equivalence was established based on non-clinical performance data. Human clinical data were not deemed necessary."

    6. Standalone (Algorithm only without human-in-the-loop) Performance

    • This is not applicable. The APRO 55 Catheter and Alembic Aspiration Tubing is a physical medical device (catheter for thrombus retrieval) and not an AI algorithm or software-only device. Therefore, standalone algorithm performance is not a relevant concept for this submission.

    7. Type of Ground Truth Used

    • For non-clinical bench testing, the "ground truth" is defined by the specifications, industry standards (e.g., ISO 10555-1, EN ISO 80369-7), and acceptable ranges or comparisons to predicate devices outlined in the acceptance criteria. For example, for "Kink Resistance," the ground truth is "Catheter shaft shall not kink at clinically relevant radii."
    • For biocompatibility testing, the ground truth is established by standardized biological response criteria (e.g., non-sensitizing, non-irritant, non-cytotoxic) defined by ISO 10993-1.

    8. Sample Size for the Training Set

    • This information is not applicable as this is a physical medical device and not an AI/ML device that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    • This information is not applicable as this is a physical medical device and not an AI/ML device.