Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K121265

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-05-22

    (26 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Amphirion™ Plus PTA Catheter is intended to dilate stenosis in the iliac, femoral, ilio-femoral, popliteal, infra-popliteal, and renal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae.

    Device Description

    The Amphirion Plus PTA Catheter (also referred to as Amphirion Plus) is an over-the-wire Percutaneous Transluminal Angioplasty (PTA) catheter consisting of a proximal hub, coaxial body catheter shaft, and a distal dilatation balloon. The Amphirion Plus device consists of a coaxial lumen. The lumen marked "WIRE" is the central lumen of the catheter, which terminates at the distal tip. This lumen is used to pass the catheter over a guidewire with a maximum diameter of 0.014". The lumen marked "BALLOON" is the balloon inflation lumen, which is used to inflate and deflate the balloon. Two radiopaque marker bands are placed under the balloon segment of the catheter shaft to provide visual reference points for balloon positioning within the vessel. The Amphirion Plus PTA Catheter is compatible with guidewires with a maximum diameter of 0.014" and with 4F introducer sheaths. The catheter is provided with a hydrophilic coating and is available in useable catheter lengths of 100cm and 150cm.

    AI/ML Overview

    This submission describes the Amphirion™ Plus PTA Catheter, an over-the-wire Percutaneous Transluminal Angioplasty (PTA) catheter. The device is intended to dilate stenosis in various arteries and treat obstructive lesions in dialysis fistulae. The 510(k) summary indicates that the device is substantially equivalent to predicate devices, with modifications primarily in balloon length, and no new safety or effectiveness issues were observed during testing.

    Here's an analysis of the provided information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission states that a battery of bench and biocompatibility tests were performed, and all pre-determined acceptance criteria were met. However, the specific quantitative acceptance criteria values for each test are not detailed in the provided text. The output primarily indicates "All of the pre-determined acceptance criteria were met and results passed." for each category.

    CharacteristicAcceptance CriteriaReported Device Performance
    Catheter Useable LengthPre-determined criteria (not specified)Met and passed
    Introducer Sheath CompatibilityPre-determined criteria (not specified)Met and passed
    Trackability-Flexibility-Kink testPre-determined criteria (not specified)Met and passed
    Torque StrengthPre-determined criteria (not specified)Met and passed
    Catheter Body Burst PressurePre-determined criteria (not specified)Met and passed
    Tensile StrengthPre-determined criteria (not specified)Met and passed
    Balloon Inflation/Deflation TimePre-determined criteria (not specified)Met and passed
    Balloon PreparationPre-determined criteria (not specified)Met and passed
    Balloon ProfilePre-determined criteria (not specified)Met and passed
    Balloon Working LengthPre-determined criteria (not specified)Met and passed
    Minimum Balloon Burst Strength (RBP)Pre-determined criteria (not specified)Met and passed. RBP of 20 ATM for all balloon sizes, within predicate range of 18-22 ATM.
    Balloon FatiguePre-determined criteria (not specified)Met and passed
    Balloon CompliancePre-determined criteria (not specified)Met and passed
    Biocompatibility Testing:
    ISO MEM Elution Cytotoxicity TestPre-determined criteria (not specified)Met and passed
    ISO Kligman Maximization TestPre-determined criteria (not specified)Met and passed
    ISO Intracutaneous Reactivity StudyPre-determined criteria (not specified)Met and passed
    ISO Acute Systemic Toxicity StudyPre-determined criteria (not specified)Met and passed
    ISO Material Mediated Rabbit Pyrogen StudyPre-determined criteria (not specified)Met and passed
    ASTM Indirect contact HemolysisPre-determined criteria (not specified)Met and passed
    ISO Thrombogenicity StudyPre-determined criteria (not specified)Met and passed
    ISO Indirect contact Complement activationPre-determined criteria (not specified)Met and passed

    2. Sample size used for the test set and the data provenance

    The document does not specify the exact sample sizes for each bench or biocompatibility test. It mentions that "The Amphirion Plus PTA Catheter was thoroughly tested on the bench" and "Biocompatibility testing was conducted on the finished Amphirion Plus PTA Catheter." These are general statements without specific numbers.

    The data provenance is not explicitly stated in terms of country of origin. The study appears to be laboratory-based (bench and biocompatibility testing) and thus would be considered pre-clinical rather than involving human subjects. Therefore, the retrospective or prospective nature of clinical data is not applicable.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable to this submission. The tests performed are engineering and biological evaluations, not clinical studies requiring expert interpretation of medical images or patient outcomes. The "ground truth" for these tests would be the established scientific and engineering standards and methods.

    4. Adjudication method for the test set

    This information is not applicable. This is not a study involving multiple human readers or interpretations needing adjudication. The tests have objective criteria.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable as the device is a PTA catheter, a physical medical device, not an AI or imaging diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable as the device is a physical medical device. Performance is assessed through bench and biocompatibility testing, not an algorithm.

    7. The type of ground truth used

    The "ground truth" for the tests performed can be considered:

    • Bench Testing: Engineering specifications, material properties, and performance standards as defined by relevant FDA guidance documents, ISO, and ASTM standards.
    • Biocompatibility Testing: Established biological safety limits and responses as defined by ANSI/AAMI/ISO10993-1, FDA 21 CFR Part 58 (GLP), and FDA guidance for PTCA catheters.

    8. The sample size for the training set

    This is not applicable. This is not a machine learning or AI device that requires a training set.

    9. How the ground truth for the training set was established

    This is not applicable.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1