Search Results
Found 8 results
510(k) Data Aggregation
(180 days)
Ask a specific question about this device
(137 days)
The AlignRT Plus system is indicated for:
- · Tracking respiratory motion throughout the simulation process to facilitate subsequent 4DCT reconstruction and coaching the patient in breathing techniques required for Deep-Inspiration Breath Hold (DIBH).
- · Verification of patient identity for their radiation treatment session.
- · Positioning and monitoring of patients during radiation delivery, relative to the setup isocenter and/or the prescribed treatment isocenter.
- · Withholding the beam automatically during radiation delivery, as well as gating the patient's respiratory motion.
- · Performing quality assurance on MV, kV imagers, room lasers, and the treatment couch.
- · Visualizing the Cherenkov signal associated with the radiation beam on entry and exit from the patient.
- · Passing and receiving information to/from other systems associated with the radiotherapy treatment.
The AlignRT Plus system (K212583) is a combination of the devices AlignRT InBore, AlignRT Offline, DoseRT, GateCT, GateRT and SimRT.
AlignRT Plus is a video-based three-dimensional (3D) surface imaging system used to monitor the patient's position in 3D before and during radiotherapy treatment. During each treatment session the patient's position is compared to the reference surface and offsets are displayed to the user. The system can be used to track pation during tumour localization in the CT scanner in order to facilitate subsequent 4D CT reconstruction. The system can be used to track the breath hold level consistency throughout the simulation process. The system can also be used to track patient motion during treatment delivery for radiation therapy procedures and hold the beam when the patient is not in position. Use of optional accessories allows the user to:
- monitor the patient's position inside bore-based linacs; and/or —
- to view the Cherenkov light emitted by the radiation beam as it enters and exits the patient's skin; and/or
- verify the patient's identity.
The system is non-invasive, does not require the use of body markers and produces no irradiation during the imaging process. The system mainly consists of advanced software. 3D cameras (≤6 cameras) and calibration tools. Each camera pod monitoring the pation will, however, project a pattern on the patient to acquire a 3D image of the patient.
This 510(k) notification is to obtain clearance for the following changes to the cleared device:
-
- Introduction of new software, version 7.3. This is an updated version of the existing cleared software, to support the:
- a. New Cherenkov imaging feature, called DoseRT.
- b. Updated respiratory gating feature, called Respiratory Module.
-
- Introduction of new hardware:
- a. DoseRT camera to enable Cherenkov imaging.
-
- Update to indications for use to include the new Cherenkov imaging feature and clarify existing indications.
The provided document describes the AlignRT Plus 7.3 system, a surface-guided radiation therapy system. However, it does not contain detailed acceptance criteria and a study dedicated to proving the device meets those criteria in the format requested.
The document focuses on demonstrating substantial equivalence to a predicate device (AlignRT Plus K212583) and a reference device (BeamSite K212606) for new features, primarily Cherenkov imaging and an updated respiratory module. It mentions "Verification tests were performed to ensure that the module works as intended and pass/fail criteria were used to verify requirements. Validation testing was performed using summative evaluation techniques per IEC 62366-1:2015. Verification and validation testing passed in all test cases." but does not provide details of these tests, specific acceptance criteria, or the results.
Therefore, many of the requested fields cannot be filled from the provided text.
Here's a summary of the information that can be extracted or inferred, and what is missing:
1. Table of Acceptance Criteria and Reported Device Performance:
| Feature/Metric | Acceptance Criteria (from document) | Reported Device Performance (from document) |
|---|---|---|
| Positioning Accuracy | Target registration errors (as measured using calibration phantom) < 1mm (0.5mm) for all couch angles. | Target registration errors (as measured using calibration phantom) < 1mm (0.5mm) for all couch angles. |
| Respiratory Tracking Accuracy (relative respiratory position) | ≤0.5mm with a confidence level of ≥95%. | Calculated to an accuracy of ≤0.5mm with a confidence level of ≥95%. |
| Cherenkov Signal Visualization | Visible for treatment beam energies of ≥6MV and ≤18MV and at treatment beam delivery rates between ≥100MU/Min and ≤2400MU/Min. | The location of the radiation dose on the patient surface during treatment can be visualized using the Cherenkov Effect for treatment at treatment beam energies of ≥6MV and ≤18MV and at treatment beam delivery rates of between ≥100MU/Min and ≤2400MU/Min. |
| Surface Displacement Tracking (RMS errors) | < 0.5mm over 10 or more breathing cycles. | < 0.5mm over 10 or more breathing cycles. |
| IEC 62366-1:2015 (Usability/Human Factors) | Device continues to meet FDA guidance on "Applying Human Factors and Usability Engineering to Medical Devices" and IEC 62366-1: 2015. | Validation testing was performed using summative evaluation techniques per IEC 62366-1:2015. Verification and validation testing passed in all test cases. |
| General Electrical Safety Standards | IEC 60601-1 compliant. | IEC 60601-1 compliant. |
| EMC Standards | IEC 60601-1-2 compliant. | IEC 60601-1-2 compliant. |
| Verification and Validation Tests | All tests passed specified pass/fail criteria. | Verification and validation testing passed in all test cases. |
2. Sample size used for the test set and the data provenance:
- Sample Size: Not specified for any of the performance claims. The document states "no clinical investigations were performed."
- Data Provenance: Not specified. It's likely internal testing data given the lack of clinical studies. Retrospective or prospective nature is not mentioned, nor is country of origin.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not specified. The document does not describe the use of experts to establish ground truth for testing.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not specified.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No MRMC comparative effectiveness study was done. The device is a patient positioning and monitoring system, not an AI-assisted diagnostic tool for human readers.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
- The document implies that the device's technical performance characteristics (e.g., positioning accuracy, respiratory tracking) were evaluated as a standalone system against predefined specifications, as it mentions "Verification tests were performed to ensure that the module works as intended and pass/fail criteria were used to verify requirements." However, no specific details like algorithms, metrics, or detailed results are provided.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
- For the positioning accuracy, the ground truth appears to be based on measurements using a "calibration phantom."
- For respiratory tracking, the ground truth for "relative respiratory position" is implied to be a measured value for comparison.
- For Cherenkov visualization, the ground truth is the actual presence and characteristics of the radiation beam.
- No expert consensus, pathology, or outcomes data is explicitly mentioned as ground truth for these technical performance tests.
8. The sample size for the training set:
- Not applicable/Not specified. The document describes a medical device system, not an AI model requiring a distinct training set in the typical sense of machine learning. The "advanced software" may incorporate algorithms, but details about training data, if any, are not present.
9. How the ground truth for the training set was established:
- Not applicable/Not specified (refer to point 8).
Ask a specific question about this device
(86 days)
The AlignRT Plus system is indicated for use to position and monitor patients relative to the prescribed treatment isocentre, and to withhold the beam automatically during radiation delivery.
For cranial treatments, a manual head adjuster is included which can be used in concert with AlignRT Plus to provide fine corrections for pitch, roll and yaw rotations.
AlignRT Plus is also used to track the patient's respiratory pattern for respiratory synchronized image acquisition, and radiation therapy treatment.
Patient contour data can be extracted and exported from the data acquired for the purpose of treatment planning.
AlignRT Plus can be calibrated directly to the treatment beam isocentre and in turn assists in performing quality assurance on MV, kV imagers, room lasers and the treatment couch.
AlignRT Plus is indicated for use during simulation, setup and stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body where radiation is indicated.
The AlignRT Plus system (K203387) is a combination of the devices AlignRT, GateCT and GateRT.
AlignRT Plus is a video-based three-dimensional (3D) surface imaging system, which is used to image the skin surface of a patient in 3D before and during radiotherapy treatment. The system consists of advanced software, a computer workstation, and one, two or three 3D camera units (each camera unit comprising a stereo pair of sensors to allow 3D surface reconstruction). The system is non-invasive, does not require the use of body markers and produces no ionizing irradiation during the imaging process.
AlignRT Plus is also able to perform both respiratory synchronised CT imaging and treatment delivery. In both instances, the system acquires a gated 3D surface model of the patient. User selected points are then tracked in real time in order to provide gating and position monitoring signals.
Real-time imaging and surface matchinq of the patient is possible during both setup and the treatment delivery to determine any patient movement. During treatment delivery, AlignRT Plus is also able to withhold the beam automatically, should the patient move outside user-defined tolerances.
Patient contour data may be extracted from surface data acquired by the system and exported for the purpose of treatment planning by radiotherapy professionals.
AlignRT Plus may be calibrated directly to the treatment beam isocentre using an optional custom designed calibration phantom and image processing software. It can analyse MV and kV digital imaging data acquired by other cleared devices. This in turn assists the user in performing quality assurance on MV, kV imagers, room lasers and the treatment couch.
The AlignRT Plus system includes the optional Head Adjuster for cranial treatments to allow for the manual, fine correction of pitch, roll and yaw in the patient's head position.
Precise isocenter calibration and the optional Head Adjuster provide improved frameless Stereotactic Radiosurgery (SRS). This is provided with the brand name "AlignRT SRS module".
The AlignRT Plus system is also provided under the brand OSMS (Optical Surface Monitoring System). This product is identical to AlignRT.
This 510(k) is requested for a modifications to FDA 510(k) cleared product "AlignRT Plus" (K203387).
This 510(k) is to obtain clearance for the following changes to the cleared device:
- Introduction of new software, SimRT 7.2. This is an updated version of the cleared software, GateCT, with a new user interface similar to the cleared AlignRT 6.2/6.3/6.4 software;
-
- Introduction of new camera pods (hardware). These function in the same way as the currently cleared hardware but are aesthetically different and have upgraded capabilities to support future product development;
-
- Improve the product's accuracy claim for Source-Surface Distancing (SSD) from <2.0mm to <0.2mm.
The provided text is a 510(k) Premarket Notification for the AlignRT Plus system. It details modifications to an already cleared device. The key information regarding acceptance criteria and device performance is found under the "System Performance and Accuracy" section within the "Summary of the Technological Characteristics" table.
Here's a breakdown of the requested information based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance:
| Feature | Acceptance Criteria (Predicate Device) | Reported Device Performance (Modified Device) | Comments |
|---|---|---|---|
| Positioning Accuracy | Target registration errors (as measured using calibration phantom) < 1mm (0.5mm) for all couch angles. | Target registration errors (as measured using calibration phantom) < 1mm (0.5mm) for all couch angles. | The system performance and accuracy have not changed. |
| Respiratory Tracking | Tracks respiratory signal from imaged surface data and sends to CT (4D CT) or to Linac or imaging device (gating). | Tracks respiratory signal from imaged surface data and sends to CT (4D CT) or to Linac or imaging device (gating). | The system performance and accuracy have not changed. |
| Surface Displacements | RMS errors < 0.5mm over 10 or more breathing cycles. | RMS errors < 0.5mm over 10 or more breathing cycles. | The system performance and accuracy have not changed. |
| Source-Surface Distancing (SSD) Accuracy | Not explicitly stated in the predicate column, but implied by the change to the modified device. | < 0.2mm (This is an improvement from a previous claim of < 2.0mm) | This is one of the key improvements being sought with this 510(k). |
2. Sample Size Used for the Test Set and Data Provenance:
The document mentions that the changes were tested with "test method and acceptance criteria as the predicate device," but it does not explicitly state the sample size used for the test set or the data provenance (e.g., country of origin, retrospective/prospective study). The study appears to be a technical verification and validation of changes to functionality and hardware.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts:
The document does not mention the use of experts to establish ground truth for the test set. The performance metrics appear to be based on physical measurements using calibration phantoms and internal system testing.
4. Adjudication Method for the Test Set:
The document does not describe an adjudication method for the test set, as the testing appears to be highly technical and objective (measurements against known standards/phantoms) rather than subjective expert review.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
The document does not indicate that a multi-reader multi-case (MRMC) comparative effectiveness study was done. The device is an objective measurement system for patient positioning and monitoring, not an image interpretation or diagnostic aid that would typically involve human readers. Therefore, an effect size of human readers improving with AI assistance is not applicable in this context.
6. Standalone (Algorithm Only) Performance:
The provided information focuses on the integrated system's performance, as the system consists of hardware (cameras) and software working together to deliver the stated functionalities. The performance metrics (positioning accuracy, surface displacement, SSD accuracy) inherently describe the algorithm's capability within the system context. However, it's not explicitly presented as a separate "algorithm-only" study with distinct metrics from the integrated device.
7. Type of Ground Truth Used:
The ground truth for the performance evaluations (positioning accuracy, surface displacements) appears to be based on physical measurements using a custom-designed calibration phantom. For respiratory tracking, it's based on the system's ability to track "user selected points" and real-time 3D surface models. For the improved SSD accuracy, it's also likely a technical measurement against a known standard.
8. Sample Size for the Training Set:
The document does not specify a training set sample size. The nature of the device (surface imaging and tracking) suggests it may rely on established physics and geometric principles rather than a large, data-driven machine learning model that requires explicit training sets in the same manner as a diagnostic imaging AI. The "software" updates mentioned (SimRT 7.2) are described as an updated version with a new user interface, not necessarily a re-trained AI model from a new dataset.
9. How the Ground Truth for the Training Set Was Established:
As no explicit training set or AI model requiring such a set is detailed, the document does not provide information on how ground truth for a training set was established.
Ask a specific question about this device
(106 days)
The AlignRT Plus system is indicated for use to position and monitor patients relative to the prescribed treatment isocentre, and to withhold the beam automatically during radiation delivery.
For cranial treatments, a manual head adjuster is included which can be used in concert with AlignRT Plus to provide fine corrections for pitch, roll and yaw rotations.
AlignRT Plus is also used to track the patient's respiratory pattern for respiratory synchronized image acquisition, and radiation therapy treatment.
Patient contour data can be extracted and exported from the data acquired for the purpose of treatment planning.
AlignRT Plus can be calibrated directly to the treatment beam isocentre and in turn assists in performing quality assurance on MV, kV imagers, room lasers and the treatment couch.
AlignRT Plus is indicated for use during simulation, setup and stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body where radiation is indicated.
The AlignRT Plus system (K193431) is a combination of the devices AlignRT, GateCT and GateRT. A special 510(K) K193431 was submitted by Vision RT Ltd. in December 2019 and this was cleared by FDA in January 2020 for the changes and improvements made in AlignRT.
AlignRT Plus is a video-based three-dimensional (3D) surface imaging system, which is used to image the skin surface of a patient in 3D before and during radiotherapy treatment. The system consists of advanced software, a computer workstation, and one, two or three 3D camera units (each camera unit comprising a stereo pair of sensors to allow 3D surface reconstruction). The system is non-invasive, does not require the use of body markers and produces no ionizing irradiation during the imaging process.
AlignRT Plus is also able to perform both respiratory synchronised CT imaging and treatment delivery. In both instances, the system acquires a gated 3D surface model of the patient. User selected points are then tracked in real time in order to provide gating and position monitoring signals.
Real-time imaging and surface matching of the patient is possible during both setup and the treatment delivery to determine any patient movement. During treatment delivery, AlignRT Plus is also able to withhold the beam automatically, should the patient move outside user-defined tolerances.
Patient contour data may be extracted from surface data acquired by the system and exported for the purpose of treatment planning by radiotherapy professionals.
AlignRT Plus may be calibrated directly to the treatment beam isocentre using an optional custom designed calibration phantom and image processing software. It can analyse MV and kV digital imaging data acquired by other cleared devices. This in turn assists the user in performing quality assurance on MV, kV imagers, room lasers and the treatment couch.
The AlignRT Plus system includes the optional Head Adjuster for cranial treatments to allow for the manual, fine correction of pitch, roll and yaw in the patient's head position.
Precise isocenter calibration and the optional Head Adjuster provide improved frameless Stereotactic Radiosurgery (SRS). This is provided with the brand name "AlignRT SRS module".
The AlignRT Plus system is also provided under the brand OSMS (Optical Surface Monitoring System). This product is identical to AlignRT.
This 510(k) is requested for a new optional InBore camera and a change in memory size for an existing FDA 510(k) cleared product "AlignRT Plus" (K193431).
The InBore camera is optional hardware for use with bore-based linear accelerators.
The increase in memory size from 32 to 64bit increases the system's processing speed, enabling a frame rate improvement.
The provided text is a 510(k) Premarket Notification for the AlignRT Plus device K203387, which describes modifications to an existing cleared device K193431. The document does not contain a detailed study that proves the device meets specific acceptance criteria with supporting data, an approach often seen for substantial equivalence claims where modifications are minor and performance remains unchanged from a previously cleared predicate. Instead, it relies on demonstrating that the modified device's performance and characteristics are the same as its predicate device.
Therefore, the following information is based on what is stated and what can be inferred from the provided text, primarily focusing on the comparison to the predicate device.
Acceptance Criteria and Reported Device Performance
The "Acceptance Criteria" for this submission are fundamentally defined by the performance of the predicate device (K193431). The "Reported Device Performance" for the modified device (K203387) is stated to be unchanged from that of the predicate.
| Acceptance Criteria (Based on Predicate Device Performance) | Reported Device Performance (Modified Device K203387) |
|---|---|
| Positioning accuracy: Target registration errors (as measured using calibration phantom) < 1mm (0.5mm) for all couch angles. | Positioning accuracy: Target registration errors (as measured using calibration phantom) < 1mm (0.5mm) for all couch angles. (Stated as "not changed") |
| Respiratory tracking: Tracks respiratory signal from imaged surface data and sends to CT (4D CT) or to Linac or imaging device (gating). Surface displacements can be tracked with RMS errors < 0.5mm over 10 or more breathing cycles. | Respiratory tracking: Tracks respiratory signal from imaged surface data and sends to CT (4D CT) or to Linac or imaging device (gating). Surface displacements can be tracked with RMS errors < 0.5mm over 10 or more breathing cycles. (Stated as "not changed") |
Study Details (Based on provided text)
Since the document asserts that the system performance and accuracy have not changed since previous clearances, a new comprehensive performance study against acceptance criteria is not explicitly provided. The submission focuses on demonstrating substantial equivalence through a comparison of the modified device to its predicate, stating that the "changes made in subject device were test method and acceptance criteria as the predicate device and the subject device is substantially equivalent to the predication and/or validation demonstrate that the device is as safe, as effective, and performs as well as or better than the predicate device."
-
Sample size used for the test set and the data provenance: Not explicitly detailed in this document. The document states that "The system performance and accuracy have not changed since previous clearances, including with the additional optional hardware (InBore camera)." This implies that the performance characteristics were established in previous submissions (e.g., K193431) and are considered to apply to the current modified device. No new test set or data provenance is detailed for K203387 for these performance metrics.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not explicitly detailed. The ground truth for the performance metrics (positioning accuracy, respiratory tracking RMS errors) would likely have been established using physical phantoms and measurement references rather than human expert consensus for the quantitative metrics stated.
-
Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable, as no clinical test set requiring human adjudication for diagnostic outcomes is described for this submission. The performance metrics are technical measurements against a known reference.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is a surface-guided radiation therapy system that primarily assists in patient positioning and monitoring, not in diagnostic interpretation by human readers. It does not involve AI in the context of improving human reader diagnostic performance.
-
If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: The "System Performance and Accuracy" section indirectly refers to standalone performance:
- "Positioning accuracy: Target registration errors (as measured using calibration phantom) < 1mm (0.5mm) for all couch angles." This is a measurement of the system's inherent accuracy using a physical phantom, which is a standalone assessment.
- "Respiratory tracking: Tracks respiratory signal from imaged surface data... Surface displacements can be tracked with RMS errors < 0.5mm over 10 or more breathing cycles." This also describes the device's intrinsic capability to track surface displacements, tested against a measurable reference.
The document states these performances "have not changed since previous clearances."
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.): For the performance characteristics mentioned, the ground truth is based on physical reference measurements using a calibration phantom. For instance, "Target registration errors (as measured using calibration phantom)" is directly specified.
-
The sample size for the training set: Not applicable based on the information provided. This device is a surface imaging and tracking system where performance is validated through precision measurements, not a machine learning model that requires a "training set" in the conventional sense for image classification or prediction tasks. The software changes mentioned are minor and related to functionality improvements (e.g., frame rate due to memory increase, new camera for bore-based linacs, user interface features), not a re-training of a core algorithmic performance.
-
How the ground truth for the training set was established: Not applicable, as there is no mention of a training set for machine learning.
Ask a specific question about this device
(24 days)
The AlignRT Plus system is indicated for use to position and monitor patients relative to the prescribed treatment isocentre, and to withhold the beam automatically during radiation delivery.
For cranial treatments, a manual head adjuster is included which can be used in concert with AlignRT Plus to provide fine corrections for pitch, roll and yaw rotations.
AlignRT Plus is also used to track the patient's respiratory pattern for respiratory synchronized image acquisition, and radiation therapy treatment.
Patient contour data can be extracted and exported from the data acquired for the purpose of treatment planning.
AlignRT Plus can be calibrated directly to the treatment beam isocentre and in turn assists in performing quality assurance on MV, kV imagers, room lasers and the treatment couch.
AlignRT Plus is indicated for use during simulation, setup and stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body where radiation is indicated.
AlignRT Plus is a video-based three-dimensional (3D) surface imaging system, which is used to image the skin surface of a patient in 3D before and during radiotherapy treatment. The system consists of advanced software, a computer workstation, and one, two or three 3D camera units (each camera unit comprising a stereo pair of sensors to allow 3D surface reconstruction). The system is noninvasive, does not require the use of body markers and produces no ionizing irradiation during the imaging process.
AlignRT Plus is also able to perform both respiratory synchronised CT imaging and treatment delivery. In both instances, the system acquires a gated 3D surface model of the patient. User selected points are then tracked in real time in order to provide gating and position monitoring signals.
Real-time imaging and surface matching of the patient is possible during both setup and the treatment delivery to determine any patient movement. During treatment delivery, AlignRT Plus is also able to withhold the beam automatically, should the patient move outside user-defined tolerances.
Patient contour data may be extracted from surface data acquired by the system and exported for the purpose of treatment planning by radiotherapy professionals.
AlignRT Plus may be calibrated directly to the treatment beam isocentre using an optional custom designed calibration phantom and image processing software. It can analyse MV and kV digital imaging data acquired by other cleared devices. This in turn assists the user in performing quality assurance on MV, kV imagers, room lasers and the treatment couch.
The AlignRT Plus system includes the optional Head Adjuster for cranial treatments to allow for the manual, fine correction of pitch, roll and yaw in the patient's head position.
Precise isocenter calibration and the optional Head Adjuster provide improved frameless Stereotactic Radiosurgery (SRS). This is provided with the brand name "AlignRT SRS module".
The AlignRT Plus system is also provided under the brand OSMS (Optical Surface Monitoring System). This product is identical to AlignRT.
This Special 510(k) is requested only for software modification for an existing 510(k) cleared product "AlignRT Plus" (K181218). No hardware or material changes have been made since its last 510(k) clearance. The two key changes made in the existing AlignRT v6.0 software which are the subject of this Special 510(k) are listed below;
- System Frame Rate improvement
- Addition of Video Postural Setup Alignment feature
I am sorry, but I lack the ability to provide a response without the full context of the document. The request references information within the document that is not provided in a conversational format. If you can provide the specific details about the acceptance criteria and study, I would be happy to help describe them.
Ask a specific question about this device
(92 days)
The AlignRT Plus system is indicated for use to position and monitor patients relative to the prescribed treatment isocentre, and to withhold the beam automatically during radiation delivery.
For cranial treatments, a manual head adjuster is included which can be used in concert with AlignRT Plus to provide fine corrections for pitch, roll and yaw rotations.
AlignRT Plus is also used to track the patient's respiratory pattern for respiratory synchronized image acquisition, and radiation therapy treatment.
Patient contour data can be extracted and exported from the data acquired for the purpose of treatment planning.
AlignRT Plus can be calibrated directly to the treatment beam isocentre and in turn assists in performing quality assurance on MV, kV imagers, room lasers and the treatment couch.
AlignRT Plus is indicated for use during simulation, setup and stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body where radiation is indicated.
The AlignRT Plus system (K123371) is a combination of the devices AlignRT, GateCT and GateRT. It is a video-based three-dimensional (3D) surface imaging system, which is used to image the skin surface of a patient in 3D before and during radiotherapy treatment. The system consists of advanced software, a computer workstation, and one, two or three 3D camera units (each camera unit comprising a stereo pair of sensors to allow 3D surface reconstruction). The system is noninvasive, does not require the use of body markers and produces no ionizing irradiation during the imaging process.
AlignRT Plus is also able to perform both respiratory synchronised CT imaging and treatment delivery. In both instances, the system acquires a gated 3D surface model of the patient. User selected points are then tracked in real time in order to provide gating and position monitoring signals.
Real-time imaging and surface matching of the patient is possible during both setup and the treatment delivery to determine any patient movement. During treatment delivery, AlignRT Plus is also able to withhold the beam automatically, should the patient move outside user-defined tolerances.
Patient contour data may be extracted from surface data acquired by the system and exported for the purpose of treatment planning by radiotherapy professionals.
AlignRT Plus may be calibrated directly to the treatment beam isocentre using an optional custom designed calibration phantom and image processing software. It can analyse MV and kV digital imaging data acquired by other cleared devices. This in turn assists the user in performing quality assurance on MV, kV imagers, room lasers and the treatment couch.
The AlignRT Plus system includes the optional Head Adjuster for cranial treatments to allow for the manual, fine correction of pitch, roll and yaw in the patient's head position.
Precise isocenter calibration and the optional Head Adjuster provide improved frameless Stereotactic Radiosurgery (SRS). This is provided with the brand name "AlignRT SRS module".
The AlignRT Plus system is also provided under the brand OSMS (Optical Surface Monitoring System). This product is identical to AlignRT.
The provided text describes the AlignRT Plus system, a video-based three-dimensional (3D) surface imaging system used for patient positioning and monitoring during radiotherapy.
Here's a breakdown of the acceptance criteria and study information:
1. Table of Acceptance Criteria and Reported Device Performance:
| Acceptance Criteria (Performance Objective) | Reported Device Performance (Modified Device) | Comments (from document) |
|---|---|---|
| Positioning Accuracy | Target registration errors < 1mm (0.5mm) for all couch angles. | The modified device has the same performance or better than the predicate. Additional bench testing using improved measurement technique has been performed to substantiate improved claims. These claims and verification data are submitted within the 510k. |
| Respiratory Tracking | Tracks respiratory signal from imaged surface data and sends to CT (4D CT) or to Linac or imaging device (gating). Surface displacements can be tracked with RMS errors < 0.5mm over 10 or more breathing cycles. | Additional bench testing using improved measurement technique has been performed to substantiate improved claims. These claims and verification data are submitted within the 510k. |
2. Sample size used for the test set and the data provenance:
The document mentions "Additional bench testing using improved measurement technique has been performed to substantiate improved claims" for both positioning accuracy and respiratory tracking. However, it does not specify the sample size used for the test set or the data provenance (e.g., country of origin, retrospective or prospective) for these tests. The testing appears to be conducted on a calibration phantom, not on human subjects.
3. Number of experts used to establish the ground truth for the test set and qualifications of those experts:
This information is not provided in the document. The testing appears to be an objective measurement against a known physical standard (calibration phantom), rather than based on expert interpretation of medical images.
4. Adjudication method for the test set:
This information is not applicable/not provided. The performance evaluation is based on direct measurements against physical standards, not on subjective assessment by multiple reviewers.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and the effect size of how much human readers improve with AI vs. without AI assistance:
A multi-reader multi-case (MRMC) comparative effectiveness study was not performed or described in this document. The AlignRT Plus is a patient positioning and monitoring system, not primarily an AI diagnostic tool for image interpretation that would involve human readers.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
The performance metrics provided for Positioning Accuracy and Respiratory Tracking (target registration errors, RMS errors) likely represent standalone algorithm performance as measured against a physical phantom. The system's function is automated measurement and feedback.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
The ground truth used for the performance testing appears to be objective physical measurements from a calibration phantom with known parameters (e.g., precise positions, controlled respiratory movements).
8. The sample size for the training set:
The document does not provide information on the sample size for the training set. As a vision-based system for physical tracking, much of its core functionality would be based on geometric algorithms rather than extensive machine learning training data in the traditional sense of medical image analysis. It's possible internal validation or calibration data was used, but details are not given.
9. How the ground truth for the training set was established:
This information is not provided in the document. Similar to the training set size, details about how any ground truth for potential internal calibration or training was established are not mentioned.
Ask a specific question about this device
(68 days)
The AlignRT Plus system is indicated for use to position and monitor patients relative to the prescribed treatment isocentre, and to withhold the beam automatically during radiation delivery. For cranial treatments, a manual head adjuster is included which can be used in concert with AlignRT Plus to provide fine corrections for pitch, roll and yaw rotations. AlignRT Plus is also used to track the patient's respiratory pattern for respiratory synchronized image acquisition therapy treatment. Patient contour data can be extracted and exported from the data acquired for the purpose of treatment planning. AlignRT Plus can be calibrated directly to the treatment beam isocentre and in turn assists in performing quality assurance on MV, kV imagers, room lasers and the treatment couch. AlignRT Plus is indicated for use during simulation, setup and stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body where radiation is indicated.
AlignRT Plus system is a combination of the predicate devices AlignRT (K052682) and GateCT-RT (K072171). It is a video-based three-dimensional (3D) surface imaging system, which is used to image the skin surface of a patient in 3D before and during radiotherapy treatment. The system consists of advanced software, a computer workstation, and one, two or three 3D camera units (each camera unit comprising a stereo pair of sensors to allow 3D surface reconstruction). The system is non-invasive, does not require the use of body markers and produces no ionizing irradiation during the imaging process.
Acceptance Criteria and Study for AlignRT Plus
The AlignRT Plus system is a video-based three-dimensional (3D) surface imaging system used to position and monitor patients during radiotherapy treatment. The following describes its acceptance criteria and the study demonstrating its performance.
1. Table of Acceptance Criteria and Reported Device Performance
| Feature/Metric | Acceptance Criteria | Reported Device Performance |
|---|---|---|
| Positioning Accuracy | Target registration errors (as measured using calibration phantom) < 1mm for all couch angles. | Target registration errors (as measured using calibration phantom) < 1mm for all couch angles. |
| Respiratory Tracking | - Tracks respiratory signal from imaged surface data.- Sends to CT (4D CT) or to Linac or imaging device (gating).- Surface displacements can be tracked with RMS errors < 0.5mm over 10 or more breathing cycles. | - Tracks respiratory signal from imaged surface data and sends to CT (4D CT) or to Linac or imaging device (gating).- Surface displacements can be tracked with RMS errors < 0.5mm over 10 or more breathing cycles. |
| System Accuracy (Predicate AlignRT K052682) | Errors along each of 3 axes of treatment: max mean: 0.27mm; max std dev: 0.65mm. | (This is the reported performance for the predicate device, against which AlignRT Plus is compared for substantial equivalence. The new device is stated to be substantially equivalent with improved features.) |
Note: The document focuses on demonstrating substantial equivalence to predicate devices (AlignRT K052682 and GateCT-RT K072171) rather than presenting a detailed, independent clinical study with specific acceptance criteria outside of the technical specifications listed above. The performance metrics listed effectively serve as the acceptance criteria for technical performance.
2. Sample Size and Data Provenance
The provided document (510(k) Premarket Notification) does not explicitly state specific sample sizes used for a test set for the AlignRT Plus device itself.
Instead, it relies on demonstrating substantial equivalence to its predicate devices. The performance figures presented are derived from:
- Calibration phantom measurements: For positioning accuracy of the AlignRT Plus. The type and number of measurements are not specified.
- Respiratory tracking performance: For AlignRT Plus, again using measurements from imaged surface data. The number of breathing cycles (10 or more) is mentioned, but not the number of patients or measurement instances.
- Predicate device performance: The system accuracy for the predicate AlignRT (K052682) is cited, presumably from the studies conducted for its prior clearance.
Data Provenance: The document originates from Vision RT Ltd. in the United Kingdom. The data presented is likely derived from internal testing and validation activities characteristic of a premarket notification, rather than a large-scale, multi-center clinical trial. It is an internal retrospective analysis of technical performance.
3. Number of Experts and Qualifications for Ground Truth
The document does not detail the use of a panel of experts to establish a "ground truth" for a specific test set in the context of a reader study. The performance metrics (positioning accuracy, respiratory tracking RMS error) are objective technical measurements performed using phantoms or direct measurements from physical surfaces/breathing cycles. Therefore, expert consensus in the traditional sense for image interpretation is not directly applicable here.
4. Adjudication Method for the Test Set
Given that the "ground truth" is based on objective physical measurements (e.g., target registration errors from a phantom, RMS errors from surface displacement tracking), an adjudication method in the sense of reconciling divergent expert opinions is not applicable or described. The measurements are presumably taken directly and automatically by the system or calibrated instruments.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No MRMC comparative effectiveness study is presented or mentioned in this 510(k) premarket notification. The document focuses on the technical capabilities and safety of the device to demonstrate substantial equivalence to previously cleared predicate devices. It does not assess the improvement of human readers with AI assistance because the device is a patient positioning and monitoring system, not an AI-assisted diagnostic or treatment planning tool that directly involves human readers' interpretation of output in the same way an imaging AI might.
6. Standalone (Algorithm Only) Performance Study
Yes, the performance data provided largely represents standalone (algorithm only) performance. The metrics of positioning accuracy and respiratory tracking capability describe the intrinsic technical accuracy of the AlignRT Plus system in performing its designated functions (e.g., identifying target registration errors, tracking surface displacements) without direct human intervention in the measurement process itself. The "user" interaction described pertains to setup (e.g., selecting tracking points) and monitoring, but the reported accuracy figures are inherent to the system's measurements.
7. Type of Ground Truth Used
The ground truth for the reported performance metrics is based on objective physical measurements rather than expert consensus, pathology, or outcomes data:
- Target registration errors: Established using a physical calibration phantom with known, precise reference points.
- Surface displacements/respiratory tracking: Established by measuring the movement of the imaged surface against known or expected movements (e.g., during breathing cycles).
8. Sample Size for the Training Set
The document does not specify a sample size for a training set. As a medical device for patient positioning and monitoring, the AlignRT Plus utilizes algorithms for 3D surface reconstruction and tracking. While such algorithms undergo internal development and validation, the 510(k) summary does not typically disclose details of internal algorithm training data, particularly for devices that are not primarily AI/ML-driven in the diagnostic sense. The underlying principles are based on established optical and image processing techniques.
9. How Ground Truth for the Training Set Was Established
Given that a specific training set size is not provided, the method for establishing ground truth for any internal training or validation of the algorithms is not detailed in this public document. It is implicitly assumed to be based on engineering validation and testing using phantoms or controlled environments to ensure the accuracy of the 3D surface reconstruction and tracking algorithms.
Ask a specific question about this device
(100 days)
The AlignRT system is used to position patients at the isocentre of the linear accelerator for radiation therapy procedures. Patient contour data can be extracted and exported from the acquired data for the purpose of treatment planning. AlignRT is used by radiotherapy professionals in appropriate hospital environments. It can be used with all types of radiotherapy patients and can image any visible anatomical region. The system is completely non-invasive and does not require the application of any external markers to the patient.
AlignRT is a video-based three-dimensional (3D) surface imaging system, which is used to image the skin surface of a patient in 3D before and during radiotherapy treatment. The system consists of advanced software, a computer workstation, two 3D camera units, cables, and templates that are used for camera calibration. The system is non-invasive, does not require the use of body markers and produces no irradiation during the imaging process. AlignRT first generates a reference surface of the optimum treatment position determined during treatment simulation. This reference image is generated by either recording the surface of a patient placed in a conventional radiotherapy simulator in which the system is installed or by importing skin contours from CT (X-Ray Computer Tomography) volumetric data generated via third party treatment planning software. Prior to each treatment session the patient's position is imaged and compared to the reference image by the system's surface matching software. Where movement from the reference position is detected the software calculates new co-ordinates to adjust the treatment couch for optimal positioning of the patient. The system is designed to interface directly with certain third party couch control systems. In order to minimise the setup errors that can result from respiratory motion, an option is available to gate the image capture of the system so that all data is acquired at a repeatable point within the breathing cycle such as the end of exhalation. Real-time imaging (1 frame/sec) of the patient is possible during the actual treatment and software verification tools are provided in order to determine any patient movement. Patient contour data may be extracted from surface data acquired by the system. In addition, lung contour data may be estimated by utilising AlignRT 3D surface data and 2D X-ray data acquired from either a conventional simulator or portal imager. These data may be exported for the purpose of treatment planning by radiotherapy professionals.
Based on the provided 510(k) summary (K052682) for the AlignRT system, the following information regarding acceptance criteria and study details can be extracted:
1. Table of Acceptance Criteria and Reported Device Performance
The provided document (K052682) is a 510(k) summary, which focuses on demonstrating substantial equivalence to predicate devices rather than providing a detailed performance study with explicit acceptance criteria and corresponding performance metrics for a novel technology. Therefore, the "acceptance criteria" listed below are inferred from the device's stated purpose and comparison to predicates, and the "reported device performance" is a general statement from the summary. Specific quantitative metrics are not detailed in this document.
| Acceptance Criteria (Inferred from Device Purpose and Comparison) | Reported Device Performance (General Statement in Summary) |
|---|---|
| Ability to image the skin surface of a patient in 3D before and during radiotherapy treatment. | "AlignRT is a video-based three-dimensional (3D) surface imaging system, which is used to image the skin surface of a patient in 3D before and during radiotherapy treatment." |
| Capability to generate a reference surface of the optimum treatment position. | "AlignRT first generates a reference surface of the optimum treatment position determined during treatment simulation." |
| Comparison of patient's current position to the reference image. | "Prior to each treatment session the patient's position is imaged and compared to the reference image by the system's surface matching software." |
| Calculation of new co-ordinates to adjust the treatment couch for optimal patient positioning where movement is detected. | "Where movement from the reference position is detected the software calculates new co-ordinates to adjust the treatment couch for optimal positioning of the patient." |
| Real-time imaging capability during treatment to detect patient movement. | "Real-time imaging (1 frame/sec) of the patient is possible during the actual treatment and software verification tools are provided in order to determine any patient movement." |
| Ability to extract patient contour data. | "Patient contour data may be extracted from surface data acquired by the system." |
| Ability to estimate lung contour data by utilizing 3D surface data and 2D X-ray data. | "In addition, lung contour data may be estimated by utilising AlignRT 3D surface data and 2D X-ray data acquired from either a conventional simulator or portal imager." |
| No new efficacy or safety concerns compared to predicate devices despite technological differences. | "performance data has been submitted to show that AlignRT achieves its intended use and that these technological differences raise no new efficacy or safety concerns." |
2. Sample Size Used for the Test Set and Data Provenance
The 510(k) summary does not explicitly state a sample size for a test set or any details about data provenance (e.g., country of origin, retrospective or prospective data) for a specific clinical study. General "performance data" is mentioned as being submitted, but the specifics are not included in this publicly available summary.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications
The 510(k) summary does not provide information regarding the number of experts, their qualifications, or how ground truth was established for any specific test set. The device's function involves comparing to a "reference surface" determined during "treatment simulation" or from "CT volumetric data," which implies clinical input from radiotherapy professionals, but specific details about establishing ground truth for validation are absent.
4. Adjudication Method for the Test Set
The 510(k) summary does not describe any adjudication method used for a test set.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
The 510(k) summary makes no mention of a Multi-Reader Multi-Case (MRMC) comparative effectiveness study, nor does it quantify any effect size of human readers improving with or without AI assistance. The AlignRT system is presented as a positioning and monitoring tool for radiotherapy, not an AI-assisted diagnostic or interpretive tool for human readers in the traditional MRMC sense.
6. Standalone (Algorithm Only) Performance Study
The document states that "performance data has been submitted to show that AlignRT achieves its intended use." This implies that a standalone performance evaluation of the system's capabilities (e.g., accuracy of 3D surface imaging, precision of position detection, correlation with actual patient movement) was likely performed and submitted to the FDA. However, the details of such a standalone study, including specific metrics and methodologies, are not present in this summary. The core of the device is its algorithm for surface matching and coordinate calculation.
7. Type of Ground Truth Used
The ground truth for the core function of the AlignRT system is implicitly related to:
- The "optimum treatment position determined during treatment simulation." This would be established by radiotherapy professionals.
- "Skin contours from CT (X-Ray Computer Tomography) volumetric data" which are considered a gold standard for anatomical representation in radiotherapy planning.
- "2D X-ray data acquired from either a conventional simulator or portal imager" when estimating lung contour data.
Therefore, the ground truth leans towards expert consensus (for optimal treatment position) and imaging modalities considered high-fidelity anatomical references (CT, X-ray). Pathology or outcomes data are not mentioned as direct ground truth for this positioning device.
8. Sample Size for the Training Set
The 510(k) summary does not provide any information regarding the sample size of a training set. Given the described functionality (3D surface imaging, matching, coordinate calculation), the system likely relies on image processing algorithms that might be trained or calibrated, but details are not disclosed.
9. How the Ground Truth for the Training Set Was Established
Since no information on a training set is provided, there is correspondingly no information on how its ground truth was established.
Ask a specific question about this device