Found 2 results

510(k) Data Aggregation

    K Number
    K043228
    Date Cleared
    2005-04-22

    (151 days)

    Product Code
    Regulation Number
    862.1117
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For in vitro diagnostic use in the quantitative determination of B-type Natriuretic Peptide (BNP) in human plasma using the ADVIA Centaur® System. This assay is indicated for the measurement of plasma BNP as an aid in the diagnosis and assessment of the severity of heart failure. In patients with acute coronary syndromes (ACS), this test, in conjunction with other known risk factors, can also be used to predict survival as well as to predict the likelihood of future heart failure. This assay is not intended for use on any other system.

    For in vitro diagnostic use in the quantitative determination of B-type Natriuretic Peptide (BNP) in human plasma using the ACS:180® Automated Chemiluminescence System. This assay is indicated for the measurement of plasma BNP as an aid in the diagnosis and assessment of the severity of heart failure. In patients with acute coronary syndromes (ACS), this test, in conjunction with other known risk factors, can also be used to predict the likelihood of future heart failure. This assay is not intended for use on any other system.

    Device Description

    The ACS:180 and ADVIA Centaur BNP assays are fully automated two-site sandwich immunoassays using direct chemiluminescent technology. The first antibody, in the Reagent, is an acridinium ester labeled monoclonal mouse anti-human antibody specific to the N-terminal portion of BNP. The second antibody, in the Solid Phase Reagent, is a biotinylated monoclonal mouse anti-human antibody specific to the C-terminal portion of BNP, which is coupled to streptavidin coated magnetic particles. Patient sample (calibrators or control materials) is incubated for 5 minutes at 37°C with the Solid Phase Reagent and the tracer antibody conjugate. Subsequently, Solid Phase Reagent is added and incubated for 2.5 minutes at 37°C. An immuno-complex is formed between the unbound antibody conjugates are washed away. Following incubation, the immuno-complex signal is measured in a luminometer. The chemiluminescent signal will have a minimum amount of bound AE label, while Samples with high levels of BNP will have maximum label complex bound. Thus, a direct relationship exists between the amount of BNP present in the patient sample and the relative light units (RLUs) detected by the system.

    AI/ML Overview

    Here's an analysis of the provided text regarding the Bayer Diagnostics ADVIA Centaur BNP Assay, broken down by your requested criteria:

    The provided document is a Summary of Safety and Effectiveness for the Bayer Diagnostics ADVIA Centaur BNP Assay and ACS:180 BNP Assay, primarily focusing on its updated indications for use and substantial equivalence to previously cleared devices. It describes the technological and performance characteristics but does not contain detailed study results like typical acceptance criteria tables for accuracy or precision studies against a defined benchmark. Instead, it states that the characteristics (like precision, measuring range, analytical sensitivity, etc.) are "Same" as the predicate device.

    Therefore, for aspects like "reported device performance" against acceptance criteria, "sample sizes used for the test set," "number of experts," "adjudication method," "MRMC study effect size," and "standalone performance," the provided document does not contain this information. It relies on the previously cleared predicate devices for establishing performance equivalency without re-presenting the detailed studies.

    Here's what can be extracted based on the information available:


    1. A table of acceptance criteria and the reported device performance

    The provided document does not explicitly present a table of acceptance criteria with corresponding device performance from new studies following the format typically seen for a new device's primary validation. Instead, it compares the proposed device's characteristics to a predicate device, largely stating they are "Same."

    However, we can infer some "performance characteristics" that are deemed acceptable by virtue of being equivalent to the predicate.

    CharacteristicAcceptance Criteria (Implied from Predicate)Reported Device Performance (Proposed Device)
    Measuring RangeADVIA Centaur: <2.0 – 5000 pg/mL ACS:180: <15 – 5000 pg/mLSame as predicate (ADVIA Centaur: <2.0 – 5000 pg/mL, ACS:180: <15 – 5000 pg/mL)
    Precision (ADVIA Centaur)Within-run 1.8 – 4.3 %CV (29.4 - 1736 pg/mL) Total 2.3 - 4.7 %CV (29.4 - 1736 pg/mL)Same as predicate
    Precision (ACS:180)Within-run 2.5 – 7.9%CV (51.5 - 1783 pg/mL) Total 3.8 - 9.9%CV (51.5 - 1783 pg/mL)Same as predicate
    Hook EffectNo high dose effect up to 100,000 pg/mLSame as predicate
    Analytical SensitivityADVIA Centaur: <2 pg/mL ACS:180: <15 pg/mLSame as predicate
    Dilution RecoveryOn-board dilution 1:2, 1:5, 1:10 with average recovery of 97% (ADVIA Centaur) and 98% (ACS:180)Same as predicate
    Interference (Hemoglobin)No interference up to 1000 mg/dLSame as predicate
    Interference (Triglycerides)No interference up to 800 mg/dLSame as predicate
    Interference (Cholesterol)No interference up to 1000 mg/dLSame as predicate
    Interference (Urea)No interference up to 200 mg/dLSame as predicate
    Interference (Creatinine)No interference up to 2.5 mg/dLSame as predicate
    Interference (Unconjugated Bilirubin)No interference up to 25 mg/dLSame as predicate
    Interference (Conjugated Bilirubin)No interference up to 25 mg/dLSame as predicate
    Interference (Human IgG)No interference up to 5.3 g/dLSame as predicate
    Interference (Pharm. Drugs)No interference from 55 commonly used drugsSame as predicate
    Sample TypeHuman plasma using EDTA as anticoagulantSame as predicate (Other types not recommended)

    The study that "proves the device meets the acceptance criteria" is implicitly the previous 510(k) clearances for the predicate devices (K031038 and K040425). The current submission (K043228) relies on substantial equivalence to these predicate devices for its performance characteristics. The document explicitly states: "The Bayer B-type Natriuretic Peptide (BNP) assays on the ACS:180® and ADVIA Centaur® are substantially equivalent to ADVIA Centaur® BNP assay." And for the technological and performance tables, it consistently lists "Same" for the proposed device's characteristics when compared to the current (predicate) one.

    2. Sample size used for the test set and the data provenance

    • Sample size: Not specified in the provided document for the current submission's validation. The document states "Same" for all performance characteristics, implying reliance on the studies for the predicate devices (K031038 and K040425).
    • Data provenance: Not specified in the provided document. As the submission relies on substantial equivalence to predicate devices, the original data provenance would be from those prior submissions.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not provided in the document as it focuses on an in vitro diagnostic assay, not an imaging device requiring expert interpretation for ground truth.

    4. Adjudication method for the test set

    This information is not provided as it is not relevant to this type of IVD device submission.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not provided. This type of study (MRMC for AI assistance) is not applicable to an in vitro diagnostic assay like a BNP test.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The device is an in vitro diagnostic assay. Its performance is inherently standalone in the sense that it measures BNP levels from a sample without human interpretive assistance for the measurement itself. The results are then used by clinicians. The performance characteristics described (precision, sensitivity, range, etc.) are its standalone performance.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For an in vitro diagnostic assay like BNP, the "ground truth" for accuracy and calibration would typically be established against:

    • A reference standard system (e.g., highly pure synthetic human BNP (amino acid 77 to 108) as mentioned for traceability).
    • Known concentrations of analytes (for linearity, sensitivity verification).
    • Clinical correlation (for assessing diagnostic utility against patient outcomes or established clinical diagnoses, such as heart failure diagnosis confirmed by other clinical methods or follow-up).
      The document mentions "Reference standard - synthetic human BNP (amino acid 77 to 108) in buffer based matrix" for traceability.

    8. The sample size for the training set

    This information is not provided. Given that this is an immunoassay and the submission is for substantial equivalence, the concept of a "training set" in the context of machine learning/AI (which your question implies) is not applicable here. The assay development would involve extensive analytical validation, but not typically a machine learning "training set."

    9. How the ground truth for the training set was established

    As above, the concept of a "training set" for AI is not applicable to this immunoassay. The ground truth for analytical validation (e.g., establishing accurate measurement, precision) is based on reference materials, calibrators, and controlled experimental conditions.

    Ask a Question

    Ask a specific question about this device

    K Number
    K040425
    Date Cleared
    2004-06-14

    (117 days)

    Product Code
    Regulation Number
    862.1117
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For in vitro diagnostic use in the quantitative determination of B-type Natriuretic Peptide (BNP) in human plasma using the ACS:180® Automated Chemiluminescence System. This assay is indicated for the measurement of plasma BNP as an aid in the diagnosis and assessment of the severity of heart failure. This test, in conjunction with other known risk factors, can also be used to predict survival in patients after myocardial infarction. This assay is not intended for use on any other system.

    For in vitro diagnostic use in the quantitative determination of B-type Natriuretic Peptide (BNP) in human plasma using the ADVIA Centaur® System. This assay is indicated for the measurement of plasma BNP as an aid in the diagnosis of and assessment of the severity of heart failure. This test, in conjunction with other known risk factors, can also be used to predict survival in patients after myocardial infarction. This assay is not intended for use on any other system.

    Device Description

    The ACS:180 and ADVIA Centaur BNP assays are fully automated two-site sandwich immunoassays using direct chemiluminescent technology, which use constant amounts of two monoclonal antibodies. The first antibody, in the Lite Reagent, is an acridinium ester labeled monoclonal mouse anti-human BNP F(ab')2 fragment specific to the ring structure of BNP. The second antibody, in the Solid Phase, is a biotinylated monoclonal mouse antihuman antibody specific to the C-terminal portion of BNP, which is coupled to streptavidin magnetic particles. Patient sample (calibrator or control materials) is incubated for 5 minutes at 37°C with the Reagent that contains the tracer antibody conjugate. Subsequently, Solid Phase reagent is added and incubated for 2.5 minutes at 37°C. An immuno-complex is formed between the BNP in the sample and the two antibody conjugates. Following incubation, the unbound antibody conjugates are washed away. The chemiluminescence of the immuno-complex signal is measured in a fuminometer. Samples with low BNP levels will have a minimum amount of bound AE label, while samples with high levels of BNP will have maximum label complex bound. Thus, a direct relationship exists between the amount of BNP present in the patient sample and the amount of relative light units (RLUs) detected by the system.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the ACS:180® and ADVIA Centaur® BNP Assays by Bayer Diagnostics. The document establishes substantial equivalence to a predicate device (ADVIA Centaur® B-Type Natriuretic Peptide (BNP) Assay, K031038) and discusses the assays' technological and performance characteristics.

    Here's an analysis of the acceptance criteria and study information provided, structured according to your request:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" for performance metrics in a pass/fail format. However, it presents the "Performance Characteristics" of the proposed devices, which implicitly serve as the demonstrated performance that FDA found substantially equivalent to the predicate device. These performance characteristics cover various analytical aspects of the assay. Since the document states "The following table compares the performance characteristics of the current and proposed Bayer ADVIA Centaur and ACS:180 BNP assays," and the values listed under "Proposed Bayer ADVIA Centaur® and ACS:180® BNP Immunoassays" are identical to "Current Bayer ADVIA Centaur® and ACS:180® BNP Immunoassays," it implies that the new devices meet the established performance of the predicate device.

    Performance CharacteristicAcceptance Criteria (Implied by Predicate Performance)Reported Device Performance (Proposed Devices)
    Precision (ADVIA Centaur)Within-run: 1.8 – 4.3 %CV (29.4 to 1736 pg/mL)Within-run: 1.8 – 4.3 %CV (29.4 to 1736 pg/mL)
    Total: 2.3 - 4.7 %CV (29.4 to 1736 pg/mL)Total: 2.3 - 4.7 %CV (29.4 to 1736 pg/mL)
    Precision (ACS:180)Within-run: 2.5 - 7.9%CV (51.5 to 1783 pg/mL)Within-run: 2.5 - 7.9%CV (51.5 to 1783 pg/mL)
    Total: 3.8 - 9.9%CV (51.5 to 1783 pg/mL)Total: 3.8 - 9.9%CV (51.5 to 1783 pg/mL)
    Hook EffectNo high dose effect up to 100,000 pg/mLNo high dose effect up to 100,000 pg/mL
    Analytical Sensitivity (ADVIA Centaur)<2 pg/mL<2 pg/mL
    Analytical Sensitivity (ACS:180)<15 pg/mL<15 pg/mL
    Dilution RecoveryOn-board dilution 1:2, 1:5, 1:10 with avg recovery of 97% (ADVIA Centaur)On-board dilution 1:2, 1:5, 1:10 with avg recovery of 97% (ADVIA Centaur)
    On-board dilution 1:2, 1:5, 1:10 with avg recovery of 98% (ACS:180)On-board dilution 1:2, 1:5, 1:10 with avg recovery of 98% (ACS:180)
    InterferenceNo interference from various substances (hemoglobin, triglycerides, cholesterol, urea, creatinine, unconjugated bilirubin, conjugated bilirubin, human IgG, 55 pharmaceutical drugs)No interference from various substances (hemoglobin, triglycerides, cholesterol, urea, creatinine, unconjugated bilirubin, conjugated bilirubin, human IgG, 55 pharmaceutical drugs)

    2. Sample Sizes Used for the Test Set and Data Provenance

    The document does not specify the exact sample sizes for the test sets or the data provenance (country of origin, retrospective/prospective) for each stated performance characteristic. It only presents the summary results. For clinical diagnostic assays like this, it's common for studies to involve various patient samples to assess precision, analytical sensitivity, dilution recovery, and interference. However, these details are not provided in this summary.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This document describes an in vitro diagnostic (IVD) assay which measures a biomarker (BNP). The "ground truth" for such assays is typically the actual concentration of the analyte, established through highly accurate and precise analytical methods, not through expert human review in the same way an imaging AI would be reviewed. Therefore, the concept of "experts" establishing ground truth for evaluating the device's analytical performance (precision, sensitivity, etc.) is not directly applicable in this context. The "ground truth" here is determined by the reference analytical methods themselves.

    4. Adjudication Method for the Test Set

    As explained above, for analytical performance characteristics of an IVD assay, there isn't a direct "adjudication method" in the sense of reconciling disagreements among human experts evaluating data. The performance is assessed against established analytical standards and reference methods.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. This type of study is relevant for medical devices that involve human interpretation (e.g., radiologists reading images and an AI assisting them). The ACS:180 and ADVIA Centaur BNP assays are entirely automated quantitative tests, so human interpretation of the assay's direct output (BNP concentration) is not the primary focus of performance evaluation in the same way. The device's output itself is a numerical value.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    Yes, the performance characteristics presented are for the standalone algorithm/device (the automated immunoassay system). The device provides a quantitative measurement of BNP without human intervention in the measurement process itself. The "without human-in-the-loop" aspect is inherent to automated diagnostic assays.

    7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)

    The ground truth for evaluating the analytical performance of these BNP assays is based on:

    • Reference Standards/Synthetic Materials: Used for calibration and establishing traceability (e.g., "synthetic human BNP (amino acid 77 to 108) in buffer-based matrix").
    • Known Concentrations: Samples with precisely known BNP concentrations are used to assess accuracy, linearity, and recovery.
    • Validated Analytical Methods: Comparison against established and validated laboratory methods or reference methods to verify analytical parameters like precision, sensitivity, and linearity.
    • Controlled Interference Studies: Samples spiked with known interfering substances at controlled concentrations to determine the device's robustness to these interferences.

    For the stated clinical indications for use (diagnosis and assessment of heart failure severity, prediction of survival after MI), the clinical utility of BNP levels is established through extensive medical literature and clinical studies that correlate BNP levels with patient outcomes, pathology (e.g., echocardiography, cardiac catheterization), and clinical diagnoses. The device's role is to accurately measure these established BNP levels. The document mentions "Decision threshold of 100 pg/mL recommended for diagnosis of heart failure. Decision threshold of 80 pg/mL recommended for prediction of survival after myocardial infarction," which are clinical ground truths, but not data used to directly evaluate the analytical performance of the device in this submission.

    8. The Sample Size for the Training Set

    The document does not provide details on specific "training sets" in the context of an algorithm that learns from data. This type of immunoassay is based on established biochemical principles and reagents, not a machine learning model that requires a distinct training phase with a dataset. The development and optimization of the assay would involve extensive R&D and testing with numerous samples, but these are not typically referred to as a "training set" in the sense of AI/ML.

    9. How the Ground Truth for the Training Set Was Established

    As this is an immunoassay and not an AI/ML device, the concept of a "training set" with associated ground truth for learning is not directly applicable. The "ground truth" for the development and validation of such an assay would be based on:

    • Precise gravimetric/volumetric measurements for reagent preparation.
    • Characterization of antibodies and their binding kinetics.
    • Use of highly purified and characterized synthetic BNP standards.
    • Comparative analysis with established reference methods during the assay development process to ensure analytical accuracy and reliability.