Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K170314

    Validate with FDA (Live)

    Device Name
    ACL AcuStar
    Date Cleared
    2017-03-03

    (30 days)

    Product Code
    Regulation Number
    864.5425
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Automated immunoassay analyzer designed specifically for in vitro diagnostic use in a clinical laboratory. The assay analysis is based on chemiluminescent technology. The system provides results for both direct measurements and calculated parameters.

    Device Description

    The AcuStar is an automated, bench-top system for lab use that measures the analyte amount in samples by: Subjecting the sample to reagents that cause a reaction with an antigen or antibody in the sample. Placing the cuvettes in a controlled environment to let the reactants bind into a complex. Separating out the complex from unused reactants. Treating this complex with a chemical that produces light in proportion to the analyte concentration. Measuring the light output to determine the amount of antibodies or antigens that were in the sample.

    AI/ML Overview

    This document is a 510(k) premarket notification for a software update to an existing medical device, the ACL AcuStar™. The submission qualifies as a "Special 510(k)" because the modification (updating the operating system from Windows XP to Windows 7) does not change the indications for use, operating principle, labeled performance claims, hardware, data reduction software, fluidic design, test parameters, calibration, quality control, consumables, or reagents. Therefore, the information provided focuses on demonstrating that the updated device is substantially equivalent to the predicate device and does not involve a new clinical study to establish acceptance criteria or device performance in the same way a de novo submission would.

    Given the nature of this Special 510(k), the prompt's questions regarding acceptance criteria, device performance, sample sizes, expert ground truth, adjudication methods, MRMC studies, and standalone performance are not directly applicable in the traditional sense for a new device evaluation. Instead, the "acceptance criteria" here relate to demonstrating that the change did not negatively impact the device's original validated performance.

    Here's how the information aligns with the prompt's request, interpreted in the context of a Special 510(k) for a software update:

    1. A table of acceptance criteria and the reported device performance

    The document does not present a table of specific performance acceptance criteria (e.g., sensitivity, specificity, accuracy) for the device itself because it's a software update to an already cleared device. Instead, the implicit acceptance criterion is that the device's performance remains unchanged despite the operating system upgrade.

    The "reported device performance" in this context is that the updated device is substantially equivalent to the predicate device, meaning its performance characteristics relevant to its intended use have not been altered.

    Acceptance Criteria (Implicit for Special 510(k) Software Update)Reported Device Performance (Conclusion)
    No change in Indications for UseSame
    No change in Operating PrincipleSame
    No change to Labeled Performance ClaimsSame (implies original performance claims are met)
    No change to HardwareSame
    No change to Data Reduction SoftwareSame
    No change to Fluidic DesignSame
    No change to Test ParametersSame
    No change to CalibrationSame
    No change to Quality ControlSame
    No change to ConsumablesSame
    No change to ReagentsSame
    Overall substantial equivalence to predicate deviceThe ACL AcuStar™ with software v3.0.1 running on Windows 7 is substantially equivalent to the cleared and currently marketed predicate device (K083518).

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    The document does not detail a specific "test set" or its sample size in the context of clinical performance evaluation for this Special 510(k). The focus is on demonstrating that the software change itself does not affect the existing device's validated performance. For a software update of this nature, testing would typically involve verification and validation (V&V) activities against the original design specifications and potentially comparing outputs from the Windows XP and Windows 7 versions using a set of representative samples, but these details are not provided in this summary. There is no mention of country of origin or retrospective/prospective data for a new performance study.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    This is not applicable to this Special 510(k). The regulatory submission for a software operating system update does not involve establishing new diagnostic ground truth with medical experts. The ground truth for the underlying assays was established during the initial clearance of the predicate device (K083518).

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    This is not applicable to this Special 510(k) as no new clinical test set requiring expert adjudication is described.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. The ACL AcuStar™ is an automated immunoassay analyzer, not a device involving human interpretation of images or data that would benefit from AI assistance in the way described for an MRMC study.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    The ACL AcuStar™ is an automated system that performs tests in a standalone manner (without continuous human intervention for each test analysis beyond loading samples and reagents). The original predicate device's performance was established as a standalone automated system. The current submission simply states that the software update does not change the operating principle, performance claims, or data reduction software, implying the standalone performance remains identical to the predicate.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for the diagnostic assays performed by the ACL AcuStar™ would have been established during the original clearance of the predicate device (K083518), likely through comparison to reference methods, clinical diagnosis, or other accepted gold standards for in vitro diagnostic tests. This Special 510(k) does not provide details on the ground truth for the original assays, as it is assumed to be established with the predicate.

    8. The sample size for the training set

    This is not applicable. The device is not an AI/ML algorithm that requires a training set in the common understanding of an "AI device." It's an automated immunoassay analyzer with a software operating system update.

    9. How the ground truth for the training set was established

    This is not applicable for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K083518

    Validate with FDA (Live)

    Date Cleared
    2009-03-13

    (107 days)

    Product Code
    Regulation Number
    864.5425
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ACL AcuStar: Automated immunoassay analyzer designed specifically for in vitro diagnostic use in a clinical laboratory. The assay analysis is based on chemiluminescent technology. The system provides results for both direct measurements and calculated parameters.
    HemosIL AcuStar D-Dimer: Fully automated chemiluminescent immunoassay for the quantitative determination of D-Dimer in human citrated plasma on the ACL AcuStar as an aid in the diagnosis of venous thromboembolism (VTE) [deep vein thrombosis (DVT) and pulmonary embolism (PE)].
    HemosIL AcuStar D-Dimer Controls: For the quality control of D-Dimer assay performed on the ACL AcuStar.

    Device Description

    ACL AcuStar: The AcuStar is an automated, bench-top system for lab use that measures the analyte amount in blood samples by: Subjecting the blood sample to reagents that cause a reaction with an antigen or antibody in the sample. Placing the cuvettes in a controlled environment to allow the reactants to bind into a complex. Separating out the complex from unused reactants. Treating this complex with a chemical that produces light in proportion to the analyte concentration. Measuring the light output to determine the amount of antibodies or antigens that were in the sample.

    HemosIL AcuStar D-Dimer: The HemosIL AcuStar D-Dimer assay is a two-step immunoassay to quantify D-Dimer in human citrated plasma using magnetic particles as solid phase and a chemiluminescent detection system. In the first step, sample, anti-D-Dimer antibody coated magnetic particles, and assay buffer are combined, and the fibrin soluble derivatives containing the D-Dimer domain present in the sample bind to the anti-D-Dimer antibody coated magnetic particles. After magnetic separation and washing, an anti-XDP antibody labeled with isoluminol is added and incubated in a second step. After a new magnetic separation and washing, two triggers are added and the resulting chemiluminescent reaction is measured as relative light units (RLUs) by the ACL AcuStar optical system. The RLUs are directly proportional to the D-Dimer concentration in the sample. The ACL AcuStar D-Dimer assay utilizes a 4 Parameter Logistic Curve (4PLC) fit data reduction method to generate a Master Curve. The Master Curve is predefined lot dependent, and is stored in the instrument through the cartridge barcode. With the measurement of calibrators, the predefined Master Curve is transformed to a new, instrument specific 4PLC Working Curve. The concentration values of the calibrators are included in the calibrator plastic tube barcodes.

    HemosIL AcuStar D-Dimer Controls: The Low, High, and Very High D-Dimer Controls are prepared by means of a dedicated process and contain different concentrations of partially purified D-Dimer obtained by digestion of Factor XIIIa cross-linked human fibrin with human plasmin.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study details for the ACL™ AcuStar™ HemosIL™ AcuStar™ D-Dimer, based on the provided 510(k) summary:

    Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific MetricAcceptance Criterion (Implicit)Reported Device Performance
    PrecisionCoefficient of Variation (CV%) for D-Dimer Controls and Calibrator 1The document does not explicitly state numerical acceptance criteria for precision (e.g., CV% < X%). However, good precision is generally expected for diagnostic assays. The reported CV% values demonstrate the reproducibility and reliability of the assay across different D-Dimer concentrations.Low D-Dimer Control:- Within run CV%: 4.0%- Total CV%: 6.8%High D-Dimer Control:- Within run CV%: 2.3%- Total CV%: 4.9%Very High D-Dimer Control:- Within run CV%: 2.5%- Total CV%: 5.6%Calibrator 1:- Within run CV%: 2.7%- Total CV%: 5.4%
    Method ComparisonCorrelation (r) and Slope when compared to predicate device (VIDAS)Not explicitly stated as a numerical threshold in the document. However, a strong correlation (r close to 1) and a slope close to 1 (or a consistent proportionality) are generally expected to demonstrate substantial equivalence to the predicate device.Slope: 1.16r (correlation coefficient): 0.888
    Clinical PerformanceSensitivity for VTE diagnosisFor a D-Dimer assay used as an aid in diagnosis and for ruling out VTE, high sensitivity is crucial to minimize false negatives. While no explicit numerical threshold is stated, 100% sensitivity is a highly desirable outcome for an exclusion assay, indicating that no true positive VTE cases were missed.Sensitivity: 100% (95% CI: 96.3%-100.0%)
    Specificity for VTE diagnosisWhile not as critical as sensitivity for exclusion assays, reasonable specificity helps reduce unnecessary further testing. No explicit numerical threshold is provided, but the reported specificity provides insight into the assay's ability to correctly identify true negatives.Specificity: 55.5% (95% CI: 49.0%-61.8%)
    Negative Predictive Value (NPV) for VTE diagnosisHigh NPV is paramount for an exclusion assay, as it indicates the probability that a patient truly does not have the condition if the test result is negative. Similar to sensitivity, 100% NPV is an ideal outcome for an exclusion assay.NPV: 100% (95% CI: 97.3%-100.0%)
    Detection LimitLower limit of D-Dimer concentration that can be reliably detectedThe detection limit is an inherent performance characteristic. A lower detection limit can offer more sensitive assessment. The document compares this to predicate devices.6.51 ng/mL
    Linear RangeRange of D-Dimer concentrations over which the assay provides accurateThis defines the measurable range of the assay. A wide linear range with auto-rerun capability is advantageous. The document compares this to predicate devices.54.3 - 1110000 ng/mL with Auto Rerun

    Study Information

    2. Sample size used for the test set and the data provenance:

    • Test Set Sample Size (Clinical Management Study): 344 frozen citrated plasma samples.
    • Data Provenance: From patients admitted to an emergency unit with suspected PE or DVT. The country of origin is not explicitly stated, but the context generally implies a clinical setting in a developed country (e.g., US or EU) for regulatory submissions of this type. The samples were retrospective, as they were "frozen citrated plasmas from patients."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • The document does not explicitly state the number of experts or their qualifications.
    • Ground Truth Establishment: "97 were confirmed as VTE positive (64 PE and 33 DVT) by standard objective tests and the remaining 247 were confirmed as negative." The "standard objective tests" would typically be interpreted and confirmed by medical specialists (e.g., radiologists for imaging, clinical physicians for diagnosis confirmation), but this is not detailed.

    4. Adjudication method for the test set:

    • The document does not specify an explicit adjudication method (e.g., 2+1, 3+1). The ground truth was established by "standard objective tests" and clinical confirmation.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done:

    • No, an MRMC comparative effectiveness study was not done. This submission focuses on the performance of a diagnostic instrument and assay (ACL AcuStar D-Dimer) as a standalone test, and its clinical utility for exclusion of VTE, rather than human reader improvement with AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Yes, the performance data presented is for the standalone device (ACL AcuStar D-Dimer assay and instrument) without human-in-the-loop assistance in its interpretation. The results (sensitivity, specificity, NPV) directly reflect the device's diagnostic capability.

    7. The type of ground truth used:

    • The ground truth was established by outcomes data and expert diagnosis based on standard objective tests:
      • VTE positive cases (PE and DVT) were "confirmed ... by standard objective tests." These typically include imaging studies (e.g., CT pulmonary angiography for PE, ultrasound for DVT) and clinical assessment.
      • VTE negative cases were "confirmed as negative" also presumably by the absence of findings on standard objective tests and clinical follow-up.

    8. The sample size for the training set:

    • The document does not specify a separate "training set" in the context of machine learning. This is a traditional in-vitro diagnostic device submission for an immunoassay. The concept of a training set for an algorithm is not directly applicable here. The "training" in this context refers to the development and optimization of the assay and its reagents during product development, which is typically done with various in-house samples and analytical studies.

    9. How the ground truth for the training set was established:

    • As noted above, the concept of a separate "training set" with ground truth in the machine learning sense is not explicitly present. For an immunoassay, the "ground truth" during development (analogous to training) would involve using characterized samples with known D-Dimer concentrations or clinical status, and optimizing the assay's reagents and calibration curve to achieve accurate measurements and clinical cut-offs. This development process for assays does not typically involve the formal "ground truth establishment" and "adjudication" methods seen in AI/ML validation studies.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1