Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K112730

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-01-13

    (115 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BIOMET 3i Dental Abutments are intended for use as accessories to endosseous dental implants to support a prosthetic device in a partially or completely edentulous patient. A dental abutment is intended for use to support single and multiple tooth prosthesis, in the mandible or maxilla. The prosthesis can be screw or cement retained to the abutment. For compatibility of BellatekTM Patient Specific Abutments, please refer to the Table below.

    Device Description

    BIOMET 3i Encode Patient Specific Abutment for Straumann Bone Level Connection can be compared to current Encode Patient Specific Abutment for Nobel Replace Connection. The change to this device is only to the connection that is intended to interface which is identical to the interface found on the STRAUMANN bone level Implant. This feature does not modify the intended functionality of the Implant. implant/abutment system. This device is also intended for dental laboratories to provide limited design input to match the anatomical requirements of the patient.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study findings for the "Encode Patient Specific Abutment for Straumann Bone Level Connection", based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Feature/TestAcceptance Criteria (Implied)Reported Device Performance
    Mechanical Integrity (Dynamic Fatigue)Meets ISO 14801:2007 "Dentistry - Dynamic Fatigue Test for Endosseous Dental Implants" standards, including worst-case scenarios (30° pre-angled abutments).All testing conducted met the acceptance criteria. Performance testing data indicates substantial equivalence to the predicate device.
    MaterialTi-6Al-4V ELI (for both abutment and screw)Ti-6Al-4V ELI
    BiocompatibilityBiocompatibleYes
    Statistical Tolerance AnalysisSuccessful statistical tolerance analysisYes
    Intended UseSame as predicate deviceSame as predicate device
    Indications for UseSame as predicate deviceSame as predicate device
    Technological CharacteristicsSimilar to predicate device, with primary difference only in the mating connection.Similar to predicate device, with primary difference only in the mating connection compatible with Straumann Bone Level Implant.
    Principles of OperationSame as predicate deviceSame as predicate device

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: The document does not explicitly state the specific number of units or samples used for the "Design Verification Testing" according to ISO 14801:2007. It only mentions that testing was conducted.
    • Data Provenance: The data is a result of prospective bench testing conducted by the manufacturer, BIOMET 3i. There is no indication of country of origin for the data provided, but the submitter is based in Palm Beach Gardens, FL, USA.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • This evaluation is based on bench testing (mechanical performance), not clinical data or expert interpretation of patient data. Therefore, the concept of "ground truth" established by experts in the context of diagnostic accuracy for the test set is not applicable here. The "ground truth" for mechanical testing is adherence to the specified ISO standard and FDA guidance.

    4. Adjudication Method for the Test Set

    • Not applicable. As this is a bench test for mechanical performance, there is no adjudication method by experts for a test set in the way it would apply to clinical or imaging studies. The evaluation is based on objective measurements against a standard.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The submission explicitly states "Clinical Data: N/A". This is a premarket notification focused on demonstrating substantial equivalence primarily through technical and performance characteristics (bench testing) to a predicate device, rather than a study involving human readers and clinical outcomes.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    • No, a standalone algorithm performance study was not done. This device is a physical dental abutment, not an AI algorithm or software.

    7. Type of Ground Truth Used

    • The "ground truth" for the performance study (Design Verification Testing) was based on compliance with established mechanical testing standards and guidance:
      • ISO 14801:2007 "Dentistry - Dynamic Fatigue Test for Endosseous Dental Implants"
      • FDA Guidance Document: "Guidance for Industry and FDA Staff: - Special Controls Class II -Standards: Root Form Endosseous Dental Abutment/ Implant"

    8. Sample Size for the Training Set

    • Not applicable. This device is a physical medical device, not an AI/ML algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable, as there is no training set for this type of device.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1