510(k) Data Aggregation

    K Number
    K222706

    Validate with FDA (Live)

    Date Cleared
    2022-12-06

    (90 days)

    Product Code
    Regulation Number
    882.1330
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    iCE-SG2 Subcutaneous Electrode Kit is intended for temporary (<14 days) use with recording and monitoring equipment for the recording and monitoring of electrical signals at the subsurface level of the brain.

    Device Description

    The iCE-SG2 Subcutaneous Electrode Kit is intended for temporary (<14 days) use with recording and monitoring equipment for the recording and monitoring of electrical signals at the subsurface level of the brain. The subject device allows for continuous electroencephalograph (EEG) monitoring in the subcutaneous space. The iCE-SG2 Subcutaneous Electrode Kit can connect to commonly used electrophysiology systems. The subject device is provided sterile and for single patient use in hospitals by healthcare professionals (HCPs).

    AI/ML Overview

    The document describes the iCE-SG2 Subcutaneous Electrode Kit and its substantial equivalence to a predicate device (iCE-SG Subcutaneous Electrode Arrays, K201678). The acceptance criteria and the study that proves the device meets them are primarily focused on biocompatibility and performance (bench), as this is a medical device for recording electrical signals from the brain.

    Here's the breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Biocompatibility: Meet ISO 10993 standards and establish biological safety.Biocompatibility evaluation conducted within risk management framework and in compliance with ISO 10993 standards. Evaluation included relevant data sources related to biological safety of finished device testing of previously cleared product (K201678) and kit components with a history of safe biological use. Conclusion: Biological safety established.
    Performance Testing (Bench): Demonstrate reproducibility in electrode insertion, stability of electrode array position, and removal from subcutaneous space. Also, pre-insertion and post-removal impedance testing should be performed.A cadaver study was conducted using ten electrodes. Functions associated with electrode array insertion, stability, and removal were shown to perform as expected. Pre-insertion and post-removal impedance testing was performed. Conclusion: All 10 out of 10 electrode arrays performed successfully based on prespecified criteria.

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: 10 electrodes were used in the cadaver study for performance testing.
    • Data Provenance: The document does not explicitly state the country of origin. The study was a "cadaver study," implying human cadavers, and is presented as part of a U.S. FDA 510(k) submission, suggesting it was conducted to U.S. regulatory standards. It is a prospective evaluation of the new device's performance.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not mention the use of experts to establish a "ground truth" in the traditional sense for the performance testing. The performance criteria (reproducibility of insertion, stability, removal, and impedance measurements) are objective and measured directly, not subject to expert interpretation for ground truth.

    4. Adjudication Method for the Test Set

    Not applicable, as a traditional ground truth based on expert consensus was not established for the performance testing. The assessment of successful performance was based on pre-specified objective criteria.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, If so, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    Not applicable. This device is an electrode kit for recording electrical signals, not an AI-powered diagnostic tool requiring human reader interpretation or MRMC studies.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable. This is a physical medical device, not an algorithm.

    7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)

    The "ground truth" for the performance study was based on objective, pre-specified criteria and measurements from the cadaver study (e.g., successful insertion, stability, removal, and impedance values). For biocompatibility, it was based on compliance with ISO 10993 standards and a history of safe biological use for components.

    8. The Sample Size for the Training Set

    Not applicable. This device is not an AI algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. This device is not an AI algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K201678

    Validate with FDA (Live)

    Date Cleared
    2021-03-08

    (262 days)

    Product Code
    Regulation Number
    882.1330
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    iCE-SG Subcutaneous Electrode Arrays are intended for temporary (<14 days) use with recording and monitoring equipment for the recording and monitoring of electrical signals at the subsurface level of the brain.

    Device Description

    The iCE-SG Subcutaneous Electrode Arrays are intended for use with recording and monitoring equipment for the purpose of recording electroencephalograph (EEG) signals. The subject device allows for continuous EEG monitoring in the subcutaneous space. The iCE-SG Subcutaneous Electrode Arrays can connect to commonly used electrophysiology systems. The subject device is provided sterile and for single patient use in hospitals by healthcare professionals (HCPs).

    A kit includes the following components:

    1. Preparation box
    2. Insertion kit box
    3. Two iCE-SG electrode boxes
    AI/ML Overview

    Here's an analysis of the acceptance criteria and supporting study for the iCE-SG Subcutaneous Electrode Arrays, based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" in a quantitative, measurable format for device performance. Instead, it details various tests to demonstrate safety and equivalence to a predicate device. The "results" column below reflects the outcomes reported for these tests, which were deemed sufficient for demonstrating substantial equivalence.

    CategorySpecific Test / CriterionReported Device Performance / Result
    BiocompatibilityCytotoxicity (L-929 cells; EMEM extract; % viability / morphological grading)Non-cytotoxic
    Sensitization (Intradermal injection and topical application in guinea pigs; sesame oil/0.9% saline / cottonseed oil/0.9% sodium chloride extracts)Non-sensitizing
    Irritation (Intracutaneous injection in New Zealand white rabbits; cottonseed oil/sodium chloride extracts)Non-irritating
    Acute Systemic Toxicity (Intraperitoneal injection of albino swiss mice with cottonseed oil/sodium chloride extracts / Intravenous injection of albino swiss mice with 0.9% sodium chloride extract)Non-toxic
    Pyrogenicity (Marginal ear vein injection of New Zealand white rabbits; 0.9% sodium chloride extract)Non-pyrogenic
    Material-Mediated Pyrogenicity / Implantation (Implantation of two articles for 28/29 days)Non-toxic
    Implantation (Four weeks, left hemisphere in New Zealand White rabbits)Non-bioreactive
    Genotoxicity (L5178Y cells; RPMIi and PEG extracts; Visual assessment; Top agar plating)Non-mutagenic
    Performance Testing (Bench)Kit component dimensions examinationDemonstrated equivalence
    Packaging opening orientation examinationDemonstrated equivalence
    Kit components colors, markings, and graphics examinationDemonstrated equivalence
    Sharp edges examinationDemonstrated equivalence
    Opacity of the packaging examinationDemonstrated equivalence
    Trocar sheath tool's penetration tip bending force enduranceDemonstrated equivalence
    Bending force endurance of the exit assist deviceDemonstrated equivalence
    Holding endurance of the posterior stopperDemonstrated equivalence
    Trocar sheath tool adhesion enduranceDemonstrated equivalence
    Adhesion of the passage assist deviceDemonstrated equivalence
    Passage assist device bending resistanceDemonstrated equivalence
    Anterior stopper enduranceDemonstrated equivalence
    Cadaver studyDemonstrated equivalence
    Performance Testing (Animal)Durability to record EEG after 14 days continuous implantation in the subcutaneous spaceDemonstrated durability
    Technological Characteristics ComparisonIndications for Use (<14 days use, recording/monitoring only, no stimulation)Substantially equivalent to predicate (shorter duration, no stimulation in subject device; deemed not to raise new safety/effectiveness questions)
    Product Codes / Regulation NumberIdentical to primary/secondary predicates
    Regulation DescriptionIdentical to primary/secondary predicates
    Use Environment (Hospitals including ICU)Substantially equivalent (same/similar environments; convenience kits expand use but no new safety questions)
    Device Components (Includes prep/insertion kits)Substantially equivalent (kits make use easier; nonclinical testing showed no new safety/effectiveness questions)
    Insertion Kit ComponentsSubstantially equivalent (trocars sold separately by predicates; no new safety/effectiveness questions)
    Location of Placement (Extracranially)Identical to primary/secondary predicates
    Duration of Use (<14 days)Substantially equivalent (shorter than predicate, but nonclinical data show no new safety/effectiveness questions)
    Contraindications (Risk for infection, unsafe/ineffective procedure)Substantially equivalent (predicate includes stimulation, subject device does not; no new safety/effectiveness questions)
    Electrode configuration (Array)Identical to primary/secondary predicates
    Single Patient Use (Yes)Identical to primary predicate
    Disposable (Yes)Identical to primary predicate
    Electrode Material (Platinum Iridium)Identical to primary predicate
    Electrode Diameter (1.12 mm)Falls within predicate range (0.86-1.96 mm); no new safety/effectiveness questions
    Maximum Stimulation Charge Density (< 30 µC/cm²) (Not applicable for subject device as it doesn't stimulate)Identical to primary predicate (not applicable to subject device but meets criteria if it were)
    Insertion Procedure (Tunneled subcutaneously, stoppers, <14 days monitoring, bedside withdrawal)Substantially equivalent (shorter duration than predicate; no new safety/effectiveness questions)
    Use of Stylet (Removed prior to passage)Identical to primary predicate
    Impedance Cutoff for Each Electrode Contact Prior to Product Release (kΩ) (0.25)Identical to primary predicate
    Sterilization (Ethylene oxide)Identical to primary predicate

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify exact sample sizes for most of the tests, only the types of animals used (e.g., L-929 cells, guinea pigs, New Zealand white rabbits, albino swiss mice).

    • Biocompatibility: Involves various cell lines and animal models as listed in the table above. Specific numbers of animals per test are not provided.
    • Performance Testing (Bench): Refers to "Examination" of various physical characteristics and "endurance" tests. No numerical sample sizes are given.
    • Performance Testing (Animal): Mentions "14 days continuous implantation" for durability in the subcutaneous space. No specific number of animals is stated for this.
    • Cadaver Study: Mentioned under Performance Testing (Bench), but no specifics on the number of cadavers or the details of the study are provided.

    The provenance of this data (e.g., country of origin, retrospective/prospective) is not explicitly stated. These are likely prospective studies conducted specifically for this submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document does not mention human experts used to establish "ground truth" for the test set in the context of diagnostic accuracy, as this device (iCE-SG Subcutaneous Electrode Arrays) is a recording electrode and not an AI-powered diagnostic tool. The "ground truth" for the electrode's performance is based on objective laboratory and animal testing results (e.g., non-cytotoxic, non-irritating, successful EEG recording durability).

    4. Adjudication Method for the Test Set

    This information is not applicable as the device is not an AI diagnostic tool requiring expert adjudication of outputs. The tests performed are objective, measurable outcomes in a laboratory or animal setting.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    An MRMC comparative effectiveness study was not performed or mentioned. This type of study is relevant for AI-assisted diagnostic devices where human reader performance (with and without AI) is evaluated. The iCE-SG Subcutaneous Electrode Arrays are recording electrodes, not a diagnostic system for interpretation.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    A standalone study of an algorithm is not applicable because the device is a hardware electrode designed for recording electrical signals, not an algorithm for data processing or diagnosis. The performance studies focus on the physical and biological compatibility of the electrode itself.

    7. Type of Ground Truth Used

    The "ground truth" for evaluating this device's safety and effectiveness is primarily based on:

    • Objective laboratory measurements: (e.g., cytotoxicity, irritation, pyrogenicity, specified physical properties like bending force, impedance cutoff).
    • Biological responses: Observed in animal models (e.g., non-sensitizing, non-bioreactive, successful EEG recording).
    • Comparison to predicate device characteristics: Demonstrating that the subject device's characteristics (e.g., dimensions, materials, use duration) are either identical, within acceptable ranges, or that any differences do not raise new questions of safety or effectiveness when compared to previously cleared devices.

    8. Sample Size for the Training Set

    This is not applicable. The device is an electrode array, not a software algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    This is not applicable as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K191868

    Validate with FDA (Live)

    Date Cleared
    2020-03-31

    (263 days)

    Product Code
    Regulation Number
    882.1400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The iCEWav Neuromonitoring Platform (iCEWav) is intended to monitor the state of the brain by recording and displaying EEG signals and can also receive and display a variety of vital signs and other measurements from third-party devices.

    The iCEWav Neuromonitoring Platform is intended for use by a physician or other qualified medical personnel. It is intended for use on patients of all ages within a hospital or medical environment including the intensive care unit, operating room, emergency room, and other clinical settings where brain monitoring may be indicated. The system provides no function that will directly prevent, mitigate, screen, treat or diagnose a specific disease or condition.

    Device Description

    The iCEWav Neuromonitoring Platform (iCEWav) is a clinical neuromonitoring system designed to streamline the collection and integration of high-fidelity electroencephalographic (EEG) data and matched systemic physiological data to assist clinicians caring for patients with brain injury.

    The iCEWav Neuromonitoring Platform includes a computer hardware element and associated touchscreen monitor that are affixed to a wheeled cart approved for use in the clinical environment. The computer hardware system is attached via USB connection to a third-party EEG amplifier system. A separate storage basket is included on the cart for storage of the EEG amplifier when not in use. The computer hardware element includes a range of standard USB. Ethernet and serial ports to allow for connection with a range of third-party physiological monitoring systems found in the clinical environment.

    The iCEWav Neuromonitoring Platform encompasses in-house iCEWav software that serves to gather EEG and systemic data, automate functions of device performance to allow ease-of-use by clinicians, combine data in a format that allows for seamless real-time and retrospective evaluation, and provide a Graphical User Interface that allows a clinician to evaluate data according to preferences of the clinician.

    EEG data collection utilizes disposable third-party clinical EEG electrodes connected to a thirdparty clinical EEG amplifier, with resulting digital signals transmitted to the iCEWav computer hardware element. In parallel, correlative physiological data (heart rate, blood pressure, etc.) from third-party monitors are transferred into the system using standard data connections (serial, Ethernet or USB). Time-locked EEG and systemic physiological data are made available for real-time display as well as local hardware-based storage. Using iCEWav, a clinician can simultaneously monitor aspects of systemic physiology and brain electrical activity to help identify and interpret ongoing clinical developments which may have negative effects on brain health.

    EEG data can be displayed as raw waveforms or can be displayed using a range of quantitative measures including Compressed Spectral Array (CSA), Total Power, Alpha/Delta Ratio, Burst Suppression Ratio, and Asymmetry Index. Data can be viewed as it is collected in real time; alternatively, stored data can be reviewed retrospectively. EEG and physiological data can be viewed in parallel fashion. Scaling parameters for each data variable as well as the time window to be reviewed can be controlled by a clinical user.

    The iCEWav system has several features to optimize ease of use. The patient Admission procedure includes a step-by-step walkthrough process to assist the user in configuring the EEG system and perform initial electrode impedance testing to ensure correct connection to the third-party EEG amplifier. The system provides several options for baseline EEG recording montages that the user can select to streamline the initiation of EEG recording. The user can also easily customize the recording montage. The system automatically detects availability of data inputs from other third-party physiological monitors that are connected to the system hardware and offers the user the list of the available inputs to be selected for display. The system offers the option for continuous impedance checking of connected electrodes through the third-party EEG amplifier, and for convenience the user may select to remove channels demonstrating poor impedance from the raw and quantitative EEG displays.

    During or following a period of recording, patient data may be accessed and reviewed from local storage.

    AI/ML Overview

    The provided text is a 510(k) summary for the iCEWav Neuromonitoring Platform. It aims to demonstrate substantial equivalence to predicate devices, rather than proving device performance against specific acceptance criteria through a clinical study.

    Therefore, the document does NOT contain information regarding:

    • A table of acceptance criteria and reported device performance: The document presents a "Substantial Equivalence Technical Characteristics" table, which compares features of the iCEWav platform to predicate devices, but it does not specify quantitative acceptance criteria for performance metrics (e.g., accuracy, sensitivity, specificity) and then report actual device performance against those criteria.
    • Sample size and data provenance for a test set: There is no mention of a test set, sample size, or whether the data was retrospective or prospective.
    • Number and qualifications of experts for ground truth: No experts are mentioned as establishing ground truth for a test set.
    • Adjudication method for the test set: Not applicable as no test set or ground truth establishment by experts is described.
    • Multi-Reader Multi-Case (MRMC) comparative effectiveness study: The document does not describe any study involving human readers or AI assistance, nor does it discuss effect sizes.
    • Standalone (algorithm only) performance: The device is a neuromonitoring platform that processes and displays data, not an algorithm providing a diagnostic output for which standalone performance would typically be evaluated in this context.
    • Type of ground truth used: No ground truth based on expert consensus, pathology, or outcomes data is described.
    • Sample size for the training set: There is no mention of a training set or corresponding sample size.
    • How ground truth for the training set was established: Not applicable as no training set is described.

    In summary, this document does not describe a study to prove the device meets acceptance criteria related to its performance in a clinical setting (e.g., diagnostic accuracy). Instead, it focuses on demonstrating substantial equivalence of the iCEWav Neuromonitoring Platform's technical characteristics and intended use to existing legally marketed predicate devices.

    The bulk of the relevant information provided in the document focuses on:

    • Device Description: The iCEWav Neuromonitoring Platform gathers, processes, and displays EEG and systemic physiological data.
    • Indications for Use: Intended to monitor the state of the brain by recording and displaying EEG signals and vital signs from third-party devices for patients of all ages in hospital settings. It explicitly states, "The system provides no function that will directly prevent, mitigate, screen, treat or diagnose a specific disease or condition."
    • Substantial Equivalence Comparison (Tables 1-4): This section directly compares the iCEWav platform with two predicate devices (Moberg Research's Component Neuromonitoring System and Persyst 13) across various features such as device class, intended user, indications for use, clinical application, hardware components, data input and display, patient admission process, data storage, and quantitative EEG parameter calculation. The "Comment" column consistently states "Equivalent," indicating that the manufacturer believes the new device performs similarly to the predicate devices in these aspects.